Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 99 - 99
1 May 2016
Ziaee H Daniel J Matharu G Pynsent P McMinn D
Full Access

Blood metal ion sampling can help detect poorly functioning metal-on-metal hip arthroplasties (MoMHA's) requiring revision. Little is known about the variation in these levels following bearing exchange. This study aimed to determine the changes that occur in blood and urine metal ion concentrations following MoMHA revision.

A single-centre prospective cohort study was undertaken between 2005 and 2012 of patients with failing large-diameter MoMHA's and high blood metal ions requiring revision to non-metal-on-metal articulations. All patients had normal renal function. Whole blood and urine were collected for metal ion analysis preoperatively and regularly following revision.

Twenty-three MoMHAs (21 hip resurfacings and 2 total hip arthroplasties; mean age 56.0 years and 65% female) were revised at a mean time of 7.9 years (range 2.0–14.5 years) from primary surgery. All revisions were performed by the senior author using primary total hip implants (12 ceramic-on-polyethylene bearings, 10 oxinium-on-polyethylene bearings, and 1 metal-on-polyethylene bearing implanted). Mean (range) metal ion concentrations pre-revision were: blood cobalt 13.9µg/l (1.32–74.7µg/l), blood chromium 8.9µg/l (1.29–57.3µg/l), urine cobalt 104.6µg/24 hours (4.35–747.3µg/24 hours), urine chromium 33.2µg/24 hours (4.39–235.4µg/24 hours). After revision the mean metal ion concentrations (percentage of pre-revision values) were: blood cobalt at 2 days=10.7µg/l (77%), 6 days=7.7µg/l (55%), 2 months=3.4µg/l (24%), 1 year=1.0µg/l (7%), 2 years=0.42µg/l (3%); blood chromium at 2 days=8.7µg/l (98%), 6 days=5.5µg/l (62%), 2 months=2.2µg/l (25%), 1 year=1.5µg/l (16%), 2 years=0.97µg/l (11%); urine cobalt at 2 days=31.9µg/24 hours (30%), 6 days=21.5µg/24 hours (21%), 2 months=6.1µg/24 hours (6%), 1 year=0.99µg/24 hours (1%), 2 years=0.61µg/24 hours (1%); urine chromium at 2 days=34.4µg/24 hours (103%), 6 days=15.8µg/24 hours (48%), 2 months=9.3µg/24 hours (28%), 1 year=2.8µg/24 hours (8%), 2 years=1.9µg/24 hours (6%).

Following MoM revision cobalt levels decline rapidly in an exponential pattern with a single rate of decay through the 2 year period, reaching reference levels within the first year. Chromium follows a similar pattern but starts lower and takes longer. Renal response to cobalt returns to reference level within days of revision.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 32 - 32
1 May 2016
Ziaee H McMinn D Daniel J
Full Access

The use of modular systems adds versatility to the implant system, better restoration of hip biomechanics and lower inventory to the hospital. There have been reports of high metal ions, ARMD reactions and high implant failure rates due to potential problems from taper failures. These are more common in metal-on-metal hip replacements, but are being also reported in other bearings.

Between 2001 and 2010, we performed 383 consecutive metal-on-metal (MoM) THRs through a posterior approach, using a BHR cup and Birmingham modular head with one of three different stems, all with 12/14 tapers. The earliest 104 hips employed a cemented MS30 stem (Zimmer GmbH, Winterthur, Switzerland). Subsequent 256 were Synergy and then 23 Anthology (both uncemented and both Smith and Nephew Orthopaedics, Memphis TN USA). There was no significant difference in the average age at surgery (65.4 years cemented vs 65.6 uncemented, p = 0.69), gender ratio (1.68 vs 1.89, p = 0.64), or bearing diameter (46.7 vs 46.8, p = 0.31). The earlier 203 Synergy stems were monoblock heads, while the remaining uncemented stems included a tapered sleeve in addition.

There were 3 deep infections and 11 debris-related failures (overall revision rate 4.9%). The revision rate from aseptic failures (ALTR, effusion, osteolysis or component loosening) is 2.87%. Kaplan-Meier analysis of the entire cohort showed a 10-year implant survival of 96.8% with revision for any reason as the end-point. Cemented stems had a 100% survival at 10 years and 98.6% at 12 years. The uncemented stems had a 93.8% survival at 10 years. Within the uncemented group, the monoblocks had a 5 and 10-year survival of 99.0% and 96.4% respectively while the sleeved had 98.7% (5 years) and 96.3% (7 years) and 82.5% at 8 years. Retreival analysis showed clear evidence of taper failure.

Our experience suggests taper failure leading to ALTRs and its sequelae. Others have reported ALTR type reactions in metal on polyethylene and ceramic on polyethylene bearing types as well in bearing diameters ranging from 28mm to 40mm. There is a need to improve taper design especially for use with large heads, and in high demand patients.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 34 - 34
1 May 2016
McMinn D Ziaee H Daniel J
Full Access

The natural knee allows multi-planar freedoms of rotation and translation, while retaining stability in the antero-posterior direction. It allows flexion with roll back, and medial, lateral and central rotation movements. The natural femoral condyles of the knee are spiral, therefore inducing a side to side translatory movement during flexion and extension. Incorporating all these features is vital in successful knee replacement design.

The different knee designs currently in use demonstrate different deficiencies in knee function. A study of 150 Posterior Cruciate (PCL) Retaining Total Knee Replacements [1] has shown that in 72% of knees direct impingement of the tibial insert posteriorly against the back of the femur was responsible for blocking further flexion. The mean pre-operative range of flexion was 105° and post-operative was 105.9°. For every 2mm decrease in posterior condylar offset, the maximum flexion was reduced by 12.2°.

The major disadvantage of the Posterior Stabilised (PS) Total Knee Replacement is gross anterior to posterior mid-flexion instability [2]. The Medial Rotation Total Knee Replacement is good in mid-flexion but not in high flexion where the femur slides forward on the tibia leading to impingement.

The Birmingham Knee Replacement (BKR) is a rotating platform knee design which is stable throughout the range of flexion. In high flexion, the BKR brings the femur to the back of the tibia. The BKR also has spiral femoral condyles, matching the natural kinematics of the knee. The combined static and dynamic effect is 10mm lateral translation of the femur in flexion and vice versa in extension.

Results for seventy nine BKRs (in seventy two patients) show the best Oxford Knee Score of 12 at follow up – excluding ten patients whose inferior scores were due to other pathologies. Knee flexion results show a 21° post-operative improvement in range of flexion.

On objective independent testing, maximum walking speed is slower for patients with a standard knee replacement (6.5km/h) and the loading through the replaced side does not match the normal side. Comparatively, patients with a BKR have a faster maximum walking speed of 11km/h and the loading closely matches that of the normal knee.

Studies based on the National Joint Register PROMs data [2] show that nearly thirty percent of Total Knee Replacement patients are not much better since their operation. A lot of improvement is needed in the design of knee replacements in order to achieve better function for knee replacement patients.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 102 - 102
1 May 2016
Daniel J Ziaee H McMinn D
Full Access

Objectives

To examine mortality and revision rates among patients with osteoarthritis undergoing hip arthroplasty and to compare these rates between patients undergoing cemented or uncemented procedures and to compare outcomes between men undergoing stemmed total hip replacements and Birmingham hip resurfacing.

Main outcome measures

Hip arthroplasty procedures were linked to the time to any subsequent mortality or revision (implant failure). Flexible parametric survival analysis methods were used to analyse time to mortality and also time to revision. Comparisons between procedure groups were adjusted for age, sex, American Society of Anesthesiologists (ASA) grade, and complexity.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 100 - 100
1 May 2016
Ziaee H Matharu G Daniel J Pynsent P McMinn D
Full Access

High short-term failure rates have been observed with a number of metal-on-metal (MoM) hip designs. Most patients require follow-up with blood metal ions, whichprovide a surrogate marker of in-vivo bearing wear. Given these results are used in clinical decision making it is important values obtained within and between laboratories are reproducible.

To assess the intra-laboratory and inter-laboratory variability of blood metal ion concentrations analysed by four accredited laboratories.

Whole blood was taken from two participants in this prospective study. The study specimen was obtained from a 42 year-old female with ceramic-on-ceramic hip arthroplasty failure resulting in unintended metal-on-ceramic wear and excessively high systemic metal ion levels. The control specimen was from a 52 year-old healthy male with no metal exposure. The two specimens were serially diluted to produce a total of 25 samples with different metal ion concentrations in two different anticoagulants each. Thus 50 samples were sent blinded in duplicate (total 100) to four accredited laboratories (A, B, C, D) to independently analyse blood metal ion concentrations. Ten commercially available reference specimens spiked with different amounts of metal ions were also obtained with known blood metal ion concentrations (range for cobalt 0.15µg/l-11.30µg/l and chromium 0.80µg/l to 37.00µg/l) and analysed by the four laboratories.

