header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

LARGE HEAD TOTAL HIP REPLACEMENT- LESSONS LEARNED

The International Society for Technology in Arthroplasty (ISTA), 28th Annual Congress, 2015. PART 3.



Abstract

The use of modular systems adds versatility to the implant system, better restoration of hip biomechanics and lower inventory to the hospital. There have been reports of high metal ions, ARMD reactions and high implant failure rates due to potential problems from taper failures. These are more common in metal-on-metal hip replacements, but are being also reported in other bearings.

Between 2001 and 2010, we performed 383 consecutive metal-on-metal (MoM) THRs through a posterior approach, using a BHR cup and Birmingham modular head with one of three different stems, all with 12/14 tapers. The earliest 104 hips employed a cemented MS30 stem (Zimmer GmbH, Winterthur, Switzerland). Subsequent 256 were Synergy and then 23 Anthology (both uncemented and both Smith and Nephew Orthopaedics, Memphis TN USA). There was no significant difference in the average age at surgery (65.4 years cemented vs 65.6 uncemented, p = 0.69), gender ratio (1.68 vs 1.89, p = 0.64), or bearing diameter (46.7 vs 46.8, p = 0.31). The earlier 203 Synergy stems were monoblock heads, while the remaining uncemented stems included a tapered sleeve in addition.

There were 3 deep infections and 11 debris-related failures (overall revision rate 4.9%). The revision rate from aseptic failures (ALTR, effusion, osteolysis or component loosening) is 2.87%. Kaplan-Meier analysis of the entire cohort showed a 10-year implant survival of 96.8% with revision for any reason as the end-point. Cemented stems had a 100% survival at 10 years and 98.6% at 12 years. The uncemented stems had a 93.8% survival at 10 years. Within the uncemented group, the monoblocks had a 5 and 10-year survival of 99.0% and 96.4% respectively while the sleeved had 98.7% (5 years) and 96.3% (7 years) and 82.5% at 8 years. Retreival analysis showed clear evidence of taper failure.

Our experience suggests taper failure leading to ALTRs and its sequelae. Others have reported ALTR type reactions in metal on polyethylene and ceramic on polyethylene bearing types as well in bearing diameters ranging from 28mm to 40mm. There is a need to improve taper design especially for use with large heads, and in high demand patients.


*Email: