header advert
Results 81 - 100 of 106
Results per page:
The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 840 - 845
1 Jun 2016
Chesser TJS Fox R Harding K Halliday R Barnfield S Willett K Lamb S Yau C Javaid MK Gray AC Young J Taylor H Shah K Greenwood R

Aims

We wished to assess the feasibility of a future randomised controlled trial of parathyroid hormone (PTH) supplements to aid healing of trochanteric fractures of the hip, by an open label prospective feasibility and pilot study with a nested qualitative sub study. This aimed to inform the design of a future powered study comparing the functional recovery after trochanteric hip fracture in patients undergoing standard care, versus those who undergo administration of subcutaneous injection of PTH for six weeks.

Patients and Methods

We undertook a pilot study comparing the functional recovery after trochanteric hip fracture in patients 60 years or older, admitted with a trochanteric hip fracture, and potentially eligible to be randomised to either standard care or the administration of subcutaneous PTH for six weeks. Our desired outcomes were functional testing and measures to assess the feasibility and acceptability of the study.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 178 - 184
1 May 2016
Dean BJF Jones LD Palmer AJR Macnair RD Brewer PE Jayadev C Wheelton AN Ball DEJ Nandra RS Aujla RS Sykes AE Carr AJ

Objectives

The PROximal Fracture of the Humerus: Evaluation by Randomisation (PROFHER) trial has recently demonstrated that surgery is non-superior to non-operative treatment in the management of displaced proximal humeral fractures. The objective of this study was to assess current surgical practice in the context of the PROFHER trial in terms of patient demographics, injury characteristics and the nature of the surgical treatment.

Methods

A total of ten consecutive patients undergoing surgery for the treatment of a proximal humeral fracture from each of 11 United Kingdom hospitals were retrospectively identified over a 15 month period between January 2014 and March 2015. Data gathered for the 110 patients included patient demographics, injury characteristics, mode of surgical fixation, the grade of operating surgeon and the cost of the surgical implants.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 191 - 197
1 May 2016
Kienast B Kowald B Seide K Aljudaibi M Faschingbauer M Juergens C Gille J

Objectives

The monitoring of fracture healing is a complex process. Typically, successive radiographs are performed and an emerging calcification of the fracture area is evaluated. The aim of this study was to investigate whether different bone healing patterns can be distinguished using a telemetric instrumented femoral internal plate fixator.

Materials and Methods

An electronic telemetric system was developed to assess bone healing mechanically. The system consists of a telemetry module which is applied to an internal locking plate fixator, an external reader device, a sensor for measuring externally applied load and a laptop computer with processing software. By correlation between externally applied load and load measured in the implant, the elasticity of the osteosynthesis is calculated. The elasticity decreases with ongoing consolidation of a fracture or nonunion and is an appropriate parameter for the course of bone healing. At our centre, clinical application has been performed in 56 patients suffering nonunion or fracture of the femur.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 116 - 121
1 Apr 2016
Leow JM Clement ND Tawonsawatruk T Simpson CJ Simpson AHRW

Objectives

The radiographic union score for tibial (RUST) fractures was developed by Whelan et al to assess the healing of tibial fractures following intramedullary nailing. In the current study, the repeatability and reliability of the RUST score was evaluated in an independent centre (a) using the original description, (b) after further interpretation of the description of the score, and (c) with the immediate post-operative radiograph available for comparison.

Methods

A total of 15 radiographs of tibial shaft fractures treated by intramedullary nailing (IM) were scored by three observers using the RUST system. Following discussion on how the criteria of the RUST system should be implemented, 45 sets (i.e. AP and lateral) of radiographs of IM nailed tibial fractures were scored by five observers. Finally, these 45 sets of radiographs were rescored with the baseline post-operative radiograph available for comparison.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 153 - 161
1 Apr 2016
Kleinlugtenbelt YV Nienhuis RW Bhandari M Goslings JC Poolman RW Scholtes VAB

Objectives

Patient-reported outcome measures (PROMs) are often used to evaluate the outcome of treatment in patients with distal radial fractures. Which PROM to select is often based on assessment of measurement properties, such as validity and reliability. Measurement properties are assessed in clinimetric studies, and results are often reviewed without considering the methodological quality of these studies. Our aim was to systematically review the methodological quality of clinimetric studies that evaluated measurement properties of PROMs used in patients with distal radial fractures, and to make recommendations for the selection of PROMs based on the level of evidence of each individual measurement property.

