header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Hip

RISK FACTORS, UNIT VARIATION AND LONG-TERM OUTCOME OF THE CORAIL/PINNACLE 36 MM METAL-ON-METAL TOTAL HIP ARTHROPLASTY: ANALYSIS FROM THE NJR DATASET OF ENGLAND AND WALES

British Hip Society meeting (BHS) March 2017



Abstract

This study utilized data from the NJR dataset on all Corail/Pinnacle total hip replacements (THR) to determine (a) the level of unit variation of the Corail/Pinnacle 36mm Metal On Metal THR within England and Wales; (b) patient, implant and surgeon factors that may be associated with higher revision rates; (c) Account for the influence of the MHRA announcement in 2010.

The national Revision Rate (RR) for the Corail / Pinnacle MOM THR was 10.77% (OR:1.46; CI:1.17–1.81). This was significantly greater than other articulation combinations (MOP 1.72%, COP 1.36%, COC 2.19%).

The 2010 MHRA announcement did not increase rate of revision (X2=1649.63, df=13, p<.001).

Patient factors associated with significantly increased revision rates included, female gender (OR 1.38 (CI 1.17–1.63, p<.001) and younger age OR 0.99 (CI 0.98–0.99), p<.001). Implant factor analysis demonstrated an inverse relationship between cup size and revision. As head length increased RR increased – highest risk of revision +12.5 (OR 1.69 (CI 1.12–2.55), p=0.13). Coxa vara, high offset stems had a higher risk of revision compared to standard offset stems (OR:1.41 (CI 1.15–1.74; p<.001). As stem size increased risk of revision decreased (OR 0.89 (CI 0.85–0.93); p<.001). Surgeon grade did not influence RR.

There was significant variation in RR between hospitals with 7 units (7/61 excluding low volume centres, <50 implants) identified as having significant higher rates of revision. However, for each of these units there was a greater proportion of higher risk patients (female, cup size 50–54, stem type).

This study has provided insight into unit variation, risk factors and the long term outcome of the Corail/Pinnacle 36mm MOMTHR. Future aims are to use these results to develop a risk stratified algorithm for the long term follow of these patients to minimize patient inconvenience and excess use of limited NHS resources.


Email: