Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVI | Pages 13 - 13
1 Apr 2012
Mangat N Langton D Joyce T Jameson S De Smet K Nargol A
Full Access

Bearing diameter and acetabular component orientation have been shown to be important variables effecting blood metal ion levels following hip resurfacing arthroplasty. So far no studies on bilateral hip resurfacings have taken into account these variables.

We examined the serum ion results of patients under the care of two experienced hip resurfacing surgeons who carry out ion analysis as part of routine post operative care. Surgeon 1: Patients were implanted exclusively with a “third generation” resurfacing device. Surgeon 2: Patients were implanted with the same “third generation” device and also a low clearance “fourth generation” resurfacing device. Only ion results from patients who were 12 months post surgery were included. Bilateral patients were matched to unilateral patients according to the surgeon performing the operation; the resurfacing system implanted and cup inclination and anteversion angles. The ion data from each bilateral group was tested against the corresponding unilateral groups using the Mann Whitney U test for non parametric data. Significance was drawn at p<0.05.

Surgeon 1: There were 310 patients with unilateral joints and 50 with bilateral joints. There were no significant differences with regard to time to follow up, activity levels, joint sizes or cup orientations. Serum chromium (Cr) and cobalt (Co) concentrations were significantly greater in the bilateral group (p<0.001). Median ion levels were greater in the bilateral group by a factor of >2 in the smallest joint sizes and <2 in the largest joints. Surgeon 2: There were 11 patients with bilateral third generation resurfacing joints and 50 with unilateral joints of the same design. The same relationship as described above was identified. There was a notable difference in the fourth generation implant group (n=13 bilateral, n=100 unilateral). Median ion levels for patients receiving bilateral joints of sizes <47mm were ten times greater than in the corresponding unilateral group.

Bearing diameter and component design are critical factors in determining metal ion levels following bilateral hip resurfacings. Surgeons must consider the potential implications of gross increases in metal ion levels prior to performing bilateral hip resurfacings in smaller patients.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 94 - 94
1 Mar 2012
Sukeik M Alshryda S Sharda P Lou T Nargol A Mason J
Full Access

Objectives

To investigate the value of tranexamic acid (TA) in reducing blood loss and blood transfusion after TKR and other clinical outcomes such as deep venous thrombosis (DVT), pulmonary embolism (PE), ischaemic heart diseases and mortality.

Methods

A systematic review and meta-analysis of published randomised and quasi-randomised trials which used TA to reduce blood loss in knee arthroplasty was conducted. The data was evaluated using the generic evaluation tool designed by the Cochrane Bone, Joint and Muscle Trauma Group.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 21 - 21
1 Mar 2012
Sharda P Alshryda S Shetty A Kalia P Nargol A Maheswaran S
Full Access

PURPOSE OF STUDY

20-70% of patients need blood transfusion postoperatively. There remain safety concerns regarding allogenic blood transfusion. Tranexamic acid (TA) is a synthetic antifibrinolytic agent that has been successfully used to stop bleeding in other specialties. We applied TA topically prior to the wound closure to find out the effect on blood loss as well as need for subsequent blood transfusion. This method of administration is quick, easy, has less systemic side effect and provides a higher concentration at the bleeding site.

MATERIALS AND METHODS

A double blind randomised controlled trial of 154 patients who underwent unilateral primary cemented total knee replacement. Patients were randomised into tranexamic acid group (1g drug mixed with saline to make up 20mls) or placebo (20ml 0.9% saline). The administration technique and drain protocol was standardised for all patients. Drain output was measured at 24 hours, and both groups compared for need of Blood transfusion. Outcome measures - blood loss, transfusion, complications, Euroqol and Oxford Knee Score.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1577 - 1585
1 Dec 2011
Alshryda S Sarda P Sukeik M Nargol A Blenkinsopp J Mason JM

We conducted a systematic review and meta-analysis of randomised controlled trials evaluating the effect of tranexamic acid (TXA) upon blood loss and transfusion in primary total knee replacement. The review used the generic evaluation tool designed by the Cochrane Bone, Joint and Muscle Trauma Group. A total of 19 trials were eligible: 18 used intravenous administration, one also evaluated oral dosing and one trial evaluated topical use. TXA led to a significant reduction in the proportion of patients requiring blood transfusion (risk ratio (RR) 2.56, 95% confidence interval (CI) 2.1 to 3.1, p < 0.001; heterogeneity I2 = 75%; 14 trials, 824 patients). Using TXA also reduced total blood loss by a mean of 591 ml (95% CI 536 to 647, p < 0.001; I2 = 78%; nine trials, 763 patients). The clinical interpretation of these findings is limited by substantial heterogeneity. However, subgroup analysis of high-dose (> 4 g) TXA showed a plausible consistent reduction in blood transfusion requirements (RR 5.33; 95% CI 2.44 to 11.65, p < 0.001; I2 = 0%), a finding that should be confirmed by a further well-designed trial. The current evidence from trials does not support an increased risk of deep-vein thrombosis (13 trials, 801 patients) or pulmonary embolism (18 trials, 971 patients) due to TXA administration.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 549 - 549
1 Nov 2011
Joyce T Langton D Lord J Grigg H Jameson S Cooke N Tulloch C Logishetty R Meek D Nargol A
Full Access

Background: There is a paucity of published data with regard to the wear of failed metal on metal (MoM) resurfacing devices.

Materials and Methods: MoM components retrieved from patients from two independent centres experiencing failure secondary to ARMD were analysed using a Mitutoyo Legex 322 coordinate measuring machine (CMM) which has an accuracy of 0.8 microns. Between 4000 – 6000 points were taken on each explant, dependent on the size of the bearing surface. Maximum wear depths and total volumetric wear were calculated. These values were compared to those from control samples retrieved following uncomplicated fractures/femoral collapse secondary to avascular necrosis (after calculating equivalent yearly wear rates).

Results: 58 ARMD components were analysed. This included 22 36mm MoM THRs (DePuy Pinnacle), 28 DePuy ASRs and 8 Zimmer Duroms. There were 30 resurfacing fracture/avascular necrosis controls. Volumetric wear rates and maximum wear depths of ARMD resurfacing components were significantly greater than the resurfacing control group for both the ASR and Duroms (p< 0.05) however 2 ARMD components exhibited similar amounts of wear compared to controls. Wear rates of the ARMD THR group were significantly lower than the ARMD resurfacing group (p< 0.05).

Conclusions: Increased articular wear is associated with an increased incidence of local adverse effects including tissue necrosis, joint effusions and fractures. However, there are a minority of patients who can develop tissue necrosis in the absence of accelerated wear, implying a spectrum of sensitivity This is reflected in the incidence of ARMD in the patient groups at the main study centre: > 5% in the ASR group and approx 1% in the THR group. We believe this indicates a failure of adequate lubrication and the resultant negative effects in larger bearing devices.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 549 - 549
1 Nov 2011
Langton D Jameson S Joyce T Nargol A
Full Access

Background: Blood metal ion levels have been shown to correlate with wear of retrieved components. Increased articular wear is associated with early failure. It is not clear what the management of patients with raised metal ions who remain asymptomatic should be.

Methods: A prospective study of the DePuy ASR resurfacing device was commenced in 2004 at our independent centre. Blood and serum metal ion analysis is carried out routinely. We assessed the clinical outcome of all patients with cobalt (Co) concentrations > 5μg/L (n=25). We consider this value to be a clear indicator of a poorly performing bearing surface. Patients who remained asymptomatic were brought back for repeat ions testing and clinical assessment. A parallel study was carried out to examine the relationship between the wear depth of retrieved explants (assessed using a coordinate measuring) and blood ion levels

Results: Increases in chromium(Cr) and Co levels from the second blood sample correlated with the first sample (p< 0.05). The four patients with the highest Cr Co levels went on to suffer spontaneous femoral neck fractures between 3 and 4 years post op (pseudotumours in two of these cases). Of the remaining 25, 5 were revised for worsening pain (large effusions and areas of tissue necrosis were seen at revision) and 10 were under investigation for new onset of pain. Wear depths correlated well with blood ion concentrations (p< 0.05) substantiating the results of another centre.

Conclusion: 60% of patients with raised metal ion levels went on to develop complications within two years of the blood test. This suggests patients may only have temporary tolerances to excessive concentrations of metal wear debris. One novel association appears to be spontaneous delayed femoral fracture. Results suggest increased ion concentrations associated with small joint size/sub optimal cup orientation will continue to rise and these patients should be followed up closely.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 549 - 549
1 Nov 2011
Langton D Jameson S Van Oursouw M De Smet K Nargol A
Full Access

Background: Definitive cup position for the reduction of blood metal ion levels has yet to be established.

Methods: Samples for serum metal ion analysis are taken routinely from patients under the care of the two senior authors of this paper. Both are high volume experienced hip resurfacing surgeons, one based in England, the other in Belgium. Metal ion results from two centres from patients with unilateral joints were correlated to size and orientation of femoral and acetabular components, UCLA activity score, age, time post surgery and post operative femoral head/neck ratios. EBRA software was used to assess cup inclination and anteversion on standing radiographs.

Cup orientation in vivo was compared to explant analysis of 60 retrieved resurfacing components using a coordinate measuring machine.

Results: Three resurfacing devices were studied. There were 620 results in total. Only femoral size and cup inclination/anteversion were found to have any effect on ion levels. In all devices, metal ion levels were inversely related to femoral size (p< 0.05). The device providing the smallest acetabular coverage arc was associated with the highest metal ion levels. Consistent throughout the implants, lowest ion levels were associated with cups with radiological inclination of 40–50° and anteversion 10–20°. Cup inclination angles lower than 40° were associated with posterior edge loading and likely sub-luxation of the femoral component.

Conclusion: The greater the coverage angle provided by the acetabular component, the greater the tolerance to suboptimal position. Lowest ion levels were found in well positioned lower clearance devices. Cup inclination/ante-version angles of 45/15+/−5° were associated with low ion levels in all three devices. We do not recommend cups to be placed with inclination angles below 40°.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 38 - 38
1 Jan 2011
Langton D Jameson S Joyce T Webb J Nargol A
Full Access

This report documents the clinical and histological findings, the operative management and the explant analyses of patients with early aseptic failure of large metal-on-metal (MonM) bearing joints.

Three hundred and fifty patients have been implanted with the ASR bearing surface (resurfacing or a modular THR) by a single surgeon at an independent centre since 2004. Six patients (all female) have been revised secondary to aseptic failure. All complained of severe groin pain exacerbated by straight leg raise and routine investigations were unable to establish a cause. Large amounts of sterile, highly viscous green fluid were aspirated from the hip joints in each case. Gross swelling of the pseudocapsule and a similar green fluid surrounding the implants were found at revision surgery. Histological examination of periprosthetic tissue samples showed changes consistent with ALVAL/metallosis, and analysis of the fluid revealed dense numbers of inflammatory cells. Symptoms in patients revised to ceramic-on-ceramic bearings improved post operatively. This was not the case with those reimplanted with MoM joints. Data from a subset of 76 patients (all unilateral resurfacings) showed that malaligned cups (anteverted > 20° +/− inclination angle > 45°) were associated with significantly higher whole blood metal ion levels than cups positioned within this range. All the patients with early aseptic failure had malaligned cups. Independent explant analysis revealed significant increases in the surface roughness values of the articular surfaces.

Our results suggest that some patients develop a significant inflammatory reaction to metal implants. This may be due to high levels of metal debris around the implant as a result of accelerated wear. Poor component alignment was found in all our patients with early aseptic failure.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 38 - 38
1 Jan 2011
Jameson S Langton D Joyce T Webb J Nargol A
Full Access

Metal ion concentrations following metal on metal hip resurfacing arthroplasty remain a concern. Variables associated with increased metal ion concentrations need to be established. This study provides metal ion data from a consecutive cohort of the first 76 patients implanted with a fourth generation hip resurfacing prosthesis.