Methods

A systematic literature search was performed in PubMed, EMbase, CINAHL and PsycINFO databases to identify relevant clinimetric studies. Two reviewers independently assessed the methodological quality of the studies on measurement properties, using the COnsensus-based Standards for the selection of health Measurement INstruments (COSMIN) checklist. Level of evidence (strong / moderate / limited / lacking) for each measurement property per PROM was determined by combining the methodological quality and the results of the different clinimetric studies.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 106 - 115
1 Apr 2016
Gruber HE Ode G Hoelscher G Ingram J Bethea S Bosse MJ

Objectives

The biomembrane (induced membrane) formed around polymethylmethacrylate (PMMA) spacers has value in clinical applications for bone defect reconstruction. Few studies have evaluated its cellular, molecular or stem cell features. Our objective was to characterise induced membrane morphology, molecular features and osteogenic stem cell characteristics.

Methods

Following Institutional Review Board approval, biomembrane specimens were obtained from 12 patient surgeries for management of segmental bony defects (mean patient age 40.7 years, standard deviation 14.4). Biomembranes from nine tibias and three femurs were processed for morphologic, molecular or stem cell analyses. Gene expression was determined using the Affymetrix GeneChip Operating Software (GCOS). Molecular analyses compared biomembrane gene expression patterns with a mineralising osteoblast culture, and gene expression in specimens with longer spacer duration (> 12 weeks) with specimens with shorter durations. Statistical analyses used the unpaired student t-test (two tailed; p < 0.05 was considered significant).


Bone & Joint Research
Vol. 5, Issue 1 | Pages 18 - 25
1 Jan 2016
Sims AL Parsons N Achten J Griffin XL Costa ML Reed MR

Background

Approximately half of all hip fractures are displaced intracapsular fractures. The standard treatment for these fractures is either hemiarthroplasty or total hip arthroplasty. The recent National Institute for Health and Care Excellence (NICE) guidance on hip fracture management recommends the use of ‘proven’ cemented stem arthroplasty with an Orthopaedic Device Evaluation Panel (ODEP) rating of at least 3B (97% survival at three years). The Thompsons prosthesis is currently lacking an ODEP rating despite over 50 years of clinical use, likely due to the paucity of implant survival data. Nationally, adherence to these guidelines is varied as there is debate as to which prosthesis optimises patient outcomes.

Design

This study design is a multi-centre, multi-surgeon, parallel, two arm, standard-of-care pragmatic randomised controlled trial. It will be embedded within the WHiTE Comprehensive Cohort Study (ISRCTN63982700). The main analysis is a two-way equivalence comparison between Hemi-Thompson and Hemi-Exeter polished taper with Unitrax head. Secondary outcomes will include radiological leg length discrepancy measured as per Bidwai and Willett, mortality, re-operation rate and indication for re-operation, length of index hospital stay and revision at four months. This study will be supplemented by the NHFD (National Hip Fracture Database) dataset.


Bone & Joint Research
Vol. 4, Issue 12 | Pages 190 - 194
1 Dec 2015
Kleinlugtenbelt YV Hoekstra M Ham SJ Kloen P Haverlag R Simons MP Bhandari M Goslings JC Poolman RW Scholtes VAB

Objectives

Current studies on the additional benefit of using computed tomography (CT) in order to evaluate the surgeons’ agreement on treatment plans for fracture are inconsistent. This inconsistency can be explained by a methodological phenomenon called ‘spectrum bias’, defined as the bias inherent when investigators choose a population lacking therapeutic uncertainty for evaluation. The aim of the study is to determine the influence of spectrum bias on the intra-observer agreement of treatment plans for fractures of the distal radius.

Methods

Four surgeons evaluated 51 patients with displaced fractures of the distal radius at four time points: T1 and T2: conventional radiographs; T3 and T4: radiographs and additional CT scan (radiograph and CT). Choice of treatment plan (operative or non-operative) and therapeutic certainty (five-point scale: very uncertain to very certain) were rated. To determine the influence of spectrum bias, the intra-observer agreement was analysed, using Kappa statistics, for each degree of therapeutic certainty.


Bone & Joint Research
Vol. 3, Issue 12 | Pages 335 - 340
1 Dec 2014
Handoll HHG Goodchild L Brealey SD Hanchard NCA Jefferson L Keding A Rangan A

Objectives

A rigorous approach to developing, delivering and documenting rehabilitation within randomised controlled trials of surgical interventions is required to underpin the generation of reliable and usable evidence. This article describes the key processes used to ensure provision of good quality and comparable rehabilitation to all participants of a multi-centre randomised controlled trial comparing surgery with conservative treatment of proximal humeral fractures in adults.

Methods

These processes included the development of a patient information leaflet on self-care during sling immobilisation, the development of a basic treatment physiotherapy protocol that received input and endorsement by specialist physiotherapists providing patient care, and establishing an expectation for the provision of home exercises. Specially designed forms were also developed to facilitate reliable reporting of the physiotherapy care that patients received.


Bone & Joint Research
Vol. 3, Issue 6 | Pages 187 - 192
1 Jun 2014
Penn-Barwell JG Rand BCC Brown KV Wenke JC

Objectives

The purpose of this study was to refine an accepted contaminated rat femur defect model to result in an infection rate of approximately 50%. This threshold will allow examination of treatments aimed at reducing infection in open fractures with less risk of type II error.

Methods

Defects were created in the stablised femurs of anaethetised rats, contaminated with Staphylococcus aureus and then debrided and irrigated six hours later. After 14 days, the bone and implants were harvested for separate microbiological analysis. This basic model was developed in several studies by varying the quantity of bacterial inoculation, introducing various doses of systemic antibiotics with and without local antibiotics.


Bone & Joint Research
Vol. 3, Issue 3 | Pages 69 - 75
1 Mar 2014
Parsons N Griffin XL Achten J Costa ML

Objectives

To study the measurement properties of a joint specific patient reported outcome measure, a measure of capability and a general health-related quality of life (HRQOL) tool in a large cohort of patients with a hip fracture.

Methods

Responsiveness and associations between the Oxford Hip Score (a hip specific measure: OHS), ICEpop CAPability (a measure of capability in older people: ICECAP-O) and EuroQol EQ-5D (general health-related quality of life measure: EQ-5D) were assessed using data available from two large prospective studies. The three outcome measures were assessed concurrently at a number of fixed follow-up time-points in a consecutive sequence of patients, allowing direct assessment of change from baseline, inter-measure associations and validity using a range of statistical methods.


Bone & Joint Research
Vol. 2, Issue 10 | Pages 210 - 213
1 Oct 2013
Griffin XL McArthur J Achten J Parsons N Costa ML

Fractures of the proximal femur are one of the greatest challenges facing the medical community, constituting a heavy socioeconomic burden worldwide. Controversy exists regarding the optimal treatment for independent patients with displaced intracapsular fractures of the proximal femur. The recognised alternatives are hemiarthroplasty and total hip replacement. At present there is no established standard of care, with both types of arthroplasty being used in many centres. The principal advantages of total hip replacement are a functional benefit over hemiarthroplasty and a reduced risk of revision surgery. The principal criticism is the increased risk of dislocation. We believe that an alternative acetabular component may reduce the risk of dislocation but still provide the functional benefit of total hip replacement in these patients. We therefore propose to investigate the dislocation risk of a dual-mobility acetabular component compared with standard polyethylene component in total hip replacement for independent patients with displaced intracapsular fractures of the proximal femur within the framework of the larger WHiTE (Warwick Hip Trauma Evaluation) Comprehensive Cohort Study.

Cite this article: Bone Joint Res 2013;2:210–13.


Bone & Joint Research
Vol. 2, Issue 10 | Pages 206 - 209
1 Oct 2013
Griffin XL McArthur J Achten J Parsons N Costa ML

Fractures of the proximal femur are one of the greatest challenges facing the medical community, constituting a heavy socioeconomic burden worldwide. Controversy exists regarding the optimal treatment for patients with unstable trochanteric proximal femoral fractures. The recognised treatment alternatives are extramedullary fixation usually with a sliding hip screw and intramedullary fixation with a cephalomedullary nail. Current evidence suggests that best results and lowest complication rates occur using a sliding hip screw. Complications in these difficult fractures are relatively common regardless of type of treatment. We believe that a novel device, the X-Bolt dynamic plating system, may offer superior fixation over a sliding hip screw with lower reoperation risk and better function. We therefore propose to investigate the clinical effectiveness of the X-bolt dynamic plating system compared with standard sliding hip screw fixation within the framework of a the larger WHiTE (Warwick Hip Trauma Evaluation) Comprehensive Cohort Study.

Cite this article: Bone Joint Res 2013;2:206–9.


Bone & Joint Research
Vol. 2, Issue 8 | Pages 162 - 168
1 Aug 2013
Chia PH Gualano L Seevanayagam S Weinberg L

Objectives

To determine the morbidity and mortality outcomes of patients presenting with a fractured neck of femur in an Australian context. Peri-operative variables related to unfavourable outcomes were identified to allow planning of intervention strategies for improving peri-operative care.

Methods

We performed a retrospective observational study of 185 consecutive adult patients admitted to an Australian metropolitan teaching hospital with fractured neck of femur between 2009 and 2010. The main outcome measures were 30-day and one-year mortality rates, major complications and factors influencing mortality.


Bone & Joint Research
Vol. 2, Issue 8 | Pages 149 - 154
1 Aug 2013
Aurégan J Coyle RM Danoff JR Burky RE Akelina Y Rosenwasser MP

Objectives

One commonly used rat fracture model for bone and mineral research is a closed mid-shaft femur fracture as described by Bonnarens in 1984. Initially, this model was believed to create very reproducible fractures. However, there have been frequent reports of comminution and varying rates of complication. Given the importance of precise anticipation of those characteristics in laboratory research, we aimed to precisely estimate the rate of comminution, its importance and its effect on the amount of soft callus created. Furthermore, we aimed to precisely report the rate of complications such as death and infection.