The intra-laboratory coefficients of variation for repeat analysis of identical patient specimens were 7.32% (laboratory A), 4.64% (B), 7.50% (C), and 20.0% (D). The inter-laboratory variability for the analysis of all 25 samples was substantial. For the unmixed study specimen the laboratory results ranged from a cobalt of 263.7µg/l (D) to 525.1µg/l (D) and a chromium of 13.3µg/l (D) to 36.9µg/l (A). For the unmixed control specimen the laboratory results ranged from a cobalt of 0.13µg/l (B) to 0.77µg/l (D) and a chromium of 0.13µg/l (D) to 7.1µg/l (A). For one of the mixed specimens the laboratory results ranged from a cobalt of 12.50µg/l (A) to 20.47µg/l (D) and a chromium of 0.73µg/l (D) to 5.60µg/l (A). Similar inter-laboratory variation was observed for the other mixed samples. The true mean (standard deviation) of the 10 commercial samples was 4.48µg/l (4.20) for cobalt and 8.97µg/l (10.98) for chromium. This was similar to the values obtained by all four laboratories: mean (standard deviation) cobalt ranged from 3.54µg/l (3.17) in laboratory A to 4.35µg/l (4.13) in laboratory D, and chromium ranged from 7.76µg/l (9.50) in laboratory B to 9.55µg/l (9.16) in laboratory A.

When testing patient samples, large variations existed both between and within four laboratories accredited to perform analysis of blood metal ion concentrations. However, this was not the case when assessing commercially spiked samples which are regularly used to validate laboratory testing. This is of great clinical concern and could lead clinicians to either recommend unnecessary revision or delay surgery, with both having the potential to adversely affect patient outcomes. It is recommended that laboratories use patient samples to assess the accuracy and reproducibility of the analyses performed. This may also assist in explaining the variations observed in this study.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 101 - 101
1 May 2016
Ziaee H Daniel J McMinn D
Full Access

Metal-metal surface replacement (MoMSRA) is increasingly used in the young. Systemic metal ion release and its effects cause concern. Do metal ions cross the placenta in pregnant women with potential mutagenic effects? The hypothesis is that metal ions pass freely through the placenta and there is no difference in maternal and cord metal levels.

This is a controlled cross-sectional study of women with MoMSRA. (n=25, mean age 32years, implantation 60months, 3 bilateral). The control group were 24 subjects with no metallic implant and not receiving cobalt/chromium supplements, mean age 31years. No patient was known to have renal failure. Whole blood specimens were obtained before delivery/ fluid-infusion and Cord blood specimens immediately after delivery.

Cobalt and chromium were detectable in all specimens in both cohorts. In the control group, the difference between maternal and cord levels was only 5 to 7% indicating free passage. Study group cord levels were significantly lower than maternal cobalt, p<0.05 and chromium p<0.0001 thus rejecting the null hypothesis.

The differences between maternal and cord metal ions in the controls indicate that normally the placenta allows an almost free passage of metal ions. The relative levels of metal ions in the maternal and cord blood in the study group reveal that the placenta exerts a modulatory effect on metal transfer.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 33 - 33
1 May 2016
McMinn D Ziaee H Daniel J
Full Access

Kinematics of the knee change during the full range of flexion [1]. The lateral femoral condyle (LFC) rolls back progressively through the entire range of flexion. The medial femoral condyle (MFC) does not move back during the first 110 degrees, but from 110 to 160 degrees it moves back by 10mm. This dual arc makes anatomical knee design a challenging task. In medial rotation, during flexion, the MFC stays in place, but the LFC moves forward in extension and backward in flexion. In lateral rotation the LFC stays still while the MFC moves back and forward in flexion and extension. During central rotation both condyles move reciprocally. However the knee is stable against an anteriorly or posteriorly directed force. It is important that all these degrees of freedom and stability are reproduced in total knee replacement (TKR) design.

Furthermore, the two femoral condyles together form a spiral. Like the threads of a screw in a nut they allow medio-lateral translation of the femur [2] in the tibial reference frame. During flexion the knee centre moves laterally nearly 20% of the width of the tibial plateau and in extension the femur translates medially. This medio-lateral translation occurs in the natural normal knee joint. This has special significance in knee design because the natural femur (along with the trochlea) moves laterally in flexion, allowing the patella to be sited laterally, while most regular TKRs drive the patella medially. In order to test this anomaly we studied patellar maltracking in vivo and in cadavers.

In vivo tracking studies using a patellar tracker during total knee replacement demonstrated average patellar maltracking of 10mm with regular TKR designs. Experiments on 22 cadaveric limbs using loaded quadriceps mechanism and trackers on the femur, tibia and patella, showed that in the natural knee the patella tracked medially by an average of 5mm. In regular TKRs the patella tracked medially. Compared to a normal knee, the patella in these designs maltracked on average by 10mm. Spiral condyles applied to the same bone cuts in the cadavers allowed the patella tracking to return to the expected lateral position in flexion.

This has application to total knee replacement design. Unless the spiral design is incorporated in the condyles, patellar maltracking is inevitable and is likely to cause lateral knee pain and stiffness post-operatively.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 98 - 98
1 May 2016
Ziaee H Daniel J McMinn D
Full Access

The modern era of hip resurfacing was initiated over two decades ago to address the poor results of existing hip replacement devices in young patients. High failure rates have been reported with certain resurfacings. This is a 1 to 17-year review of a single surgeon series of resurfacings in patients under the age of 50 years.

Between July 1997 and June 2014, 3627 hip resurfacing arthroplasties were implanted at our Centre in 2878 patients using a posterior approach. Of these 863 patients (1063 hips, 754 in men and 309 in women) were under the age of 50 years at the time of operation. They were followed up with postal questionnaires for up to 10 years through independent Outcomes Centres initially and are currently followed up by our own Centre.

18 patients (24 hips) died 5.9 (0.02 – 11) years after surgery due to unrelated causes, including one patient (1 hip) who was revised and died 5 years after revision. Mean follow-up is 11.9 years (0.8 to 17.8 years). There were 22 revisions altogether (2.1%) at a mean of 6.2 years (0.01 to 14.6 years) including one malpositioned cup in a female patient with developmental dysplasia who dislocated post-operatively and had to be repositioned. 9 hips failed from collapse of the femoral-head and 3 hips from femoral-neck fractures, giving rise to 12 femoral failures in all. There were six deep infections and three bearing-related failures including one pseudotumour, one for osteolysis and one for unexplained pain with neither metallosis nor pseudotumour.

With revision for any reason as the end-point Kaplan-Meier survival analysis showed 98.9%, 98.6% and 97.2% implant survival at 5, 10 and 17 years. Men had better survival (99.1% and 98.1% at 10 and 17 years respectively) than women (97.3% and 95.3%).

17-year cumulative revision rates were higher in patients with a pre-operative diagnosis of dysplasia (6.1%) and AVN (7.6%) compared to all other diagnoses combined (1.5%). Patients with osteoarthritis had the best results (99.5%, 99.3%, 98.4% at 5, 10 and 17 years respectively).

Our study shows that resurfacing arthroplasty is a viable option for hip arthritis in the young with a low incidence of wear-related failures in the long-term.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 25 - 25
1 Nov 2015
Matharu G Daniel J Ziaee H Pynsent P McMinn D
Full Access

Introduction

Little is known about variations in blood metal ions following bearing exchange for poorly functioning metal-on-metal hip replacements (MoMHRs). This study aimed to determine the changes that occur in blood and urine metal ion concentrations following MoMHR revision.

Methods

Between 2005–2012, a single-centre prospective cohort study was undertaken of patients with large-diameter MoMHRs and high blood metal ions requiring revision. Whole blood and urine were collected both pre-revision and post-revision for metal ion analysis.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 156 - 156
1 Mar 2013
Ziaee H Pradhan C Daniel J McMinn D
Full Access

Introduction

Metal-metal surface replacement (MoMSRA) continues to be used in young women. Systemic metal ion release and its effects cause concern. Do metal ions crossing the placenta in pregnant women have potential mutagenic effects? The hypothesis is that metal ions pass freely through the placenta and there is no difference in maternal and cord metal levels.

Methods

This is a controlled cross-sectional study of women with MoMSRA. (n = 25, 3 bilateral, mean age 32 years, time from implantation to delivery 60 months). The control group consisted of 24 subjects, mean age 31 years, with no metallic implant and not receiving cobalt/chromium supplements. No patient was known to have renal failure. Whole blood specimens were obtained before delivery and before any infusion or transfusion, and cord blood specimens immediately after delivery.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 47 - 47
1 Mar 2013
Daniel J Ziaee H Pradhan C McMinn D
Full Access

Introduction

Large diameter metal-on-metal hip arthroplasty (LDMMTHA) provides benefits of reduced dislocation rates and low wear. The use of modular systems allows better restoration of hip biomechanics. There have been reports of modular LDMMTHAs with tapered sleeves generating excessively high metal ions, due to possible mismatch between the titanium stem and the cobalt-chrome sleeve and the dual Morse tapers involved. We evaluated metal ion levels in LDMMTHA patients with and without a cobalt-chrome (CoCr) tapered sleeve.

Methods

A cross-sectional series of 91 patients with proximal porous titanium alloy stem LDMMTHA with identical design CoCr bearings, attending a 1 to 2-year review were assessed with routine clinical and radiographic examinations, hip scores and metal ion analysis. Of these 65 had a single Morse taper between monoblock CoCr heads and the stems. Twentysix had a tapered cobalt-chrome sleeve in addition, with the resultant dual taper. Mean bearing diameter was 46 mm in both groups and mean age was 58 years in the monoblocks and 66 years in the tapered sleeve group.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 155 - 155
1 Mar 2013
Ziaee H Daniel J Pradhan C McMinn D
Full Access

Introduction

Modern metal-on-metal bearing resurfacings have been in use for nearly two decades. Local and systemic metal ion exposure continues to cause concern. We could not find a prospective metal ion study in such patients with a 10-year follow-up. This is the first ten year prospective study of metal ion levels in blood and their release in urine following hybrid fixed metal-on-metal surface arthroplasty.

Methods

Twenty six patients were included in an ongoing longitudinal metal ion study of patients with unilateral metal-on-metal hip resurfacings. Three of them were excluded due to subsequent contralateral resurfacing and one has relocated abroad. Cobalt and chromium levels were assessed in 12 hour urine collections before and periodically after operation (5 days to 10 years) using high resolution plasma mass spectrometry. Mean age at operation was 53 years and mean BMI 27.9.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 49 - 49
1 Mar 2013
Pradhan C Daniel J Ziaee H McMinn D
Full Access

Introduction

This is a retrospective review of the incidence of deep venous thrombosis (DVT) in 679 consecutive unilateral primary hip arthroplasty procedures performed between January 2007 and December 2010 managed with no anticoagulants. Mean age at operation was 58 years. Mean BMI was 26. The prophylaxis regimen included hypotensive epidural anesthesia, compression stockings, intermittent calf compression, early mobilization and an antiplatelet agent.

Methods

562 hybrid hip resurfacing procedures and 117 uncemented THRs, all performed through a posterior incision were included. Doppler ultrasound screening for DVT was performed in all patients between the fourth and sixth post-operative days. Patients were reviewed clinicoradiologically 6 to 10 weeks after operation and with a postal questionnaire at the end of 12 weeks to detect symptomatic VTE incidence following discharge. 14 patients with pre-existent VTE, coagulation disorders or cardiac problems requiring anticoagulant usage were excluded.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 48 - 48
1 Mar 2013
Daniel J Pradhan C Ziaee H McMinn D
Full Access

Introduction

High early failure rates have been reported with certain metal-metal surface arthroplasties and good results have been reported with others. This is a minimum 10-year review of the first 1000 consecutive resurfacings including all ages and diagnoses from one centre.

Methods

The first 1000 surface arthroplasties (892 patients) were followed-up with postal questionnaires. Of these the first 402 hips (350 patients) were also invited for a clinico-radiological review. 54 patients (63 hips) died 6.7 years (0.7–12.6) later due to unrelated causes. Mean follow-up is 12.2 years (range 10.8–13.7). Radiographs were assessed independently by a senior musculoskeletal radiologist.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 153 - 153
1 Mar 2013
Daniel J Ziaee H Pradhan C McMinn D
Full Access

Introduction

Wear and corrosion of metal-on-metal (MM) bearings releases (a) soluble metal ions which collect locally and pass into the systemic circulation and (b) insoluble particles which undergo local deposition and lymphoreticular dissemination. Debris-related failures from osteolysis, metallosis and pseudotumours warrants revision of these MM bearing devices to non-MM bearing arthroplasties with the expectation that both the systemic and local effects will be reversed with time since the source of metal ion release is removed.

The purpose of the present study is to determine (a) whether metal ion levels in blood and urine decrease after revision of a MM bearing arthroplasty to a non-MM bearing device and (b) the rate at which this decrease is effected.

Methods

Blood and urine levels of cobalt and chromium ions are studied prospectively over two years in 15 patients whose MM resurfacings were revised to cross-linked polyethylene containing total hip replacements (THRs). Specimen collection was started before and periodically after the revision at 2, 4 and 6 days and 2 months, 6, 12 and 24 months after operation. None of the patients had other MM devices or compromised renal function.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 154 - 154
1 Mar 2013
Daniel J Ziaee H Pradhan C McMinn D
Full Access

Introduction

Hip simulator studies show that metal-on-metal bearing wear can be reduced by reducing the diametral clearance of the bearing. We present the six-year follow-up results of a prospective clinico-radiological and metal ion study in patients with a low clearance metal-metal surface arthroplasty. The results are compared to published results of similar design bearings with conventional clearance.

Methods

Twentysix male patients (mean age 55 years, mean BMI 26) who received a 50 mm bearing resurfacing (radial clearance 50μm) were included in an ongoing clinico-radiological and metal ion study. Urine/blood specimens were obtained before and periodically after hip resurfacing. Patients were also assessed with Oxford Hip Scores and Harris Hip Score questionnaires. Two hips were excluded during follow-up, one for revision and another for contralateral hip arthroplasty.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IV | Pages 7 - 7
1 Mar 2012
Daniel J Pradhan C Ziaee H McMinn D
Full Access

Introduction

The results of the Birmingham Hip Resurfacing (BHR) device in several series reveal that the predominant mode of failure is femoral neck fracture or femoral head collapse and that careful patient selection and precise operative technique are vital to the success of this procedure. In this report we consider the results of BHR in patients with severe arthritis secondary to femoral head AVN.

Methods

This was a single-surgeon consecutive series of BHRs with a minimum follow-up of 5 years. Fifty-nine patients with Ficat-Arlet grade III or IV femoral head AVN (66 hips) and treated with BHRs at a mean age of 43.9 years (range 19 to 67.7 years) were followed up for 5.4 to 9.6 years (mean 7.1 years). No patient died and none was lost to follow-up. Revision for any reason was the end-point and unrevised patients were assessed with Oxford hip scores. They were also reviewed clinically and with AP and lateral radiographs.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IV | Pages 28 - 28
1 Mar 2012
Ziaee H Daniel J Pradhan C McMinn D
Full Access

Introduction

Large diameter metal-metal total hip replacements (MM THRs) offer the advantages of low wear and low dislocation risk and are being increasingly used in high-demand patients whose bone quality rules out the possibility of a hip resurfacing. However suggests that large headed MM devices may result in greater systemic metal exposure compared to small diameter bearings. This raises fresh concerns of elevated systemic metal levels.

Methods

Whole blood concentrations and daily output of cobalt and chromium in 28 patients with unilateral large diameter MM THRs (42 to 54mm bearings) were studied at 1-year follow-up. These were compared with the whole blood levels in 20 patients at 1 year and daily output of metal ions in 28 patients with 28mm MM THRs at 1 to 3 years. Both bearings are made of high carbon cobalt-chrome alloy, the larger bearing is as-cast alloy and the smaller is wrought alloy. High resolution inductively coupled plasma mass spectrometry was used for analysis. None of the patients had other metal devices or compromised renal function. They had either a cemented polished tapered stainless steel stem or a cementless porous ingrowth titanium alloy stem.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_III | Pages 145 - 145
1 Feb 2012
Pradhan C Daniel J Ziaee H Pynsent P McMinn D
Full Access

Introduction

Secondary osteoarthritis in a dysplastic hip is a surgical challenge. Severe leg length discrepancies and torsional deformities add to the problem of inadequate bony support available for the socket. Furthermore, many of these patients are young and wish to remain active, thereby jeopardising the long-term survival of any arthroplasty device.

For such severely dysplastic hips, the Birmingham Hip Resurfacing (BHR) device provides the option of a dysplasia component, a hydroxyapatite-coated porous uncemented socket with two lugs to engage neutralisation screws for supplementary fixation into the solid bone of the ilium more medially. The gap between the superolateral surface of the socket component and the false acetabulum is filled with impacted bone graft.

Methods and results

One hundred and thirteen consecutive dysplasia BHRs performed by the senior author (DJWM) for the treatment of severely arthritic hips with Crowe grade II and III dysplasia between 1997 and 2000 have been reviewed at a minimum five year follow-up. There were 106 patients (59M and 47F). Eighty of the 113 hips were old CDH or DDH, 29 were destructive primary or secondary arthritis with wandering acetabulae and four were old fracture dislocations of the hip. Mean age at operation was 47.5 years (range 21 to 68 years – thirty-six men and forty-four women were below the age of 55 years).

There were two failures (1.8%) out of the 113 hips at a mean follow-up of 6.5 years (range 5 to 8.3 years). One hip failed with a femoral neck fracture nine days after the operation and another failed due to deep infection at 3.3 years.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 450 - 451
1 Nov 2011
Daniel J Kamali A Ziaee H McMinn D
Full Access

Evidence with respect to conventional hip arthroplasties suggests that device wear is related to patient activity rather than duration of usage. Activity level questionnaires appear to suggest that subjects with resurfacing arthroplasties continue to remain active after the procedure. However there is a paucity of objective evidence relating to the step rates of these patients in their daily lives and its effect on metal ions generated. The aim of this investigation is to assess

the activity levels of hip resurfacing patients as follow up progresses and

if there is any correlation between activity and metal ions generated.

Twenty-five consecutive male patients (average age 56 years) who underwent a unilateral 50 mm diameter hip resurfacing carried out by a single surgeon (DJWM) were recruited after informed consent. Patient step activity (Step Activity Monitor, SAM, Cymatech. Seattle WA, USA) was recorded at 1, 2, and 4-year follow-up stages and at the same time patient whole blood samples were collected and analysed using High Resolution Inductively Coupled Mass Spectroscopy (HR-ICPMS).

All patients in this study had well functioning hips at the four year follow up stage. All femoral components implanted were within the desired range of neutral to 10°. The mean acetabular component inclination angle was 42° (33° – 55°). Patient overall step activity remains unchanged up to the 4-year follow-up period.

At one year follow up, the whole blood cobalt and chromium concentrations show no correlation to mean number of steps taken per day by each patient (r2=0.02).

The correlation between whole blood cobalt and chromium concentration versus a function of body weight and peak index is not significant (r2=0.11).

This study provides objective evidence of the activity rates of patients at different stages of follow-up after a MoM surface replacement arthroplasty. It should be emphasised that the walking speeds of these patients on average was significantly slower than 1 Hz, which is generally used in laboratory hip simulator studies.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 39 - 39
1 Jan 2011
Daniel J Ziaee H Pradhan C McMinn D
Full Access

It has been suggested that metal ion level elevations in certain bilateral MM bearing arthroplasties were overwhelming the renal threshold for metal excretion leading to systemic build-up of metal ions above the expected levels. In order to investigate this we studied renal concentrating efficacy through concurrent specimens of urine and whole blood over a range of metal levels.

Concurrent specimens from 305 unselected patients were obtained. They include preoperative patients (76) and those with unilateral and bilateral hip arthroplasties (229) through to ten years after operation. 39 pre-operative specimens and 4 follow-up specimens had blood levels at or below the detection limit for cobalt or chromium and were excluded. The ratio of urine to whole blood concentration was 0.78 in the pre-operative patients. In patients with MM arthroplasties the different ratios in patients increased (from 3.1 to 9.2) with increasing urine concentrations.

The ratio of urine cobalt concentration to blood cobalt concentration is a measure of renal concentrating efficacy. Amongst pre-operative controls, this ratio is 0.78, indicating that there is renal conservation of cobalt. In terms of cobalt, these patients’ urine was dilute in comparison to whole blood. In patients with MM bearings, the ratio went up to 4.8 indicating that the kidneys were now actively excreting against a concentration gradient in an attempt to maintain internal milieu. If renal threshold was being breached at higher levels, then the ratio should progressively fall at higher concentrations. The trend in this experiment shows quite the contrary effect and the ratio reached 9.2 in those with urine cobalt > 15 μg/l, demonstrating that renal clearance efficiency holds up even against this steep gradient and that the threshold is not breached within clinically relevant levels.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 29 - 29
1 Jan 2011
Daniel J Ziaee H Pradhan C McMinn D
Full Access

This is the first six-year report of a prospective longitudinal study of daily output of cobalt and chromium in urine and their levels in blood following hybrid metal-on-metal surface arthroplasty.

Urine and whole blood specimens were analyzed before and periodically after hip resurfacing in 26 male patients after informed consent (mean age 52.9 years, mean BMI 27.9). Two of these patients have undergone contralateral hip resurfacings for progressive pain from end-stage arthritis and had to be excluded.

All patients were found to have well-functioning resurfacings at 6-year follow-up. No patient complained of persistent pain or disability. The median 6-year Oxford hip score was 12. Urine chromium and cobalt at six years were 3.9 and 7.8 μg/24 hrs and blood levels were 1.11 and 1.17μg/l respectively. Both urine and blood levels show a statistically significant early increase reaching a peak six months to one year postoperatively followed by a steady decrease over the following five years, although the individual reductions are not statistically significant, except for blood chromium where the 4 and 6 year levels were significantly lower than the 1-year level.

Elevated systemic metal exposure following MM bearing arthroplasty continues to cause concern. Our results show that metal release in these bearings shows a reducing trend after an initial peak dispelling the fear that a steady build-up of in vivo metal occurs with progressively increasing blood levels. However, as long as the significance of these elevated levels remains unknown, the need for continued vigilance persists.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 66 - 66
1 Jan 2011
Ziaee H Daniel J Pradhan C Pynsent P McMinn D
Full Access

Introduction: Metal-metal (MM) surface replacement arthroplasty of the hip is showing promise as an effective and enduring solution in the management of hip arthritis in the young, and is being increasingly used. The main concern is the release of metal ions into the systemic circulation and their long-term effects. There is concern that metal ions cross the placenta in pregnant women with MM bearing arthroplasties with the potential for mutagenic effects in the offspring. The hypothesis is that metal ions pass freely through the placenta and that there is no difference in the maternal and cord metal ion levels.

Methods: This is a controlled cross-sectional study of women of child-bearing age with MM bearing surface replacement arthroplasties. (n= 25, mean age: 32 years, mean duration after resurfacing 60.3 months, 3 bilateral). The control group consisted of 24 subjects who do not have a metallic implant, were not taking any supplements containing cobalt or chromium salts and were registered to undergo an elective LSCS at the regional hospital. Mean age of the control group was 31.3 years. No patient in either group was known to have renal failure. Whole blood specimens were obtained from the mothers just before delivery and before infusion of any fluids, and cord blood specimens were obtained immediately after delivery. High resolution inductively couples mass spectrometry (HRICPMS) was used for metal ion analysis.

Results: Cobalt and chromium were detectable in all specimens in both the study patients and controls. In the study group, mean cord metal ion levels were significantly lower than the maternal cobalt (p < 0.05) and chromium (p < 0.0001) levels thus rejecting the null hypothesis. In the control group, the mean difference between maternal and cord metal ion levels was only 5 to 7% and was statistically not significant (p > 0.5).

The mean difference in cord chromium between the study and control groups is not statistically significant (p > 0.05), although cord cobalt in the study group is significantly higher (p < 0.01) than that in controls. Whilst there is a four-fold elevation of maternal cobalt in the study patients and an almost 7-fold increase in maternal chromium levels as compared to the control group, the elevation in the cord cobalt and chromium in the study group are smaller.

Discussion and Conclusion: The differences between maternal and cord metal ions in the control patients indicate that normally the placenta allows an almost free passage of metal ions. This is understandable when we realize that these elements are also required by the developing foetus for its cellular and metabolic functions. The relative levels of metal ions in the maternal and cord blood in the study group patients reveal that the placenta does exert a modulatory effect on metal transfer. The mean cord levels of cobalt and chromium in these patients are only 57% and 24% of the maternal blood levels respectively. There is a continuing need for efforts to reduce metal ion release from artificial joints and also to monitor metal ion levels in patients with MM devices.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 401 - 401
1 Jul 2010
Ziaee H Daniel J Pradhan C McMinn D
Full Access

Introduction: Metal-on-metal (MM) bearing wear releases soluble metal ions which enter the systemic circulation and insoluble metal particles which collect in the periprosthetic tissues and disseminate through the lymphoreticular system. Disseminated particles also release ions through corrosion. The rate of metal ion level reduction following revision of a MM bearing offers insights into the relative contribution of metal ions from the bearing and from disseminated particles.

Methods: Whole blood concentrations and daily output of metal ions were studied, prospectively over a period of one year, in seven patients whose MM resurfacings were revised to metal–polythylene THRs. None of the patients had other metal devices or compromised renal function.

Results: Preoperative levels in these patients were highly elevated as expected from a failing device. Thereafter there is a trend of reducing metal levels in whole blood and urine in a biphasic manner. Over the first four weeks there is a rapid decline, followed by a period of slow decrease over the next twelve months (figure).

Discussion: The steep reduction of cobalt release immediately following revision supports the reported short half-life of cobalt ions. The later protracted trend can only be accounted for through progressive corrosion from previously worn particles. However this trend is also not sustained indefinitely and tends to approach control levels eventually.

Some authors have suggested that metal wear in patients with well-functioning MM bearings occurs only during the run-in wear phase and that continued corrosion of metal particles released during that period is responsible for metal level elevation later on. However the reducing trend in the later phase following revision in this study suggests that metal ion elevation from corrosion is not sustained indefinitely and therefore cannot by itself account for the persistent elevation of systemic metal levels throughout. Bearing wear continues to occur throughout bearing life.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 391 - 391
1 Jul 2010
Daniel J Ziaee H Pradhan C McMinn D
Full Access

Introduction: End-stage hip arthritis secondary to femoral head avascular necrosis (AVN) in young patients is a therapeutic challenge. Hip resurfacing (HR) has been showing excellent medium-term results in patients with osteoarthritis. Destructive changes in a large segment of the femoral head from AVN can increase the risk of postoperative femoral neck fracture or femoral head collapse following a resurfacing procedure. Careful patient selection and precise operative technique are vital to success. This is a study of the results of HR in patients with arthritis secondary to femoral head osteonecrosis.