All patients agreed to post-operative blood metal ion sampling at a minimum of one year. Post-operative radiographic measurements of cup inclination and anteversion were obtained using the EBRA software. Mean whole blood chromium (Cr) and cobalt (Co) concentrations in patients receiving the smallest femoral implants (Ł51mm) were greater than in the patients implanted with the largest prostheses (ł53mm) by a factor of 3 and 9 respectively. Ion concentrations in the small femoral group were significantly related to acetabular inclination (R=0.439, P< 0.001 for Cr, R=0.372, P=0.004 for Co) and anteversion (R=0.330, P=0.010 for Cr, R=0.338, P=0.008 for Co). This relationship was not significant in the large implant group. Mean Cr and Co concentrations in patients with accurately orientated cups (inclination < 45°, anteversion < 20°) were 3.7μg/l and 1.8 μg/l respectively, compared to 9.1μg/l and 17.5μg/l in malaligned cups.

A reduced surface contact area caused by cup malalignment may increase contact stresses, resulting in a high wear rate if fluid film lubrication is inadequate. Improved fluid film lubrication has previously been found in larger heads in vitro. Accurate acetabular component positioning is essential in order to reduce metal ion concentrations following hip resurfacing.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_IV | Pages 542 - 542
1 Oct 2010
Sukeik M Alshryda S Mason J Nargol A
Full Access

Since the era of total knee replacement (TKR) began in the late 1960s, total knee replacement has become one of the commonest operations in orthopaedic practice.

TKR is frequently associated with transfusion of allogenic blood Benoni G 1995; Seppo T 1997. In our centre, 30 % of patients who had undergone TKR received allogenc blood transfusion perioperatively. Although, serological screening has reduced the risk for viral infection to a very low levelKlein HG 1995; Schreiber GB 1996, the public is still concerned about this potential serious complication. Allogenic blood transfusion can be also associated with other non infectious complications such as haemolysis, immunosuppression, transfusion-related acute lung injury and even death.Madjdpour C 2005 Therefore, further refinement of strategies to avoid exposure to allogeneic blood is needed.

Amongst the technologies to minimise the need for blood transfusion is the use of the anti-fibrinolytic drugs aprotinin, tranexamic acid (TXA), and epsilon amino-caproic acid (EACA).New Reference

Objectives: The purpose of this review is to investigate the evidence for the efficacy of Tranexamic acid in reducing peri-operative blood loss and blood transfusion after total knee replacement, and the evidence for any effect on clinical outcomes such as reduction in re-operation rates or increase in complication rates (e.g. deep venous thrombosis, pulmonary embolism, ischaemic heart diseases and mortality).

Method: Systematic review and metananalysis based on Cochrane guidelines of all randomised and quasirandomised trials.

Results: Fiften RCTs were included in the study; there has been a significant reduction of blood loss (P value 0.00001, I2 = 89%), blood transfusion without increase in systeanatic side effects such as ischaemic heart diseases, DVT, pulmondary embolisms. There was no singnificant difference in length of stay.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 398 - 398
1 Jul 2010
Bhatnagar S Langton D Aval S Holland J Nargol A
Full Access

Background: Resurfacing cups may produce significant clinical problems when placed at the extremes of version, including increased production of metal debris and psoas tendonitis.

Methods: We obtained the femoral and acetabular components of two unused ASR (Articular Surface Replacement) and BHR (Birmingham Hip Resurfacing) devices. The components were placed in moulds at varying degrees of inclination and anteversion in a grid to simulate pelvic landmarks and AP radiographic images were captured. The true radiological anteversion was determined by rotating the x-ray beam so that it was perpendicular to the acetabular axis and measuring the angle subtended by the cup rim and the vertical axis.

Five blinded orthopaedic registrars then used EBRA (Einzel-Bild-Roentgen-Analysis, University of Inns-bruck, Austria) software to determine the radiological anteversion from the AP films. Twenty-five ASR and twenty-five BHR images were analysed.

At the same time each observer was asked to grade the cups as “1” (< 10°) “2” (10–20°) “3” (20–30°) or “4” (> 30°) depending on the appearances of the cup vertices.

Results: Mean error for each observer was −0.7 (minimum) to 1.6° (maximum). The range of standard deviations of error for each observer was +/− 2.2 (minimum) to 3.5° (maximum). Retroverted cups were not identified in the majority of cases.

Cups graded as “1” or “4” showed high sensitivity and specificity for the true grade as determined on the lateral radiographs.

