header advert
Results 1 - 9 of 9
Results per page:
Bone & Joint Research
Vol. 10, Issue 12 | Pages 797 - 806
8 Dec 2021
Chevalier Y Matsuura M Krüger S Traxler H Fleege† C Rauschmann M Schilling C

Aims

Anchorage of pedicle screw rod instrumentation in the elderly spine with poor bone quality remains challenging. Our study aims to evaluate how the screw bone anchorage is affected by screw design, bone quality, loading conditions, and cementing techniques.

Methods

Micro-finite element (µFE) models were created from micro-CT (μCT) scans of vertebrae implanted with two types of pedicle screws (L: Ennovate and R: S4). Simulations were conducted for a 10 mm radius region of interest (ROI) around each screw and for a full vertebra (FV) where different cementing scenarios were simulated around the screw tips. Stiffness was calculated in pull-out and anterior bending loads.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 12 - 12
1 Dec 2021
Samsami S Pätzold R Winkler M Herrmann S Müller PE Chevalier Y Augat P
Full Access

Abstract

Objective

Bi-condylar tibia plateau fractures are one of challenging injuries due to multi-planar fracture lines. The risk of fixation failure is correlated with coronal splits observed in CT images, although established fracture classifications and previous studies disregarded this critical split. This study aimed to experimentally and numerically compare our innovative fracture model (Fracture C), developed based on clinically-observed morphology, with the traditional Horwitz model (Fracture H).

Methods

Fractures C and H were realized using six samples of 4th generation tibia Sawbones and fixed with Stryker AxSOS locking plates. Loading was introduced through unilateral knee replacements and distributed 60% medially. Loading was initiated with six static ramps to 250 N and continued with incremental fatigue tests until failure. Corresponding FE models of Fractures C and H were developed in ANSYS using CT scans of Sawbones and CAD data of implants. Loading and boundary conditions similar to experimental situations were applied. All materials were assumed to be homogenous, isotropic, and linear elastic. Von-Mises stresses of implant components were compared between fractures.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 72 - 72
1 Apr 2018
Santos I Mahmoud M Thorwächter C Bourgeois A Müller P Pietschmann M Chevalier Y
Full Access

Background

While total shoulder arthroplasty (TSA) is a generally successful procedure, glenoid loosening remains a common complication. Though the occurrence of loosening was related to patient-specific factors, biomechanical factors related to implant features may also affect the fixation of the glenoid component, in particular increased glenohumeral mismatch that could result in eccentric loads and translations. In this study, a novel test setup was used to quantify glenohumeral pressures for different motion patterns after TSA.

Methods

Six cadaveric human shoulders were implanted with total shoulder replacements (Exactech, Inc., USA) and subjected to cyclic internal-external, flexion-extension and abduction-adduction rotations in a passive motion testing apparatus. The system was coupled to a pressure sensor system (Tekscan, Inc., USA) to acquire joint loads and to a Zebris system (Zebris Medical, GmbH, Germany) to measure joint kinematics. The specimens were subjected to a total of 2160 cycles and peak pressures were compared for each motion pattern.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 75 - 75
1 Apr 2018
Matsuura M Schmidutz F Sprecher C Müller P Chevalier Y
Full Access

Introduction

Stemless shoulder implants have recently gained increasing popularity. Advantages include an anatomic reconstruction of the humerus with preservation of bone stock for upcoming revisions. Several implant designs have been introduced over the last years. However, only few studies evaluated the impact of the varying designs on the load transfer and bone remodeling. The aim of this study was to compare the differences between two stemless shoulder implant designs using the micro finite element (µFE) method.

Materials and Methods

Two cadaveric human humeri (low and high bone mineral density) were scanned with a resolution of 82µm by high resolution peripheral quantitative computer tomography (HR-pQCT). Images were processed to allow virtual implantation of two types of reverse-engineered stemless humeral implants (Implant 1: Eclipse, Arthrex, with fenestrated cage screw and Implant 2: Simpliciti, Tornier, with three fins). The resulting images were converted to µFE models consisting of up to 78 million hexahedral elements with isotropic elastic properties based on the literature. These models were subjected to two loading conditions (medial and along the central implant axis) and solved for internal stresses with a parallel solver (parFE, ETH Zurich) on a Linux Cluster. The bone tissue stresses were analysed according to four subregions (dividing plane: sagittal and frontal) at two depths starting from the bone-implant surface and the distal region ending distally from the tip of Implant 1 (proximal, distal)


