Vol. 97-B, Issue 7 | Pages 890 - 898
1 Jul 2015
Renkawitz T Weber M Springorum H Sendtner E Woerner M Ulm K Weber T Grifka J

We report the kinematic and early clinical results of a patient- and observer-blinded randomised controlled trial in which CT scans were used to compare potential impingement-free range of movement (ROM) and acetabular component cover between patients treated with either the navigated ‘femur-first’ total hip arthroplasty (THA) method (n = 66; male/female 29/37, mean age 62.5 years; 50 to 74) or conventional THA (n = 69; male/female 35/34, mean age 62.9 years; 50 to 75). The Hip Osteoarthritis Outcome Score, the Harris hip score, the Euro-Qol-5D and the Mancuso THA patient expectations score were assessed at six weeks, six months and one year after surgery. A total of 48 of the patients (84%) in the navigated ‘femur-first’ group and 43 (65%) in the conventional group reached all the desirable potential ROM boundaries without prosthetic impingement for activities of daily living (ADL) in flexion, extension, abduction, adduction and rotation (p = 0.016). Acetabular component cover and surface contact with the host bone were > 87% in both groups. There was a significant difference between the navigated and the conventional groups’ Harris hip scores six weeks after surgery (p = 0.010). There were no significant differences with respect to any clinical outcome at six months and one year of follow-up. The navigated ‘femur-first’ technique improves the potential ROM for ADL without prosthetic impingement, although there was no observed clinical difference between the two treatment groups.

Cite this article: Bone Joint J 2015; 97-B:890–8.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 306 - 311
1 Mar 2015
Weber M Lechler P von Kunow F Völlner F Keshmiri A Hapfelmeier A Grifka J Renkawitz T

Femoral stem version has a major influence on impingement and early post-operative stability after total hip arthroplasty (THA).

The main objective of this study was to evaluate the validity of a novel radiological method for measuring stem version. Anteroposterior (AP) radiographs and three-dimensional CT scans were obtained for 115 patients (female/male 63/72, mean age 62.5 years (50 to 75)) who had undergone minimally invasive, cementless THA. Stem version was calculated from the AP hip radiograph by rotation-based change in the projected prosthetic neck–shaft (NSA*) angle using the mathematical formula ST = arcos [tan (NSA*) / tan (135)]. We used two independent observers who repeated the analysis after a six-week interval. Radiological measurements were compared with 3D-CT measurements by an independent, blinded external institute.

We found a mean difference of 1.2° (sd 6.2) between radiological and 3D-CT measurements of stem version. The correlation between the mean radiological and 3D-CT stem torsion was r = 0.88 (p < 0.001). The intra- (intraclass correlation coefficient ≥ 0.94) and inter-observer agreement (mean concordance correlation coefficient = 0.87) for the radiological measurements were excellent.

We found that femoral tilt was associated with the mean radiological measurement error (r = 0.22, p = 0.02).

The projected neck–shaft angle is a reliable method for measuring stem version on AP radiographs of the hip after a THA. However, a highly standardised radiological technique is required for its precise measurement.

Cite this article: Bone Joint J 2015; 97-B:306–11.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLIV | Pages 9 - 9
1 Oct 2012
Haimerl M Dohmen L Gneiting S Sendtner E Wörner M Springorum R Grifka J Renkawitz T
Full Access

There is a complex interaction among acetabular component position and the orientation of the femoral component in determining the maximum, impingement-free prosthetic range of motion (ROM) in total hip arthroplasty (THA). Regarding restrictions in ROM, femoral antetorsion is one of the most important parameters. But, ROM is also influenced by parameters like the deviation between the femoral shaft and the mechanical axis in a sagittal projection. This deviation is best described as “Femoral Tilt” (FT). This study analysis the incidence of FT in clinical practice and its consequences on post-operative ROM. Based on these results, the effects of changes in FT on ROM-based cup optimisation are assessed by a using a virtual ROM analysis.

For studying the incidence of FT, 40 (16 male, 24 female) postoperative computerised tomography (CT) scans were analysed using a 3D CT planning software. The implant models were superimposed onto the image data to determine their exact position. The anatomical orientations were determined by planning anatomical landmarks and coordinate directions (i.e. mechanical axis, posterior condyle axis). Descriptive statistics were calculated for FT. Effects of changes in FT and CCD on ROM were analysed by calculating zones of compliance. FT was varied between 2.1° and 9.3° for 135°.

The overall range of post-operative values for femoral tilt was 5.7° ± 1.8° (mean ± standard deviation, minimum 1.7°, maximum 10.2°). The zone of compliance significantly depended on FT (difference more than 200%). The optimum cup position changed from 35° radiographic inclination/30° anteversion to 39°/30° when FT was increased from 2.1° to 9.3°.

Within this study, it was demonstrated that FT has a significant effect on postoperative ROM in THAs. First of all, it was shown that clinically FT values lie in a range between 2.1° and 9.3° (95% CI), where we used a long-shaft stem type with a relatively low possibility to influence sagittal tilt angles. FT may significantly change zones of compliance up to 200% as well as optimised cup positions. Thus, standard combined anteversion formulas, which were proposed in the literature to implement femur first approaches for THA, do only particularly address an optimisation of post-operative ROM. Instead, a sophisticated virtual ROM analysis based on a navigated femur-first approach would enable accurate ROM estimations as parameters like FT are hard to be assessed intra-operatively.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLIV | Pages 55 - 55
1 Oct 2012
Haimerl M Poitzsch L Gneiting S Schubert M Sendtner E Wörner M Springorum R Renkawitz T
Full Access

Incorrect restoration of leg length (LL) and offset is a major source of patient dissatisfaction and dysfunction after total hip arthroplasties (THAs). Evaluations on anterior-posterior x-ray images are state-of-the-art to assess the accuracy of such techniques. However, x-ray based measurements of LL and offset are challenging and limited in terms of accuracy. Within this study, we evaluated the accuracy of such measurements by analysing a series of clinical data. We evaluated the results on the non-treated side, since we know that there should be no significant difference between pre- and postoperative measurements on this side.

A series of 44 consecutive patients was analysed regarding changes in the difference between pre- and post-operative LL and offset measurements. Anterior-posterior x-rays were taken pre- (pre-OP) and post-operatively (post-OP) with a calibration by a scaling ruler (pre-OP) or implant size (post-OP). The LL and offset measurements were performed with a digital planning software based on the teardrop and transischial line. The distance between the teardrop/transischial line and the trochanter minor was measured to assess LL differences. Femoral offset (FO) was calculated as the orthogonal distance between the centre of the femoral head and the proximal shaft axis. Global offset (GO) was calculated as the distance between the inferior aspect of the teardrop figure and the shaft axis along the teardrop line. Descriptive statistics (mean value ± standard deviation) were calculated for the different types of measurements. Statistically significant differences were checked according to a student's t-test (α = 0.05).

The differences between the pre-and post-operative situation was 0.8±3.2 mm for LL, 0.2±3.5 mm for GO, and −0.5±2.5 mm for FO when referencing to the teardrop line and 0.9±4.0 mm (LL) and −0.3±2.7 mm (FO) for the transischial line. The error distributions did not show statistically significant differences when referencing to the teardrop or transischial line. But high differences (0.1±6.6 mm) were found when comparing the LL values (teardrop vs. transischial) case-by-case.

Within this study we demonstrated that x-ray based offset and LL measurements show reasonable inaccuracies. X-ray based evaluations of navigation-based techniques to assist LL and offset restoration cannot produce significantly better results than these analysed limits. That is, even if the navigation technique would be perfectly accurate, the evaluation would not achieve better accuracies than approximately ±3.5 mm for LL, ±3.5 mm for GO, and ±2.5 mm for FO.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLIV | Pages 32 - 32
1 Oct 2012
Dohmen L Haimerl M Gneiting S Schubert M Buchele C Sendtner E Wörner M Springorum R Renkawitz T
Full Access

Limited postoperative range-of-motion (ROM) can lead to patient dissatisfaction and dislocation in total hip arthroplasties (THAs). To avoid this, femur first approaches have been developed which optimise particular aspects of ROM by using a virtual analysis of ROM. This study analysis whether it is possible to accurately assess ROM based on an intra-operative acquisition of anatomical structures by using an image-free navigation system. It compares the outcome of a collision detection algorithm when using 3d models from computerised tomography (CT) scans on the one side and intra-operatively acquired 3D models on the other side within a cadaver study. It focuses on peri-acetabular impingements.

