header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

ASSESSMENT OF POSTOPERATIVE RANGE-OF-MOTION IN TOTAL HIP ARTHROPLASTIES BY MEANS OF IMAGE FREE NAVIGATION

The International Society for Computer Assisted Orthopaedic Surgery (CAOS)



Abstract

Limited postoperative range-of-motion (ROM) can lead to patient dissatisfaction and dislocation in total hip arthroplasties (THAs). To avoid this, femur first approaches have been developed which optimise particular aspects of ROM by using a virtual analysis of ROM. This study analysis whether it is possible to accurately assess ROM based on an intra-operative acquisition of anatomical structures by using an image-free navigation system. It compares the outcome of a collision detection algorithm when using 3d models from computerised tomography (CT) scans on the one side and intra-operatively acquired 3D models on the other side within a cadaver study. It focuses on peri-acetabular impingements.

During the cadaver session 14 hips (7 cadavers) were treated surgically by using press-fit implants. 3D models of the pelvis and femora were generated based on segmented pre-operative CT data sets. Intra-operative data acquisition was performed by using a CT-free navigation software. Beside standard landmarks, points at the acetabular rim and femoral resection plane were acquired. For assessing ROM, a 3D model of the pelvis was generated. The information about the femoral resection plane was directly entered into the collision detection algorithm. 3D Computer Aided Design (CAD) models provided by the implant manufacturer were used for the implants. Based on this setup, the ROM values for flexion (FLEX), external rotation at 0° flexion (EXT), and internal rotation at 90° flexion (INTROT90) were compared. Differences within intended ROM were considered relevant, since the goal was to enable the prevention of clinically relevant ROM limitations.

The average difference between the CT based and navigation data based ROM analysis was 2.13° ± 3.11° for FLEX, 3.33° ± 5.51° for EXT, and 1.6° ± 3.66° INTROT90. The values reduce to 1.58° ± 2.78° (FLEX) and 0.91° ± 3.77° (INTROT90) when only ROM values within the intended ROM are considered. For EXT all ROM values lied above the threshold for intended ROM. Thus, no relevant differences were found for this motion direction.

In this study, a real-time collision detection based approach was developed and evaluated, which allows to virtually detect prosthetic and bony impingements. It was shown that ROM can be assessed accurately based on an image-free navigation technique. This information can be used intra-operatively to adjust the position of the implants and thus avoid postoperative ROM limitations. In particular, it enables a comprehensive femur first approach which allows us to optimise the post-operative results regarding functional parameters like ROM.