Methods: This is a single-surgeon consecutive series with a 4 to 14-year (mean 8.6) follow-up. 95 patients (104 hips) with Ficat-Arlet grade III or IV osteonecrosis and treated with HR at a mean age of 43 (range 18 to 68) years. Two patients died due to unrelated causes and none is lost to follow-up. Revision for any reason was the end-point. Unrevised patients were assessed clinically and with Oxford hip scores and AP and lateral radiographs.

Results: Nine failures (1 fracture, 5 femoral head collapse, 2 infections, 1 cup loosening) give a failure rate of 8.7% and a cumulative survivorship of 89% at 14 years. In one further patient the femoral component has tilted into varus. He is asymptomatic but knows that he may need a revision if symptoms develop. No other patient shows clinical or radiological adverse signs.

Discussion and Conclusion: Several studies suggest that the results of arthroplasty are generally worse in AVN compared to those in osteoarthritis. HR has demosntrated good results in young patients with good quality femoral head bone. Reviewing the above results it appears to us that the relatively poorer cumulative survival observed in patients with a diagnosis of AVN (89%) makes AVN a relative contraindication to hip resurfacing.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 400 - 400
1 Jul 2010
Ziaee H Daniel J Pradhan C McMinn D
Full Access

Introduction: Systemic metal ion elevation continues to cause concern with metal-on-metal (MM) bearings, particularly in young people, in view of their expected long life-time usage. Reducing bearing clearance is claimed to be a means of reducing metal ion release.

Methods: 26 consecutive male patients (mean age 55 years, mean BMI 26) who fulfilled the inclusion criteria and received a 50mm bearing (diametral clearance 100 μm) were included. Clinico-radiographic review and urine and blood specimens were obtained before and periodically after the procedure. Two hips were excluded during follow-up, (one revision and another contralateral hip arthroplasty). Results were compared with a similar design bearing, 50 or 54 mm diameter and conventional clearance.

Results: At the four-year stage all patients had excellent hip function. However three patients had progressive acetabular radiolucent lines. Cobalt and chromium in both cohorts at all follow-up levels were significantly higher than the preoperative levels. Compared to the conventional clearance (CC) group, the pre-operative urine chromium and 6M to 48M urine cobalt and chromium were significantly lower in the LC group (p < 0.005). Blood metal levels were lower in the LC group at 1-year follow-up but showed a converging trend thereafter. At 4-year follow-up, the differences are considerably less, with no significant difference in blood cobalt (figure).

Discussion: Under ideal conditions, closely matched components (lower clearance) would lead to a thicker fluid film and less wear. However a larger clearance than ideal is needed to allow for asphericities, surface roughness, deformation and the evolution of in vivo lubricant. Peri-acetabular radiolucent lines cause concern. Attempts to reduce systemic metal exposure should not adversely affect other bearing characteristics such as friction. The search for a bearing which would generate low wear without producing a detrimental effect on other bearing attributes, such as friction, should continue.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 391 - 391
1 Jul 2010
Ziaee H Pradhan C Daniel J McMinn D
Full Access

Introduction: Metal-metal (MM) hip resurfacing is being increasingly used in the young. The main concern is the invariable systemic metal ion release. In young women the concern is that metal ions cross the placenta in pregnant women with MM bearing arthroplasties. We earlier presented an interim report on this subject the results of which ate established in a larger cohort

Methods: This is a controlled cross-sectional study of women of child-bearing age with MM resurfacings. (n= 22, mean age: 32 years, mean duration after resurfacing 60.3 months, 3 bilateral). The control group consisted of 24 pregnant subjects who did not have a metallic implant (mean age 31.3 years). Whole blood specimens were obtained from the mothers and umbilical cords at delivery.

Results: None of the babies had a congenital anomaly. Cobalt and chromium were detectable in all specimens including all controls. In the study group, mean cord metal ion levels were significantly lower than the maternal cobalt (p < 0.05) and chromium (p < 0.0001). In the control group, the mean cord blood metal levels differed very little from the maternal levels (p > 0.5). The mean difference in cord chromium between the study (0.33 μg/l) and control groups (0.21 μg/l) was not statistically significant, although the difference in cord cobalt was significant (0.41 μg/l).

Discussion: The differences between maternal and cord metal ions in the control patients is very small indicating that, under these circumstances the placenta allows an almost free passage of metal ions. The relative levels in the study group reveal that the placenta exerts a modulatory effect on metal transfer when maternal levels are above normal. Cobalt and chromium cross the placenta, irrespective of the presence of metal devices and therefore there is a need to continue efforts to reduce metal ion release.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 400 - 400
1 Jul 2010
Daniel J Ziaee H Pradhan C McMinn D
Full Access

Introduction: In vitro simulation experiments and in vivo metal ion studies have been used to investigate metal-metal bearing wear. In vitro studies demonstrate an early high wear phase followed by a rapid decline to a significantly lower steady state phase. Clinical metal ion studies have never shown such a significant fall in later years although they reveal early high wear. This study compares in vitro and in vivo wear rates.

Methods: In vivo measurements were obtained from daily cobalt excretion in 26 patients with 50 and 54mm resurfacings up to 4 years. Their activity averaged 2Mcyc-per-yr. In vitro measurements were obtained from gravimetric wear rates (Prosim hip simulator) of ten 50 mm diameter resurfacings of the same design. Diluted calf serum was the lubricant.

Results: Simulator results, shown in fig 1, are wear per day equivalent. In fig 2 it is seen that during the first year simulator results predict wear that exceeds metal ion output. This can be accounted for by postulating that particulate debris is higher during the early years. Subsequently the plots converge showing that particulate debris release is progressively reduced in comparison to metal ion release. At 3 years the simulator predicts lower wear than that observed in the metal ion study. This can be accounted for by postulating that corrosion of previously shed particles is responsible for the difference.

Discussion: From these results it can be stated that during the run-in period, 4/5ths of bearing wear occurs as insoluble particles and the rest is soluble metal ions. This relationship progressively changes through the steady state phase. At around the 3-year stage, even if we assume that most bearing wear releases soluble metal ions, nearly a fifth (2.8/14.4) can only be accounted for through passive corrosion of wear particles.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 391 - 391
1 Jul 2010
Daniel J Ziaee H Pradhan C McMinn D
Full Access

Introduction: Modern metal-metal (MM) Hip Resurfacing (HR) was developed as a conservative option for young patients with severe arthritis. Whilst some centres have reported excellent early results, other series have found a high incidence of osteolysis and still others reported soft tissue necrosis and periarticular changes. These are not always detectable with conventional imaging. This is the first 10 year clinico-radiological and multi-slice CT assessment of hip resurfacings.

Methods: The study includes 124 consecutive single-surgeon HRs (113 patients), mean age 52.8 years (27 to 75), mean follow-up 10.6 (10.4 to 10.8 years). Diagnoses include primary osteoarthritis (102), osteonecrosis (6), dysplasia (12) and others (4). Five patients (7 hips) died 5 to 10.3 years later from unrelated causes. Unrevised patients are reviewed with questionnaires, conventional radiographs and CT assessment.

Results: With revision for any reason as the end-point there were seven failures 0.4 to 9.7 years after operation (one failed from femoral neck fracture, four due to femoral head collapse and two were deep infections, 94% 10-year cumulative survival. Five cases showed osteolysis and four had neck thinning. No aseptic loosening, migration or malorientation is found. No patient is awaiting a revision.

Discussion and Conclusion: The performance of MMHR continues to be good at 10 years. Arthroplasty devices are known to manifest two phases of failure, one during early years and another in later years. Early failure with this device has been low. The interim years continue to be promising and we are yet to find out when the late failures are likely to occur.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_III | Pages 552 - 552
1 Aug 2008
Kamali A Daniel JT Javid SF Youseffi M Band T Ashton R Hussain A Li C Daniel J McMinn D
Full Access

Introduction: Cementless cup designs in metal-on-metal (MoM) hip resurfacing devices generally depend on a good primary press-fit fixation which stabilises the components in the early post-operative period. Pressfitting the cup into the acetabulum generates non-uniform compressive stresses on the cup and consequently causes non-uniform cup deformation. That in turn may result in equatorial contact, high frictional torque and femoral head seizure. It has been reported that high frictional torque has the potential to generate micromotion between the implant and its surrounding bone and as a result adversely affect the longevity of the implant. The aim of this study was to investigate the effects of cup deformation on friction between the articulating surfaces in MoM bearings with various clearances.

Materials and methods: Six Birmingham Hip Resurfacing (BHR) devices with various clearances (80 to 306 μm) were tested in a hip friction simulator to determine the friction between the bearing surfaces. The components were tested in clotted blood which is the primary lubricant during the early post-operative period. The joints were friction tested initially in their pristine conditions and subsequently the cups were deflected by 25– 35 μm using two points pinching action before further friction tests were carried out.

Results and Discussions: It has been reported that reduced clearance results in reduced friction. However, none of the previous studies have taken cup deflection into consideration nor have they used physiologically relevant lubricant. The results presented in this study show that for the reduced clearance components, friction was significantly increased when the cups were deflected by only 30 μm. However, for the components with higher clearance the friction did not change before and after deflection. It is postulated that the larger clearances can accommodate for the amount of distortion introduced to the cups in this study.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 314 - 314
1 Jul 2008
Daniel J Pradhan C Salama A Ziaee H McMinn D
Full Access

Introduction: The increasing use of metal-metal bearings in hip arthroplasty raises concerns of elevated systemic metal levels. Lubrication theory and in-vitro studies suggest that larger diameter bearings generate less wear. Does this hold true in real life?

Methods: This is a cross-sectional study of 181 patients with either a Birmingham Hip Resurfacing or a 28mm Metasul THR. Whole blood levels (at 1year) and urinary metal ion output in the early (1 to 3) years and medium term (4 to 6years) in patients with these bearings were compared. None of the patients had other metal devices or compromised renal function. Both devices used a high-carbon cobalt-chrome bearing with no post-casting heat treatment. Cementless porous-ingrowth titanium cups and cemented polished tapered stainless steel stems or cementless porous-ingrowth titanium alloy stems were used in THRs.

Results: Whole blood metal levels at 1-year and daily output of metal ions in the early years and medium term (figure 1) in both bearings were in the same range and without a significant difference.

Discussion: Metal ion monitoring is the best way to estimate bearing wear in vivo and the best measure of device safety in the long term. Total wear over a given time period is best estimated from timed metal ion excretion rates. Blood levels represent a balance between the release of metal from the device and its renal clearance.

An earlier study (Clarke et al JBJS(Br) 2003) suggests that smaller bearings generate less wear. In that study bearings with different metallurgy and wear properties were grouped together, a potential confounding factor. The present study does not suffer from that error and our findings do not support the view that a larger bearing diameter leads to either an increase or decrease in metal ion generation.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 314 - 314
1 Jul 2008
Daniel J Pradhan C Ziaee H Pynsent P McMinn D
Full Access

Introduction. Painful post-collapse femoral head osteonecrosis (AVN) continues to be a therapeutic challenge. Joint preserving surgery does not produce satisfactory results after femoral head collapse, making an arthroplasty almost inevitable. Does metal-metal resurfacing offer a conservative option that matches the consistent results of a stemmed THA in these patients?

Methods. 104 consecutive resurfacings (94 patients) performed for Ficat stage III/IV AVN were reviewed clinically, radiologically and with Oxford hip scores. Mean age at operation was 43.9 years. Aetiology included trauma (20%), steroids/chemotherapy (25%), alcohol abuse (8%), AVN secondary to Perthes’/SUFE (4%) and idiopathic (43%). Two types of devices were used a) McMinn Resurfacing Arthroplasty, HA-coated smooth uncemented cup and cemented femoral component and b) Birmingham Hip Resurfacing, porous HA cup and cemented femoral component.

Results. At 2 to 11.5 years (mean 7.7 years), there were seven (6.7%) failures: four from further femoral head collapse, two infections and one osteolysis, aseptic loosening. The cumulative survival at 11years is 90% (figure 1). Mean Oxford score at follow-up is 15.4. 55% participate in moderate to heavy work or participate in regular sporting activity.

Discussion. The results of any treatment modality in AVN are not as good as those in osteoarthritis. One reason is that the etiopathological factors that caused non-traumatic AVN (steroids etc) have the potential to cause continued femoral head damage. Post-traumatic patients need complex reconstructions and risk a higher failure rate. Those on immunosuppressants are prone to infections. The results of hip resurfacing in AVN in this series are no different from THRs in many published series. Metal-metal resurfacing is a good conservative option for post-collapse femoral head AVN and gives acceptable results if strict patient selection criteria are followed.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 305 - 305
1 Jul 2008
Daniel J Pradhan C Ziaee H McMinn D
Full Access

Introduction: Dislocation rates with THA vary from 3% to 15%. One specialist centre reported a 6.4% early dislocation rate with a 28mm ceramic on polyethylene THA in young patients (mean age 56 years) in a single surgeon series. Although young patients have the advantage of better soft tissues, their greater mobility demands increase dislocation risk.

Dislocation rates in large headed metal-on-metal resurfacings are extremely low. However, many patients are unsuitable for resurfacing and need a replacement. In such cases, it is attractive to transfer the large-headed metal-metal bearing advantage to replacement arthroplasty in order to reduce wear and dislocation rates. Does large diameter metal-metal total hip replacement really reduce the early dislocation rate?

Methods: 206 consecutive primary metal-metal THRs (189 patients) were included. The device consists of an uncemented cup, a matching modular cobalt chrome head (head diameter ranged 38 – 58mm) fixed on a stem through a 12/14 cone. Cemented stems were used in 107 procedures and 99 were proximal-porous uncemented stems.

Age at operation ranged from 37 to 83 years. Thirty patients were 55 years or under, eighty one were 56 – 65 years and ninety five were over 65 years. There were 122 females and 67 males. Posterior approach was used in all.

Results: There were no dislocations in these 206 consecutive procedures.

Discussion: Metal-metal hips have lower dislocation rates than hips containing polyethylene (0.9% against 6.4% in a matched series). This is attributed to the suction-fit effect of metal-metal bearings. Large diameter bearings have the additional benefit of having to translate a greater jump distance before a dislocation. This dual advantage leading to extremely low dislocation rates was first noted in metal-metal resurfacings. In large headed metal-metal THRs, the head-neck ratio is even more favourable and these devices appear to eliminate early dislocation as a major complication.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 299 - 299
1 Jul 2008
Daniel J Pradhan C Ziaee H Pynsent P McMinn D
Full Access

Introduction: Hip resurfacing is a bone conserving option that offers a better revision prospect for young and active patients. Encouraging results from several centres prove that they function well in the early years. Their long-term survival will be known from continued monitoring of early resurfacings.

Methods: This is a retrospective study of two cohorts of young (< 55 years) patients of osteoarthritis treated with hybrid-fixed metal-metal resurfacings. The cohorts are a) consecutive patients treated by the senior author in 1994 and 95 with a hydroxyapatite-coated smooth uncemented cup and a cemented femoral component and b) consecutive patients treated with hydroxyapatite-coated porous uncemented cup and a cemented femoral component since 1997 with a minimum follow-up of 5 years. 420 resurfacings (360 patients, 287 males and 73 females) were reviewed with Oxford hip scores and activity level monitoring (UCLA scale). Mean age at operation was 48.3 years.

Results: Ten patients (11 hips) died from unrelated causes. Out of the remaining 409 hips (350 patients) at a follow-up of 5 to 11.5 years (mean 7.1 years), there was one failure (cumulative failure rate 0.25% at 11 years) from avascular necrosis of the femoral head. The mean Oxford score of the 350 patients is 13.4. 87% had a UCLA score of 7 and above. 55% participated in impact sports or were involved in heavy occupational work.

Discussion: In the present study, with no loss to follow-up, excellent hip survival (99.75%) and activity level were seen. Young patients regard return to activities as one of their highest priorities. None of these patients were advised to change their activities at work or leisure.

The extremely low failure rate in the medium term proves the suitability of resurfacing in young active patients. However, caution needs to be exercised until long term results are available.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 299 - 299
1 Jul 2008
Ziaee H Daniel J Pradhan C McMinn D
Full Access

Introduction. The usage of metal-metal bearings in young patients has rekindled the debate about the potential adverse effects of mutagenecity on offspring born to them. This question could be answered in part if it was known whether metal ions are transferred to the developing foetus. One recent study seems to suggest that such transfer does not occur [Brodner et al, J Arthroplasty2004; 19 Suppl(3) p102–107]. Unfortunately the instrument used there was not sensitive enough (5 out of 6 analyses were below the limit of detection), leaving the question of transplacental metal ion transfer unanswered. The present study uses a more powerful analytic technique.

Methods. After informed consent, whole blood specimens were obtained at the time of delivery from five patients who had undergone a Birmingham Hip Resurfacing and from their babies’ umbilical cords. High resolution plasma mass spectrometry (HRICPMS) was used for analysis.

Results. Cobalt and chromium ions have been detected in all the specimens obtained so far. The cord blood cobalt levels were lower than the mothers blood levels in all the specimens. A similar relationship was found in all but one individual chromium measurement. The mean (±95% CI) of the two groups are shown in figure 1.

Discussion and Conclusion. The present study shows that with the use of whole blood specimens rather than serum; and the use of a better analytic method than previously used in the only other study on the topic, metal ions can be detected in all specimens of patients with metal-metal devices and in the cord blood of babies born to them.

