Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 17 - 17
2 May 2024
Whitehouse M Patel R French J Beswick A Navvuga P Marques E Blom A Lenguerrand E
Full Access

Hip bearing surfaces materials are typically broadly reported in national registry (metal-on-polyethylene, ceramic-on-ceramic etc). We investigated the revision rates of primary total hip replacement (THR) reported in the National Joint Registry (NJR) by detailed types of bearing surfaces used.

We analysed THR procedures across all orthopaedic units in England and Wales. Our analyses estimated all-cause and cause-specific revision rates. We identified primary THRs with heads and monobloc cups or modular acetabular component THRs with detailed head and shell/liner bearing material combinations. We used flexible parametric survival models to estimate adjusted hazard ratios (HR).

A total of 1,026,481 primary THRs performed between 2003–2019 were included in the primary analysis (Monobloc cups: n=378,979 and Modular cups: n=647,502) with 20,869 (2%) of these primary THRs subsequently undergoing a revision episode (Monobloc: n=7,381 and Modular: n=13,488).

Compared to implants with a cobalt chrome head and highly crosslinked polyethylene (HCLPE) cup, the overall risk of revision for monobloc acetabular implant was higher for patients with cobalt chrome or stainless steel head and non-HCLPE cup. The risk of revision was lower for patients with a delta ceramic head and HCLPE cup implant, at any post-operative period.

Compared to patients with a cobalt chrome head and HCLPE liner primary THR, the overall risk of revision for modular acetabular implant varied non-constantly. THRs with a delta ceramic or oxidised zirconium head and HCLPE liner had a lower risk of revision throughout the entire post-operative period.

The overall and indication-specific risk of prosthesis revision, at different time points following the initial implantation, is reduced for implants with a delta ceramic or oxidised zirconium head and a HCLPE liner/cup in reference to THRs with a cobalt chrome head and HCLPE liner/cup.


Implants in total hip replacement (THR) are associated with different clinical and cost-effectiveness profiles,. We estimate the costs and outcomes for NHS patients in the year after THR associated with implant bearing materials using linked routinely collected data.

We linked NJR primary elective THR patients for osteoarthritis to HES and National PROMs. We estimated health care costs, health-related quality of life indices, and revision risks, in the year after primary and revision THRs overall. We used generalised linear models adjusting for patient and hospital characteristics and estimated 10-year cumulative probability of revision. We imputed utilities using chained equations for half the sample with missing PROMS.

We linked 577,973 elective primary THRs and 11,812 subsequent revisions. One year after primary THR, patients with the cemented THRs using cobalt chrome or stainless steel head with HCLPE liner/cup cost the NHS, on average, £13,101 (95%CI £13,080,£13,122), had an average quality-of-life score of 0.788 (95%CI 0.787,0.788), and a 10-year revision probability of 1.9% (95%CI 1.6,2.3). Compared to the reference, patients receiving a cemented THR with delta ceramic head and HCLPE liner/cup, hybrid THR with delta ceramic head and HCLPE liner/cup, and hybrid THR with alumina head and HCLPE liner/cup had lower 1-year costs (-£572 \[95% CI -£775,-£385\], -£346 \[-£501,-£192\], -£371 \[-£574,-£168\] respectively), better quality of life (0.007 \[95% CI 0.003,0.011\], 0.013 \[0.010,0.016\], 0.009 \[0.005,0.013\] respectively), and lower 10-year revision probabilities (1.4% \[1.03,2.0\], 1.5 \[1.3,1.7\], 1.6%\[1.2,2.1\] respectively).

Implant bearing materials are associated with varying mean costs and health outcomes after primary THR. Ours is the first study to derive costs and health outcomes from large, linked databases using multiple imputation methods to deal with bias. Our findings are useful for commissioning and procurement decisions and to inform a subsequent cost-effectiveness model with more granular detail on THR implant types.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 48 - 48
23 Feb 2023
Patel R Elliott R
Full Access

Regional anaesthesia is integral to best practice analgesia for patients with neck of femur fractures (NOFFs). These patients are generally frail and are vulnerable to side effects of opioid analgesia. Femoral nerve block (FNB) or fascia-iliaca block (FIB) can reduce opioid requirement. Literature supports good efficacy for extra-capsular NOFFs however it is acknowledged to be suboptimal for intracapsular fractures.

We present a novel technique, using point of care ultrasound guidance to perform hip ultrasound guided haematoma (HUSH) aspiration, and injection of local anaesthetic (block) for intracapsular NOFFs.

This a case control series. A consecutive series of cognitively intact patients, with an isolated intra-capsular NOFF, received a HUSH block using 10mls of 0.75% Ropivicaine. Haematoma was aspirated and volume recorded. This was performed in addition to standard NOFF pathway analgesia that includes a FIB and multimodal analgesia including opioids. Visual Analogue Scale (VAS)pain scores at rest and on movement were recorded pre and post procedure as well as combined morphine equivalent units administered post HUSH block. The control arm was a retrospective group of similar patients who followed the routine care pathway including a FIB. VAS pain scores from observation charts and usage of morphine equivalent units were calculated.

Ten patients consented to receive HUSH blocks and we included thirty-eight patients in our control series. The HUSH block group showed mean VAS pain score of 4.2/10 at rest and 8.6 on movement prior to block. In the time after the block, VAS pain scores reduced to 1.5 at rest (p=0.007) and 3.1 on movement (p=0.0001) with a mean total morphine equivalent use of 8.75mg. This is significantly different from the control group's mean VAS pain at rest score 6.9 (p=0.0001) and 24.1mg total morphine equivalent (p=0.07).

HUSH Block in addition to fascia iliaca block appears to significantly better pain relief in intracapsular neck of femur fracture patients when compared to fascia iliaca block alone. We believe it is relatively easy to perform with readily available ultrasound scanners in emergency departments.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_11 | Pages 1 - 1
1 Nov 2022
Patel R
Full Access

Abstract

Aims

The aim of this study was to evaluate the indications for patients presenting with knee pain undergoing magnetic resonance imaging of the knee prior to referral to the orthopaedic department and to ascertain whether plain radiograph imagining would be more beneficial prior to an MRI scan.

Method

A retrospective review of all referrals received by the hospital over a 6-month period was performed. Patients with knee pain that underwent an MRI scan were classified into two age groups, under 50 years and over 50 years old. Patients having undergone Magnetic resonance imaging (MRI) prior to referral were identified, and findings of the scan were recorded. These patients were reviewed further to see if a plain radiograph had been completed prior to or after the MRI.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 53 - 58
1 Jan 2022
Tai DBG Wengenack NL Patel R Berbari EF Abdel MP Tande AJ

Aims

Fungal and mycobacterial periprosthetic joint infections (PJI) are rare events. Clinicians are wary of missing these diagnoses, often leading to the routine ordering of fungal and mycobacterial cultures on periprosthetic specimens. Our goal was to examine the utility of these cultures and explore a modern bacterial culture technique using bacterial blood culture bottles (BCBs) as an alternative.

Methods

We performed a retrospective review of patients diagnosed with hip or knee PJI between 1 January 2010 and 31 December 2019, at the Mayo Clinic in Rochester, Minnesota, USA. We included patients aged 18 years or older who had fungal, mycobacterial, or both cultures performed together with bacterial cultures. Cases with positive fungal or mycobacterial cultures were reviewed using the electronic medical record to classify the microbiological findings as representing true infection or not.


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1119 - 1126
1 Jun 2021
Ivy MI Sharma K Greenwood-Quaintance KE Tande AJ Osmon DR Berbari EF Mandrekar J Beauchamp CP Hanssen AD Abdel MP Lewallen DG Perry K Block DR Snyder MR Patel R

Aims

The aim of this study was to determine the diagnostic accuracy of α defensin (AD) lateral flow assay (LFA) and enzyme-linked immunosorbent assay (ELISA) tests for periprosthetic joint infection (PJI) in comparison to conventional synovial white blood cell (WBC) count and polymorphonuclear neutrophil percentage (PMN%) analysis.

Methods

Patients undergoing joint aspiration for evaluation of pain after total knee arthroplasty (TKA) or total hip arthroplasty (THA) were considered for inclusion. Synovial fluids from 99 patients (25 THA and 74 TKA) were analyzed by WBC count and PMN% analysis, AD LFA, and AD ELISA. WBC and PMN% cutoffs of ≥ 1,700 cells/mm3 and ≥ 65% for TKA and ≥ 3,000 cells/mm3 and ≥ 80% for THA were used, respectively. A panel of three physicians, all with expertise in orthopaedic infections and who were blinded to the results of AD tests, independently reviewed patient data to diagnose subjects as with or without PJI. Consensus PJI classification was used as the reference standard to evaluate test performances. Results were compared using McNemar’s test and area under the receiver operating characteristic curve (AUC) analysis.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 345 - 351
1 Mar 2020
Pitts C Alexander B Washington J Barranco H Patel R McGwin G Shah AB

Aims

Tibiotalocalcaneal (TTC) fusion is used to treat a variety of conditions affecting the ankle and subtalar joint, including osteoarthritis (OA), Charcot arthropathy, avascular necrosis (AVN) of the talus, failed total ankle arthroplasty, and severe deformity. The prevalence of postoperative complications remains high due to the complexity of hindfoot disease seen in these patients. The aim of this study was to analyze the relationship between preoperative conditions and postoperative complications in order to predict the outcome following primary TTC fusion.

Methods

We retrospectively reviewed the medical records of 101 patients who underwent TTC fusion at the same institution between 2011 and 2019. Risk ratios (RRs) associated with age, sex, diabetes, cardiovascular disease, smoking, preoperative ankle deformity, and the use of bone graft during surgery were related to the postoperative complications. We determined from these data which pre- and perioperative factors significantly affected the outcome.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 11 - 11
1 Oct 2019
Wyles CC Paradise CR Masters TL Patel R Abdel MP Trousdale RT Sierra RJ
Full Access

Introduction

Adverse local tissue reactions (ALTR) can result in devastating soft tissue and osseous destruction, while potentially increasing the risk of concomitant periprosthetic joint infection (PJI). The aims of this study were to evaluate cobalt (Co) and chromium (Cr) levels generated in simulators from metal-on-polyethylene (MoP) and ceramic-on-polyethylene (CoP) constructs, and determine their impact on native tissues and PJI risk through evaluation of human adipose-derived mesenchymal stem cells (AMSCs) and Staphylococcus epidermidis isolates.

Methods

Ten hip simulator constructs were assembled with 36-mm high-offset femoral heads, highly cross-linked polyethylene liners, and titanium stems. Five constructs used CoCr femoral heads and five used ceramic. Constructs were submerged in bovine serum (BS) and run for 1,000,000 cycles. Samples of BS were collected and evaluated for CoCr concentration. Various concentrations of CoCr were chosen for further assessment of cytotoxicity and growth impact on AMSCs and S. epidermidis and compared to inert SiO2.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 58 - 58
1 Mar 2017
Noble P Patel R Jones H Kim R Gold J Ismaily S
Full Access

INTRODUCTION

Stable fixation of cementless tibial trays remains a challenge due bone density variability within the proximal tibia and the spectrum of loads imposed by different activities. This study presents a novel approach to measuring the interface motion of cementless tibial components during functional loading and tests whether interface motion of cementless tibial trays varies around the implant periphery.

METHODS

We developed a method to measure relative displacement of a tibial tray relative to the underlying bone using 3D digital image correlation (DIC) and multi-camera stereo photogrammetry. A clinically successful design of cementless total knee prosthesis (Zimmer Inc, Warsaw, IN) was implanted in 6 fresh cadaveric knees. A black-on-white stochastic pattern was applied to the outer surface of the tibia and the cementless prosthesis. High resolution digital images were prepared of the interface region and divided into 25 × 25 pixel regions of interest (ROI). Stereo images of the same ROI were generated using two cameras angled at 60 degrees using image correlation techniques. All specimens were mounted in a custom-built functional activity simulator and loaded with the forces and moments recorded during three common functional activities (standing from a seated position, walking, and stair descent), as reported in the Orthoload database, scaled by 50% for application to cadaveric bone. Prior to functional testing, each implant-tibia construct was preconditioned with 500 cycles of flexion from 5–100 degrees under a vertical tibial load of 1050 N at a frequency of 0.2 Hz. During loading, image data was acquired simultaneously (±20 μs) from the entire circumference of the tibial interface forming 4 stereo images using 8 cameras spaced at 90 degree intervals (Allied Vision Technologies, Exton, PA) using custom image acquisition software (Mathworks, Natick, MA) (Figure 1). The multiple stereo images were registered using the surface topography of each specimen as measured by laser scanning (FARO Inc., Montreal) (Figure 2). During post-processing, the circumferential tray/tibia interface was divided into 10 zones for subsequent analysis (Figure 3). Interface displacements were measured on a point-to-point basis at approximately 700 sites on each specimen using commercial DIC software (Dantec Dynamics, Skovlunde, Denmark) (Figure 4).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 57 - 57
1 Mar 2017
Noble P Gold J Patel R Lenherr C Jones H Ismaily S Alexander J
Full Access

