Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 456 - 456
1 Sep 2012
Grammatopoulos G Pandit H Mellon S Glyn-Jones S Gundle R Mclardy-Smith P Murray D Gill H
Full Access

INTRODUCTION

Studies have suggested that there is a reduction in head-neck-ratio (HNR) associated with MoMHRA. A reduction in HNR at operation would decrease range of movement and increase impingement risk. Impingement could lead to 20 edge loading, increasing wear. Serum ion levels of Chromium (Cr) and Cobalt (Co) are surrogate markers of wear. Although acetabular component orientation has been shown to contribute to wear and PT development, the role of a decrease in HNR has only been highlighted in PT development. This study aimed to measure changes in HNR that occur at resurfacing and determine any gender- and component size-specific differences. In addition it aimed to determine whether changes in HNR could be associated with increased wear.

METHODS

84 patients (56M: 28F) with unilateral MoMHRA were included. The mean age at surgery was 57 years. The mean femoral component was 49mm. Components were considered small if <45mm, average if between 45–50mm and large if >50mm. Three designs were implanted; BHR, C+ and Recap. The average follow up was 4 years. All patients had Cr/Co levels measured at follow up. Patients were considered to have high ions if Cr and Co levels were 5.1ppb and 4.4ppb respectively.

Pre-operative HNR (HNRpre) and the post-operative HNR (HNRpost) were made from the respective pelvic radiographs. Assuming a 2mm thick cartilage layer, the HNR based on the diameter of the articular cartilage pre-operatively (HNRart) was calculated.

The immediate changes in HNR as a result of the operation were expressed relative to articular HNR pre-op:

HNRartpost=HNRpost–HNRart


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 578 - 578
1 Sep 2012
Grammatopoulos G Judge A Pandit H Mclardy-Smith P Glyn-Jones S Desmet K Murray D Gill H
Full Access

INTRODUCTION

Although simulation studies have shown superior wear properties of metal-on-metal articulations, increased concern exists regarding the excess in-vivo wear of a small number of Metal-on-Metal-Hip-Resurfacing (MoMHRA) implants. Serum ion levels of Chromium (Cr) and Cobalt (Co) are surrogate markers of wear. Risk factors associated with increased wear include female gender, small components, dysplasia, cup orientation outside safe zone and femoral head downsize during surgery with an associated decrease in Head-Neck-Ratio (HNR). However, these factors are interlinked. This study aims to identify the factors that are most important for subsequent wear of MoMHRA, by performing a multivariate analysis.

METHODS

206 patients (124M: 82F) with unilateral MoMHRA were included in this study. The average follow up was 3.3 years. All patients had Cr/Co levels measured at follow up. Inclination and anteversion of each cup were measured using EBRA. Cups were analysed as being within or outside the previously defined optimum-zone. HNR measurements were made from pre-operative (HNRpre) and post-operative (HNRpost) radiographs. The immediate changes in HNR (downsize/upsize of femoral head) as a result of the operation were expressed as:

HNRprepost=HNRpost–HNRpre

Multivariate linear regression modelling was used to explore the association between measures of ions with the following predictor variables (gender, age, diagnosis, femoral component size, orientation of the acetabular component, head/neck ratio and position of femoral stem). Analyses were carried out separately for each outcome (Cr and Co). Classification and Regression Tree (CART) models were fitted as a complimentary approach to regression modelling.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 584 - 584
1 Sep 2012
Grammatopoulos G Thomas G Pandit H Glyn-Jones S Gill H Beard D Murray D
Full Access

INTRODUCTION

The introduction of hard-on-hard bearings and the consequences of increased wear due to edge-loading have renewed interest in the importance of acetabular component orientation for implant survival and functional outcome following hip arthroplasty. Some studies have shown increased dislocation risk when the cup is mal-oriented which has led to the identification of a safe-zone1. The aims of this prospective, multi-centered study of primary total hip arthroplasty (THA) were to: 1. Identify factors that influence cup orientation and 2. Describe the effect of cup orientation on clinical outcome.

METHODS

In a prospective study involving seven UK centers, patients undergoing primary THA between January 1999 and January 2002 were recruited. All patients underwent detailed assessment pre-operatively as well as post-op. Assessment included data on patient demographics, clinical outcome, complications and further surgery/revision. 681 primary THAs had adequate radiographs for inclusion. 590 hips received cemented cups. The primary functional outcome measure of the study was the change between pre-operative and at latest follow up OHS (OHS). Secondary outcome measures included dislocation rate and revision surgery. EBRA was used to determine acetabular inclination and version.

The influence of patient's gender, BMI, surgeon's grade and approach on cup orientation was examined. Four different zones tested as possibly ± (Lewinnek Zone, Callanan's described zone and zones ± 5 and ±10 about the study's mean inclination and anteversion) for a reduced dislocation risk and an optimal functional outcome.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 68 - 68
1 Aug 2012
Tu Y Xue H Liu X Cai M Xia Z Murray D
Full Access

Great interest in unicompartmental knee arthroplasty (UKA) for medial osteoarthritis has rapidly increased following the introduction of minimally invasive UKA (MI-UKA). This approach preserves the normal anatomy of knee, causes less damage to extensor mechanism and results in a more rapid post-operative recovery. However, experience with this approach is limited in China. The aim of this report was to determine the short-term clinical and radiographic outcomes of MI-UKA in the Chinese, and to identify any features that are unique to this population. Fifty two knees, in forty-eight patients, with medial compartmental osteoarthritis treated by MI-UKA via C-arm intensifier guide (CAIG) from May 2005 to January 2009 were reviewed. Pain and range of motion (ROM) was assessed using the HSS scoring system before and after surgery. Pre- and postoperative alignment of the lower limbs was measured and compared. The mean follow up time was 24 months (12-42 months). In all cases the pain over medial compartment of the knees was relieved or subsided. The post-operative ROM was 0-136 degree (mean 122degree), and the mean alignment was 2degree varus (0- 7degree varus). The HSS score increased from 72(61-82) to 92(72-95). 93% of the postoperative scores were good or excellent. Interestingly, the distribution of femoral component sizes of these patients was XS 2%, Small 83%, Medium 15%, Large 0%, XL 0%; whereas tibial component size was AA 27%, A 55%, B 15%, C 3%, D 0%, E 0%, and F 0%. The optimal fitted match between tibial and femoral size was: tibia AA and A with XS and small femur, tibia B and C with medium femur. The estimated match was: tibia D and E with large femur, tibia F with XL femur. In contrast to the Oxford report, the sizes of these components are smaller and not in correlation with the height, weight and BMI of the patients. We conclude that MI-UKA is an effective method for treating medial compartmental osteoarthritis of the knee in the Chinese population. CAIG is a feasibly intraoperative measure to predict femoral component sizes. However, component sizes and combinations are different from the Oxford guideline.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 47 - 47
1 Aug 2012
Merle C Waldstein W Pegg E Streit M Gotterbarm T Aldinger P Murray D Gill H
Full Access

In pre-operative planning for total hip arthroplasty (THA), femoral offset (FO) is frequently underestimated on AP pelvis radiographs as a result of inaccurate patient positioning, imprecise magnification, and radiographic beam divergence. The aim of the present study was to evaluate the reliability and accuracy of predicting three-dimensional (3-D) FO as measured on computed tomography (CT) from measurements performed on standardised AP pelvis radiographs.

In a retrospective cohort study, pre-operative AP pelvis radiographs and corresponding CT scans of a consecutive series of 345 patients (345 hips, 146 males, 199 females, mean age 60 (range: 40-79) years, mean body-mass-index 27 (range: 29-57) kg/m2) with primary end-stage hip osteoarthritis were reviewed. Patients were positioned according to a standardised protocol and all images were calibrated. Using validated custom programmes, FO was measured on corresponding AP pelvis radiographs and CT scans. Inter- and intra-observer reliability of the measurement methods were evaluated using intra-class correlation coefficients (ICC). To predict 3-D FO from AP pelvis measurements, the entire cohort was randomly split in two groups and gender specific linear regression equations were derived from a subgroup of 250 patients (group A). The accuracy of the derived prediction equations was subsequently assessed in a second subgroup of 100 patients (group B).

In the entire cohort, mean FO was 39.2mm (95%CI: 38.5-40.0mm) on AP pelvis radiographs and 44.6mm (95%CI: 44.0-45.2mm) on CT scans. FO was underestimated by 14% on AP pelvis radiographs compared to CT (5.4mm, 95%CI: 4.8-6.0mm, p<0.001) and both parameters demonstrated a linear correlation (r=0.642, p<0.001). In group B, we observed no significant difference between gender specific predicted FO (males: 48.0mm, 95%CI: 47.1-48.8mm; females: 42.0mm, 95%CI: 41.1-42.8mm) and FO as measured on CT (males: 47.7mm, 95%CI: 46.1-49.4mm, p=0.689; females: 41.6mm, 95%CI: 40.3-43.0mm, p=0.607).

The results of the present study suggest that femoral offset can be accurately and reliably predicted from AP pelvis radiographs in patients with primary end-stage hip osteoarthritis. Our findings support the surgeon in pre-operative templating and may improve offset and limb length restoration in THA without the routine performance of CT.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 28 - 28
1 Aug 2012
Xia Z Murray D
Full Access

Metal and their alloys have been widely used as implantable materials and prostheses in orthopaedic surgery. However, concerns exist as the metal nanoparticles released from wear of the prostheses cause clinical complications and in some cases result in catastrophic host tissue responses. The mechanism of nanotoxicity and cellular responses to wear metal nanoparticles are largely unknown. The aim of this study was to characterise macrophage phagocytosed cobalt/chromium metal nanoparticles both in vitro and in vivo, and investigate the consequent cytotoxicity. Two types of macrophage cell lines, murine RAW246.7 and human THP-1s were used for in vitro study, and tissues retrieved from pseudotumour patients caused by metal-on-metal hip resurfacing (MoMHR) were used for ex vivo observation. Transmission electron microscopy (TEM), scanning electron microscopy (SEM) in combination with backscatter, energy-disperse X-ray spectrometer (EDS), focused ion beam (FIB) were employed to characterise phagocytosed metal nanoparticles. Alamar blue assay, cell viability assays in addition to confocal microscopy in combination with imaging analysis were employed to study the cytotoxiticy in vitro. The results showed that macrophages phagocytosed cobalt and chromium nanoparticles in vitro and the phagocytosed metal particles were confirmed by backscatter SEM+EDS and FIB+EDS. these particles were toxic to macrophages at a dose dependent manner. The analysis of retrieved tissue from revision of MoMHR showed that cobalt/chromium metal nanoparticles were observed exclusively in living macrophages and fragments of dead macrophages, but they were not seen within either live or dead fibroblasts. Dead fibroblasts were associated with dead and disintegrated macrophages and were not directly in contact with metal particles; chromium but not cobalt was the predominant component remaining in tissue. We conclude that as an important type of innate immune cells and phagocytes, macrophages play a key role in metal nanoparticles related cytotoxicity. Metal nanoparticles are taken up mainly by macrophages. They corrode in an acidic environment of the phagosomes. Cobalt that is more soluble than chromium may release inside macrophages to cause death of individual nanoparticle-overloaded macrophages. It is then released into the local environment and results in death of fibroblasts and is subsequently leached from the tissue.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 78 - 78
1 Aug 2012
Merle C Waldstein W Gregory J Goodyear S Aspden R Aldinger P Murray D Gill H
Full Access

In uncemented total hip arthroplasty (THA), the optimal femoral component should allow both maximum cortical contact with proximal load transfer and accurate restoration of individual joint biomechanics. This is often compromised due to a high variability in proximal femoral anatomy. The aim of this on-going study is to assess the variation in proximal femoral canal shape and its association with geometric and anthropometric parameters in primary hip OA.

In a retrospective cohort study, AP-pelvis radiographs of 98 consecutive patients (42 males, 56 females, mean age 61 (range:45-74) years, BMI 27.4 (range:20.3-44.6) kg/m2) who underwent THA for primary hip OA were reviewed. All radiographs were calibrated and femoral offset (FO) and neck-shaft-angle (NSA) were measured using a validated custom programme. Point-based active shape modelling (ASM) was performed to assess the shape of the inner cortex of the proximal femoral meta- and diaphysis. Independent shape modes were identified using principal component analysis (PCA). Hierarchical cluster analysis of the shape modes was performed to identify natural groupings of patients. Differences in geometric measures of the proximal femur (FO, NSA) and demographic parameters (age, height, weight, BMI) between the clusters were evaluated using Kruskal-Wallis one-way-ANOVA or Chi-square tests, as appropriate.

In the entire cohort, mean FO was 39.0 mm, mean NSA was 131 degrees. PCA identified 10 independent shape modes accounting for over 90% of variation in proximal femoral canal shape within the dataset. Cluster Analysis revealed 6 shape clusters for which all 10 shape modes demonstrated a significantly different distribution (p-range:0.000-0.015). We observed significant differences in age (p=0.032), FO (p<0.001) and NSA (p<0.001) between the clusters. No significant differences with regard to gender or BMI were seen.

Our preliminary analysis has identified 6 different patterns of proximal femoral canal shape which are associated with significant differences in femoral offset, neck-shaft-angle and age at time of surgery. We are currently evaluating the entire dataset of 345 patients which will allow a comprehensive classification of variation in proximal femoral shape and joint geometry. The present data may optimise preoperative planning and improve future implant design in THA.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIX | Pages 102 - 102
1 Jul 2012
van Duren B Pandit H Tilley S Price M Gill H Murray D Thomas N
Full Access

Introduction

Traditional TKR designs exhibit abnormal and unpredictable kinematics: with posterior subluxation in extension and anterior slide with flexion. These can contribute to restricted knee flexion and reduced quadriceps efficiency. Newer designs attempt to provide “guided motion” with the aim of mimicking normal knee kinematics. The Journey (Smith & Nephew) BCS TKR incorporates both an anterior and a posterior cam/post mechanism while Triathlon PS TKR (Stryker) incorporates a posterior cam/post mechanism. This study compares the in-vivo kinematics of these two designs and compares it with normal knee.

Methods

Knee kinematics of 10 patients with Journey-BCS TKR and 11 patients with Triathlon PS TKR; all with excellent clinical outcome (average age: 65) were analysed. Patients underwent fluoroscopic assessment of the knee during a step-up and deep knee bend exercise. 2D fluoroscopic images were recorded. Data was analysed for patella tendon angle (PTA) and contact points using a 3D model fitting technique. This data was compared to normal knee kinematics (n=20).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXII | Pages 21 - 21
1 Jul 2012
Monk A Grammatopoulos G Chen M Gibbons C Beard D Murray D Gill H
Full Access

A femoral head/neck ratio (HNR) of less than 1.27 is associated with an increased risk of arthritis. The aim of this study was to establish whether there is evolutionary evidence that the homonin, bipedal stance has led to alterations in HNR that predispose humans to osteoarthritis (OA).

Specimens provided by The Natural History Museums of London, Oxford and the Department of Zoology, University of Oxford were grouped according to gait pattern, HAKF (Hip and knee flexed), Arboreal (ability to stand with hip and knee joints extended) and homonin/bi-pedal. Specimens included those from Devonion, Triassic, Jurrasic, Cretaceous, Miocene, Paleolithic, Pleistocene periods to modern day. Three-dimensional skeletal geometries were segmented using CT images and HNR measurements were taken from coronal views. These were compared with the HNR of 119 asymptomatic human volunteers and 210 patients that had a hip joint replacement for primary OA.

Species of the HAKF group had the smallest HNR (1.10, SD:0.09). Species of the Arboreal group had significantly higher HNR (1.63, SD:0.15) in comparison to the Bipedal group (1.41, SD:0.04) (p=0.006), Human (1.33, SD:0.08) and the OA group (1.3, SD:0.09).

The range of movement associated with arboreal habitat caused an associated change in HNR. This study would suggest that the HNR peaked in the Miocene period with species that ambulated on both ground and trees. More recent homonin gait appears to have developed a smaller HNR and humans have the smallest amongst their close ancestors. Evolutionary theory would suggest that modern environmental pressures might pre-dispose future hominin evolution to OA, secondary to a further reduction in HNR.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIX | Pages 92 - 92
1 Jul 2012
Mehmood S Batta V Gulati A Pandit H Bottomley N Gil H Beard D Dodd C Jackson W Murray D Price A
Full Access

INTRODUCTION

Establishing a full-thickness cartilage in the lateral compartment and functionally intact ACL is vital before proceeding with unicompartmental knee replacement (UKR). The aim of this study is to assess whether MRI is a useful adjunct in predicting suitability for UKR, as compared to standard and stress radiographs.

