Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 86 - 86
1 Mar 2017
Weijia C Nagamine R Osano K Takayama M Todo M
Full Access

Introduction

The effect of each step of medial soft tissue releases on the external rotation angle of the femoral component was assessed during posterior stabilized total knee arthroplasty (PS-TKA) with modified gap control technique.

Methods

Consecutive 840 knees were assessed. During PS-TKA, medial soft tissue release was done to obtain rectangular gap in extension using tensors/balancers. The deep fiber of medial collateral ligament (MCL) was released in all cases. No more release was done in 464 knees. Only anterior fiber of superficial MCL was released in 49 knees, and only posterior fiber of superficial MCL was released in 129 knees. Both fibers were released in 169 knees. Additional pes anserinus was released in 29 knees. Rotation angle of the femoral component was decided based on the flexion gap angle. The angle was compared among the five groups.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 46 - 46
1 Mar 2017
Nagamine R Weijia C Todo M Osano K Takayama M Kawasaki M Kaneyama R
Full Access

Introduction

The hip-knee-ankle (HKA) angle between the mechanical axis of the femur (FM) and the mechanical axis of the tibia (TM) is the standard parameter to assess the coronal alignment of the lower extremity. TM is the line between the center of the tibial spines notch (Point T) and the center of the tibial plafond. However, this theory is based on the premise that TM coincides the anatomical axis of the tibia (TA). Fig.1a shows typical varus knee with medial shift of the tibial articular surface. In this case, TM does not coincide TA. Fig. 2 demonstrates the error of HKA angle when Point T locates medial to TA. Fig.2a shows normal alignment. Fig.2b shows varus alignment. Fig. 2c shows the tibia with medial shift of the tibial articular surface. The tibia has 7 degrees varus articular inclination in Fig.2b and 2c. However, HKA angle is 0 degree in Fig.2c. HKA angle underestimates varus deformity in knees with medial shift of the tibial articular surface. However, the degree of medial shift of the tibial articular surface is obscure. In this study, detailed anatomical configuration of the proximal tibia was evaluated. The effect of the value of HKA angle on the coronal alignment in TKA was then discussed.

Methods

This study consists of 117 knees. On the AP view radiograph of the tibia, three distance and two angle parameters were measured. Those were tibial articular surface width, distance between medial edge of the tibial articular surface and Point T, distance from TA to Point T. Angle between TM and TA, and the varus inclination angle of the tibial articular surface relative to the perpendicular line to TA.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 46 - 46
1 Feb 2017
Kaneyama R Higashi H Yoshii H Shiratsuchi H Sasho T Suzuki T Matsuno Y Nagamine R Weijia C
Full Access

INTRODUCTION

The extension and flexion gaps are affected by different factors in total knee arthroplasty (TKA). Flexion but not extension gap measurements are influenced by posterior cruciate ligament (PCL) preservation or resection and patella reduction or eversion and thigh weight. If the flexion gap is measured with the thigh placed on the tibia, the measurement results must include the thigh weight; nevertheless, there is no detailed report regarding the thigh weight influence on the flexion gap. In this study, we investigated how thigh weight affected flexion gap measurement.

METHODS

Four knees of whole-body fresh-frozen cadavers (Mongolian race) were investigated. The femur and tibia were dissected with a standard measured resection technique. After the femoral component was set, the flexion gap was measured with a knee balancer. The distraction force of 20, 30, and 40 pounds were loaded at the joint level. For each measurement, the influences of the patella reduced or everted (PR or PE) and the PCL preserved or resected (CR or PS) were estimated. The flexion gap was measured five times in four different categories (CR/PR, CR/PE, PS/PR, PS/PE) and the thigh weight was reduced by weights (0, 0.5, 1.0, 2.0, 3.0 kg) using a string and pulley system. During measurement, the femur was just placed on the tibia, and the knee flexion angle was maintained at 90 degrees with a goniometer. After all measurements, the lower limbs were resected, and the thighs were weighed with a scale. Steel-Dwasstest (non-parametric multiple comparison test) were performed for statistical analysis, and p < 0.05 was considered significant.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 126 - 126
1 May 2016
Weijia C Nagamine R
Full Access

Purpose

Factors influencing flexion angle of the knee before and after PS-TKA were assessed.

Methods

In 368 PS-TKA cases (71 males and 297 females) by means of modified gap control technique with Stryker NRG system, multi-variance analysis was performed to assess factors influencing flexion angle before TKA and flexion angle 3 weeks after TKA. Their mean age was 74.1 years old. Operative techniques and angle of the components were included as the factors.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 72 - 72
1 Jan 2016
Nagamine R Weijia C Patil S D'Lima D Todo M
Full Access

Introduction

The effect of each step of medial soft tissue release was assessed taking the expansion strength and patellar condition into account in five fresh frozen normal cadaver specimens.

Methods

In each cadaver specimen, only proximal tibia was cut. Then, ACL was cut, and deep MCL fiber was released. This condition was set as “the basic”. Joint gap distance and angle were measured at full extension, 30°, 60°, 90°, 120° flexion and in full flexion. The measurement was firstly done with the standard tensor/balancer with the patella everted, and the next with the offset tensor/balancer with the patella reduced. The torque of 10, 20 and 30 inch-pounds were applied through the specialized torque wrench. After the measurement in “the basic”, PCL, MCL superficial fibres, pes anserinus and semi-membranosus were released step by step. Measuring the joint gap distance and angle with the same scheme above were conducted after the each step.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 446 - 446
1 Dec 2013
Nagamine R Weijia C Todo M Hirokawa S Kondo K Kawasaki M Osano K
Full Access

Purpose:

Differences in the sizes of femoral and tibial components between females and males, between osteoarthritis (OA) and rheumatoid arthritis (RA), and between measured bone resection and the gap control technique during TKA were assessed.

Method:

500 PS-TKAswith the Stryker NRG system in 408 cases were assessed. There were 83 male knees and 417 female knees, and 472 OA knees and 28 RA knees. This study was performed in Japan, and almost all OA knees had varus deformities. In each case, the sizes of the femoral and tibial components were measured on radiographs. The measured sizes represented those of the measured bone resection. TKA was performed by the gap control technique using a tensor/balancer with 30 inch-pounds expansion strength, and the sizes of the femoral and tibial components (used size) were recorded.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 366 - 366
1 Dec 2013
Kaneyama R Nagamine R Weijia C Shiratsuchi H
Full Access

Objective:

Accurate measurement of the extension and flexion gap is important in total knee arthroplasty (TKA). Particularly, the flexion gap may be influenced by several factors; therefore, tension of the posterior cruciate ligament (PCL), knee extensor mechanism, and the thigh weight may need to be considered while estimating the flexion gap. However, there is no comprehensive study on the flexion gap, including an assessment of the influence of gravity on the gap. The purpose of this study is to investigate the influence of PCL, knee extensor mechanism, and thigh weight on the flexion gap by using a fresh frozen cadaver.

Methods:

A fresh frozen lower limb that included the pelvis was used for the assessments. The knee was resected by a measured resection technique and a femoral component was implanted to estimate the component gap. The knee was flexed by precisely 90 degrees using a computer navigation system. The flexion gap was measured in different situations: group A, PCL preserved and patella reduced; group B, PCL preserved and patella everted; group C, PCL resected and patella reduced; and group D, PCL resected and patella everted. In each group, the measurements were obtained under 3 different conditions: 1, knee flexed and the lower limb on the operation table under gravity, as is usually done in TKA; 2, hip and knee flexed 90 degrees to avoid the influence of gravity; and 3, knee set in the same position as in condition 1 and the thigh was held by hand to reduce the influence of the thigh weight.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 367 - 367
1 Dec 2013
Kawasaki M Nagamine R Kondo K Weijia C Osano K
Full Access

Introduction

Following total knee arthroplasty (TKA), some patients show patella baja. It is possible that patella baja after posterior stabilized (PS)-type TKA causes the patellar clunk syndrome and limitation of flexion. The purpose of this study was to examine patellar height before and after PS-type TKA and identify the factors related to the change in patellar height.