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 74 - 74
1 Apr 2018
Chevalier Y Chamseddine M Santos I Müller P Pietschmann M
Full Access

Introduction

Glenoid loosening, still a main complication for shoulder arthroplasty, was suggested to be related implant design, surgical aspects, and also bone quality. However, typical studies of fixation do not account for heterogeneity in bone morphology and density which were suggested to affect fixation failure. In this study, a combination of cyclic rocking horse tests on cadaver specimens and microCT-based finite element (microFE) analysis of specimens of a wide range of bone density were used to evaluate the effects of periprosthetic bone quality on the risks of loosening of anatomical keeled or pegged glenoid implants.

Methods

Six pairs of cadaveric scapulae, scanned with a quantitative computer tomography (QCT) scanner to calculate bone mineral density (BMD), were implanted with either cemented anatomical pegged or keeled glenoid components and tested under constant glenohumeral load while a humeral head component was moved cyclically in the inferior and superior directions. Edge displacements were measured after 1000, 4000 and 23000 test cycles, and tested for statistical differences with regards to changes or implant design. Relationships were established between edge displacements and QCT-based BMD below the implant. Four other specimens were scanned with high-resolution peripheral QCT (82µm) and implanted with the same 2 implants to generate virtual models. These were loaded with constant glenohumeral force, varying glenohumeral conformity and superior or inferior load shifts while internal stresses at the cement-bone and implant-cement interfaces were calculated and related to apparent bone density in the periprosthetic zone.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 27 - 27
1 Jan 2017
Chevalier Y Pietschmann M Thorwaechter C Chechik O Adar E Dekel A Mueller P
Full Access

Treatment of massive rotator cuff tears can be challenging. Previous studies with irreparable rotator cuff tears showed good clinical results of tendon healing with the arthroscopic insertion of a protective biodegradable spacer balloon filled with saline solution between the repaired tendon and the acromion [1,2], but so far no scientific evidence has showed how the device alters pressures over the repaired tendon. This biomechanical study investigated the effects of a spacer inserted in the subacromial space on pressures over the repaired rotator cuff tendon in passive motion cycles typical for post-operative rehabilitation routines.

Six human cadaveric shoulders were prepared with the humerus cut 15cm below the joint and embedded in a pot, while the scapula fixed at three points on a plate. A rotator cuff tear was simulated and repaired using a suture anchor and a Mason-Allen suture. The specimens were then mounted on a custom-made pneumatic testing rig to induce passive motion cycles of adduction-abduction (90–0°) and flexion-extension (0–40°) with constant glenohumeral and superior loads and tension is exerted on the supraspinatus tendon with weights. A pressure sensor was placed between the supraspinatus tendon and the acromion. After pressure measurements for 15 cycles of each motion type, the InSpace balloon (OrthoSpace, Inc, Israel) was inserted and the specimens tested and pressure measured again for 15 cycles. Statistically significant changes in peak pressures were then measured before and after balloon.

Peak pressures were measured near 90 degrees abduction. No statistical differences were observed for internal-external rotation before and after balloon-shaped subacromial spacer was inserted. Mean pressures in abduction-adduction were significantly reduced from 121.7 ± 9.5 MPa to 51.5 ± 1.2 MPa. Peak pressures after repair were 1171.3 ± 99.5 MPa and 1749.6 ± 80.7 MPa in flexion-extension and abduction-adduction motion, respectively, and significantly decreased to 468.7 ± 16.0 MPa and 535.1 ± 27.6 MPa after spacer insertion (p<0.0001).

The use of the spacer above the repaired tendon reduced peak pressures and distributed them more widely over the sensor during both abduction-adduction and flexion-extension motions and therefore can reduce the stress on the rotator cuff repair. The InSpace system may reduce the pressure on the repaired tendon, thus potentially protecting the repair. Further studies to investigate this phenomenon are warranted, in particular relating these changes to shoulder kinematics following tear repair and spacer insertion.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 75 - 75
1 May 2016
Chevalier Y Santos I Mueller P Pietschmann M
Full Access

Introduction

Glenoid loosening, still a main complication in shoulder arthroplasty, could be related to glenohumeral orientation and conformity, cementing techniques, fixation design and periprosthetic bone quality [1,2]. While past numerical analyses were conducted to understand the relative role of these factors, so far none used realistic representations of bone microstructure, which has an impact on structural bone properties [3]. This study aims at using refined microFE models including accurate cortical bone geometry and internal porosity, to evaluate the effects of fixation design, glenohumeral conformity, and bone quality on internal bone tissue and cement stresses under physiological and pathological loads.