METHODS

We identified 50 patients with a knee found suitable for UKR based on their standard and stress radiographs (full-thickness cartilage on lateral side). These patients underwent an additional cartilage-specific MRI scan to identify the status of ACL and the lateral compartment. The final decision regarding the suitability for UKR was based on the intra-operative observation.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIX | Pages 43 - 43
1 Jul 2012
Price A Jackson W Field R Judge A Carr A Arden N Murray D Dawson J Beard D
Full Access

Purpose

The Oxford Knee Score (OKS) is a validated and widely used PROM that has been successfully used in assessing the outcome of knee arthroplasty (KA). It has been adopted as the nationally agreed outcome measure for this procedure and is now routinely collected. Increasingly, it is being used on an individual patient basis as a pre-operative measure of osteoarthritis and the need for joint replacement, despite not being validated for this use. The aim of this paper is to present evidence that challenges this new role for the OKS.

Method

We have analysed pre-operative and post-operative OKS data from 3 large cohorts all undergoing KA, totalling over 3000 patients. In addition we have correlated the OKS to patient satisfaction scores. We have validated our findings using data published from the UK NJR.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIV | Pages 24 - 24
1 Jul 2012
Mehmood S Gill H Murray D Glyn-Jones S
Full Access

Stem geometry is known to influence the outcome in THA; however it is unknown whether the material properties, stiffness in particular can influence the stem stability and outcome. The aim of this study was to measure the influence of stem material properties on micromotion and migration using Roentgen Stereophotogrammetric Analysis (RSA) system.

41 patients were implanted with a collarless polished tapered (CPT) femoral stem (Zimmer, Warsaw, Indiana), which was made of either cobalt-chromium (CoCr) (n=21) or stainless steel (n=20). RSA was used to measure dynamically inducible micromotion (DIMM: difference in stem position in going from double-leg stance (DLS) to single leg stance (SLS)), prosthesis bending (difference in the head-tip distance when going from DLS to SLS), and mean migration of the head, tip and the cement restrictor. DIMM and bending were measured at 3 months, migration at 6, 12 and 24 months. All analyses were carried out using SPSS for windows (v.15.0.0, Chicago. IL, USA). Results were reported as mean ± 95% confidence interval (CI) and regarded as significant when p < 0.05.

Preliminary analysis showed that total head DIMM was significantly (p = 0.02) greater for CoCr (0.97mm ± 0.6mm) than stainless steel (0.27mm ± 0.6mm). The mean stem bending for CoCr was 0.08mm (± 0.06mm) and for stainless steel 0.15mm (± 0.06mm) (p =0.77). Both implants heads migrated posteriorly, medially and distally. The mean subsidence for the cobalt-chromium and stainless steel stems was 1.02mm (± 0.19mm) (p < 0.001) and 1.12mm (± 0.34mm) (p=0.001) (p= 0.07) at 24 months.

It was interesting to note that the dynamically induced micromotion was greater for the stiffer stem, however there were no differences in terms of overall migration, indicating that survival (in terms of loosening) should be the similar for both steel and CoCr versions of this implant.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 4 - 4
1 May 2012
Simpson D Kendrick B Hughes M Rushforth G Gill H Murray D
Full Access

Introduction

Primary mechanical stability is important with uncemented THR because early migration is reduced, leading to more rapid osseointegration between the implant and bone. Such primary mechanical stability is provided by the design features of the device. The aim of this study was to compare the migration patterns of two uncemented hip stems, the Furlong Active and the Furlong HAC stem; the study was designed as a randomised control trial. The implants were the Furlong HAC, which is an established implant with good long term results, and the Furlong Active, which is a modified version of the Furlong HAC designed to minimise stress concentrations between the implant and bone, and thus to improve fixation.

Materials and methods

The migration of 43 uncemented femoral components for total hip replacement was measured in a randomised control trial using Roentgen Stereophotogrammetric Analysis (RSA) over two years. Twenty-three Furlong HAC and twenty Furlong Active stems were implanted into 43 patients. RSA examinations were carried out post-operatively, and at six months, 12 months and 24 months post-operatively. The patients stood in-front of a purpose made calibration frame which contained accurately positioned radio-opaque markers. From the obtained images, the 3-D positions of the prosthesis and the host bone were reconstructed. Geometrical algorithms were used to identify the components of the implant. These algorithms allowed the femoral component to be studied without the need to attach markers to the prosthesis. The migration was calculated relative to the femoral coordinate system representing the anterior-posterior (A-P), medial-lateral (M-L) and proximal-distal (P-D) directions respectively. Distal migration was termed subsidence.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 47 - 47
1 May 2012
Bottomley N McNally E Jones L Javaid M Arden N Gill H Dodd C Murray D Beard D Price A
Full Access

Introduction

Anteromedial osteoarthritis of the knee (anteromedial gonarthrosis-AMG) is a common form of knee arthritis. In a clinical setting, knee arthritis has always been assessed by plain radiography in conjunction with pain and function assessments. Whilst this is useful for surgical decision making in bone on bone arthritis, plain radiography gives no insight to the earlier stages of disease. In a recent study 82% of patients with painful arthritis had only partial thickness joint space loss on plain radiography. These patients are managed with various surgical treatments; injection, arthroscopy, osteotomy and arthroplasty with varying results. We believe these varying results are in part due to these patients being at different stages of disease, which will respond differently to different treatments. However radiography cannot delineate these stages. We describe the Magnetic Resonance Imaging (MRI) findings of this partial thickness AMG as a way of understanding these earlier stages of the disease.

Method

46 subjects with symptomatic partial thickness AMG underwent MRI assessment with dedicated 3 Tesla sequences. All joint compartments were scored for both partial and full thickness cartilage lesions, osteophytes and bone marrow lesions (BML). Both menisci were assessed for extrusion and tear. Anterior cruciate ligament (ACL) integrity was also assessed. Osteophytes were graded on a four point scale in the intercondylar notch and the lateral margins of the joint compartments. Scoring was performed by a consultant radiologist and clinical research fellow using a validated MRI atlas with consensus reached for disagreements. The results were tabulated and relationships of the interval data assessed with linear by linear Chi2 test and Pearson's Correlation.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVII | Pages 11 - 11
1 May 2012
Hossain M Parfitt D Beard D Darrah C Nolan J Murray D Andrew J
Full Access

Introduction

Preoperative psychological distress has been reported to predict poor outcome and patient dissatisfaction after total hip replacement (THR). We investigated this relationship in a prospective multi-centre study between January 1999 and January 2002.

Methods

We recorded the Oxford Hip Score (OHS) and SF36 score preoperatively and up to five years after surgery and a global satisfaction questionnaire at five year follow up for 1039 patients. We dichotomised the patients into the mentally distressed (Mental Health Scale score - MHS <50) and the not mentally distressed (MHS (50) groups based on their pre-operative MHS of the SF36. 776 (677 not distressed and 99 distressed) out of 1039 patients were followed up at 5 years.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 12 - 12
1 May 2012
Simpson DJ Kendrick B Price AJ Murray D Gill H
Full Access

Introduction

Unicompartmental Knee Replacement (UKR) is an appealing alternative to Total Knee Replacement (TKR) when the patient has isolated compartment osteoarthritis (OA). A common observation post-operatively is radiolucency between the tibial tray wall and the bone. In addition, some patients complain of persistent pain following implantation with a UKR; this may be related to elevated bone strains in the tibia. The aim of this study was to investigate the mechanical environment of the tibia bone adjacent to the tray wall, following UKR, to determine whether this region of bone resorbs, and how altering the mechanical environment affects tibia strains.

