header advert
Results 1 - 32 of 32
Results per page:
Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 79 - 79
1 Dec 2022
Langohr GD Mahaffy M Athwal G Johnson JA
Full Access

Patients receiving reverse total shoulder arthroplasty (RTSA) often have osseous erosions because of glenohumeral arthritis, leading to increased surgical complexity. Glenoid implant fixation is a primary predictor of the success of RTSA and affects micromotion at the bone-implant interface. Augmented implants which incorporate specific geometry to address superior erosion are currently available, but the clinical outcomes of these implants are still considered short-term. The objective of this study was to investigate micromotion at the glenoid-baseplate interface for a standard, 3 mm and 6 mm lateralized baseplates, half-wedge, and full-wedge baseplates. It was hypothesized that the mechanism of load distribution from the baseplate to the glenoid will differ between implants, and these varying mechanisms will affect overall baseplate micromotion.

Clinical CT scans of seven shoulders (mean age 69 years, 10°-19° glenoid inclinations) that were classified as having E2-type glenoid erosions were used to generate 3D scapula models using MIMICS image processing software (Materialise, Belgium) with a 0.75 mm mesh size. Each scapula was then repeatedly virtually reconstructed with the five implant types (standard,3mm,6mm lateralized, and half/full wedge; Fig.1) positioned in neutral version and inclination with full backside contact. The reconstructed scapulae were then imported into ABAQUS (SIMULIA, U.S.) finite element software and loads were applied simulating 15°,30°,45°,60°,75°, and 90° of abduction based on published instrumented in-vivo implant data. The micromotion normal and tangential to the bone surface, and effective load transfer area were recorded for each implant and abduction angle. A repeated measures ANOVA was used to perform statistical analysis.

Maximum normal micromotion was found to be significantly less when using the standard baseplate (5±4 μm), as opposed to the full-wedge (16±7 μm, p=0.004), 3 mm lateralized (10±6 μm, p=0.017), and 6 mm lateralized (16±8 μm, p=0.007) baseplates (Fig.2). The half-wedge baseplate (11±7 μm) also produced significantly less micromotion than the full-wedge (p=0.003), and the 3 mm lateralized produced less micromotion than the full wedge (p=0.026) and 6 mm lateralized (p=0.003). Similarly, maximum tangential micromotion was found to be significantly less when using the standard baseplate (7±4 μm), as opposed to the half-wedge (12±5 μm, p=0.014), 3 mm lateralized (10±5 μm, p=0.003), and 6 mm lateralized (13±6 μm, p=0.003) baseplates (Fig.2). The full wedge (11±3 μm), half-wedge, and 3 mm lateralized baseplate also produced significantly less micromotion than the 6 mm lateralized (p=0.027, p=012, p=0.02, respectively). Both normal and tangential micromotion were highest at the 30° and 45° abduction angles (Fig.2). The effective load transfer area (ELTA) was lowest for the full wedge, followed by the half wedge, 6mm, 3mm, and standard baseplates (Fig.3) and increased with abduction angle.

Glenoid baseplates with reduced lateralization and flat backside geometries resulted in the best outcomes with regards to normal and tangential micromotion. However, these types of implants are not always feasible due to the required amount of bone removal, and medialization of the bone-implant interface. Future work should study the acceptable levels of bone removal for patients with E-type glenoid erosion and the corresponding best implant selections for such cases.

For any figures or tables, please contact the authors directly.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 85 - 85
1 Dec 2022
Fleet C McNeil D Trenholm JAI Johnson JA Athwal G
Full Access

Massive irreparable rotator cuff tears often lead to superior migration of the humeral head, which can markedly impair glenohumeral kinematics and function. Although treatments currently exist for treating such pathology, no clear choice exists for the middle-aged patient demographic. Therefore, a metallic subacromial implant was developed for the purpose of restoring normal glenohumeral kinematics and function. The objective of this study was to determine this implant's ability in restoring normal humeral head position. It was hypothesized that (1) the implant would restore near normal humeral head position and (2) the implant shape could be optimized to improve restoration of the normal humeral head position.

A titanium implant was designed and 3D printed. It consisted of four design variables that varied in both implant thickness (5mm and 8mm) and curvature of the humeral articulating surface (high constraint and low constraint. To assess these different designs, these implants were sequentially assessed in a cadaver-based biomechanical testing protocol. Eight cadaver specimens (64 ± 13 years old) were loaded at 0, 30, and 60 degrees of glenohumeral abduction using a previously developed shoulder simulator. An 80N load was equally distributed across all three deltoid heads while a 10N load was applied to each rotator cuff muscle. Testing states included a fully intact rotator cuff state, a posterosuperior massive rotator cuff tear state (cuff deficient state), and the four implant designs. An optical tracking system (Northern Digital, Ontario, Canada) was used to record the translation of the humeral head relative to the glenoid in both superior-inferior and anterior-posterior directions.

Superior-Inferior Translation

The creation of a posterosuperior massive rotator cuff tear resulted in significant superior translation of the humeral head relative to the intact cuff state (P=0.016). No significant differences were observed between each implant design and the intact cuff state as all implants decreased the superior migration of the humeral head that was observed in the cuff deficient state. On average, the 5mm low and high constraint implant models were most effective at restoring normal humeral head position to that of the intact cuff state (-1.3 ± 2.0mm, P=0.223; and −1.5 ± 2.3mm, P=0.928 respectively).

Anterior-Posterior Translation

No significant differences were observed across all test states for anterior-posterior translation of the humeral head. The cuff deficient on average resulted in posterior translation of the humeral head, however, this was not statistically significant (P=0.128). Both low and high constraint implant designs were found to be most effective at restoring humeral head position to that of the intact cuff state, on average resulting in a small anterior offset (5mm high constraint: 2.0 ± 4.7mm, P=1.000; 8mm high constraint: 1.6 ± 4.9mm, P=1.000).

The 5mm high constraint implant was most effective in restoring normal humeral head position in both the superior-inferior and anterior-posterior directions. The results from this study suggest the implant may be an effective treatment for restoring normal glenohumeral kinematics and function in patients with massive irreparable rotator cuff tears. Future studies are needed to address the mechanical efficiency related to arm abduction which is a significant issue related to patient outcomes.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_13 | Pages 37 - 37
1 Dec 2022
Fleet C de Casson FB Urvoy M Chaoui J Johnson JA Athwal G
Full Access

Knowledge of the premorbid glenoid shape and the morphological changes the bone undergoes in patients with glenohumeral arthritis can improve surgical outcomes in total and reverse shoulder arthroplasty. Several studies have previously used scapular statistical shape models (SSMs) to predict premorbid glenoid shape and evaluate glenoid erosion properties. However, current literature suggests no studies have used scapular SSMs to examine the changes in glenoid surface area in patients with glenohumeral arthritis. Therefore, the purpose of this study was to compare the glenoid articular surface area between pathologic glenoid cavities from patients with glenohumeral arthritis and their predicted premorbid shape using a scapular SSM. Furthermore, this study compared pathologic glenoid surface area with that from virtually eroded glenoid models created without influence from internal bone remodelling activity and osteophyte formation. It was hypothesized that the pathologic glenoid cavities would exhibit the greatest glenoid surface area despite the eroded nature of the glenoid and the medialization, which in a vault shape, should logically result in less surface area.