Methods

Four cadaveric scapulae were scanned at 82µm resolution with a high resolution peripheral quantitative computer tomography (XtremeCT Scanco). Images were processed and virtually implantated with two anatomical glenoid replacements (UHMWPE Keeled and Pegged designs, Exactech). These images were converted to microFE models consisting of nearly 43 million elements, with detailed geometries of compact and trabecular bone, implant, and a thin layer of penetrating cement through the porous bone. Bone tissue, implant and cement layer were assigned material properties based on literature. These models were loaded with a central load at the glenohumeral surface, with the opposite bone surface fully constrained. Effects of glenohumeral conformity were simulated with increases of the applied load area from 5mm-radius to a fully conformed case with the entire glenoid surface loaded. The models were additionally subjected to a superiorly shifted load mimicking torn rotator cuff conditions. These models were solved and compared for internal stresses within the structures (Figure 1) with a parallel solver (parFE, ETH Zurich) on a computation cluster, and peak stresses in each region compared by design and related to apparent bone density.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 100 - 100
1 Sep 2012
Labey L Chevalier Y Fukagawa S Innocenti B Okon T Bellemans J Kowalczewski J
Full Access

Introduction

Optimal knee joint function obviously requires a delicate balance between the osseous anatomy and the surrounding soft tissues, which is distorted in the case of joint line elevation (JLE). Although several studies have found no correlation between JLE and outcome, others have linked JLE to inferior results. The purpose of this in vitro investigation was to evaluate the effect of JLE on tibiofemoral kinematics and collateral ligament strains.

Materials and Methods

Six cadaver knees were equipped with reflective markers on femur and tibia and CT scans were made. A total knee arthroplasty (TKA) was performed preserving the native joint level. The knees were then tested in passive flexion-extension and squatting in a knee kinematics simulator while marker positions were recorded with an optical system. During squatting quadriceps forces were measured as well as tibio-femoral contact pressures. Finally, a revision TKA was performed with JLE by 4 mm. The femoral component was downsized and a thicker insert was used. The knees were again tested as before.

Based on the bony landmarks identified in the CT scans and the measured trajectories of the markers, relative tibiofemoral kinematics could be calculated as well as distance changes between insertions of the collateral ligaments.

Statistical tests were carried out to detect significant differences in kinematic patterns, ligaments elongation, tibiofemoral contact pressures and quadriceps forces between the primary TKA and after JLE.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 372 - 372
1 Mar 2004
Nau T Chevalier Y Duval N deGuise J
Full Access

Aims: It was the aim of the present study to evaluate the resulting 3D kinematics following different surgical techniques of reconstruction in a combined posterior cruciate ligament (PCL)/posterolateral structures (PLS) injury model. Methods: In nine human cadaveric knees, 3D kinematics were recorded during the path of ßexion-extension using a computer based method. Additional laxity tests were conducted at 30¡ and 90¡ of ßexion. Testing was performed before and after cutting the PLS and PCL, followed by PCL reconstruction alone. Reconstructing the posterolateral corner, three surgical techniques were compared: 1) biceps tenodesis (BT), 2) posterolateral corner sling procedure (PLCS), and 3) bone patellar-tendon bone allograft reconstruction (BPTB). Results: Posterior as well as rotational laxity were closely restored to intact values by all tested procedures. Compared to the intact knee, 3D kinematics revealed signiþcant internal tibial rotation for 1) BT (mean = 3.9¡, p = 0.043) and for 3) BPTB allograft (mean = 4.3¡, p = 0.012). 2) PLCS demonstrated a tendency to internal tibial rotation between 0¡ and 60¡ of ßexion (mean = 2.2¡, p = 0.079). Varus/valgus rotation as well as anterior/posterior translation did not show signiþcant differences for any of the tested techniques. Conclusion: The present study showed that despite satisfying results in static laxity testing, pathological 3D knee kinematics were not restored to normal, demonstrated by a non-physiological internal tibial rotation during the path of ßexion-extension.