Conclusions: EBRA software can be used to calculate the anteversion of resurfacing cups to a clinically acceptable degree. The clinician must be aware of the limitations of the software most notably the difficulty in identifying a retroverted cup and errors arising from poor quality radiographs in terms of pelvic rotation. The presented clinical grading system can be used as a rapid assessment tool to identify cups at the extremes of anteversion.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 390 - 390
1 Jul 2010
Langton D Sprowson A Jameson S Joyce T Reed M Partington P Carluke I Nargol A
Full Access

Background: There are no large comparative metal ion studies of commercially available hip resurfacing devices which have taken into account the effects of femoral size and cup inclination and anteversion.

Patients and methods: Metal ion analysis is carried out routinely at our independent centre. We present the metal ion results of 95 unilateral ASR patients and 70 unilateral BHR patients. For all patients, acetabular cup orientation was assessed using EBRA software. Patients with other metallic implants and those within 12 months of surgery were excluded.

Results: Whole blood/serum chromium (Cr) and cobalt (Co) concentrations were inversely related to femoral component size in both the ASR and BHR group (p< 0.05). Cr and Co levels were only seen to increase in the BHR group when the cup was implanted with an inclination greater than 55°. A significant relationship was identifed between the anteversion of the BHR cup and Cr and Co (p< 0.05 for Co, Spearman Rank correlation), with an increase in ions observed at anteversion angles > 17°. Cr and Co were more strongly influenced by cup position in the case of the ASR, with an increase in metal ions observed at inclinations greater than 45° and anteversion angles of < 10° and > 20°.

Discussion: The increased tolerance of the BHR cup to inclinations between 45–55° is likely due to the larger BHR cup providing greater protection against edge loading. When the cohort was divided by gender, the median Cr concentrations of the male ASR patients were significantly lower than those of the BHR males (p< 0.001). This suggests that in larger components positioned at more satisfactory angles of inclination and anteversion, the lower clearance of the ASR proves more significant than the extra coverage provided by the BHR cup. The BHR appears to be more sensitive to changes in anteversion than inclination.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 397 - 398
1 Jul 2010
Langton D Joyce T Jameson S Nargol A
Full Access

Background: There is a paucity of published literature with regard to metal ion levels following bilateral hip resurfacings.

Method: Recent research has identified femoral component size and acetabular cup anteversion and inclination as important variables involved in metal ion release into the blood. We compared 13 patients with bilateral joints resurfaced using the ASR and 11 patients with bilateral BHRs to those with unilateral joints of similar size and cup inclinations/anteversions. Patients were excluded if the last procedure was within 12 months of blood sampling and if they had other metallic implants. Analysis of standing radiographs to determine cup orientation was carried out using EBRA software. Blood samples were analysed using ICPMS.

Results: Median whole blood Cr and Co values were higher in the BHR bilateral group when compared to the unilateral BHR group by a factor of 1.12 and 1.24 respectively. (5.17 vs 4.59 for Cr, 2.40 vs 1.93 for Co). The difference was significant for Co (p=0.030) but not for Cr (p=0.136). For the large ASR joints (53mm), median Cr and Co values were higher in the bilateral group by a factor of 1.5 and 1.85 respectively. (4.55 vs 2.97 (Cr) 2.83 vs 1.53 (Co)). The difference was significant for both Cr (p=0.001) and Co (p=0.022). For the small ASR joints (< 53mm), median Cr Co values were higher in the bilateral group by a factor of 1.95 and 2.30 respectively. (8.29 vs 4.25 (Cr), 6.78 vs 2.94 (Co)) (p=0.019 Co) (p=0.007 Co).

Conclusion: Metal ion concentrations are significantly greater in patients with bilateral resurfacings. The Cr Co concentrations observed in our patients with bilateral small ASR joints are double those in the published literature on bilateral 28mm metal on metal(MoM) joints implying that the lubrication achieved by small ASRs is sub optimal.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 390 - 391
1 Jul 2010
Langton D Jameson S Joyce T Natu S Logishetty R Tulloch C Nargol A
Full Access

In our independent centre, in the period from January 2003 to august 2008, over 1100 36mm MoM THRs have been implanted as well as 155 Birmingham Hip Resurfacing procedures, 402 ASR resurfacings and 75 THRs using ASR XL heads on SROM stems.