Computer tomography (CT) scans from 20 patients exhibiting type A2 glenoid erosion according to the Walch classification [Walch et al., 1999] were obtained. A scapular SSM was used to predict the premorbid glenoid shape for each scapula. The scapula and humerus from each patient were automatically segmented and exported as 3D object files along with the scapular SSM from a pre-operative planning software. Each scapula and a copy of its corresponding SSM were aligned using the coracoid, lateral edge of the acromion, inferior glenoid tubercule, scapular notch, and the trigonum spinae. Points were then digitized on both the pathologic humeral and glenoid surfaces and were used in an iterative closest point (ICP) algorithm in MATLAB (MathWorks, Natick, MA, USA) to align the humerus with the glenoid surface. A Boolean subtraction was then performed between the scapular SSM and the humerus to create a virtual erosion in the scapular SSM that matched the erosion orientation of the pathologic glenoid. This led to the development of three distinct glenoid models for each patient: premorbid, pathologic, and virtually eroded (Fig. 1). The glenoid surface area from each model was then determined using 3-Matic (Materialise, Leuven, Belgium).

Figure 1. (A) Premorbid glenoid model, (B) pathologic glenoid model, and (C) virtually eroded glenoid model.

The average glenoid surface area for the pathologic scapular models was 70% greater compared to the premorbid glenoid models (P < 0 .001). Furthermore, the surface area of the virtual glenoid erosions was 6.4% lower on average compared to the premorbid glenoid surface area (P=0.361).

The larger surface area values observed in the pathologic glenoid cavities suggests that sufficient bone remodelling exists at the periphery of the glenoid bone in patients exhibiting A2 type glenohumeral arthritis. This is further supported by the large difference in glenoid surface area between the pathologic and virtually eroded glenoid cavities as the virtually eroded models only considered humeral anatomy when creating the erosion.

For any figures or tables, please contact the authors directly.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_6 | Pages 43 - 43
1 Jul 2020
Berkmortel C Johnson JA Langohr GD King GJ DeDecker S
Full Access

Hemiarthroplasty is a common procedure that is an attractive alternative to total arthroplasty because it conserves natural tissue, allows for quicker recovery, and has a lower cost. One significant issue with hemiarthroplasties is that they lead to accelerated wear of the opposing native cartilage, likely due to the high stiffness of the implant. The purpose of this study was to investigate the range of currently available biomaterials for hemiarthroplasty applications. We employed a finite-element (FE) model of a radial head implant against the native capitellum as our joint model.

The FE model was developed in ABAQUS v6.14 (Dassault Systèmes Simulia Corp., Providence, RI, USA). A solid axisymmetric concave implant with seven different materials and the native radial head were evaluated, six modelled as elastic materials with different Young's moduli (E) and Poisson's Ratios (ν), and one modelled as a Mooney-Rivlin hyperelastic material. The materials investigated were CoCr (E=230 GPa, ν = 0.3), PEEK (E=3.7 GPa, ν = 0.36), HDPE (E=2.7 GPa, ν = 0.42), UHMWPE (E=0.69 GPa, ν = 0.49), Bionate 75D (E=0.288 GPa, ν = 0.39), Bionate 55D (E=0.039 GPa, ν = 0.45), and Bionate 80A (modelled as a Mooney-Rivlin hyperelastic material). A load of 100 N was applied to the radius through the center of rotation representing a typical load through the radius. The variable of interest was articular contact stress on the capitellum.

The CoCr implant had a maximum contact stress over 114% higher than the native radial head. By changing the material to lower the stiffness of the implant, the maximum contact stress was 24%, 70%, 105%, 111%, 113%, and 113% higher than the native radial head for Bionate 80A, Bionate 55D, Bionate 75D, UHMWPE, HDPE, and PEEK respectively.

This work shows that lowering implant stiffness can reduce the contact stress on cartilage in hemiarthroplasty implants. By changing the material below a Young's modulus of ∼100 MPa elevated stresses on the capitellum can be markedly reduced and hence potentially reduce or prevent degenerative changes of the native articulating cartilage. Low stiffness implant materials are not a novel concept, but to date there have been few that investigate materials (such as Bionate) as a potential load bearing material for implant applications. Further work is required to assess the efficacy of these materials for articular bearing applications.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1383 - 1387
1 Oct 2013
Lanting BA Ferreira LM Johnson JA Athwal GS King GJW

We measured the tension in the interosseous membrane in six cadaveric forearms using an in vitro forearm testing system with the native radial head, after excision of the radial head and after metallic radial head replacement. The tension almost doubled after excision of the radial head during simulated rotation of the forearm (p = 0.007). There was no significant difference in tension in the interosseous membrane between the native and radial head replacement states (p = 0.09). Maximal tension occurred in neutral rotation with both the native and the replaced radial head, but in pronation if the radial head was excised. Under an increasing axial load and with the forearm in a fixed position, the rate of increase in tension in the interosseous membrane was greater when the radial head was excised than for the native radial head or replacement states (p = 0.02). As there was no difference in tension between the native and radial head replacement states, a radial head replacement should provide a normal healing environment for the interosseous membrane after injury or following its reconstruction. Load sharing between the radius and ulna becomes normal after radial head Replacement. As excision of the radial head significantly increased the tension in the interosseous membrane it may potentially lead to its attritional failure over time.

Cite this article: Bone Joint J 2013;95-B:1383–7.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 13 - 13
1 Sep 2012
Glennie RA Giles JW Athwal GS Johnson JA Faber KJ
Full Access

Purpose

Glenoid component loosening is a common reason for failed total shoulder arthroplasty. Multiple factors have been suggested as causes for component loosening that may be related to cement technique. The purpose of the study was to compare the load transfer across a polyethylene glenoid bone construct with two different cementing techniques.

Method

Eight cadaveric specimens underwent polyethylene glenoid component implantation. Four had cement around the pegs only (CPEG) and four had cement across the entire back (CBACK) of the implant including around the pegs. Step loading was performed with a pneumatic actuator and a non-conforming humeral head construct capable of applying loads at various angles. Strain gauges were placed at the superior and inferior poles of the glenoid and position trackers were applied to the superior and inferior aspects polyethylene component. Micro CT data were obtained before and after the loading protocol.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 85 - 85
1 Sep 2012
Alolabi B Gray A Ferreira LM Johnson JA Athwal GS King GJ
Full Access

Purpose

The coronoid and collateral ligaments are key elbow stabilizers. When repair of comminuted coronoid fractures is not possible, prosthetic replacement may restore elbow stability. A coronoid prosthesis has been designed with an extended tip in an effort to augment elbow stability in the setting of residual collateral ligament insufficiency. The purpose of this biomechanical study, therefore, was to compare an anatomic coronoid replacement with an extended tip implant both with and without ligament insufficiency.

Method

Two coronoid prostheses were designed and developed based on CT-derived images adjusted for cartilage thickness: an anatomical implant and an extended-tip implant. Passive elbow extension was performed in 7 cadaveric arms in the varus and valgus positions. Varus-valgus laxity of the ulna relative to the humerus was quantified with a tracking system with an intact coronoid, a 40% coronoid deficiency, an anatomical prosthesis, and an extended prosthesis, with the collateral ligaments sectioned and repaired.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 150 - 150
1 Sep 2012
Giles JW Elkinson I Boons HW Ferreira LM Litchfield R Johnson JA Athwal GS
Full Access

Purpose

The management of moderate to large engaging Hill-Sachs lesions is controversial and surgical options include remplissage, allograft reconstruction, and partial resurfacing arthroplasty. Few in-vitro studies have quantified their biomechanical characteristics and none have made direct comparisons. The purpose of this study was to compare joint stability and range of motion (ROM) among these procedures using an in-vitro shoulder simulator. It was hypothesized that all procedures would prevent defect engagement, but allograft and partial resurfacing would most accurately restore intact biomechanics; while remplissage would provide the greatest stabilization, possibly at the expense of motion.