INTRODUCTION

Cementless tibial trays commonly fail through failure of fixation due to excessive interface motion. However, the specific combination of axial and shear forces precipitating implant failure is unknown. This has led to generic loading profiles approximating walking to perform pre-clinical assessment of new designs, even though telemetric data demonstrates that much larger forces and moments are generated during other functional activities. This study was undertaken to test the hypotheses: (i) interface motion of cementless tibial trays varies as a function of specific activities, and (ii) the response of the cementless tibial interface to walking loading is not representative of other functional activities.

MATERIALS and METHODS

Six fresh-frozen cadaveric tibias were tested using a custom designed functional activity simulator after implantation of a posterior stabilized total knee replacement (NexGen LPS, Zimmer, Warsaw IN). Activity scenarios were selected using force (Fx, Fy, Fz) and moment (Mx, My, Mz) data from patients with instrumented tibial trays (E-tibia) published by Bergmann et al. A pattern of black and white spray paint was applied to the surface of the specimen including the tibial tray and bone. Each specimen was preconditioned through application of a vertical load of 1050N for 500 cycles of flexion-extension from 5–100°. Following preconditioning, each tibia was loaded using e-tibia values of forces and moments for walking, stair-descent, and sit-to-stand activities. The differential motion of the tibial tray and the adjacent bony surface was monitored using digital image correlation (DIC) (resolution: 1–2 microns in plane; 3–4 microns out-of-plane). Four pairs of stereo-images of the tray and tibial bone were prepared at sites around the circumference of the construct in both the loaded and unloaded conditions: (i) before and after pre-conditioning and (ii) before and after the 6 functional loading profiles. The images were processed to provide circumferential measurements of interface motion during loading. Differences in micromotion and migration were evaluated statistically using step-wise multivariate regression.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 135 - 135
1 Feb 2017
Varadarajan KM Patel R Zumbrunn T Rubash H Malchau H Freiberg A Muratoglu O
Full Access

Introduction

Dual-mobility (DM) liners provide increased range of motion and stability. However, large head diameters have been associated with anterior hip pain due to impingement with surrounding soft-tissues, particularly the iliopsoas. Further, during hip extension the liner can get trapped due to anterior soft-tissue impingement that resists rotation being imparted to the liner from posterior stem-liner contact. Over time this can cause liner rim damage, leading to intra-prosthetic dislocation of the small diameter inner head. To address this, an anatomically contoured dual mobility (ACDM) liner was designed to reduce the volume of the liner below the equator that can interact with soft-tissues (Fig. 1). In this study, we utilized finite element analysis to evaluate tendon-liner contact pressure and tendon stresses with ACDM and conventional designs during hip extension, wherein the posterior edge of liner is in contact with the stem while the anterior edge is exposed to the soft-tissue.

Methods

The average uniaxial stiffness (350 N/mm), and average dimensions (width × thickness = 14mm × 4mm) of 10 cadaver psoas tendon samples were determined in a separate study. The iliopsoas tendon was modelled as a Yeoh hyper-elastic material, and the material constants were tuned to match the experimental uniaxial test data. Cadaver specific FEA models were created for 5 specimens (10 hips) using computed tomography (CT) scans. The implant components were modeled as being rigid relative to the iliopsoas tendon. The iliopsoas tendon was modelled as extending from its insertion point on the lesser trochanter to the psoas notch on the pelvis for hip flexion angles of −15°, 0°, 15° and 30°. Appropriately sized DM components were implanted virtually for each specimen. Once placed in its proper position, the liner was rotated about the flexion axis until it contacted the stem posteriorly to represent its orientation during hip extension (Fig. 2). A 500N tensile load was applied to the iliopsoas tendon and the average/max stresses within the tendon, and average/max contact pressures between the tendon and liner were measured.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 15 - 15
1 May 2016
Varadarajan K Zumbrunn T Duffy M Patel R Freiberg A Malchau H Rubash H Muratoglu O
Full Access

Introduction

Dual Mobility (DM) implants have gained popularity for the treatment and prevention of hip dislocation, with increased stability provided by a large diameter mobile liner. However, distal regions of the liner can impinge on soft-tissues like hip capsule and iliopsoas, leading to anterior hip pain. Additionally, soft-tissue impingement may trap the mobile liner, leading to excessive loading of the liner rim, from engagement with the femoral stem, and subsequent intra-prosthetic dislocation. The hypothesis of this study was that reducing the liner profile below the equator (contoured design) can mitigate soft-tissue impingement without compromising inner-head pull-out resistance and overall hip joint stability (Fig. 1).

Methods

The interaction of conventional and contoured liners with anterior soft-tissues was evaluated in 10 cadaveric hips (5 specimens; 2 male, 3 female; age 65 ± 10 yrs; liner diameter 42–48mm) via visual observation and fluoroscopic imaging. A metal wire was sutured to the deep fibers of the iliopsoas tendon/muscle, and metal wires were embedded in the mobile liners for fluoroscopic visualization (Fig. 2). All soft-tissue except the anterior hip capsule and iliopsoas was removed, and a rope was attached to the iliopsoas to apply tension along its natural orientation.

Resistance to inner-head pull-out was evaluated via Finite Element Analysis (FEA) by simulating a full cycle of insertion of the inner head into the mobile liner and subsequent pullout. The femoral head, acetabular shell, and stem were modeled as rigid, while the mobile liner was modeled as plastically deformable. Hip joint stability was evaluated by dynamic simulations in for two dislocation modes: (A) Posterior dislocation (at 90° hip flexion) with internal hip rotation; (B) Posterior dislocation (starting at 90° flexion) with combined hip flexion and adduction. A 44 mm diameter conventional and a 44 mm contoured liner were evaluated during these tests.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 122 - 122
1 May 2016
Patel R Zumbrunn T Varadarajan K Freiberg A Rubash H Muratoglu O Malchau H
Full Access

Introduction

Dual-mobility (DM) liners have increased popularity due to the range of motion and stability provided by these implants. However, larger head diameters have been associated with anterior hip pain, due to surrounding soft-tissue impingement, particularly the iliopsoas. To address this, an anatomically contoured dual mobility (ACDM) liner was designed by reducing the volume of the liner below the equator (Fig1). Previous cadaver studies have shown that the ACDM significantly reduces iliopsoas tenting and trapping of the liner compared to conventional designs. We created a finite element study based on previous cadaver testing to further analyze the effectiveness of the ACDM design in reducing soft-tissue impingement, specifically the tendon-liner contact pressure and the tendon stress.

Methods

The finite element model was developed within COMSOL 4.3b. The psoas tendon was modelled as a Yeoh hyper-elastic Material, which uses 3 constants (c1-c3), density (1.73g/cm3) and a bulk modulus (26GPa)[Hirokawa,2000]. In a previous, separate study, the average stiffness of 10 psoas tendon samples (5 cadavers), were measured to be 339[N/mm] in the linear region with average width and thickness of 14mmX4mm. The 3 constants were tuned to match experimental uniaxial test data, and were 5[GPa], 0[Gpa], and 46[GPa] for c1, c2, and c3 respectively.

The implant components were rigidly modeled relative to the psoas. Cadaver specific CT models were used to create the FEA geometry. The insertion points for the Psoas were digitally determined on the proximal end of the lesser trochanter, and the psoas notch on the pelvis for hip flexion angles of −15°, 0°, 15° and 30°. These insertion points determined the length of the psoas and its relative position to the femoral head in 3D. The specific liner size and position for each cadaver was determined by implant planning with the CT models. In this abstract, we only present data for 2 specimens (left/right hips) with 44mm conventional DM, and 44mm ACDM, matching specimen anatomy. A 500N tensile load was applied to the psoas tendon proximally to simulate moderate physiological loading, the average/max stresses and contact pressures between the psoas and the two liner designs were determined.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 87 - 87
1 May 2016
Saied F Patel R Ismaily S Harrington M Landon G Parsley B Noble P
Full Access

Summary

There is tremendous variability amongst surgeons' ability to reference anatomic landmarks. This may suggest the necessity of other objective methods in determining femoral alignment and rotation.