During this period we have experienced a number of failures with patients complaining of worsening groin pain at varying lengths of time post operatively. Aspiration of the hip joints yielded a large sterile effusion on each occasion. At revision, there were copious amounts of green grey fluid with varying degrees of necrosis. There were 11 failures of this nature in patients with ASR implants (10 females) and 2 in the 36 MoM THR group (one male one female).

Tissue specimens from revision surgery showed varying degrees of ‘ALVAL’ as well as consistently high numbers of histiocytes. Metal debris was also a common finding.

A fuller examination of our ASR cohort as a whole has shown that smaller components placed with inclinations > 45° and anteversions < 10 or > 20° are associated with increased metal ion levels. The 11 ASR failed joints were all sub optimally positioned (by the above definition), small components.

Explant analysis using a coordinate measuring machine and out of roundness device confirmed greater than expected wear of each component. The lower number of failures in the 36mm MoM group, as well as the equal sex incidence, suggests that the majority of these failures are due to the instigation of an immune reaction by large amounts of wear debris rather than adverse reactions to well functioning joints. It is likely that small malpositioned ASRs function in mixed to boundary lubrication, and this, combined with the larger radius of these joints compared to the 36mm MoM joints, results in more rapid wear.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 389 - 389
1 Jul 2010
Jameson S Langton D Nargol A
Full Access

Introduction: Excellent medium term results have been reported with the BHR hip resurfacing implant. A number of modifications have been made to the latest designs in an attempt to preserve acetabular bone stock, improve function and prolong survival. We present the clinical and radiological results from the first independent series of ASR resurfacings.

Methods: The first consecutive series of 214 ASR hip resurfacings (192 patients) were followed up prospectively. No patients were lost to follow-up. The mean age of patients at implantation was 56 years and 40% were female. All patients had pre- and 2-year post-operative Harris Hip Scores (HHS) and UCLA activity scores. Radiographic and implant survival analysis was performed at 24–54 months following implantation.

Results: One hundred and seventy-three hips (87%) had an excellent Harris Hip Score (90 or above). Mean postoperative UCLA activity score was 7 and 92% were highly satisfied with the outcome. There were eleven revisions (5.1%). Four (1.9%) had femoral neck fractures (three had a femoral neck notch), two (0.9%) collapsed secondary to avascular necrosis and five (2.3%) were revised because of ongoing pain, as a result of metal wear debris.

Discussion: Although the fracture rate is similar to reports in the literature, the overall revision rate was higher. Rates of wear debris-related failure is concerning. Further investigation of specific implant failure is necessary.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 301 - 302
1 May 2010
Jameson S Ramisetty N Langton D Webb J Logishetty R Nargol A
Full Access

Introduction: There are ongoing concerns regarding metal wear debris following the use of metal-on-metal (MonM) bearings for hip surface and total arthroplasty. A Type IV Hypersensitivity reaction to MonM articulations has previously been identified (aseptic lymphocyte dominated vasculitis associated lesion, ALVAL) but little is known of its incidence, diagnosis or management. Persisting groin pain in MonM patients may be undiagnosed ALVAL. At our single centre we have reviewed and compared three types of MonM articulations to examine the incidence of ALVAL and to identify trends.

Methods: The resurfacing group comprised 250 patients with the ASR prosthesis. In the resurfacing hybrid total hip replacement (THR) group there were 86 patients implanted with an ASR head on a stem. The final group comprised of 625 patients with a MonM THR using a 36mm Pinnacle head. Both the S-ROM and the Corail stems were used in the THR groups. Patients with persisting and activity-restricting groin pain had tests for infection. Patients were counselled and revision was offered if ALVAL was suspected from the clinical picture, blood results and the aspiration result. Specimens for microbiological and histological analysis were taken at the time of revision.

Results: We found 5 cases of histologically proven ALVAL in the absence of infection in 961 patients. The incidence was: 1.2% in the resurfacing group, 2.3% of Resurfacing Hybrid THR group and 0 in the 36 mm THR group. All 5 cases were in female patients. Only 1 case had any radiological abnormality. One patient was initially revised from a resurfacing to a 36mm MonM THR without clinical success. All patients have now been revised to ceramic-on-ceramic bearings with improvements in outcome.

Discussion: ALVAL may be under-diagnosed. The 5 patients we describe showed good clinical recovery following their primary procedure. However, activity levels decreased and pain increased at 6–12 months post-op. All described non-specific systemic symptoms. On examination, a painful straight leg raise was a characteristic finding. This may result from the significant effusion found around the hip at each revision. Fluid aspirated from these hips was of a characteristic colour (green grey) and viscosity. The failure of the revision of a resurfacing to a smaller MonM bearing highlights the problem of sensitisation to the metal debris. Any subsequent revision to a MonM bearing is unlikely to improve clinical outcome. This finding is consistent with previous reports in the literature. Our results suggest the incidence of ALVAL may be higher that previously thought. We suggest all patients with significant groin pain should have inflammatory markers tests and a hip aspiration performed. In the absence of infection, revision to an alternative bearing surface may be indicated.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 302 - 302
1 May 2010
Jameson S Webb J Langton D Ramisetty N Logishetty R Nargol A
Full Access

Introduction: Potential problems resulting from metal-on-metal hip wear debris are well known. Previous studies have shown an association between high cup angles and raised metal ion levels, but a link to clinical outcome has yet to be established. We aim to show the relationship between high cup angles, raised metal ion levels and pain following hip resurfacing.

Methods: This analysis comprised the first 250 Depuy ASR hip resurfacings performed by a single surgeon (senior author) over a 3 yr period at an independent centre. Patients were followed up, examined and x-rayed at regular intervals. We measured Harris Hip Score (HHS) and pain. The acetabular cup inclination angles were measured from standardised pelvic radiographs. A subgroup of 80 patients had serum and whole blood chromium and cobalt measurements performed. This subgroup comprised patients with a similar sex split, age, activity level, follow-up and cup angle profile to the parent group.

Results: HHS improved from 51.6 pre-operatively to 94.6 post-operatively. Males had a significantly higher post-op HHS (97.1) compared with the females (91.0). 2.1% of male patients had pain compared with 8.7% of females. High cup angles were associated with pain in females. This relationship did not occur in males. In patients with cup angles of 48° and above there were no males with pain compared to 15.4% of females. Females under 48° had a HHS of 93.7 compared to 88.0 for those over 48°. When we analysed metal ion levels there was a similar relationship. Patients with cup angles of 48° and above had significantly higher serum and whole blood chromium and cobalt levels compared with lower cup angles. Three patients with cup angles over 50° developed ALVAL (aseptic lymphocyte dominated vasculitis associated lesion) based on histological findings. All were female.

Discussion: High cup angles are associated with increased pain and elevated metal ion levels. Women seem to be more intolerant of an higher cup angle than men. This may be a consequence of the smaller prostheses in females. Cups implanted lower than 48° had a better clinical outcome and less metal ion wear. Like other 4th generation designs, the ASR cup is not a complete hemisphere (unlike the BHR) and may be susceptible to edge loading at lower inclination angles than previously thought. We believe the current recommendation of 45° +/−5° is too high based on our findings. We recommend an inclination angle that does not exceed 48°.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 61 - 61
1 Mar 2010
Jameson* S Langton D Nargol A
Full Access

Optimal cup orientation for metal-on-metal hip resurfacing has yet to be established. Guidance is based on hip replacement data and in vitro studies. We sought to determine the influence of component size and positioning on early clinical outcome.

This study comprises a consecutive series of 200 hip resurfacings. All had Harris Hip Scores (HHS) at one-year review. Acetabular inclination angles were measured on pre-operative radiographs, and cup inclination/anteversion angles on 3-month post-operative films using EBRA. Restoration of anatomy was defined as placement of the cup within +/−5 degrees of pre-operative inclination. The difference between pre-operative acetabular and post-operative cup inclination was termed cup-angle difference (CAD).