Method

Eight cadaveric shoulders were tested on an active in-vitro shoulder simulator. Each specimen underwent testing in 11 conditions: intact, Bankart lesion, Bankart repair, and two unrepaired Hill-Sachs lesions (30% & 45%) which were then treated with each of the three techniques. Anterior joint stability, ROM in extension and internal-external rotation, and glenohumeral engagement were assessed. Stability was quantified as resistance, in N/mm, to an anteriorly applied load of 70N.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 155 - 155
1 Sep 2012
Elkinson I Giles JW Faber KJ Boons HW Ferreira LM Johnson JA Athwal GS
Full Access

Purpose

The remplissage procedure may be performed as an adjunct to Bankart repair to address an engaging Hill-Sachs defect. Clinically, it has been reported that the remplissage procedure improves joint stability but that it may also restrict shoulder range of motion. The purpose of this biomechanical study was to examine the effects of the remplissage procedure on shoulder motion and stability. We hypothesized that the remplissage procedure would improve stability and prevent engagement but may have a deleterious effect on motion.

Method

Eight cadaveric forequarters were mounted on a custom biomechanical testing apparatus which applied simulated loads independently to the rotator cuff muscles and to the anterior, middle and posterior deltoid. The testing conditions included: intact shoulder, Bankart defect, Bankart repair, 2 Hill-Sachs defects (15%, 30%) with and without remplissage. Joint range of motion and translation were recorded with an optical tracking system. Outcomes measured were internal-external rotation range of motion in adduction and 90 combined abduction, extension range of motion and stability, quantified in terms of joint stiffness and engagement, in abduction.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 18 - 18
1 Sep 2012
Elkinson I Giles JW Faber KJ Boons HW Ferreira LM Johnson JA Athwal GS
Full Access

Purpose

The remplissage technique of insetting the infraspinatus tendon and posterior joint capsule into an engaging Hill-Sachs lesion has gained in popularity. However, a standardized technique for suture anchor and suture placement has not been defined for this novel procedure. The purpose of this biomechanical study was to compare three remplissage techniques by evaluating their effects on joint stiffness and motion.

Method

Cadaveric forequarters (n=7) were mounted on a custom active biomechanical shoulder simulator. Three randomly ordered techniques were conducted: T1- anchors in the valley of the defect, T2- anchors in the rim of the humeral head; T3- anchors in the valley with medial suture placement. The testing conditions included: intact, Bankart, Bankart repair, and 15% & 30% HS lesions with repairs (T1, T2, T3). Outcome measures including internal-external range of motion and stability were recorded. Stability was quantified in terms of glenohumeral joint stiffness against an externally applied anterior force of 70N.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 82 - 82
1 Sep 2012
Gray A Alolabi B Ferreira LM Athwal GS King GJ Johnson JA
Full Access

Purpose

The coronoid process is an integral component for elbow stability. In the setting of a comminuted coronoid fracture, where repair is not possible, a prosthetic device may be beneficial in restoring elbow stability. The hypothesis of this in-vitro biomechanical study was that an anatomic coronoid prosthesis would restore stability to the coronoid deficient elbow.

Method

A metal coronoid prosthesis was designed and developed based on CT-derived images adjusted for cartilage thickness. The kinematics and stability of eight fresh-frozen male cadaveric arms (mean age 77.4 years, range 69–92 years) were quantified in the intact state; after collateral ligament sectioning and repair (control state); after a simulated 40% transverse coronoid fracture; and after implantation of a coronoid prosthesis. Elbow flexion was simulated passively with the arm oriented in the varus position and the forearm in pronation. Varus-valgus angulation (VV) and internal-external rotation (IE) of the ulna relative to the humerus were quantified with an electromagnetic tracking system (Flock of Birds, Ascension Technologies, Burlington, VT, static accuracy: 1.8mm position, 0.5 orientation).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 16 - 16
1 Sep 2012
Sabo MT Shannon H Ng J Ferreira LM Johnson JA King GJ
Full Access

Purpose

Capitellum hemiarthroplasty is an emerging concept. The current metallic capitellar implants have spherical surface shapes, but the native capitellum is not spherical. This study evaluated the effect of capitellar implant shape on the contact mechanics of the radiocapitellar joint when articulating with the native radial head.

Method

Eight paired radii and humeri were potted in a custom jig. Articular casts were made with medium-viscosity resin while 85 N of axial load was applied to the reduced radiocapitellar joint at 0, 45, and 90 of elbow flexion, and at neutral, 50 pronation and 50 supination at each flexion angle. The native radiocapitellar articulation was compared to capitellar hemiarthroplasties of two surface designs (anatomical and spherical). Contact area and shape (circularity) were determined. Circularity was defined as the ratio of the minor axis and major axis of the shape.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 554 - 554
1 Nov 2011
Giles JW Glennie A Ferreira LM Athwal G Faber KJ Johnson JA
Full Access

Purpose: Loosening of glenoid components in total shoulder arthroplasty is a common clinical problem which can necessitate revision surgery. The mechanism of loosening is poorly understood and may relate to implant design, component fixation techniques, and interfacial tensile stresses. We are unaware of any studies that have examined the fundamental aspects of load transfer to bone for various joint loading configurations. Hence, the objective of this study was to investigate the effect of joint loading on bone strain adjacent to a poly-ethylene glenoid implant.

Method: Five specimens (4 males; avg age: 59.5 yrs) implanted with a cemented, all polyethylene component (Anatomical Shoulder; Zimmer) were tested using an apparatus capable of producing loading vectors with various angles, magnitudes and directions. Each specimen was tested using a ramp load of 0–150 N (at 10N/sec) in two directions (superior and inferior) and with six angles of load application. A uniaxial strain gauge was placed in each of the four quadrants of the glenoid, approximately 1 mm medial to the glenoid rim. The primary axis of each strain gauge was oriented medio-laterally to record bone strains. The humeral head was simulated by a custom steel ball with a radius of curvature consistent with a nonconforming humeral prosthesis.

Results: The relationship between strain and applied force was not linear (superior quadrant at 40o: linear fit R2=0.96; quadratic fit R2=0.999; p< 0.0005), and was dependent on the loading angle. During pure compressive loading, tension was observed in the superior and inferior quadrants of the glenoid; while less consistent results in the anterior and posterior quadrants revealed variable tension and compression. Superior and inferior loading each caused increasing ipsilateral tension, occurring from 0–30o and 0–20o, respectively.

Conclusion: The current study is thought to be the first to directly measure load transfer at the implant-bone interface. We demonstrated load transfer nonlinearities between a surgically implanted glenoid component and the underlying bone in all locations and for a wide range of loading conditions. This has important implications towards the modeling of these constructs using finite element analyses. The results also illustrate tensile loading during compressive and small eccentricity loading cases. These results suggest a polyethylene flexure, causing the periphery of the glenoid implant to flex upwards placing the cement mantle and underlying bone in tension. Tensile loads that are linked to cement mantle fracture and implant loosening are produced under loading conditions associated with activities of daily living. This study has provided insight into the mechanisms of load transfer between a cemented polyethylene glenoid implant and the underlying bone. Reduction or elimination of these interfacial tensile stresses around the glenoid periphery should be considered when developing novel methods for component fixation.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 556 - 556
1 Nov 2011
King GJ Greeley GS Beaton BJ Ferreira LM Johnson JA
Full Access

Purpose: This in-vitro study examined the effect of simulated Colles fractures on load transmitted to the distal ulna, using an in-line load cell. Our hypothesis was distal radial fracture malposition will increase distal radial ulnar joint (DRUJ) load relative to the native position of the radius.

Method: Eight fresh frozen upper-extremities were mounted in a motion simulator which enabled active forearm rotation. An osteotomy was performed just proximal to the distal radioulnar joint, and a 3-degree of freedom modular appliance was implanted which simulated Colles type distal radial fracture deformities. This device allowed for accurate adjustment of dorsal angulation and translation (0, 10, 20 and 30 degrees dorsal angulation and 0, 5 and 10mm dorsal translation both isolated and in combination). A 6-DOF load cell was inserted in the distal ulna 1.5 cm proximal to the ulnar head to quantify DRUJ joint forces. Distal ulnar loading was measured following simulated distal radial deformities with both an intact and sectioned triangular fibrocartilage complex (TFCC).

Results: The maximum resultant transverse distal ulnar load occurred during active forearm pronation and supination. Increasing magnitudes of dorsal angulation and translation of the distal radius increased loading in the distal ulna. For pronation with the ligaments intact, the transverse resultant load for the non-fracture, native positioning was significantly lower (p< 0.05) than the majority of malpositioned cases except for the translations only (not combined with angulation). However, all fracture orientations for supination had an increased effect on the resultant loading (p< 0.05) when ligaments were intact. Greater forces were measured in the distal ulna when the TFCC intact relative to TFCC sectioning. Sectioning the TFCC eliminated the effect of fracture malposition for both pronation and supination. The range of maximum transverse force for intact pronation and supination was between 118& #61617;34N and 130& #61617;39N, respectively. Similarly, for sectioned pronation and supination, the maximum transverse forces were and 93& #61617;40N and 89& #61617;24N, respectively.

Conclusion: Malpositioning of distal radial fractures in dorsal translation and angulation was found to increase forces in the distal ulna, which may be an important source of residual pain following malunion of Colles fractures. Healing of the distal radius in an anatomic position resulted in the least forces. Sectioning the TFCC released the tethering effect of the radius on the ulna, decreasing DRUJ force. This is the first study of its kind to attempt to quantify the forces at the DRUJ as a result of Colles fractures, and these early findings provide important baseline information related to the biomechanics of the DRUJ.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 556 - 556
1 Nov 2011
Ng J Lalone EA McDonald CP Ferreira LM King GJ Johnson JA
Full Access

Purpose: The identification of anatomical landmarks is an important aspect of joint surgery, to ensure proper placement and alignment for implants and other reconstructive procedures. At the elbow, the center of the capitellum (derived via a digitization of the surface and subsequent sphere fitting) has been well established as a key landmark to identify the axis of rotation of the joint. For some cases, and in particular minimally invasive surgery, only small regions of the capitellum may be exposed which may lead to errors in determining the centre. The purpose of this study was to identify the optimal location of digitizations of the capitellum.

Method: Twenty-five fresh frozen cadaveric distal humeri (19 left, 6 right) were studied. Using an x-ray computed tomography scanner, volumetric images of each specimen were acquired and used to reconstruct a 3-dimensional digital model of the specimen using the Visualization Toolkit (VTK). A sphere-fit algorithm was used to determine the centre of the spherical capitellum based on manually chosen (digitized) points across the 3D capitellar surface. The true geometric centre was located by digitizing points across the entire capitellar surface. Three sub-regions of the capitellum, commensurate with typical surgical approaches with minimal dissection, were then digitized. These were superior anterior lateral (SAL), inferior anterior lateral (IAL) and a combination of these two regions. These regions were compared to the true center using a 1-way Repeated Measures ANOVA with significance set to p = 0.05.

Results: Digitizations of only SAL and IAL sub-regions resulted in the largest differences relative to the true centre: SAL = 3.9±3.4 mm, IAL = 4.2±3.4 mm, (p < 0.0005). There was no difference between SAL and IAL (p = 1.0). Digitization of the combined SAL + IAL regions, while significantly different from the entire capitellum, resulted in the smallest mean difference of 0.87±0.84 mm.

Conclusion: These data show that the region of digitization affects the accuracy of predicting the capitellum centre. In a previous study by our group, we showed that an accurate determination of the centre of a sphere can be achieved with a small surface area of digitization. In the current study, the large errors that occurred when a small surface was digitized (i.e. SAL and IAL alone), are in all likelihood, due the non-spherical nature of the capitellum. In summary, while the most precise method in locating the true centre is to digitize the entire capitellar surface where possible, an alternative approach is to digitize both the superior and inferior anterior lateral regions.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 551 - 552
1 Nov 2011
Ferreira LM Bell TH Johnson JA King GJ
Full Access

Purpose: Most displaced olecranon fractures can be treated with ORIF. However with severe comminution or bone loss, excision of the fragments and repair of the triceps to the ulna is recommended. The triceps can be reattached to either the anterior or posterior aspect of the ulna. The purpose of this in-vitro study was to determine the effect of triceps repair technique on elbow laxity and extension strength in the setting of olecranon deficiency.

Method: Eight unpreserved cadaveric arms were used (age 75 ± 11 years). Surface models were generated from CT images and sequential olecranon resections in 25% increments were performed using real-time navigation. Muscle tendons (biceps, brachialis, brachioradialis and triceps) were sutured to actuators of an elbow motion simulator, which produced active extension. A tracking system recorded kinematics in the varus and valgus positions. A triceps advancement was performed using either an anterior or posterior repair to the remaining olecranon in random order. Triceps extension strength was measured in the dependent position with the elbow flexed 90° using a force transducer located at the distal ulnar styloid, while triceps tension was increased from 25–200 N. Outcome variables included maximum varus-valgus elbow laxity and triceps extension strength. Two-way repeated measures ANOVAs were performed for laxity comparing resection level and repair method. Three-way repeated measures ANOVAs were performed for triceps extension strength comparing triceps tension, resection level and repair method. Significance was set at p < 0.05.

Results: Progressive olecranon resection increased elbow laxity (p < 0.001). Although the posterior repair produced slightly greater laxity for all but the 50% resection, this difference was not significant (p = 0.2). The posterior repair provided greater extension strength than the anterior repair at all applied triceps tensions and for all olecranon resections (p = 0.01). The initial 0% resection reduced extension strength for both repairs (p < 0.01), however, there was no effect of progressive olecranon resections (p = 0.09).

Conclusion: There was no significant difference in laxity between the anterior and posterior repairs. Thus even for large olecranon resections, the technique of triceps repair does not have significant influence on joint stability. Extension strength was not reduced by progressive olecranon resections, perhaps due to wrapping of the triceps tendon around the trochlea putting it in-line with the ulna and giving it a constant moment arm. Triceps extension strength was higher for the posterior repair. This is likely due to the greater distance and hence moment arm of the posterior repair to the joint rotation center. Conversely, the anterior repair brings the triceps insertion closer to the joint center, reducing the moment arm. Since there was no significant difference in laxity between the repairs, the authors favour the posterior repair due to its significantly higher triceps extension strength.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 552 - 553
1 Nov 2011
Ferreira LM Fay KE Lalone EA Johnson JA King GJ
Full Access

Purpose: Techniques to quantify soft-tissue forces in the upper extremity are not well described. Consequently, ligament forces of the elbow joint have not been reported. Knowledge of the magnitudes of tension of the primary valgus stabilizer, the anterior bundle of the medial collateral ligament (AMCL), would allow for an improved understanding of the load bourne by the ligament. The purpose of this in vitro study was to quantify the magnitude of tension in the native AMCL throughout flexion with the arm in the valgus orientation. We hypothesized that tension in the AMCL would increase with flexion.