Methods

We tested a rat model of femoral fracture on 84 rats based on Bonnarens’ original description. We used a proximal approach with trochanterotomy to insert the pin, a drop tower to create the fracture and a high-resolution fluoroscopic imager to detect the comminution. We weighed the soft callus on day seven and compared the soft callus parameters with the comminution status.


Bone & Joint Research
Vol. 2, Issue 6 | Pages 112 - 115
1 Jun 2013
Ismail HD Phedy P Kholinne E Kusnadi Y Sandhow L Merlina M

Objectives

Nonunion is one of the most troublesome complications to treat in orthopaedics. Former authors believed that atrophic nonunion occurred as a result of lack of mesenchymal stem cells (MSCs). We evaluated the number and viability of MSCs in site of atrophic nonunion compared with those in iliac crest.

Methods

We enrolled five patients with neglected atrophic nonunions of long bones confirmed by clinical examinations and plain radiographs into this study. As much as 10 ml bone marrow aspirate was obtained from both the nonunion site and the iliac crest and cultured for three weeks. Cell numbers were counted using a haemocytometer and vitality of the cells was determined by trypan blue staining. The cells were confirmed as MSCs by evaluating their expression marker (CD 105, CD 73, HLA-DR, CD 34, CD 45, CD 14, and CD 19). Cells number and viability were compared between the nonunion and iliac creat sites.


Bone & Joint Research
Vol. 2, Issue 5 | Pages 79 - 83
1 May 2013
Goffin JM Pankaj P Simpson AHRW Seil R Gerich TG

Objectives

Because of the contradictory body of evidence related to the potential benefits of helical blades in trochanteric fracture fixation, we studied the effect of bone compaction resulting from the insertion of a proximal femoral nail anti-rotation (PFNA).

Methods

We developed a subject-specific computational model of a trochanteric fracture (31-A2 in the AO classification) with lack of medial support and varied the bone density to account for variability in bone properties among hip fracture patients.


Bone & Joint Research
Vol. 2, Issue 3 | Pages 51 - 57
1 Mar 2013
Sullivan MP Torres SJ Mehta S Ahn J

Neurogenic heterotopic ossification (NHO) is a disorder of aberrant bone formation affecting one in five patients sustaining a spinal cord injury or traumatic brain injury. Ectopic bone forms around joints in characteristic patterns, causing pain and limiting movement especially around the hip and elbow. Clinical sequelae of neurogenic heterotopic ossification include urinary tract infection, pressure injuries, pneumonia and poor hygiene, making early diagnosis and treatment clinically compelling. However, diagnosis remains difficult with more investigation needed. Our pathophysiological understanding stems from mechanisms of basic bone formation enhanced by evidence of systemic influences from circulating humor factors and perhaps neurological ones. This increasing understanding guides our implementation of current prophylaxis and treatment including the use of non-steroidal anti-inflammatory drugs, bisphosphonates, radiation therapy and surgery and, importantly, should direct future, more effective ones.


Bone & Joint Research
Vol. 2, Issue 2 | Pages 26 - 32
1 Feb 2013
Neumann H Schulz AP Gille J Klinger M Jürgens C Reimers N Kienast B

Objectives

Osteochondral injuries, if not treated adequately, often lead to severe osteoarthritis. Possible treatment options include refixation of the fragment or replacement therapies such as Pridie drilling, microfracture or osteochondral grafts, all of which have certain disadvantages. Only refixation of the fragment can produce a smooth and resilient joint surface. The aim of this study was the evaluation of an ultrasound-activated bioresorbable pin for the refixation of osteochondral fragments under physiological conditions.

Methods

In 16 Merino sheep, specific osteochondral fragments of the medial femoral condyle were produced and refixed with one of conventional bioresorbable pins, titanium screws or ultrasound-activated pins. Macro- and microscopic scoring was undertaken after three months.


Bone & Joint Research
Vol. 1, Issue 11 | Pages 289 - 296
1 Nov 2012
Savaridas T Wallace RJ Muir AY Salter DM Simpson AHRW

Objectives

Small animal models of fracture repair primarily investigate indirect fracture healing via external callus formation. We present the first described rat model of direct fracture healing.

Methods

A rat tibial osteotomy was created and fixed with compression plating similar to that used in patients. The procedure was evaluated in 15 cadaver rats and then in vivo in ten Sprague-Dawley rats. Controls had osteotomies stabilised with a uniaxial external fixator that used the same surgical approach and relied on the same number and diameter of screw holes in bone.