header advert
Results 1 - 4 of 4
Results per page:
The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 33 - 41
1 Jan 2018
Matharu GS Hunt LP Murray DW Howard P Pandit HG Blom AW Bolland B Judge A

Aims

The aim of this study was to determine whether the rates of revision for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle components varied according to the year of the initial operation, and compare these with the rates of revision for other designs of MoM THA.

Patients and Methods

Data from the National Joint Registry for England and Wales included 36 mm MoM THAs with Pinnacle acetabular components which were undertaken between 2003 and 2012 with follow-up for at least five years (n = 10 776) and a control group of other MoM THAs (n = 13 817). The effect of the year of the primary operation on all-cause rates of revision was assessed using Cox regression and interrupted time-series analysis.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 50 - 50
1 Jun 2017
Bolland B Cook E Tucker K Howard P
Full Access

This study utilized data from the NJR dataset on all Corail/Pinnacle total hip replacements (THR) to determine (a) the level of unit variation of the Corail/Pinnacle 36mm Metal On Metal THR within England and Wales; (b) patient, implant and surgeon factors that may be associated with higher revision rates; (c) Account for the influence of the MHRA announcement in 2010.

The national Revision Rate (RR) for the Corail / Pinnacle MOM THR was 10.77% (OR:1.46; CI:1.17–1.81). This was significantly greater than other articulation combinations (MOP 1.72%, COP 1.36%, COC 2.19%).

The 2010 MHRA announcement did not increase rate of revision (X2=1649.63, df=13, p<.001).

Patient factors associated with significantly increased revision rates included, female gender (OR 1.38 (CI 1.17–1.63, p<.001) and younger age OR 0.99 (CI 0.98–0.99), p<.001). Implant factor analysis demonstrated an inverse relationship between cup size and revision. As head length increased RR increased – highest risk of revision +12.5 (OR 1.69 (CI 1.12–2.55), p=0.13). Coxa vara, high offset stems had a higher risk of revision compared to standard offset stems (OR:1.41 (CI 1.15–1.74; p<.001). As stem size increased risk of revision decreased (OR 0.89 (CI 0.85–0.93); p<.001). Surgeon grade did not influence RR.

There was significant variation in RR between hospitals with 7 units (7/61 excluding low volume centres, <50 implants) identified as having significant higher rates of revision. However, for each of these units there was a greater proportion of higher risk patients (female, cup size 50–54, stem type).

This study has provided insight into unit variation, risk factors and the long term outcome of the Corail/Pinnacle 36mm MOMTHR. Future aims are to use these results to develop a risk stratified algorithm for the long term follow of these patients to minimize patient inconvenience and excess use of limited NHS resources.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 8 - 8
1 Jun 2016
Cook R Bolland B Tilley S Latham J
Full Access

Using a femoral head from one manufacturer on the stem of another manufacturer poses the risk that the taper interface between the components do not contact correctly and the performance of the joint will be impaired. The cohorts in this study are a combination of modular Birmingham Hip Resurfacing (BHR) and Adept femoral heads on CPT stems. The study reviews the geometry of the taper interfaces to establish if the taper clearance angles was outside of the normal range for other taper interfaces. In addition the rates of material loss from the bearings and taper and a ranking of the stem damage were reviewed to determine if the levels of loss were above that seen for other similar joints.

The material loss analysis demonstrated that the rates or levels of loss from the bearings, taper and stem are no different to levels published for manufacturer matched joints and in many cases are lower. The results demonstrate that the taper clearance angles for the mixed manufacturer joints (BHR-CPT: 0.067 to −0.116, Adept-CPT: 0.101 to −0.056) were within the range of other studies and manufacturer matched clearances (0.134 to −0.149).

Using components from different manufacturers has not in this instance increased the level of material loss from the joints, when compared to other similar manufacturer matched joints.


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 463 - 472
1 Apr 2015
Panagiotidou A Meswania J Osman K Bolland B Latham J Skinner J Haddad FS Hart A Blunn G

The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper interface of a modular femoral component and to investigate whether different combinations of material also had an effect. The combinations we examined were 1) cobalt–chromium (CoCr) heads on CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic heads on CoCr stems.

In test 1 increasing torque was imposed by offsetting the stem in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14 Nm and 18 Nm.

In test 2 we investigated the effect of increasing the bending moment by offsetting the application of axial load from the midline in the mediolateral plane. Increments of offset equivalent to head + 0 mm, head + 7 mm and head + 14 mm were used.

Significantly higher currents and amplitudes were seen with increasing torque for all combinations of material. However, Ti stems showed the highest corrosion currents. Increased bending moments associated with using larger offset heads produced more corrosion: Ti stems generally performed worse than CoCr stems. Using ceramic heads did not prevent corrosion, but reduced it significantly in all loading configurations.

Cite this article: Bone Joint J 2015;97-B:463–72.