Method: Five fresh-frozen cadaveric upper extremities (mean age 72 ± 10 years) were tested. To produce active muscle loading in a motion simulator, cables were affixed to the distal tendons of the brachialis, biceps brachii, triceps brachii, and brachioradialis and attached to actuators. The wrist was fixed in neutral flexion/extension and the forearm in neutral rotation. The arm was orientated in the valgus gravity-loaded position. A custom designed ligament load transducer was inserted into the AMCL. Active simulated flexion was achieved via computer-controlled actuation while passive elbow flexion was achieved by an investigator manually guiding the arm through flexion. Motion of the ulna relative to the humerus was measured using a tracking device.

Results: Both the active and passive motion pathways showed an increase in AMCL tension with increasing angles of elbow flexion (p < 0.05). There was no difference in AMCL tension levels between active and passive elbow flexion (p = 0.20). The mean maximum tension achieved was 97±33N and 94±40 N for active and passive testing respectively.

Conclusion: AMCL tension levels were observed to increase with elbow flexion, indicating that other structures (such as the joint capsule and the shape of the articulation) are likely more responsible for joint stability near full extension, and that the AMCL is recruited at increased angles of elbow flexion. With respect to load magnitudes, Regan et al. found the maximum load to failure of the AMCL was 261 N, while Armstrong et al. reported a failure load of 143 N in cyclic testing. The maximum AMCL tension level observed in this study was 160 N. Failure of the AMCL was not observed, which may be due to differences in specimen size, age, or the method of load application. In summary, this in vitro cadaveric study has provided a new understanding of the magnitudes of AMCL tension through the arc of elbow flexion, and this has important implications with respect to the desired target strength of repair and reconstruction techniques. These findings will also assist in the development and validation of computational models of the elbow.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 269 - 270
1 Jul 2011
Bell TH King GJ Johnson JA Ferreira LM McDonald CP
Full Access

Purpose: The purpose of this study was to determine the effect of serial olecranon resections on elbow stability.

Method: Eight fresh, previously frozen cadaveric arms underwent CT scanning. The specimens were mounted in an in-vitro motion simulator, and kinematic data was obtained using an electromagnetic tracking system. Simulated active and passive flexion was produced with servo-motors and pneumatic pistons attached to specific muscles. Flexion was studied in the dependent, horizontal, varus, and valgus positions. Custom computer navigation software was utilized to guide serial resection of the olecranon in 12.5% increments. A triceps advancement repair was performed following each resection.

Results: Serial olecranon resections resulted in a significant increase in valgus-varus (V-V) laxity for both passive (p< 0.001) and active (p=0.04) flexion. For passive motion this increase reached statistical significance following the 12.5% resection. This corresponded to an increase in V-V laxity of 1.4 ± 0.1o and a total laxity of 7.5 ± 1.0o. For active flexion this increase reached significance following the 62.5% resection. This corresponded to an increase in V-V laxity of 5.6 ± 1.1o and a total laxity of 11.2 ± 1.5. There was no significant effect of sequential olecranon excision on elbow kinematics or stability with the elbow in the vertical or horizontal positions. The elbows became grossly unstable after resection of greater than 75% of the olecranon.

Conclusion: A progressive increase in the varus-valgus laxity of the elbow was seen with sequential excision of the olecranon. Laxity of the elbow was increased with excision of 75% of the olecranon, likely due to the loss of the bony congruity and attachment site of the posterior band of the medial collateral ligament. Gross instability resulted when 87.5% or greater was removed, likely due to damage to the anterior band of the medial collateral ligament as it inserts on the sublime tubercle of the ulna. Rehabilitation of the elbow with the arm in the dependant position should be considered following excision of the olecranon; varus and valgus orientations should be avoided. The contribution of the olecranon to elbow stability may be even more important in patients with associated ligament injuries or fractures of the elbow.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 249 - 249
1 Jul 2011
Sabo M Fay K Ferreira L McDonald C Johnson JA King GJ
Full Access

Purpose: Osteochondritis dissecans (OCD) of the capitellum most commonly affects adolescent pitchers and gymnasts, and presents with pain and mechanical symptoms. Fragment excision is the most commonly employed surgical treatment; however, patients with larger lesions have been reported to have poorer outcomes. It’s not clear whether this is due to increased contact pressures on the surrounding articular surface, or if fragment excision causes instability of the elbow. The purpose of this study was to determine if fragment excision of simulated OCD lesions of the capitellum alters kinematics and stability of the elbow.

Method: Nine fresh-frozen cadaveric arms were mounted in an upper extremity joint motion simulator, with cables attaching the tendons of the major muscle tendons to motors and pneumatic actuators. Electromagnetic receivers attached to the radius and ulna enabled quantification of the kinematics of both bones with respect to the humerus. Three-dimensional CT scans were used to plan lesions of 12.5% (mean 0.8cm2), 25%, 37.5%, 50%, and 100% (mean 6.2cm2) of the capitellar surface, which were marked on the capitellum using navigation. Lesions were created by burring through cartilage and subchondral bone. The arms were subjected to active and passive flexion in both the vertical and valgus-loaded positions, and passive forearm rotation in the vertical position.

Results: No significant differences in varus-valgus or rotational ulnohumeral kinematics were found between any of the simulated OCD lesions and the elbows with an intact articulation with active and passive flexion, regardless of forearm rotation and the orientation of the arm (p> 0.7). Radiocapitellar kinematics were not significantly affected during passive forearm rotation with the arm in the vertical position (p=0.07–0.6).

Conclusion: In this in-vitro biomechanical study even large simulated OCD lesions of the capitellum did not alter the kinematics or laxity of the elbow at either the radiocapitellar or ulnohumeral joints. These data suggest that excision of capitellar fragments not amenable to fixation can be considered without altering elbow kinematics or decreasing stability. Further study is required to examine other factors, such as altered contact stresses on the remaining articulation, that are thought to contribute to poorer outcomes in patients with larger lesions.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 248 - 248
1 Jul 2011
McDonald CP Johnson JA Peters TM King GJ
Full Access

Purpose: This study evaluated the accuracy of humeral component alignment in total elbow arthroplasty. An image-based navigated approach was compared against a conventional non-navigated technique. We hypothesized that an image-based navigation system would improve humeral component positioning, with navigational errors less than or approaching 2.0mm and 2.0°.

Method: Eleven cadaveric distal humeri were imaged using a CT scanner, from which 3D surface models were reconstructed. Non-navigated humeral component implantation was based on a visual estimation of the flexion-extension (FE) axis on the medial and lateral aspects of the distal humerus, followed by standard instrumentation and positioning of a commercial prosthesis by an experienced surgeon. Positioning was based on the estimated FE axis and surgeon judgment. The stem length was reduced by 75% to evaluate the navigation system independent of implant design constraints. For navigated alignment, the implant was aligned with the FE axis of the CT surface model, which was registered to landmarks of the physical humerus using the iterative closest point algorithm. Navigated implant positioning was based on aligning a 3D computer model calibrated to the implant with a 3D model registered to the distal humerus. Each alignment technique was repeated for a bone loss scenario where distal landmarks were not available for FE axis identification.

