Vol. 100-B, Issue SUPP_4 | Pages 11 - 11
1 Apr 2018
Kwong L Billi F Keller S Kavanaugh A Luu A Ward J Salinas C Paprosky W
Full Access

Introduction

The objective of this study was to compare the performance of the Explant Acetabular Cup Removal System (Zimmer), which has been the favored system for many surgeons during hip revision surgery, and the new EZout Powered Acetabular Revision System (Stryker).

Methods

54mm Stryker Trident® acetabular shells were inserted into the foam acetabula of 24 composite hemi-pelvises (Sawbones). The hemi-pelvises were mounted on a supporting apparatus enclosing three load cells. Strain gauges were placed on the hemipelvis, on the posterior and the anterior wall, and on the internal ischium in proximity to the acetabular fossa. A thermocouple was fixed onto the polar region of the acetabular component. One experienced orthopaedic surgeon and one resident performed mock revision surgery 6 times each per system.