Introduction

Despite the durability of total knee arthroplasty, there is much room for improvement with regards to functional outcome and patient satisfaction. One important factor contributing to poor outcomes after TKA is malrotation of the femoral component. It has been postulated that this is due to failure of surgeons to correctly reference bony landmarks, principally the femoral epicondyles, however, this is unproven. The purpose of this study was to evaluate the accuracy of joint surgeons and trainees in identifying anatomic landmarks for positioning the femoral component and to determine the effect of prior training and experience.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 91 - 91
1 Jan 2016
Derasari A Gold J Alexander J Kim SW Patel R Parekh J Incavo S Noble P
Full Access

Introduction

Mechanically-assisted corrosion of the head-neck junction present a dilemma to surgeons at revision THR whenever the femoral component is rigidly fixed to the femur. Many remove the damaged femoral head, clean the femoral taper and fix a new head in place to spare the patient the risks associated with extraction and replacement of the well-functioning femoral stem. This study was performed to answer these research questions:

Will new metal heads restore the mechanical integrity of the original modular junction after impaction on corroded tapers?

Which variables affect the stability of the new interface created at revision THR?

Materials and Methods

Twenty-two tapers (CoCr, n=12; TiAlV, n=10) were obtained for use in this study. Ten stems were in pristine condition, while 12 stems had been retrieved at revision THR and with corrosion damage to the trunnion (Goldberg scale 4). Twenty-two new metal heads were obtained for use in the study, each matching the taper and manufacturer of the original component. The following test states were performed using a MTS Machine: 1. Assembly, 2. Disassembly, 3. Assembly, 4. Toggling and 5. Disassembly. All head assemblies were performed wet using 50% calf serum in accordance to ISO 7206-10. During toggling, each specimen's loading axis was aligned 25° to the trunnion axis in the frontal plane and 10° in the sagittal plane (Figure 1). Toggling was performed at 1Hz for 2,000 cycles with a sinusoidal loading function (230N–4300N). During loading, 3D motion of the head-trunnion junction was measured using a custom jig rigidly attached to the head and the neck of each prosthesis. Relative displacement of the head with respect to the neck was continuously monitored using 6 high resolution displacement transducers with an accuracy of ±0.6µm. Displacement data was independently validated using FEA models of selected constructs.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_5 | Pages 18 - 18
1 Mar 2014
Al-hadithy N Furness N Patel R Crockett M Anduvan A Jobbaggy A Woods D
Full Access

Cementless surface replacement arthroplasty (CSRA) is an established treatment for glenohumeral osteoarthritis. Few studies however, evaluate its role in cuff tear arthopathy. The purpose of this study is to compare the outcomes of CSRA for both glenohumeral osteoarthritis and cuff tear arthopathy.

42 CSRA with the Mark IV Copeland prosthesis were performed for glenohumeral osteoarthritis (n=21) or cuff tear arthopathy (n=21). Patients were assessed with Oxford and Constant scores, patient satisfaction, range of motion and radiologically with plain radiographs.

Mean follow-up and age was 5.2 years and 74 years in both groups. Functional outcomes were significantly higher in OA compared with CTA with OSS improving from 18 to 37.5 and 15 to 26 in both groups respectively. Forward flexion improved from 60° to 126° and 42° to 74° in both groups. Three patients in the CTA group had a deficient subscapularis tendon, two of whom dislocated anteriorly.

Humeral head resurfacing arthroplasty is a viable treatment option for glenohumeral osteoarthritis. In patients with CTA, functional gains are limited. We suggest CSRA should be considered in low demand patients where pain is the primary problem. Caution should be taken in patients with a deficient subscapularis due to the high risk of dislocation.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 456 - 456
1 Dec 2013
Borque K Gold J Ismaily S Patel R Incavo S Noble P
Full Access

Purpose:

Knee pain and instability during high demand activities such as stair descent are reported by patients after TKA. Previous studies theorized that this pain is from increased demand on the quadriceps required to stabilize the femur on the tibia. In this study we explore the relationship between implant design, the posterior cruciate ligament (PCL), and AP stability of the knee during stair descent.