HHS inversely correlated with CAD (P=0.023) and anteversion (P=0.003), and directly correlated with femoral head size (P< 0.001). In patients with restoration of inclination anatomy mean HHS at one year was significantly higher at 98.7 compared with cups placed outside the normal anatomy restoration limits (93.8, P=0.003). Patients with anteversion > 20 degrees had a significantly lower HHS (P=0.010) compared with cups anteverted < 20 degrees. 96% of patients with HHS < 90 had malaligned cups (inclination over 45 degrees, anteversion over 20 degrees).

Restoring pre-operative cup inclination, anteverting the cup < 20 degrees and using large femoral heads improves early clinical outcome following MonM hip resurfacing. We recommend accurate pre-operative planning and meticulous attention to intra-operative cup positioning with these results in mind.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 57 - 57
1 Mar 2010
Joyce* T Langton D Jameson S Nargol A
Full Access

Resurfacing metal-on-metal hip arthroplasty is currently showing promising clinical results. However there are concerns related to such implants, including the elevated levels of metal ions typically seen in patients. Valuable data can be obtained from explanted prostheses but due to their recent introduction few retrieval studies on resurfacing hip prostheses have been published.

Five ASR hip resurfacing prostheses were revised due to pain. From two patients, head and cup were available for independent explant analysis. In the other three cases only femoral components were available. All were removed from female patients and all were revised to ceramic-on-ceramic hip prostheses. Post-operative radiographic measurements of cup inclination and ante-version were obtained using the EBRA software. The surface roughness values of the articulating surfaces of the explants were measured using a non-contacting profilometer. A co-ordinate measuring machine was used to measure the diameter of the head and the cup and thus the diametral clearance. The same measurements were then taken from a new unused ASR prosthesis and compared. Using elastohydrodynamic theory the minimum effective film thickness of the implant was calculated. In turn this allowed the lubrication regime to be determined.

The average roughness values of the head and the cup of one implant were found to be 0.135microns and 0.058microns respectively, with a diametral clearance of 110microns. These results indicated that, at the time of removal, the prosthesis would have operated in the boundary lubrication regime. Other explants showed evidence of localised contact between the head and the rim of the acetabular cup, and these showed articulating surfaces with typical roughness values of between 0.025microns and 0.050microns. The new ASR had head and cup surface roughness values of 0.010microns and 0.012microns respectively and a diametral clearance of 87microns, implying that a new implant would operate under fluid film lubrication. All cups five were implanted with inclination angles over 45 degrees and anteversion over 25 degrees.

These results suggest that components with high inclination and anteversion angles display greater than expected wear and may operate in boundary rather than fluid film lubrication which may eventually lead to early failure.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 406 - 406
1 Sep 2009
Jameson S Langton D Nargol A
Full Access

Introduction: Patient selection has been critical to the excellent medium-term clinical results following hip resurfacing. Hypersensivity to metal ion debris has been described in previous generations of metal-on-metal bearings. This may also be a problem that affects modern designs. Characteristic histological changes have been identified (ALVAL). There are few studies that include large female numbers, and show separate outcome and implant survival.

Methods: Eighty-one female hips at a mean of 23 months (11–43) had an ASR hip resurfacing procedure at a single-surgeon independent centre. Mean age was 55 years (28–69). Harris Hip Scores (HHS) were recorded at one-year follow-up. Failures were analysed.

Results: HHS improved from 46.4 (11–77) to 90.2 (27–100). Overall, there was a 7.4 % revision rate. There were 3 femoral neck fractures. In the entire series of 98 female patients there were 3 cases of severe pain requiring revision (3.1%). All three patients had HHS < 50 at one year follow-up. Patients had groin pain, reduced flexion and a painful straight leg raise. Blood results were not suggestive of infection. Aspiration of the hip joint in each case revealed copious amounts of milky green grey aseptic fluid. All had similar macroscopic changes at revision. There were characteristic histological changes in keeping with ALVAL. All 3 patients were revised to THRs with ceramic bearings.

Discussion: The failure rate of 7.4% in the older female group is poor at this early stage following hip resurfacing. The incidence of metal hypersensitivity in our series suggests this complication may be more common than previously thought. Patients with persisting pain of unknown aetiology following resurfacing at other centres may have metal hypersensitivity.