SERUM METAL ION CONCENTRATIONS FOLLOWING BILATERAL MOM HIP RESURFACING ARTHROPLASTY: JOINT SIZE IS CRITICAL

Indian Orthopaedic Society UK (IOS)



Abstract

Bearing diameter and acetabular component orientation have been shown to be important variables effecting blood metal ion levels following hip resurfacing arthroplasty. So far no studies on bilateral hip resurfacings have taken into account these variables.

We examined the serum ion results of patients under the care of two experienced hip resurfacing surgeons who carry out ion analysis as part of routine post operative care. Surgeon 1: Patients were implanted exclusively with a “third generation” resurfacing device. Surgeon 2: Patients were implanted with the same “third generation” device and also a low clearance “fourth generation” resurfacing device. Only ion results from patients who were 12 months post surgery were included. Bilateral patients were matched to unilateral patients according to the surgeon performing the operation; the resurfacing system implanted and cup inclination and anteversion angles. The ion data from each bilateral group was tested against the corresponding unilateral groups using the Mann Whitney U test for non parametric data. Significance was drawn at p<0.05.

Surgeon 1: There were 310 patients with unilateral joints and 50 with bilateral joints. There were no significant differences with regard to time to follow up, activity levels, joint sizes or cup orientations. Serum chromium (Cr) and cobalt (Co) concentrations were significantly greater in the bilateral group (p<0.001). Median ion levels were greater in the bilateral group by a factor of >2 in the smallest joint sizes and <2 in the largest joints. Surgeon 2: There were 11 patients with bilateral third generation resurfacing joints and 50 with unilateral joints of the same design. The same relationship as described above was identified. There was a notable difference in the fourth generation implant group (n=13 bilateral, n=100 unilateral). Median ion levels for patients receiving bilateral joints of sizes <47mm were ten times greater than in the corresponding unilateral group.

Bearing diameter and component design are critical factors in determining metal ion levels following bilateral hip resurfacings. Surgeons must consider the potential implications of gross increases in metal ion levels prior to performing bilateral hip resurfacings in smaller patients.