There is sufficient evidence in this study to prove that metal ions do cross the placenta. There is therefore a continuing need for vigilance on the possible effects on the offspring born to patients with metal-metal devices.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 317 - 317
1 Jul 2008
Daniel J Kamali A Saravi S Youseffi M Ashton R Band T McMinn D
Full Access

Introduction: modern cementless joints depend on bony ingrowth for durable long term fixation. Increased friction and micromotion in the early weeks can prevent ingrowth and affect long-term success.

Most friction studies are conducted in a bovine serum- carboxymethylcellulose (bs-cmc) medium. Following implantation however, the joint is bathed in blood which contains macromolecules and cells. The effect of these on friction is not fully understood.

A progressive radiolucent line (fig 1) observed in some low clearance resurfacings raises the concern that increased friction may be affecting component fixation. The purpose of this investigation was to study the effect of clearance on friction for a given bearing diameter in the presence of blood as lubricant.

Methods: Six Birmingham Hip Resurfacing devices with a nominal diameter of 50mm each and a range of diametral clearances (80, 135, 175, 200, 243 and 306μm) were used. Frictional measurements were carried out on a Prosim Hip Friction Simulator (Simsol Simulation Solutions, Stockport, UK). The test was conducted sequentially with whole blood (viscosity 0.009Pas) and a BS-CMC mixture as the lubricants (viscosity 0.01Pas).

Results: Low clearance devices (80–175μm) generated higher friction with blood than with BS-CMC (fig 2). With blood as the lubricant, low clearance devices generated much higher friction than higher clearance devices (200–306μm).

Discussion: Ongoing research into the in vitro performance of bearings is performed in hip simulators with lubricants that are believed to simulate joint fluid in terms of viscosity. However these lubricants are unable to simulate the friction effects of macromolecules.

The results of this study suggest that reduced clearance bearings have the potential to generate higher friction when blood is the lubricant. this higher friction in the low clearance bearings may produce micromotion in the early postoperative period and hamper bony ingrowth resulting in impaired fixation with long-term implications for survival.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 309 - 309
1 Jul 2008
Daniel J Pradhan C Ziaee H McMinn D
Full Access

Introduction: Hip resurfacing is a good conservative option for young patients with arthritis. Resurfacings risk two unique failure mechanisms that do not occur in THA, i.e. femoral neck fracture and femoral head collapse.

Old age, osteopaenia, alcohol abuse, and large cysts are risk factors for fractures. It has been suggested that performing a bilateral resurfacing puts the first side at risk of fracture from the force used in implanting the second resurfacing. Is this a true risk or a sampling error?

Methods: Out of 2576 consecutive resurfacings performed by the senior author (July 1997 – May 2005), 191 patients (382 hips, 14.8% of all resurfacings) presented with bilateral arthritis and had both hips operated in the same hospital admission. 133 patients had the two operations a week apart and 58 had both the same day. A posterior approach was used in all cases with the patient in the lateral position on the contralateral side.

Results: Of the 382 resurfacings, only two failed from a femoral neck fracture. Both had the second operation a week after the first. A 35-year lady (rheumatoid arthritis) sustained a femoral neck fracture of the first hip following a fall nine weeks after the operation. A 57-year man (osteoarthritis) fractured his femoral neck at 3.5 months. He fractured the side operated second.

Discussion: The incidence of femoral neck fracture in the author’s series of 2576 resurfacings is 0.4%. Patients who present with bilateral severe arthritis are more likely to have non-primary OA such as inflammatory arthritis. It is difficult to conclude if such bilateral cases are more predisposed to a fracture by virtue of the pathology itself.

The low incidence of fractures (2/382, 0.5%) in this bilateral resurfacing series does not support the view that there is an increased risk of fracture from a bilateral procedure.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 316 - 316
1 Jul 2008
Daniel J Ziaee H Pynsent P McMinn D
Full Access

Introduction: Metal ions generated from metal-metal joints are transported in plasma and within cells. Whole blood (WB) analysis is a good measure of systemic metal exposure because it includes both intracellular and extracellular compartments. Analysis of whole blood is technically more challenging than serum. But can serum levels be considered equivalent to whole blood levels in the assessment of systemic metal exposure?

Methods: 262 concurrent WB and serum specimens from patients after metal-metal hip arthroplasty were analyzed for metal levels by high-resolution mass-spectrometry (reporting limits 0.06μg/l for serum and 0.1μg/l for WB). The relationship between serum and WB levels were studied using four methods of agreement testing: paired t-test, Pearson correlation, Bland-Altman agreement and variability on normalized-scatter.

Results: Mean WB and serum chromium concentrations were 2.2 and 4.2μg/l (p< 0.001). Corresponding values for cobalt were 2.4 and 3.2 μg/l (p< 0.001). The normalized scatter (fig 1B) shows poor congruence between whole blood and serum levels especially at lower concentrations. Pearson correlation confirms poor agreement at lower concentrations even after excluding outliers.

Bland-Altman analysis (Figure 1A) shows the limits of agreement between serum and WB are unacceptably wide (1.7 to -5.1 for chromium) suggesting poor agreement.

Discussion: The proportion of metal ion levels in serum and blood cells shows great variability. With advances made inmass-spectrometry, poor instrument sensitivity is a non-issue. Is there then a case for continued use of serum as a measure of systemic metal ion exposure? It can be justified only if serum levels show good agreement with WB levels throughout the range of measurements. Concurrent analysis of serum and whole blood metal ion concentrations in the present study, do not show such an agreement with four methods of testing, suggesting that serum metal concentrations cannot be reliably used as an instrument to measure systemic metal ion exposure.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 15 - 15
1 Mar 2006
McMinn D Daniel J Pradhan C Pynsent P Ziaee H Band T Ashton R
Full Access

Hip Resurfacing has always been an attractive concept for the treatment of hip arthritis in younger patients. Introduction of modern metal-on-metal hip resurfacing in 1991 in Birmingham, UK made this concept a reality.

In the early years, resurfacings were used only by a few experienced surgeons. From 1997, Birmingham Hip Resurfacings (BHRs) are being widely used in younger and more active patients. A breakdown of the ages at operation in the regional NHS hospital in Birmingham during the period April 1999 to March 2004 show that the mean age of metal-metal resurfacings is 51 years and the mean age of total hip replacements is 70 years.

At a 3.7 to 10.8 year follow-up (mean follow-up 5.8 years), the cumulative survival rate of metal-metal resurfacing in young active patients with osteoarthritis is 99.8%. In the long term, none of these patients were constrained to change their occupational or leisure activities as a result of the procedure. The overall revision rate of BHRs in all ages and all diagnoses is also very low (19 out of 2167 [0.88%] with a maximum follow-up of 7.5 years).

Improvements in instrumentation and a minimally invasive approach developed by the senior author have made this successful device even more attractive. Although objective evidence does not support the fact that the longer approach was any more invasive than the minimal route, patient feedback shows that it is very popular with them. While minimal approach is indeed appealing, it has a steep learning curve. In the early phase of this curve, care should be taken to avoid the potential risk of suboptimal component placement which can adversely affect long-term outcome.

It is true that metal-metal bearings are associated with elevated metal ion levels. In vitro studies of BHRs show that they have a period of early run-in wear. This is not sustained in the longer term. These findings are found to hold true in vivo as well, in our studies of 24- hour cobalt output and whole blood metal ion levels. Epidemiological studies show that historic metal-metal bearings are not associated with carcinogenic effects in the long-term.

Metal ion levels in patients with BHRs are in the same range as the levels found in those with historic metal-metal total hip replacements.


Orthopaedic Proceedings
Vol. 87-B, Issue SUPP_III | Pages 355 - 355
1 Sep 2005
O’Hara J McMinn D
Full Access

Introduction and Aims: The authors present their results following treatment of 15 patients with complex hip deformities by this new combination of operations.

Method: Fifteen patients aged 14 to 36 years (one male) were treated by contemporaneous metal-on-metal hip resurfacing and rotation osteotomy of the femur to nor-malise anteversion over a five-year period (1996–2001). The resufacing was performed in the usual way; anteversion was corrected at the end of the operation where limited internal or external rotation (< 20deg) was felt likely to interfere with the foot progression angle. The plate was removed about one year later.

Results: There were no peri-operative complications. Weightbearing was restricted until bone healing was complete (8–13 weeks). Thereafter patients mobilised normally. One patient had her plate removed at six months, as there was sleep disturbance due to local tenderness. At review, all patients were pleased with the outcome. Pre-operative HHS was 65–72: at review it was 89–96. There were no complications in the medium term. All patients had an abnormal foot progression angle pre-operatively (14 had fixed internal rotation, one external rotation). At review, in extension all fell within the physiological range IR50/ER50.