Methods:

CTs of 6 fresh-frozen human cadaveric knees (average age: 61 ± 6.5 years) with functioning cruciates were prepared. All specimens were mounted in a computer controlled, 6 DOF simulator programed to apply physiologic muscle loads and flexion/extension moments simulating the highest demand phase of stair descent (terminal swing to initial contact). A contemporary design of TKA was implanted in each specimen by an experienced surgeon. Testing was repeated after implantation of tibial inserts of the CR, CS with and without a PCL and PS designs.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 84 - 84
1 Dec 2013
Ismaily S Patel R Suarez A Incavo S Bolognesi MP Noble P
Full Access

Introduction

Malpositioning of the tibial component is a common error in TKR. In theory, placement of the tibial tray could be improved by optimization of its design to more closely match anatomic features of the proximal tibia with the motion axis of the knee joint. However, the inherent variability of tibial anatomy and the size increments required for a non-custom implant system may lead to minimal benefit, despite the increased cost and size of inventory.

This study was undertaken to test the hypotheses:

That correct placement of the tibial component is influenced by the design of the implant.

The operative experience of the surgeon influences the likelihood of correct placement of contemporary designs of tibial trays.

Materials and Methods

CAD models were generated of all sizes of 7 widely used designs of tibial trays, including symmetric (4) and asymmetric (3) designs. Solid models of 10 tibias were selected from a large anatomic collection and verified to ensure that they encompassed the anatomic range of shapes and sizes of Caucasian tibias. Each computer model was resected perpendicular to the canal axis with a posterior slope of 5 degrees at a depth of 5 mm distal to the medial plateau. Fifteen joint surgeons and fourteen experienced trainees individually determined the ideal size and placement of each tray on each resected tibia, corresponding to a total of 2030 implantations. For each implantation we calculated: (i) the rotational alignment of the tray; (ii) its coverage of the resected bony surface, and (iii) the extent of any overhang of the tray beyond the cortical boundary. Differences in the parameters defining the implantations of the surgeons and trainees were evaluated statistically.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 138 - 138
1 Dec 2013
Noble P Patel R Ashfaq K Bernstein D Ismaily S Incavo S
Full Access

Introduction

After TKR, excessive tension within the lateral retinaculum can lead to joint instability, component wear, stiffness and pain. The spatial distribution of strain in the lateral retinculum is unknown, both in the native knee and after TKR. In this study we measure the magnitude and distribution of mechanical strain in the lateral retinaculum with knee flexion, both in the native knee and after TKR. We hypothesize that:

Strain in the lateral retinaculum will increase as a function of flexion.

Some regions of the lateral retinaculum experience greater strain than others.

TKR will affect the magnitude and location of strain during knee flexion.

Materials and Methods

A fiduciary grid of approximately 40–70 markers was attached to the exposed lateral retinacula of five fresh frozen cadaveric knees in order to allow tracking of soft-tissue deformation. Each knee was flexed from 0–120° in a 6 degree-of-freedom custom activity simulator that physiologically loaded the knee during a squatting maneuver. During simulation, the displacement of each fiduciary point was measured using visible-light stereo-photogrammetry. The fiduciary grid divided into four distinct regions for strain analysis. Using the grid of the native knee in full extension as the initial state, the average principal strain in each region was calculated as a function of flexion. Measurements were repeated after TKR was performed using a contemporary implant system.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 148 - 148
1 Dec 2013
Incavo S Noble P Gold KBJ Patel R Ismaily S
Full Access

Introduction

Increasing attention to the functional outcome of total knee arthroplasty (TKA) has demonstrated that many patients experience limitations when attempting to perform demanding activities that are normal for age-matched peers, primarily because of knee symptoms. Episodes of instability following TKA are most commonly reported during activities in which significant transverse or torsional forces are supported by the joint with relatively low joint compression forces, including stair-descent and walking on sloped or uneven surfaces. This study was performed to examine the influence of conformity between the femoral and tibial components on the Antero-Posterior (AP) stability of knee during stair descent.

Methods

Six cadaveric knees were loaded in a six degree-of-freedom joint simulator, with the application of external forces simulating the action of the quadriceps and hamstring muscles and the external loads and moments occurring during stair descent, including the stages of terminal swing phase, weight-acceptance phase (prior to and after quadriceps contraction) and mid-stance. During these manoeuvres, the displacement and rotation of the femur and the tibia were measured with a multi-camera high resolution motion analysis system (Fig. 1). Each knee was tested in the intact and ACL deficient condition – and after implantation of total knee prosthesis with Cruciate-Retaining (CR), Cruciate-Sacrificing with an intact PCL (CS + PCL), Cruciate-Sacrificing with an absent PCL (CS-PCL) and Posterior-Stabilizing (PS) tibial inserts (Figs 2 and 3).