Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 28 - 28
1 Feb 2020
Kamada K Takahashi Y Tateiwa T Shishido T Masaoka T Pezzotti G Yamamoto K
Full Access

Introduction

Highly crosslinked, ultra-high molecular weight polyethylene (HXLPE) acetabular liners inherently have a risk of fatigue failure associated with femoral neck impingement. One of the potential reasons for liner failure was reported as crosslinking formulations of polyethylene, increasing the brittleness and structural rigidity. In addition, the acetabular component designs greatly affect the mechanical loading scenario, such as the offset (lateralized) liners with protruded rim above the metal shells, which commonly induce a weak resistance to rim impingement. The purpose of the present study was to compare the influence of the liner offset length on the impingement resistance in the annealed (first generation) and vitamin E-blended (second-generation) HXLPE liners with a commercial design.

Materials and Methods

The materials tested were the 95-kGy irradiated annealed GUR1020, and the 300-kGy irradiated vitamin E-blended GUR1050 HXLPE offset liners, which were referred to as “20_95” and “50E_300”, respectively. These liners had 2, 3, 4-mm rim offset, 2.45-mm rim thickness, and 36-mm internal diameter. Their rims were protruded above the metal rim at 2, 3, 4mm. Rim impingement testing was performed using an electrodynamic axial-torsional machine. The cyclic impingement load of 25–250N was applied on the rims through the necks of the femoral stems at 1Hz. The rotational torque was simultaneously generated by swinging the stem necks on the rims at 1Hz and its rotational angle was set at the range of 0–10˚. The percent crystallinity was analyzed on the as-received (intact) and impinged HXLPE acetabular rims by confocal Raman microspectroscopy.