Conclusion: This new combination of established operations facilitates the bone conserving benefits of the metal-on-metal resurfacing with corrective rotational osteotomy in patients with complex hip deformity. We have avoided the use of expensive custom protheses and have allowed patients the benefits of a prosthesis minimising bone resection and retaining the physiological modulus of elasticity


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 158 - 162
1 Feb 2005
Itayem R Arndt A Nistor L McMinn D Lundberg A

The Birmingham hip resurfacing metal-on-metal arthroplasty was introduced in 1997 and has shown promising short- to mid-term results. We used radiostereophotogrammetric analysis (RSA) to study the stability of 20 resurfacing arthroplasties over a follow-up period of 24 months.

Radiological examinations for RSA were performed immediately after surgery and at two, six, 12 and 24 months after operation. Precision and detection of migration thresholds (non-zero movement) were calculated.

All the results corresponded well to those found in similar experimental arrangements with standard hip prostheses. Migration of the cup and vertical and mediolateral migration of the head were calculated. The values were low at two years compared with those of earlier studies of cemented femoral components in conventional total hip replacements indicating that there was no evidence of excessive early migration or loosening of the components.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 422 - 423
1 Apr 2004
Daniel A Pynsent P McMinn D
Full Access

The growing interest in the use of metal on metal bearings in the treatment of end stage hip arthritis in the young has raised the need to monitor metal ion levels in these patients. Blood levels are good indices of the safety of these devices. Total wear/corrosion over a given time period is best estimated from 12 or 24-hour urine collections. Whole blood samples of 52 patients with metal on metal hip arthroplasties were analysed for levels of chromium, cobalt and molybdenum using high-resolution induction coupled plasma mass-spectrometry. Fifteen of these were in an older age group (mean age 75.3years) and had metal on metal hip replacements performed 19–28 years before the assessment. The rest (37) were younger (mean age 52.9 years) and had hip resurfacings performed in the last 4 years. The results show higher levels of ions than those found in the normal population.

This agrees with other studies. However we found no significant difference between the levels in the older patients with THR and the younger patients with hip resurfacing. In another study, urine collections from 78 patients who had hip resurfacings in the past and are now at different postoperative durations (2, 4, 6, 8, 10, 12, 18, 24 and 36-month periods) were analysed for metal ions using the same technique. The mean levels show an increase compared to the preoperative values but this increase is not statistically significant.

The urinary excretion of cobalt in these groups have also been compared with the expected cobalt release due to corrosion in this alloy. They were found to rise above the corrosion levels only in the 8–12 month period, possibly due to “run-in” of the bearing. It can be concluded that the ion levels can be attributed mainly to corrosion and there is minimal contribution from wear.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 423 - 423
1 Apr 2004
McMinn D
Full Access

Development of the metal/ metal hip resurfacing began in 1989 with the first patient implantation in February 1991. In the first three years three methods of implant fixation were employed in a pilot study of 119 resurfacings and the optimum fixation was identified as Hydroxyapatite Coated (HA) uncemented cups and cemented femoral components.

From March 1994 until February 2001 the author has performed 1503 hybrid fixed resurfacing implants, 294 of the McMinn type (HA on Smooth Metal Cup) and 1209 Birmingham Hip resurfacings (BHR; HA on Porous Metal Cup). Mean age was 52.9 years, diagnosis – Osteoarthritis 78%; Dysplasia 7%; Inflammatory 2%; Avascular necrosis 5%; other 8%. In the 294 McMinn Resurfacings 8 have failed and 7 have been revised. Reasons for failure; 5 cup loosenings, two infections and one femoral head collapse in a patient with pre-existing avascular necrosis. High wear of the metal/ metal bearings, with altered metallurgy, is now presenting as a cause for concern with these McMinn resurfacings performed in 1996 where the metal was Solution Heat Treated (SHT) and Hot Isostatically Pressed (HIP). Patients from 1996 are exhibiting asymptomatic radiographic osteolysis and at revision metallosis is seen with high wear measured on the retrieved components. Laboratory studies have shown that SHT and HIP alters the carbide microstructure and increases wear on abrasive wear testing.

In the 1209 BHR’s 7 failures have occurred requiring revision. Reasons for revision – 4 femoral neck fractures, 2 infections and 1 collapsed femoral head from avascular necrosis. Implant fixation is reliable as judged by very low migration on an R.S.A. study. Bearing wear is very low as judged from wear analysis of explants as would be expected from the historically proven “As Cast” metal microstructure. Function is good as judged by the Harris Hip Score (mean 96.3 at 3years). Cumulative survival on all Hybrid fixed implants is 97% at 7–8 years.

Hip resurfacing, using metal on metal bearings with historically proven metallurgy using hybrid fixation, offers a viable treatment alternative for the young and more active patient with hip arthritis.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 242 - 242
1 Mar 2004
Kulkarni A Pynsent P McMinn D
Full Access

Aims: The rate of polyethylene wear has been reported to be considerably higher with uncemented metal backed cups articulating with cemented as well as uncemented femoral components. There are many reports comparing wear between cemented all polyethylene and uncemented metal backed cups from different manufacturers, however there are no reports comparing wear of polyethylene manufactured in identical method.

Patients and methods: 361 consecutive hips between 1988 and 1995 were operated at our hospital by using either cemented all polyethylene Zimmer Flanged (ZF) or uncemented Harris-Galante (HG) acetabular component rticulated with Exeter polished tapered cemented stainless steel stem with a 26mm size head. A total of 107 hips (75 ZF and 32 HG) were excluded. A custom-made computer programme was used to assess polyethylene wear (by using the method of Livermore etal). A statistical analysis was used to find average annual wear and significance of difference in the wear of ZF cups and HG cups using student’s t-test.

Results: There were 87 male and 167 were female patients with an average age of 63.92 years (18 – 86). The average follow-up for ZF cups was 3.85 years (2.5 – 5.4). and for HG cups was 5.3 years (2.5 – 9.3). The mean annual wear rate of ZF cups was 0.127mm (SD 0.088). For HG cups the mean was 0.1388mm (SD 0.086).

Conclusion: Polyethylene ZF cups have wear characteristics similar to HG cups at a mean follow-up of 4.6 years.


Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_III | Pages 232 - 232
1 Nov 2002
McMinn D
Full Access

Most total knee replacement designs incorporate the views of the designer relating to the function of the normal knee. A video of the normal cadaverie knee will be shown and depending on the loading regime the following movements can be clearly demonstrated: femoral roll-back, femoro-tibial rotation, medial pivot and lateral pivot. Knee replacement designs which do not allow all of the above movements risk Kinematic conflict. A knee replacement design will be shown which permits all of the normal knee movements.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 559 - 559
1 May 1998
McMinn D


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 3 | Pages 371 - 374
1 May 1993
Glithero P Grigoris P Harding L Hesslewood McMinn D

We report the results of imaging with labelled white cells in 52 patients before the revision of 54 cemented joint prostheses at which the diagnosis of infection was made from biopsies. Twenty-five hips were imaged with 111In-oxine-labelled cells; 20 hips and 11 knees were imaged with 99mTc-hexamethylpropylene-amineoxime-labelled cells. Of these, 13 hips and five knees proved to be infected. The scans taken together had an accuracy of 82%, a sensitivity of 44% and a specificity of 100%. Indium scans gave 37% sensitivity, 99mTc labelling 50% sensitivity. Infected arthroplasties with positive scans had presented significantly earlier than those with negative scans, the time after the original insertion being 1.1 years for the true-positive scans and 6.1 years for the false-negative scans. The value of labelled white-cell scans in the detection of infection in failed joint replacements is dependent on the activity of the infection. There is reduced sensitivity to the more insidious infections which affect arthroplasties and aspiration under controlled conditions remains an important investigation.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 1 | Pages 25 - 27
1 Jan 1993
Grigoris P Roberts P McMinn D Villar R

We describe a simple, retroperitoneal approach for the removal of acetabular components that have migrated into the pelvis. The dense fibrous tissue layer which surrounds the implant protects the iliac vessels during removal of the cup by this method.


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 3 | Pages 472 - 473
1 May 1992
Grigoris P Treacy R McMinn D


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 2 | Pages 265 - 269
1 Mar 1992
Roberts P Walters A McMinn D

In 78 consecutive patients, we performed fine-needle aspiration of the hip before revision surgery. At the revision operation biopsies were taken from the capsule and joint tissues. The aspirate and the homogenised soft-tissue specimens were cultured using a radiometric technique. Fifteen hips proved to be infected and 63 sterile on culture of the tissues taken during revision surgery; 94% of the aspiration results were correct. The sensitivity of the technique was 87% and the specificity 95%. Fine-needle aspiration combined with a radiometric culture technique is a simple and reliable method of diagnosing infection of hip prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 6 | Pages 899 - 901
1 Nov 1991
McMinn D Roberts P Forward G

A new approach to the hip is described that allows better exposure for difficult revision surgery without the need for trochanteric osteotomy. A V-shaped myofascial flap consisting of the proximal part of vastus lateralis and its fascia with gluteus medius and minimus, is reflected off the greater trochanter and proximal femur giving access to the entire capsule of the hip. If the leg has been significantly lengthened by the revision procedure, abductor tension can be adjusted at closure by altering the reattachment of the flap, analogous to a V-Y-plasty.