During the cadaver session 14 hips (7 cadavers) were treated surgically by using press-fit implants. 3D models of the pelvis and femora were generated based on segmented pre-operative CT data sets. Intra-operative data acquisition was performed by using a CT-free navigation software. Beside standard landmarks, points at the acetabular rim and femoral resection plane were acquired. For assessing ROM, a 3D model of the pelvis was generated. The information about the femoral resection plane was directly entered into the collision detection algorithm. 3D Computer Aided Design (CAD) models provided by the implant manufacturer were used for the implants. Based on this setup, the ROM values for flexion (FLEX), external rotation at 0° flexion (EXT), and internal rotation at 90° flexion (INTROT90) were compared. Differences within intended ROM were considered relevant, since the goal was to enable the prevention of clinically relevant ROM limitations.

The average difference between the CT based and navigation data based ROM analysis was 2.13° ± 3.11° for FLEX, 3.33° ± 5.51° for EXT, and 1.6° ± 3.66° INTROT90. The values reduce to 1.58° ± 2.78° (FLEX) and 0.91° ± 3.77° (INTROT90) when only ROM values within the intended ROM are considered. For EXT all ROM values lied above the threshold for intended ROM. Thus, no relevant differences were found for this motion direction.

In this study, a real-time collision detection based approach was developed and evaluated, which allows to virtually detect prosthetic and bony impingements. It was shown that ROM can be assessed accurately based on an image-free navigation technique. This information can be used intra-operatively to adjust the position of the implants and thus avoid postoperative ROM limitations. In particular, it enables a comprehensive femur first approach which allows us to optimise the post-operative results regarding functional parameters like ROM.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 502 - 502
1 Sep 2012
Renkawitz T Koller M Bonnlaender G Drescher J Riederer T Grifka J Schaumburger J Lechler P
Full Access

Background

For some time, optimized perioperative pathway protocols have been implemented in orthopedic surgery. In our hospital an accelerated clinical pathway has been successfully in effect for several years, focused on safely decreasing patients' length of stay and increasing their function at the time of discharge. The aim of the present project was to evaluate whether a further optimization is even more promising regarding early postoperative outcome parameters.

Materials and Methods

Prospective, parallel group design in an Orthopaedic University Medical Centre. 143 patients, scheduled for unilateral primary total knee replacement (TKR) under perioperative regional analgesia were included. 76 patients received a Standard Accelerated Clinical Pathway (SACP). 67 patients received an Optimized Accelerated Clinical Pathway (OACP) including patient-controlled regional analgesia pumps, ultra-early/doubled physiotherapy and motor driven continuous passive motion machine units. Main measures were early postoperative pain on a visual analogue scale, consumption of regional anaesthetics, knee range of motion, time out of bed, walking distance/stair climbing, circumference measurements and Knee Society Scores of the operated leg. Patients in both groups were checked for a possible discharge by a blinded orthopedic surgeon on the 5th and 8th postoperative (po) day, using a discharge checklist including the KATZ Index of Independence in Activities of Daily Living, standard requirements for pain at rest/mobilization, walking distance and regular wound healing. A potential discharge was only approved if the patient was able to meet all six criteria from the discharge checklist. Re-admission within 6 weeks after discharge from hospital was registered.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1021 - 1026
1 Aug 2011
Kalteis T Sendtner E Beverland D Archbold PA Hube R Schuster T Renkawitz T Grifka J

Orientation of the native acetabular plane as defined by the transverse acetabular ligament (TAL) and the posterior labrum was measured intra-operatively using computer-assisted navigation in 39 hips. In order to assess the influence of alignment on impingement, the range of movement was calculated for that defined by the TAL and the posterior labrum and compared with a standard acetabular component position (abduction 45°/anteversion 15°).

With respect to the registration of the plane defined by the TAL and the posterior labrum, there was moderate interobserver agreement (r = 0.64, p < 0.001) and intra-observer reproducibility (r = 0.73, p < 0.001). The mean acetabular component orientation achieved was abduction of 41° (32° to 51°) and anteversion of 18° (−1° to 36°). With respect to the Lewinnek safe zone (abduction 40° ±10°, anteversion 15° ±10°), 35 of the 39 acetabular components were within this zone. However, there was no improvement in the range of movement (p = 0.94) and no significant difference in impingement (p = 0.085).

Alignment of the acetabular component with the TAL and the posterior labrum might reduce the variability of acetabular component placement in total hip replacement. However, there is only a moderate interobserver agreement and intra-observer reliability in the alignment of the acetabular component using the TAL and the posterior labrum. No reduction in impingement was found when the acetabular component was aligned with the TAL and the posterior labrum, compared with a standard acetabular component position.