Methods

Lateral X-ray films were taken at 90 degrees flexion before and after TKA using fluoroscopy in 87 patients (95 knees) (Fig. 1a, b). The components and surgical technique for TKA were Scorpio NRG (Stryker) and the modified gap control technique, respectively. The Insall-Salvati ratio (ISR) and the Labelle-Laurin method (LL) were measured as parameters of patellar height (Fig. 1c, d). Posterior condylar offset (PCO) (Fig. 1e), the distance from the anterior femoral line to the tibial tuberosity (TA), and the distance from the tibial tuberosity to the posterior condyle of the femur [TP; {TA-F (the length of the femoral condyle)}] (Fig. 1f) were examined as parameters that could be associated with the change in patellar height. All parameters were divided by patellar length to compensate for the expansion rate in each photograph. The mean LL/P, PCO/P, TA/P, and TP/P before TKA were set at 100%.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 582 - 582
1 Dec 2013
Weijia C Nagamine R Kondo K Osano K
Full Access

INTRODUCTION:

In varus knee, posterior cruciate ligament (PCL) release has been reported to result in the increase of the flexion gap without significant effect on the extension gap. However, the effect of release on gap angle is still obscure. On the other hand, gap angle and distance measured with the tension devices may vary due to different distraction forces. In this study, difference of gap angle and distance before and after PCL resection in knee extension and 90° flexion was inspected. Effect of different distraction force on gap was also assessed.

OBJECTIVES:

Fifty cases with medial osteoarthritis undergoing PS-TKA were included in the study. PCL of all the cases were identified intact before resection.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 581 - 581
1 Dec 2013
Weijia C Nagamine R
Full Access

Objective:

Total elbow joint arthroplasty has limited longevity and is therefore not appropriate for younger rheumatoid arthritis patients. Arthroplasty using an inter-positional membrane may be another surgical option for this population. However, clinical results for joint arthroplasty using the inter-positional membrane have not traditionally been favorable because rheumatoid activity could not be controlled. Today, rheumatoid activity can be controlled with biologics; therefore, the utility of the inter-positional membrane procedure was re-evaluated.

Methods:

An 8×6 cm sheet of fascia was detached from the patient's tensor fascia lata muscle to produce a JK membrane. The fascia was stretched on a frame and kept in a 2% chromic acid potassium solution for 24 hours. Then, the fascia was exposed to direct sunlight in order to reduce the dichromic acid. The fascia was washed out in running water for 24 hours and was then stored in phenol with the addition of 70% alcohol. Elbow arthroplasties were performed on three elbow joints in two young female patients. The first case had a significantly damaged right elbow joint with severe joint dysfunction. A JK membrane arthroplasty was done for the first case in 2003, when this patient was 34 years old. Biologics were administered with methotrexate after the surgery. The second case demonstrated bilateral ankylosed elbows due to idiopathic juvenile arthritis. Bilateral JK membrane arthroplasties were performed in 2010, when this patient was 32 years old. Several operative and manual manipulations were necessary in order to increase the range of motion following surgery. Biologics were administered with cyclosporine.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 277 - 277
1 Mar 2013
Nagamine R Hirokawa S Todo M Weijia C Kondo K
Full Access

Introduction

Reliability of a gap control technique with the tensor/balancer during PS-TKA was assessed by means of fluoroscopic images after TKA.

Methods

Thirty-one subjects were selected for assessment. The mean age of the subjects was 73.0 years old. During PS-TKA, a parapatellar approach was used. Cruciate ligaments were excised, and distal femoral and proximal tibial cuts were made. After all osteophytes were removed, the joint gap angle and distance were measured in full extension and at 90° flexion using a tensor/balancer. Medial soft tissue releases were performed and soft tissue balancing was obtained in full extension so that the joint gap angle was 3° or less than 3°. The joint gap angle and distance between femoral and tibial cut surfaces in full extension, and between a tangent to the posterior femoral condyles and tibial cut surface at 90° flexion were measured. The external rotation angle of the anterior and posterior cuts of the femur was decided based on the joint gap angle at 90° flexion. The size of the femoral component was decided based on the joint gap distance in full extension and at 90° flexion. Then only the trial femoral component was inserted. The joint gap angle and distance between the tangent to the condyles of the trial femoral component and tibial cut surface in full extension and at 90° flexion were measured.

More than one month after TKA, the fluoroscopic images of the prostheses were taken during knee extension/flexion. Then, a torque of about 5 Nm was applied to the lower leg in order to assess the varus/valgus flexibility during flexion. The pattern matching method was used to measure the 3D movements of the prostheses from the fluoroscopic images. The joint gap angle was calculated in full extension and at 90° flexion. The varus/valgus flexibility at each flexion angle was also assessed.