Results: Implant alignment error was significantly lower using navigation (P< 0.001). Navigated implant alignment error was 1.2±0.3 mm in translation and 1.3±0.3° in rotation for the intact scenario, and 1.1±0.5 mm and 2.0±1.3° for the bone loss scenario. Non-navigated alignment error was 3.1±1.3 mm and 5.0±3.8° for the intact scenario, and 3.0±1.6 mm and 12.2±3.3° for the bone loss scenario. Without navigation, 5 implants were aligned outside 5° for intact bone while 9 were aligned outside 10° for the bone loss scenario.

Conclusion: Image-based navigation improved the accuracy of humeral component placement to less than 2.0 mm and 2.0°. Further, outliers in implant positioning were reduced using image-based navigation, particularly in the presence of bone loss. Implant malalignment may well increase the likelihood of early implant wear, instability and loosening. It is likely that improved implant positioning will lead to fewer implant related complications and greater prosthesis longevity.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 247 - 247
1 Jul 2011
Ferreira LM King GJ Johnson JA
Full Access

Results: Repeatability of creating motion-based JCS was less than 1 mm and 1° in all directions. The inter-specimen standard-deviations of position and orientation measurements were smaller for the motion-based than for the anatomy-based JCS in every direction and for every specimen (p< 0.006). The ulno-humeral varus angle and internal/external rotation kinematics of active flexion showed less inter-specimen variability when calculated using motion-based JCS (p< 0.05).


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 248 - 248
1 Jul 2011
McDonald CP Johnson JA Peters TM King GJ
Full Access

Purpose: While computer-assisted techniques can improve the alignment of the implant articulation with the native structure, stem abutment in the intramedullary canal may impede achievement of this alignment. In the current study, the effect of a fixed valgus (6 degree) stemmed humeral component on the alignment of navigated total elbow arthroplasty was investigated. Our hypothesis was that implantation of a humeral component with a reduced stem length would be more accurate than implantation of the humeral component with a standard length stem.

Method: Thirteen cadaveric distal humeri were imaged using a CT scanner, and a 3D surface model was reconstructed from each scan. Implantation was performed using two implant configurations. The first set was unmodified (Regular) while the second set was modified by reducing the length of the humeral stem to 25% of the original stem (Reduced). A surface model of the humeral component was aligned with the flexion-extension (FE) axis of the CT-based surface model, which was registered to the landmarks of the physical humerus using the iterative closest point algorithm. Navigated implant positioning was based on aligning a 3D computer model calibrated to the implant with a 3D model registered to the distal humerus.

Results: Implant alignment error was significantly lower for the Reduced implant, averaging 1.3±0.5 mm in translation and 1.2±0.4° in rotation, compared with 1.9±1.1 mm and 3.6±2.1° for the Regular implant. Abutment of the implant stem with the medullary canal of the humerus prevented optimal alignment of the Regular humeral component as only four of the 13 implantations were aligned to within 2.0° using navigation.

Conclusion: These results demonstrate that a humeral component with a fixed valgus angulation cannot be accurately positioned in a consistent fashion within the medullary canal of the distal humerus without sacrificing alignment of the FE axis due to stem abutment. Improved accuracy of implant placement can be achieved by introducing a family of humeral components, with three valgus angulations of 0°, 4° and 8°. Based on humeral morphology for these specimens, 12 of the 13 implants may be positioned to within 2° of the native FE axis using one of these 3 valgus angulations.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 269 - 269
1 Jul 2011
Sabo M Fay K Ferreira LM McDonald CP Johnson JA King GJ
Full Access

Purpose: Coronal shear fractures of the humerus include the Kocher-Lorenz fracture, an osteochondral fracture of the capitellar articular surface, the Hahn-Steinthal fracture, a substantial shear fragment, extension into the trochlea, and complete involvement of the capitellum and trochlea. If the fracture proves irreparable, it is not known what the impact of fragment excision would have on the biomechanics of the elbow. The purpose of this study was to examine the effect of the sequential loss of the capitellum and trochlea on the kinematics and stability of the elbow.

Method: Eight fresh-frozen cadaveric arms were mounted in an upper extremity joint testing system, with cables attaching the tendons of the major muscles to motors and pneumatic actuators. Electromagnetic receivers attached to the radius and ulna enabled quantification of the kinematics of both bones with respect to the humerus. The distal humeral articular surface was sequentially excised to replicate clinically relevant coronal shear fractures while leaving the collateral ligaments intact. Active flexion in both the vertical and valgus-loaded positions, and passive rotation in the vertical position was conducted for each excision.

Results: Excision of the capitellum had no effect on ulnohumeral stability or kinematics in both the vertical or valgus positions (p=1.0). Excision of the entire capitellum and trochlea led to significant valgus instability with the arm in the valgus position (p=0.01), while excision of the lateral trochlea led to increased valgus instability with pronated flexion in the valgus position (p=0.049). Progressive loss of the articular surface led to posterior, inferior, and medial displacement of the radial head with respect to the capitellum and increased external rotation of the ulna with respect to the humerus in the vertical position (p< 0.05).

Conclusion: Excision of the capitellum did not result in valgus or rotational instability, while excision of the trochlea resulted in multiplanar instability. The radial head displaced medially because it is constrained to the ulna by the annular ligament, and the ulna pivoted into valgus and external rotation on the residual trochlea and medial collateral ligament. In patients with coronal shear fractures, the trochlea must be reconstructed to prevent instability and the potential for secondary degenerative change.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 14 - 14
1 Mar 2010
McDonald CP King GJW Peters TM Johnson JA
Full Access

Purpose: The successful placement of elbow prostheses, external fixators and ligament reconstructions is dependent on the accurate identification of the elbow’s flexion-extension (FE) axis. In the case of periarticular bone loss, the FE axis must be visually estimated, as the necessary anatomical landmarks may not be available. Hence, referencing the uninjured elbow anatomy may prove beneficial in accurately defining this axis. However, this is contingent on the morphological features being similar between the two sides. Our objective was to compare distal humeral morphology between paired specimens. Our hypothesis was that anthropometric measurements from the distal humerus would be similar to the contralateral side.

Method: CT Images of 25 paired distal humeri were obtained. A right-to-left surface registration was then performed on each pair using the iterative closest point (icp) least-squares algorithm, thus placing each specimen in the same coordinate system.. Anthropometric characteristics measured (and compared between the left and right sides) included the angles of the FE and epicondylar axes in both the coronal and transverse planes, the anterior offset of the FE axis with respect to the humeral shaft axis, the length of the FE axis and the radius of curvature of the capitellum and trochlea.

Results: There was no statistically significant difference between the left and right humeri for the eight anthropometric characteristics studied (p > 0.05). The mean difference in magnitude for the FE axis angle was approximately 1.0° in both the coronal and transverse planes and the difference in magnitude for 80% of the paired specimens was less than 1.5°.

Conclusion: The anthropometric features of the distal humerus that are typically employed during elbow surgery are similar from side to side. Preoperative imaging of the contralateral normal elbow should be considered in patients with periarticular bone loss where referencing anatomical landmarks of the injured side is not possible. This information can be used as part of a preoperative plan to determine the ideal position of the implant, ligament reconstruction or external fixator during surgery. Contralateral imaging should be particularly useful when combined with computer-assisted elbow surgery.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 34 - 35
1 Mar 2010
Athwal GS Faber KJ Johnson JA Frank SG King GJW
Full Access

Purpose: Unrepairable fractures of the radial head are often treated with radial head arthroplasty. Insertion of a radial head prosthesis that is too thick, or overstuffed, is believed to be a common complication that may result in pain, arthrosis, capitellar wear and decreased elbow range of motion. The purpose of this study was to develop guidelines for determining the appropriate thickness of radial head implants. We hypothesized that

radiographic incongruity of the medial facet of the ulnohumeral joint and that

the macroscopic presence of a gap in the lateral facet of the ulnohumeral joint correlate with radial head overstuffing.

Method: Six human cadaveric upper extremities were used to evaluate the clinical and radiographic effects of overstuffing of a radial head arthroplasty. Each specimen received an anatomic radial head replacement and then underwent overstuffing with +2 mm, +4 mm, +6 mm and +8 mm lengths. Gross lateral ulnohumeral joint spaces were measured, and anteroposterior radiographs were taken of the elbow from which radiographic medial and lateral ulnohumeral joint spaces were measured.

Results: Intraoperative gapping of the lateral ulnohumeral facet was shown to be highly reliable for detecting radial head overstuffing, increasing from a mean of 0.0 mm at standard length to 1.0 mm with 2 mm overstuffing (p < 0.05). Radiographically, the congruity of the lateral ulnohumeral facet was significantly different with 2 mm of overstuffing as compared to the anatomic length (p < 0.05). The congruity of the medial ulnohumeral facet only became significantly different with +6 mm of overstuffing as compared to the anatomic length (p < 0.05).

Conclusion: Radiographic incongruity of the medial facet of the ulnohumeral joint was an unreliable indicator of radial head overstuffing. Radiographic gapping of the lateral ulnohumeral facet demonstrated sufficient sensitivity to diagnose radial head overstuffing when compare to the standard length implant radiographs. Visual gapping of the lateral ulnohumeral facet on the cadaver specimens reliably indicated radial head overstuffing and should be a useful anatomic feature to assess intraoperatively.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 14 - 14
1 Mar 2010
Brownhill JR Beaton BJ Ferreira LM Johnson JA King GJW
Full Access

Purpose: Aseptic loosening is one of the leading causes of failure in total elbow arthroplasty. It is logical to postulate that incorrect implant positioning and alignment may lead to excessive loading and wear which can induce the loosening cascade. However, the effect of implant malalignment on wear inducing loads in the elbow is not yet known. This in-vitro study determined the effect of anterior malpositioning, and varus-valgus (VV) and internal-external (IE) malrotations on humeral stem loading in total elbow arthroplasty.

Method: The humeral, ulnar, and radial components of a linked total elbow arthroplasty were optimally positioned using computer navigation in eight cadaveric elbows, mounted in a load/motion control elbow simulator (age 75yrs, range 42–93; 5 male). A modular, humeral component was employed to generate implant malpositioning errors of ±6° VV, ±8° IE, and 5mm anterior. The implant was instrumented with strain gauges to quantify VV and IE bending loads during elbow flexion with the forearm in supination. Load output was combined using a sum-of-squares technique. Passive flexion was performed with the arm in the varus and valgus orientations; passive and active flexion were performed with the arm in the vertical orientation.

Results: With the arm (humerus) in the vertical orientation, bending loads increased between 418Nmm and 1618Nmm for all malaligned implant positions (p< 0.05). Passive flexion (1354±859Nmm) produced higher resultant loads for the optimally positioned implant than active (819±891Nmm) flexion (p< 0.05). Although it varied during flexion, loading with the arm in varus (2928±1273Nmm) or valgus (2494±743Nmm) orientations resulted in up to a three-fold increase in loading when compared to the vertical orientation (p< 0.01).

Conclusion: These data demonstrate that humeral component malpositioning increases loading in the implant, however further studies are required to determine the long term effect on polyethylene wear and component loosening. Prosthesis designs that replicate the native flexion-extension axis and make use of sophisticated instrumentation or computer assistance to achieve precise positioning during implantation should lead to improved arthroplasty durability. Also, loading was higher with the arm in varus or valgus orientations, suggesting that patients should avoid activities post-operatively that require their elbow to be positioned in this way.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 14 - 14
1 Mar 2010
Brownhill JR Pollock JW Ferreira LM Johnson JA King GJW
Full Access

Purpose: Ligaments and osseous constraints are the only static stabilizers in a healthy elbow. Following arthroplasty, the use of semi-constrained, or linked, implants provides a potential third static stabilizer. However, this constraint may increase loading on the prosthesis, and hence accelerate polyethylene wear. The presence of competent collateral ligaments and the radial head would be expected to improve elbow stability and decrease loading on the ulnohumeral articulation. This in vitro study determined the effects of the collateral ligaments, radial head, and implant linkage on kinematics and wear-inducing loads in total elbow arthroplasty.

Method: Eight cadaveric upper extremities (age 73.5yrs; 5 male), were tested using an elbow motion simulator. Humeral, ulnar, and radial components of an elbow arthroplasty were positioned using a computer-assisted technique. Varus-valgus and internal-external bending loads were measured during flexion using an instrumented humeral component. A tracking receiver attached to the ulna recorded its position during active and passive flexion in the vertical orientation, and passive flexion in the varus and valgus orientations. Kinematics and loading were measured with and without implant linkage, with an intact, resected and replaced radial head, and before and after sectioning of the collateral ligaments.

Results: There were no differences in the bending loads with the arm in the vertical orientation regardless of the status of the ligaments, radial head or implant linkage (p> 0.2). Radial head excision produced an increase in valgus angulation of the ulna (6.7±6.4°) but did not influence bending loads in the vertical orientation (p< 0.05). Loading was lowest with the unlinked implant, and with ligaments and radial head intact, with the arm in the valgus (1065±466Nmm) (p< 0.01) and varus (1333±698Nmm) (p< 0.05) orientations.

Conclusion: Our results show that the radial head is an important valgus stabilizer for the prosthesis employed in this investigation. Linkage of the articulation increases implant loading during passive flexion with the arm in the varus and valgus orientations, which may increase implant wear. This suggests that, when using prostheses of this design, linkage of the articulation may be unnecessary if adequate bone stock and ligaments are available, whilst preserving or repairing the collateral ligaments and preserving or replacing the radial head.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_II | Pages 224 - 224
1 May 2009
McDonald C Beaton BJB Johnson JA King GJW Peters TM
Full Access

Accurate implant alignment with the flexion-extension axis of the elbow is likely critical for optimal function and durability following elbow replacement arthroplasty. Implant alignment can be optimised by imaging the contralateral normal elbow prior to surgery and transferring this information to the diseased elbow in the operating room through registration. Successful registration is dependent on the presence of unique anatomical landmarks. Bone loss can create a challenge for registration as key anatomical landmarks are absent, limiting the number of sampling areas. This study investigated the effect of intraoperative sampling area on registration accuracy. We hypothesised that a low registration error can be achieved by acquiring surface data from areas unlikely compromised due to injury and readily available to the surgeon during typical surgical exposures.

CT images of twenty cadaveric distal humeri were acquired. Surface data was acquired from nineteen anatomical landmarks of the distal humerus using a hand-held laser scanner (FastSCANTM, Polhemus). Registration to the CT image was performed for thirty-nine landmark combinations. Only six combinations are discussed for succinctness.

Combining data from the anterior articular surface and humeral shaft, the lowest registration error was achieved in translation (0.8±0.3 mm) and rotation (0.3±0.2°). However, using data from the posterior shaft and proximal medial supracondylar column, a registration error of 1.1±0.2 mm and 0.4±0.2° was achieved.

Based on the results of this study, a low registration error can be achieved by acquiring data from two areas that are located proximal to the articular surface (the proximal medial supracondylar column and posterior humeral shaft), readily available surgically, and unlikely compromised due to distal humeral fractures, non-unions or bone loss due to severe erosive arthritis. Registration error was similar to the reported resolution of the laser scanner. Overall, this study demonstrates the promise for a successful registration of the contralateral normal elbow to physical surface data of the diseased or injured elbow using only a small portion of undamaged bone structure.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_II | Pages 243 - 243
1 May 2009
Fraser GS Pichora JE Ferreira LM Brownhill JR Johnson JA King GJW
Full Access

This in-vitro study evaluated the influence of ligament tensioning and the effectiveness of lateral collateral ligament (LCL) repair using transosseous sutures on the initial kinematics and stability of the elbow.

Six fresh upper-extremities were mounted in a motion simulator with tracking system, which enabled both passive and simulated active elbow flexion. The intact elbow was tested then the LCL was sectioned from its humeral origin and repaired with a transosseous suture technique. Locking sutures were placed in the LCL and passed through a humeral bone tunnel entering at the centre of curvature of the capitellum with exit holes in the lateral epicondyle. An actuator pulled on the sutures to achieve 20, 40 and 60 N of LCL repair tension and the sutures were then secured. The dependent variable of this study was the motion pathways of the ulna relative to the humerus. The data were analyzed using a two-way, repeated-measures ANOVA with relevant post-hoc paired t-tests.

With the arm oriented in the horizontal position under varus gravity loading, the repairs tracked in greater valgus than the intact LCL regardless of the repair tension. The larger the initial repair tension, the more the elbows tracked in valgus. Initial tension of 60 N was statistically different than the intact LCL with the forearm in pronation (p=0.04). Both the 40 and 60 N initial tensions were statistically different than the intact LCL with the forearm in supination (p< 0.01).

Repair of the LCL using transosseous sutures effectively restores the varus stability of the elbow. The initial tension of LCL repairs affects the kinematics of the elbow, with a tendency to over-tighten the ligament and pull the elbow into valgus. These data suggest that acute repair of the LCL should be performed using a transosseous suture technique, and that a tension of 20N or perhaps less is sufficient to restore stability.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_II | Pages 243 - 243
1 May 2009
Brownhill JR Pollock J Ferreira LM Johnson JA King GJW
Full Access

A primary mode of failure for total elbow arthroplasty is osteolysis caused by wear debris. Loading of the polyethylene components by off-axis bearing loads is the likely cause of this debris. Load transfer at the elbow is affected by many factors, including the state of the radial head. New implant designs provide the option to use the intact, resected, or implant reconstructed radial head. However, the effect of the radial head state on stability and loading has not yet been investigated in these new implant designs.

We postulated that the presence of the native or prosthetic radial head would reduce the wear-inducing loading patterns experienced by the humeral component and improve joint stability compared to when the radial head is resected.

Seven cadaveric upper extremities, amputated at the mid humerus, were tested in a joint motion simulator equipped with an electromagnetic tracking system to quantify motion. Simulated active flexion was tested with the arm in the dependent position. Passive elbow flexion was conducted with the arm in the varus and valgus gravity-loaded orientations. After testing the intact elbow, the collateral ligaments were sectioned and a linked Latitude ulno-humeral joint replacement was performed (Tornier, Stafford, TX). The humeral component was instrumented with strain gauges for measuring varus-valgus bending and internal-external torsion. Ulno-humeral kinematics and humeral component loading were measured when the radial head was intact, resected, and following radial head arthroplasty.

An increase in varus-valgus laxity was noted following replacement of the ulno-humeral joint with the prosthesis (p< 0.05). There was no difference in joint laxity between the intact radial head, radial head excision or radial head arthroplasty (p> 0.05). Torsion moments increased, while bending loads decreased in the humeral component following radial head excision and were restored following radial head arthroplasty (p< 0.05).

No significant effect of radial head state on varus-valgus joint laxity was observed for the linked ulno-humeral prosthesis. In the absence of collateral ligaments, the observed post-operative increase in varus-valgus laxity can be attributed to the difference in laxity between the native joint and the articular components of the linked implant. Load transfer was altered by radial head excision, which may affect the magnitude of bearing wear and the incidence of aseptic loosening. Further studies are required to determine whether these changes in load transfer influence wear of the polyethylene components or implant loosening.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1256 - 1259
1 Sep 2008
Kedgley AE DeLude JA Drosdowech DS Johnson JA Bicknell RT

This study compared the effect of a computer-assisted and a traditional surgical technique on the kinematics of the glenohumeral joint during passive abduction after hemiarthroplasty of the shoulder for the treatment of fractures. We used seven pairs of fresh-frozen cadaver shoulders to create simulated four-part fractures of the proximal humerus, which were then reconstructed with hemiarthroplasty and reattachment of the tuberosities. The specimens were randomised, so that one from each pair was repaired using the computer-assisted technique, whereas a traditional hemiarthroplasty without navigation was performed in the contralateral shoulder. Kinematic data were obtained using an electromagnetic tracking device.

The traditional technique resulted in posterior and inferior translation of the humeral head. No statistical differences were observed before or after computer-assisted surgery.

Although it requires further improvement, the computer-assisted approach appears to allow glenohumeral kinematics to more closely replicate those of the native joint, potentially improving the function of the shoulder and extending the longevity of the prosthesis.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_III | Pages 445 - 445
1 Oct 2006
Deshpande S Mackenzie G Kedgley A Johnson JA Chess DG
Full Access

Optimal soft tissue tension maximises function after total knee arthroplasty (TKA). Excessive tension may lead to stiffness and or pain, while inadequate tension can lead to instability. Composite component thickness is a prime determinant of this soft tissue tension. The thickness provided by polyethylene inserts currently allows for a 2–3 mm incremental change. This study analyses the effect of incremental change in polyethyl-ene thickness on soft tissue tension.

Computer assisted (Stryker Knee Nav) TKA was performed on 8 cadaveric knee specimens (4 pairs). Kinematic data was collected through the navigation software. The soft tissue tension was analysed by measuring compartmental loads. A validated load cell instrumented tibial insert was used to measure medial and lateral compartmental loads independently. The effect of 1mm increments in polyethylene thickness on compartmental loads was evaluated.

We measured an increase in compartmental loads with increasing insert thickness. The peak loads in each compartment showed different behaviour reflecting varying tension in the medial and lateral sides. The peak loads generated showed a reduction after reaching a maximal level with further increase in insert thickness. With a one mm increase in insert thickness, 75 % of specimens showed greater than 200 % increase in the peak loads in the lateral compartment. Similarly the medial loads showed a greater than 100% increase. Individual specimens showed a high variability in loading patterns.

Our study highlights high variation of knee loads present between subjects. The compartmental loads vary as a function of insert thickness. The high sensitivity of compartmental loads with a 1mm increment is significant and has not been previously appreciated, especially intraoperatively. The currently available TKA inserts with 2–3 mm increments may make obtaining optimal soft tissue tension difficult. In addition to the current focus of obtaining accurate leg alignment, further computer aided techniques are required to address soft tissue tension.