Materials and methods

A finite element (FE) model of a cadaver tibia implanted with an Oxford UKR was used in this study, based on a validated model. A single static load, measured in-vivo during a step-up activity was used. There was a 1 mm layer of cement surrounding the keel in the cemented UKR, and the cement filled the cement pocket. In accordance with the operating procedure, no cement was used between the tray wall and bone. For the cementless UKR a layer of titanium filled the cement pocket. An intact tibia was used to compare to the cemented and cementless UKR implanted tibiae. The tibia was sectioned by the tray wall, defining the radiolucency zone (parallel to the vertical tray wall, 2 mm wide with a volume of 782.5 mm3), corresponding to the region on screened x-rays where radiolucencies are observed. Contact mechanics algorithms were used between all contacting surfaces; bonded contact was also introduced between the tray wall and adjacent bone, simulating a mechanical tie between them. Strain energy density (SED), was compared between the intact and implanted tibia for the radiolucency zone. Equivalent strains were compared on the proximal tibia between the intact and implanted tibia models. Forty patients (20 cemented, 20 cementless) who had undergone UKR were randomly selected from a database, and assessed for radiolucency.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 24 - 24
1 May 2012
Bottomley N Javaid M Gill H Dodd C Murray D Beard D Price A
Full Access

Introduction

Anteromedial gonarthrosis is a common well described pattern of knee osteoarthritis with cartilage wear beginning in the anteromedial quadrant of the medial tibial plateau in the presence of an intact and functioning ACL. It is well known that mechanical factors such as limb alignment and meniscal integrity affect the progression of arthritis and there is some evidence that the morphology of the tibial plateau may be a risk factor in the development of this disease. The extension facet angle is the angle of the downslope of the anterior portion of the medial tibial plateau joint surface in relation to the middle portion on a sagittal view. If this is an important factor in the development of AMG there may be potential for disease modifying intervention.

This study investigates if there is a significant difference in this angle as measured on MRI between a study cohort with early AMG (partial thickness cartilage damage and intact ACL) and a comparator control cohort of patients (no cartilage damage and ACL rupture).

Methods

3 Tesla MRI scans of 99 patients; 54 with partial thickness cartilage damage and 44 comparitors with no cartilage damage (acute ACL rupture) were assessed. The extension facet angle was measured (Osirix v3.6) using a validated technique on two consecutive MRI T2 sagittal slices orientated at the mid-coronal point of the medial femoral condyle. (InterClass Correlation 0.95, IntraClass Correlation 0.97, within subject variation of 1.1° and coefficient of variation 10.7%). The mean of the two extension angle values was used. The results were tabulated and analysed (R v2.9.1).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 19 - 19
1 Mar 2012
Bottomley N Kendrick B Ferguson J Al-Ali S Dodd C Murray D Beard D Price A
Full Access

Introduction

Total knee arthroplasty (TKA) accounts for 84% of all knee replacement surgery in the UK (NJR 2009) despite published epidemiological data showing that single compartment disease is most prevalent. We investigated this incompatibility further by describing the compartmental pattern and stage of cartilage loss of all patients with osteoarthritis (OA) presenting to a specialist knee clinic over one year.

Methods

All new primary referrals in a calendar year by local General Practitioners to knee clinic at a United Kingdom Hospital were assessed. Tertiary referrals and second opinions were excluded. The final diagnosis after all imaging was recorded and tabulated. The standing AP, lateral and skyline radiographs of all cases of arthritis were scored to assess the pattern of disease.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 25 - 25
1 Mar 2012
Pandit H Jenkins C Gill H Beard D Price A Dodd C Murray D
Full Access

Introduction

The results of the mobile bearing Oxford unicompartmental knee replacement (UKR) in the lateral compartment have been disappointing with a five year survival of 82%. Therefore, it is recommended that mobile bearings should not be used for lateral UKR. This low survivorship is primarily due to high dislocation rate, all occurring in the first year. A detailed analysis of the causes of bearing dislocation confirmed the elevated lateral tibial joint line to be a contributory factor. A new surgical technique was therefore introduced in which care was taken neither to remove too much bone from distal femur nor to over tighten the knee and thus ensure that the tibial joint line was not elevated. Other modifications to the technique were also introduced including use of a domed tibial component.

Aim

The aim of this study is to compare the outcome of these iterations: the original series [series I], Series II with improved surgical technique and the domed tibial component [Series III].


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 412 - 418
1 Mar 2012
Judge A Arden NK Kiran A Price A Javaid MK Beard D Murray D Field RE

We obtained information from the Elective Orthopaedic Centre on 1523 patients with baseline and six-month Oxford hip scores (OHS) after undergoing primary hip replacement (THR) and 1784 patients with Oxford knee scores (OKS) for primary knee replacement (TKR) who completed a six-month satisfaction questionnaire.

Receiver operating characteristic curves identified an absolute change in OHS of 14 points or more as the point that discriminates best between patients’ satisfaction levels and an 11-point change for the OKS. Satisfaction is highest (97.6%) in patients with an absolute change in OHS of 14 points or more, compared with lower levels of satisfaction (81.8%) below this threshold. Similarly, an 11-point absolute change in OKS was associated with 95.4% satisfaction compared with 76.5% below this threshold. For the six-month OHS a score of 35 points or more distinguished patients with the highest satisfaction level, and for the six-month OKS 30 points or more identified the highest level of satisfaction. The thresholds varied according to patients’ pre-operative score, where those with severe pre-operative pain/function required a lower six-month score to achieve the highest levels of satisfaction.

Our data suggest that the choice of a six-month follow-up to assess patient-reported outcomes of THR/TKR is acceptable. The thresholds help to differentiate between patients with different levels of satisfaction, but external validation will be required prior to general implementation in clinical practice.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 34 - 34
1 Mar 2012
Ferguson J Pandit H Price A Marks B Gill H Murray D Dodd C
Full Access

Introduction

Obesity has been considered a relative contra-indication in unicompartmental knee arthroplasty (UKA) due to fear of high wear rates, loosening and tibial collapse. The aim of this study was to investigate the impact of high body mass index (BMI) on ten-year survivorship and five-year functional outcome after Oxford UKA, a fully congruous mobile bearing design with large contact area and low wear rate.

Methods

This prospective study examines a consecutive series of 595 knees (mean age 66 years, range: 33-88) undergoing Oxford UKA with a minimum 5-year follow-up. Patients were divided into three groups; Group I (Normal body weight), BMI <25 (n=171), Group II (overweight), BMI 25- 30 (n=264), and Group III (Obese), BMI ≥30 (n=160). The survivorship and functional outcome (as assessed by change in Oxford Knee Score [DeltaOKS]) and Knee Society Score (KSS) for all three groups were compared.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 32 - 32
1 Mar 2012
Kendrick B Simpson D Bottomley N Kaptein B Garling E Gill H Dodd C Murray D Price A
Full Access

Purpose of study

To investigate the linear penetration rate of the polyethylene bearing in unicompartmental knee arthroplasty at twenty years.

Introduction

The Phase 1 Oxford medial UKR was introduced in 1978 as a design against wear, with a fully congruous articulation. In 1987 the Phase 2 implant was introduced with new instrumentation and changes to the bearing shape. We have previously shown a linear penetration rate (LPR) of 0.02 mm/year at ten years in Phase 2, but that higher penetration rates can be seen with impingement. The aim of this study was to determine the 20 year in-vivo LPR of the Oxford UKR, using Roentgen Stereophotogrammetric Analysis (RSA).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 16 - 16
1 Mar 2012
Rout R Mcdonnell SM Jayadev C Xia Z Murray D Hulley PA Price AJ
Full Access

Aim

This aim of this study was to investigate apoptosis, reactive oxygen species (ROS), and their upstream markers in Anteromedial Gonarthrosis (AMG).

Methods

Ten resection specimens, from patients undergoing unicompartmental knee replacement for AMG, and ten control specimens, collected from vascular disease patients undergoing above knee amputation, were used. Routine histology and immunohistochemical studies were conducted for Terminal deoxynucleotidyl transferase dUTP nick end labeling (TUNEL), Active Caspase 3, Cytochrome C, Active Bax, Bim, 3-Nitrotyrosine and Forkhead box O3A (FOXO 3A).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 27 - 27
1 Mar 2012
Pandit H Jenkins C Weston-Simons S Ferguson J Popat BM Gibbons M Price A Gill H Dodd C Murray D
Full Access

Introduction

Kozinn and Scott have made recommendations about contra-indications for unicompartmental knee replacement (UKR). They suggest that patients younger than 60, weight > 82 kilograms, patients with exposed bone in patella-femoral compartment or patients who are physically active/perform heavy labour should not be offered a UKR. In addition, chondrocalcinosis is a contra-indication. These strict selection criteria are based on the experience with fixed bearing UKAs and are more intuitive than evidence based. The Oxford UKR has a fully congruous mobile bearing and has been shown to have minimal wear. Over the past 25 years, the Oxford Group has followed a standardised protocol for patient selection for UKR. We ignore patella-femoral joint pathology, chondrocalcinosis, patient's age, weight and activity level when deciding the suitability for UKR.

Methods

Using the standardised indications, more than 1100 Oxford UKRs have been performed to date over the last 10 years. These patients are assessed pre-operatively and at regular intervals post-operatively in a dedicated research clinic. We present the results of these consecutive cases. Patients were classified into two groups: group I (satisfy Kozinn-Scott recommendations) and group II (outside recommendations).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IV | Pages 26 - 26
1 Mar 2012
Steffen R O'Rourke K Murray D Gill H
Full Access

In 12 patients, we measured the oxygen concentration in the femoral head-neck junction during hip resurfacing through the anterolateral approach. This was compared with previous measurements made for the posterior approach. For the anterolateral approach, the oxygen concentration was found to be highly dependent upon the position of the leg, which was adjusted during surgery to provide exposure to the acetabulum and femoral head. Gross external rotation of the hip gave a significant decrease in oxygenation of the femoral head. Straightening the limb led to recovery in oxygen concentration, indicating that the blood supply was maintained. The oxygen concentration at the end of the procedure was not significantly different from that at the start.

The anterolateral approach appears to produce less disruption to the blood flow in the femoral head-neck junction than the posterior approach for patients undergoing hip resurfacing. This may be reflected subsequently in a lower incidence of fracture of the femoral neck and avascular necrosis.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 31 - 31
1 Mar 2012
Kendrick B Pandit H Jenkins C Beard D Gill H Price A Dodd C Murray D
Full Access

Purpose of Study

To assess the incidence of radiolucency in cemented and cementless Oxford unicompartmental knee replacement at two years.

Introduction

Most unicompartmental knee replacements (UKRs) employ cement for fixation of the prosthetic components. The information in the literature about the relative merits of cemented and cementless UKR is contradictory, with some favouring cementless fixation and others favouring cemented fixation. In addition, there is concern about the radiolucency that frequently develops beneath the tibial component with cemented fixation. The exact cause of the occurrence of radiolucency is unknown but it has been hypothesised that it may suggest suboptimal fixation.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_II | Pages 13 - 13
1 Feb 2012
Steffen R Smith S Gill H Beard D McLardy-Smith P Urban J Murray D
Full Access

This study aims to investigate femoral blood flow during Metal-on-Metal Hip Resurfacing (MMHR) by monitoring oxygen concentration during the operative procedure.

Patients undergoing MMHR using the posterior approach were evaluated. Following division of fascia lata, a calibrated gas-measuring electrode was inserted into the femoral neck, aiming for the supero-lateral quadrant of the head. Baseline oxygen concentration levels were detected after electrode insertion 2-3cm below the femoral head surface and all intra-operative measures were referenced against these. Oxygen levels were continuously monitored throughout the operation. Data from ten patients are presented.

Oxygen concentration dropped most noticeably during the surgical approach and was reduced by 62% (Std.dev +/-26%) following dislocation and capsulectomy. Insertion of implants resulted in a further oxygenation decrease by 18% (Std.dev +/-28%). The last obtained measure before wound closure detected 22% (Std.dev +/-31%) of initial baseline oxygen levels. Variation between subjects was observed and three patients demonstrated a limited recovery of oxygen levels during implant insertion and hip relocation.

Intra-operative measurement of oxygen concentration in blood perfusing the femoral head is feasible. Results in ten patients undergoing MMHR showed a dramatic effect on the oxygenation in the femoral head during surgical approach and implant fixation. This may increase the risk of avascular necrosis and subsequent femoral neck fracture. Future experiments will determine if less invasive procedures or specific positioning of the limb can protect the blood supply to femoral neck and head.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 546 - 546
1 Nov 2011
Grammatopoulos G Pandit H Taylor A Whitwell D Glyn-Jones S Gundle R McLardy-Smith P Murray D Gill H
Full Access

Introduction: Metal-on-metal-hip-resurfacing-arthroplasty (MoMHRA) has been associated with the development of inflammatory pseudotumours(IP), especially in females. IPs have been linked to wear debris, which can be related to metal-ion blood levels. Acetabular component position has been shown to influence wear. We have identified an optimum component orientation minimising IP risk around an inclination/anteversion of 40°/20°±10°. Our aim was to see if this optimal position results in lower metal ions and to identify the boundary of an optimal placement zone for low wear.

Methods: A cohort of 104 patients(60M:44F) with unilateral MoMHRA was studied. Blood tests were obtained at a mean follow up of 3.9 years and serum Co/Cr levels were measured(ICPMS). High metal ion concentrations were defined as Co> 4.1ppb and Cr> 5.2ppb. Radiographic cup inclination/anteversion were measured using EBRA. The differences in ion levels between different orientation zones were investigated. Three orientation zones were defined centered on 40°/20°: Z1-within ±5°, Z2-outside ±5°/within ±10° and Z3-within ±10°.

Results: There was a wide range of cup placements. Females had significantly (p< 0.001) smaller components(mean:51, 44–60) than males(mean: 56, 52–64). Cr levels, but not Co, were higher in females(p=0.002) and those with small femoral components(< 50mm, p =0.03). Patients with cups within Z1 (n=13) had significantly lower Co(p=0.005) and Cr(p=0.001). Males with cups within Z3(n=27) had lower ion levels in comparison to those outside, which were significantly lower for Co(p=0.049) but not Cr(p=0.084). Females had similar ion levels within and out of Z3(Cr/Co: p=0.83/0.84). Co levels were significantly lower in Z1(n=13) in comparison to Z2(n=33)(p=0.048) but Cr levels were not different (p=0.06).

Discussion: MoMHRA cups placed within ±5° of the optimum(40°/20°) had significantly lower metal ions indicating lower wear within this narrow zone. This safe zone, could extend to ±10° for males only. The narrower safe zone coupled with smaller components implanted are possible factors contributing to the increased IP incidence seen in females.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 550 - 550
1 Nov 2011
Glyn-Jones S Roques A Esposito C Walter W Tuke M Murray D
Full Access

Introduction: Metal on metal hip resurfacing arthroplasty-induced pseudotumours are a serious complication, which occur in 1.8% of patients who undergo this procedure in our institution. The aim of this study was to measure the 3D in vivo wear on the surface of resurfacing components revised for pseudotumour, compared to a control group.

Method: Thirty-six hip resurfacing implants were divided into two groups; 18 patients with a clinical and histopathological diagnosis of pseudotumour and 18 controls (revised for femoral neck fracture and infection). Three dimensional contactless metrology (Redlux Ltd) was used to scan the surface of the femoral and acetabular components, to a resolution of 20 nanometers. The location, depth and area of the wear scar were determined for each component. A separate blinded analysis to determine the presence of absence of impingement was performed by one of the authors.

Results: The volumetric wear rate for femoral component of the pseudotumour group was 3.29 mm3/yr (SD5.7) and 0.79 mm3/yr (SD1.2) for the control group (p=0.005). In the pseudotumour group, the volumetric wear rate of the acetabular component was 2.5 mm3/yr (SD6.9) compared to 0.36 mm3/yr (SD0.80) for the control group (p=0.008). Edge-wear was detected in 89% of acetabular components in the pseudotumour group and 21% of those in the control group (p=0.01). Anterior or posterior edge-wear, consistent with impingement was present on the femoral components of 73% of patients in the pseudotumour group and 22% in the control group (p=0.01).

Discussion: This work demonstrates that implants revised for pseudotumour have significantly higher volumetric wear rates than controls. They also have a significantly higher incidence of edge-wear than controls. We suggest that a significant proportion of pseudotumours are associated with high concentrations of metal wear debris; however a minority may result from a hypersensitivity reaction to metal ions.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 547 - 547
1 Nov 2011
Thomas G Simpson D Taylor A Whitwell D Gibbons C Gundle R Mclardy-smith P Gill H Glyn-jones S Murray D
Full Access

Introduction: The use of highly cross-linked polyethylene (HXLPE) is now commonplace for total hip arthroplasty, however there is no long-term data to support its use. Hip simulator studies suggest that the wear rate of some types of HXLPE is ten times less than conventional polyethylene (UHMWPE). The outcomes of hip simulator studies are not always reproduced in vivo and there is some evidence that HXLPE wear may increase between 5 and 7 years.

Method: A prospective double blind randomised control trial was conducted using Radiostereometric Analysis (RSA). Fifty-four subjects were randomised to receive hip replacements with either UHMWPE liners or HXLPE liners. All subjects received a cemented CPT stem and uncemented Trilogy acetabular component (Zimmer, Warsaw, IN, USA). The 3D penetration of the head into the socket was determined to a minimum of 7 years.

Results: The total liner penetration was significantly different at 7 years (p=0.005) with values of 0.33 mm (SE 0.05 mm) for the HXLPE group and 0.55 mm (SE 0.05 mm) for the UHMWPE group. The steady state wear rate from 1 year onwards was significantly lower for HXLPE (0.005 mm/yr, SE 0.007 mm/yr) than for UHMWPE (0.037 mm/yr, SE 0.009 mm/yr) (p=0.007). The direction of wear was supero-lateral.

Discussion: We have previously demonstrated that the penetration in the first year is creep-dominated, from one year onwards the majority of penetration is probably due to wear. This study confirms the predictions from hip simulator studies which suggest that the wear rate of this HXLPE approaches that of metal-on-metal and ceramic-on-ceramic articulations. HXLPE may have the potential to reduce the need of revision surgery, due to wear debris induced osteolysis. It may also enable surgeons to use larger couples, thus reducing the risk of impingement and dislocation.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 546 - 546
1 Nov 2011
Grammatopoulos G Pandit H Taylor A Whitwell D Glyn-Jones S Gundle R McLardy-Smith P Gill H Murray D
Full Access

Introduction: Metal on metal hip resurfacing arthroplasty(MoMHRA) is an alternative option to THR in the treatment of young adults with OA. A recognised MoMHRA complication is the development of an inflammatory pseudotumour(IP). Diagnosis is made with the aid of US and/or MRI. To-date, no radiographic indication of the presence of IP has been identified. Neck thinning is a recognised phenomenon in MoMHRA hips not associated with any adverse clinical events. Its pathogenesis is considered multi-factorial. Our aim was to establish whether excessive neck narrowing is associated with the presence of a pseudotumour.

Methods: Twenty-seven hips (26 patients) with IP confirmed clinically, radiologically, intra-operatively and histologically were matched for sex, age, pre-operative diagnosis, component size and follow-up with an asymptomatic MoMHRA cohort (Control n=60). For all patients, prosthesis-neck-ratio(PNR) was measured on plain AP pelvic radiographs post-operatively and at follow-up as previously described and validated.

Results: All IP patients (4M:23F) and all (12M:48F) but two controls had a posterior approach at the time of MoMHRA. Post-operatively, there was no difference in the PNR between the two groups (p=0.19). At an average follow up of 3.5 years (range:0.7–8.3), IP patients(mean 1.26, 1.10–1.79) had a significantly higher (p< 0.0001) PNR in comparison to their controls(mean 1.14, 1.03–1.35). Greater neck narrowing occurred in both genders. IP necks had narrowed by an average of 8% (range:3–23). The degree of neck narrowing was correlated with length of survival of implant (p=0.001).

Discussion: This study shows a strong association between IP and neck narrowing. Processes such as impingement and increased wear are considered to be involved in the pathogenesis of both IP and neck narrowing. Furthermore, the presence of an IP, could lead to altered vascularity via a mass effect and further contribute to neck narrowing. Neck narrowing in symptomatic MOMHRA patients should alert surgeons of the possible presence of IP.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 298 - 298
1 Jul 2011
Glyn-Jones S Pandit H Doll H McLardy-Smith P Gundle R Gibbons M Athanasou N Ostlere S Whitwell D Taylor A Gill R Murray D
Full Access

Metal on metal hip resurfacing (MMHR) is a popular procedure for the treatment of osteoarthritis in young patients. Several centres have observed masses, arising from around these devices, we call these inflammatory pseudotumours. They are locally invasive and may cause massive soft tissue destruction. The aim of this study was to determine the incidence and risk factors for pseudotumours that are serious enough to require revision surgery.

In out unit, 1,419 MMHRs were performed between June 1999 and November 2008. All revisions were identified, including all cases revised for pseudotumour. Pseudotumour diagnosis was made by histological examination of samples from revision. A Kaplan-Meier survival analysis was performed, Cox regression analysis was used to estimate the independent effects of different factors.

The revision rate for pseudotumour increased with time and was 4% (95% CI: 2.2% to 5.8%) at eight years. Female gender was a strong risk factor: at eight years the revision rate for pseudotumours in men was 0.5% (95% CI 0% to 1.1%), in women over 40 it was 6% (95% CI 2.3% to 10.1%) and in women under 40 it was 25% (95% CI 7.3% to 42.9%) (p< 0.001). Other factors associated with an increase in revision rate were, small components (p=0.003) and dysplasia (p=0.019), whereas implant type was not (p=0.156).

We recommend that resurfacings are undertaken with caution in women, especially those younger than 40 years of age, but they remain a good option in men. Further work is required to understand the patho-aetiology of pseudotumours so that this severe complication can be avoided.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 171 - 171
1 May 2011
Kwon Y Mellon S Murray D Gill H
Full Access

Introduction: Edge-loading, a phenomenon whereby the femoral component comes into contact with the edge of the acetabular component, has been suggested to increase wear in metal-on-metal hip resurfacing arthroplasty (MoMHRA). Pseudotumours (soft-tissue mass relating to the hip joint) have been associated with elevated serum and hip aspirate metal ion levels. This study aimed to investigate in vivo edge-loading in MoMHRA patients with pseudotumours by quantifying dynamic loci of the hip joint segment force relative to the acetabular component during functional activities.

Materials and Methods: A total of 21 MoMHRA patients (30 hips) in two groups were investigated in this Ethics approved study:

6 patients with pseudo-tumours detected using ultrasound/MRI;

15 patients without pseudotumours.

Three-dimensional lower limb motion analysis (12 camera Vicon System) was performed to estimate hip joint segment force during walking, chair-rising and stair-climbing. CT scans were used to determine each patient’s specific hip joint centre and acetabular component orientation. Edge-loading was defined to occur when a hip joint segment force vector/ cup intersection was located within 10% of the cup radius from the edge of the cup. Serum cobalt and chromium levels were analysed using Inductively-Coupled Plasma Spectrometer.

Results: Edge-loading in the pseudotumour group occurred with significantly (p=0.02) longer (4-fold increase) duration as well as greater magnitude (7-fold increase) of force, compared to the non-pseudotumour group. The duration and force of the edge-loading were activity-dependent, with proportionally greater difference observed during stair climbing. The acetabular cup orientation values in the pseudotumour group were found within the safe zone of Lewinnek in one third of the hips with the remaining two thirds outside the safe zone. The presence of pseudotumour was associated with:

significantly higher median serum cobalt levels: 14.3ug/l (range 10.6–64.1) vs. 1.9ug/l (range 1.2–5.0), p< 0.001;

significantly higher median serum chromium levels: 21.2ug/l (range 13.8–45.2) vs. 1.8ug/l (range 0.7–7.6), p< 0.001.

Discussion: Edge-loading in MoMHRA patients with pseudotumours occurred in vivo with significantly longer duration and greater magnitude of force impulse compared to the patients with a well functioning MoMHRA during activities of daily living. This suggests that edge-loading may be an important mechanism that leads to localised high wear, with subsequent elevation of metal ion levels in MoMHRA patients with pseudotumours. Although the acetabular component malposition, such as increase in both inclination and anteversion angles, appears to be an important factor in edge-loading, the aetiology of edge-loading is likely to be multi-factorial.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 173 - 173
1 May 2011
Simpson D Kueny R Murray D Zavatsky A Gill H
Full Access

Introduction: A unique failure mode of hip resurfacing is femoral neck fracture. These tend to occur early after surgery during normal activities. One theory regarding fracture occurrence includes the introduction of stress magnifiers in the form of notches on the superior neck. The presence of a notch can arise from reaming or from removal of osteophytes during surgery. The aim of the present study was to investigate the effect of notching the femoral neck, following resurfacing by using a finite element (FE) model.

Methods: A physiological load case was simulated in the FE model of a femur, implanted with a cemented hip resurfacing system. Twelve implant alignments were modelled: an ideal implant alignment with no notch, and a 1 mm, 3 mm, 5 mm and 7 mm superior notch; 5° anteversion, 5° and 10° degrees retroversion; 5° and 10° degrees in varus and valgus. These models were compared to that of an intact femur for baseline analysis.

The intact femur geometry was derived from a CT dataset of a cadaveric femur and CT numbers were converted into a realistic distribution of material properties. The FE intact mesh was based on an experimentally validated mesh of a human femur. The femur was segmented into 22 neck sections.

The loading condition was modelled to represent an instant at 10% of gait where all muscle forces were included. The femoral neck regions were compared between the models to evaluate the effect of notch sizes on stress distribution. Maximum tensile stresses were compared to the ultimate tensile stress (UTS) of cortical and cancellous bone.

Results: As the notch size increased the peak and average 1st (tensile) and 3rd (compressive) principal stress increased along the superior portion of the femoral neck. For the 5 mm superior notch, the maximum 1st principal stress increased by 283% and 154% when compared to that of the ideally aligned implant and the intact femur respectively. The largest increase of tensile stress was observed when the implant was mal-aligned in 10° of varus; this resulted in a 768% increase in stress compared to the ideally implanted model.

Discussion: The introduction of a superior notch causes a stress concentration on the femoral neck. Although the stress concentration is pronounced, a notch on the superior aspect of the femoral neck may not lead to fracture following resurfacing; the UTS of cortical bone is 100MPa, and the UTS of cancellous bone is between 2MPa and 20MPa. Peak stresses in the model are well below the UTS of cortical bone, and for damage to accumulate in cancellous bone, energy absorption in the ‘honey-comb’ structure of trabecular bone must be considered. Varus mal-alignment resulted in the largest increase in tensile stress on the superior aspect of the neck, and has been associated with femoral neck fracture; this type of mal-alignment may be critical when considering femoral neck fractures.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 210 - 210
1 May 2011
Simpson D Kendrick B O’Connor J Pandit H Dodd C Murray D
Full Access

Introduction: The results of the mobile bearing Oxford partial knee replacement (PKR) in the lateral compartment have been disappointing with a five year survival of 82%. Bearing dislocation is a particular concern, and to address this issue a new domed implant was introduced with a modified surgical technique. The aim of this study was to compare the risk of dislocation between a domed and flat lateral PKR.

Methods: Separate kinematic models were generated for the domed and flat bearings. The femoral component, tibial tray and bearing were aligned in a neutral position; the flat bearing was positioned centrally on the tibial tray and 2 mm from the side wall; the domed bearing was placed concentrically on the domed tibial tray. Dislocation in the Posterior (A-P), Lateral (M-L) and Medial against the tray wall (L-M-wall) were investigated. For each dislocation the tibial tray was restrained in all degrees of freedom (DOF) and the femoral component was restrained in five DOF; A-P and M-L displacements; A-P, M-L and Superior-Inferior (S-I) rotations. The bearing was restrained from rotating about the S-I axis for each dislocation. For the L-M-wall dislocation the underside of the bearing was held in contact with the tibial tray wall such that the lowest S-I displacement of the femoral component was achieved. The least amount of distraction required for bearing dislocation to occur was calculated for the seven bearing sizes available. The effect of medial-lateral positioning of the femur on dislocation was investigated.

Results: The minimum femur distraction to cause A-P flat and domed bearing dislocation ranged from 4.68mm to 3.91mm and 6.29mm to 5.59mm respectively as the bearing thickness increased from 3.5mm to 9.5mm. The minimum femur distraction to cause L-M-wall flat and domed bearing dislocation ranged from 3.42mm to 4.16mm and 4.55mm to 5.44mm respectively as the bearing thickness increased from 3.5 mm to 9.5 mm. The femur distraction required for L-M-wall bearing dislocation increased from 4.55mm to 6.3mm with a 2 mm medial movement of the femoral component. A 4 mm lateral movement of the femoral component decreased the distraction from 4.55mm to 2.35mm.

Discussion: A domed bearing can lead to an increased femoral distraction of between 25% and 37%, significantly reducing the likelihood of dislocation. This may be significant during everyday activities and demonstrates that the new domed design should reduce the incidence of bearing dislocation by increasing the amount of entrapment; our current series of 200 patients has no dislocations. Increasing the thickness of the bearing has a small effect on the distraction required to allow bearing dislocation. The medial-lateral placement of the femoral component has a pronounced effect on the femoral distraction required for bearing dislocation over the tray wall; medial placement of the femoral component is advisable.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 96 - 96
1 May 2011
Bottomley N Javaid M Judge A Gill H Murray D Beard D Price A
Full Access

Introduction: Anteromedial gonarthrosis is a common well described pattern of knee osteoarthritis with cartilage wear beginning in the anteromedial quadrant of the medial tibial plateau in the presence of an intact and functioning ACL. It is well known that mechanical factors such as limb alignment and meniscal integrity affect the progression of arthritis and there is some evidence that the morphology of the tibial plateau may be a risk factor in the development of this disease. The extension facet angle is the angle of the downslope of the anterior portion of the medial tibial plateau joint surface in relation to the middle portion on a sagittal view. If this is an important factor in the development of AMG there may be potential for disease modifying intervention.

This study investigates if there is a significant difference in this angle as measured on MRI between a study cohort with early AMG (partial thickness cartilage damage and intact ACL) and a comparator control cohort of patients (no cartilage damage and ACL rupture).

Methods: 3 Tesla MRI scans of 99 patients; 54 with partial thickness cartilage damage and 44 comparitors with no cartilage damage (acute ACL rupture) were assessed. The extension facet angle was measured (Osirix v3.6) using a validated technique on two consecutive MRI T2 sagittal slices orientated at the mid-coronal point of the medial femoral condyle. (InterClass Correlation 0.95, IntraClass Correlation 0.97, within subject variation of 1.1° and coefficiant of variation 10.7%). The mean of the two extension angle values was used. The results were tabulated and analysed (R v2.9.1).

Results: Of the 99 knees, 38 were female and 61 male; 44 left knees and 55 right. The mean extension facet angle for the partial thickness group was 12.7° (SD 3.35) and for the comparator group 8.7° (SD 3.09). There was a significant difference between these 2 groups (Mann Whitney U, p< 0.001). Although there were significantly more men than women in the comparator group, stratification analysis showed that there was no effect of gender on the mean extension facet angle.

Discussion: There is a significance difference in the extension facet angle between patients with AMG with only partial thickness cartilage loss and a comparator group. This has not been shown in a study group of this size before. Since none of the subjects had full thickness cartilage loss it is unlikely that this difference is due to bone attrition changing the angle as part of the disease process but this is an important area for further study. We believe that a higher medial tibial extension facet angle alters the mechanics within the medial compartment, placing these patients at higher risk of developing AMG. This may present an opportunity for risk factor modification, for example osteotomy.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 221 - 222
1 May 2011
Kwon Y Glyn-Jones S Simpson D Kamali A Counsell L Mclardy-Smith P Beard D Gill H Murray D
Full Access

Introduction: Pseudotumours (soft-tissue masses relating to the hip joint) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated serum and hip aspirate metal ion levels, suggesting that pseudotumours occur when there is increased wear. This study aimed to quantify in vivo wear of implants revised for pseudotumours and a control group of implants revised for other reasons of failure.

Methods: A total of 30 contemporary MoMHRA implants in two groups were investigated in this Institutional Review Board approved study:

8 MoMHRA implants revised due to pseudotumour;

22 MoMHRA implants revised due to other reasons of failure (femoral neck fracture and infection).

The linear wear of retrieved implants was measured using a Taylor-Hobson Roundness machine. The average linear wear rate was defined as the maximum linear wear depth divided by the duration of the implant in vivo.

Results: In comparison with the non-pseudotumour implant group, the pseudotumour implant group was associated with:

significantly higher median linear wear rate of the femoral component: 8.1um/year (range 2.75–25.4um/year) vs. 1.79um/year (range 0.82–4.15um/year), p=0.002; and

significantly higher median linear wear rate of the acetabular component: 7.36um/year (range1.61–24.9um/year) vs. 1.28um/year (range 0.18–3.33um/year), p=0.001.

Similarly, differences were also measured in absolute wear values. The median absolute linear wear was significantly higher in the pseudotumour implant group:

21.05um (range 2.74–164.80um) vs. 4.44um (range 1.50–8.80um) for the femoral component, p=0.005; and

14.87um (range 1.93–161.68um) vs. 2.51um (range 0.23–6.04um) for the acetabular component, p=0.008.

Wear on the acetabular cup components in the pseudotumour group always involved the edge, indicating edge-loading of the bearing. In contrast, edge-loading was observed in only one acetabular component in the non-pseudotumour group of implants. The deepest wear was observed well within the bearing surface for the rest of the non-pseudotumour group. The difference in the incidence of edge-loading between the two groups was statistically significant (Fisher’s exact test, p=0.03).

Discussion: Significantly greater linear wear rates of the MoMHRA implants revised due to pseudotumour support the in vivo elevated metal ion concentrations in patients with pseudotumours. This study provides the first direct evidence to confirm that pseudotumour is associated with increased wear at the MoM articulation. Furthermore, edge-loading with the loss of fluid film lubrication may be the dominant wear generation mechanism in patients with pseudotumour.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 208 - 208
1 May 2011
Kendrick B Simpson D Gill H Valstar E Kaptein B Dodd C Murray D Price A
Full Access

Introduction: Approximately 20% of unicompartmental knee replacement (UKR) revisions are related to polyethylene wear. The Phase 1 Oxford UKR was introduced as a design against wear, with a fully congruent mobile bearing. The Phase 2 implant was introduced with new instrumentation (femoral mill) and changes to the bearing shape (lower anterior wall) to reduce the incidence of anterior impingement. We have previously shown that the Oxford UKR has a wear rate of 0.02 mm/year at ten years, in well functioning devices, but that higher wear rates can be seen with impingement or if the congruous articulation is lost. The aim of this study was to determine the 20 year in-vivo wear of the Oxford Phase 1 and Phase 2 UKR, using Roentgen Stereophotogrammetric Analysis (RSA).

Method: We measured the in-vivo wear of 6 Phase 1 (5 patients, mean age 65.24 years) and 7 Phase 2 (4 patients, mean age 63.43) Oxford UKR bearings. Average time since surgery was 22.37 years and 19.46 years for the Phase 1 and Phase 2 implants respectively. Selection criteria included patients who were mobile, with an exercise tolerance greater than 100m as per the American Knee Society Score (AKSS) functional questionnaire. RSA x-rays were taken with the knee in the normal anatomical position on standing and with the knee flexed to 30o. The Oxford knee score (OKS) and AKSS were gained at the RSA examination. Phase 1 and 2 components were reverse engineered by laser scanning, and converted to CAD models. The CAD models of the tibia and femur were pose-estimated in the RSA software (Medis Specials, Leiden, Netherlands). A sphere was fit to the femoral component and the minimum bearing thickness was determined by measuring the shortest perpendicular distance between the sphere and the plane contained on the tibial tray articular surface. The linear wear for each bearing was calculated by subtracting the measured thickness from the corrected nominal bearing thickness. Non-parametric statistics were used to compare the two Phases.

Results: There was no significant difference in age, OKS and AKSS between the two groups. The median wear rate was 0.078 mm/year for Phase 1 and 0.023 mm/year for Phase 2. This difference was statistically significant (p = 0.027).

Discussion: The difference in wear rate is explained by impingement in Phase 1, which was reduced by design changes with the introduction of Phase 2; the Phase 2 is designed to avoid impingement between the femur and the bearing. This study demonstrates that very low wear rates can be maintained with the Phase 2 implant to the end of the second decade after implantation. This is of particular importance when the device is used in younger patients and demonstrates that the Oxford UKR can be a definitive implant for the treatment of isolated compartmental osteoarthritis.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 182 - 182
1 May 2011
Simpson D Kendrick B Gill H Pandit H Dodd C Price A Murray D
Full Access

Introduction: Partial Knee Replacement (PKR) is an appealing alternative to Total Knee Replacement (TKR) when the patient has isolated compartment osteoarthritis (OA). In nearly all cases there is a radiolucency observed between the tibial tray wall and the boney interface. The reasons why radiolucencies appear are unknown, but the bone will adapt to its altered mechanical environment by bone remodelling in accordance with ‘Wollf’s Law’. The aim of this study was to investigate the mechanical environment of the tibia bone adjacent to the tray wall, following cemented and cementless PKR, in order to determine whether this region of bone resorbs.

Methods: A validated finite element (FE) model of a cadaver tibia implanted with an Oxford PKR was used in this study. Kinematic data from fluoroscopy measurements during a step-up activity were used to determine the relative tibio-femoral positioning for the Oxford PKR model. Load data were adapted from the in-vivo measured loads using an instrumented implant during a step-up activity. The standard operating protocol was simulated for the Oxford PKR FE models, with the tibial tray implanted in a neutral position. The tibia was sectioned around the tray. Zone 7 was defined as parallel to the vertical tray wall, corresponding to the region on screened x-rays where radiolucencies are observed. It was assumed that the bone in the implanted tibia will attempt to normalise its stress-strain patterns locally to its equilibrium state, the intact tibia, for the same loading conditions. Forty patients (20 cemented, 20 cementless) who had undergone PKR were randomly selected from a database, and their screened x-rays assessed for radiolucency in region 7.

Results: The SED in region 7 was 80% lower in the cemented and cementless tibia, compared to the intact tibia (Figure 2). The maximum tensile stress was 63% lower in the cemented and cementless tibia, compared to the intact tibia. The corresponding maximum compressive stress was 52% lower. Radiolucency was observed in all forty radiographs in region 7.

Discussion: After implantation with a cemented or cementless PKR the bone strains and SED in region 7 are reduced. This reduction may provide the signal for adaptive bone remodelling and bone will be resorbed from this region, decreasing the volume and increasing the SED. Bone resorption will continue until the equilibrium state is reached. If a ‘lazy’ zone between 35% and 50% of the remodelling signal is considered, bone resorption will still occur due to the large decrease in SED for this region. For region 7 to return its SED to the equilibrium state, its volume will need to be reduced by 80%. This is likely to be the reason why a radiolucency is observed clinically in this region in almost every case, whether a cemented or cementless implant is used.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 220 - 220
1 May 2011
Thomas G Simpson D Gill H McLardy-Smith P Murray D Glyn-Jones S
Full Access

Introduction: The use of second generation highly cross-linked polyethylene (HXLPE) is now commonplace for total hip arthroplasty, however there is no long-term data to support its use. Hip simulator studies suggest that the wear rate of HXLPE is ten times less than conventional polyethylene (UHMWPE). The outcomes of hip simulator studies are not always reproducible in vivo. Long term clinical data is required, as there is emerging clinical data, which suggests that some types of second generation HXLPE may have increased wear after 5 years.

Method: A prospective double blind randomised control trial was conducted using Radiostereometric analysis (RSA). Fifty-four subjects were randomised to receive hip replacements with either UHMWPE liners or HXLPE liners. All subjects received a cemented CPT stem and uncemented Trilogy acetabular component (Zimmer, Warsaw, IN, USA). The 3D penetration of the head into the socket was determined to a minimum of 7 years.

Results: The total liner penetration was significantly different at 7 years (p=0.01) with values of 0.33mm (SD 0.17mm) for the HXLPE group and 0.51mm (SD 0.14mm) for the UHMWPE group. The steady state wear rate from 1 year onwards was significantly lower for HXLPE (0.003 mm/yr, SD 0.04 mm/yr) than for UHMWPE (0.03 mm/yr, SD 0.03 mm/yr) (p=0.01). The direction of wear was in the antero-medial direction in both groups.

Conclusion: We have previously demonstrated that the penetration in the first year is creep-dominated, from one year onwards the majority of penetration is due to wear. The wear rate of this second generation HXLPE approaches that of metal on metal bearings. Second-generation HXLPE may have the potential to reduce the risk of revision surgery, due to wear debris induced osteolysis.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 221 - 221
1 May 2011
Glyn-Jones S Roques A Esposito C Gill H Walter W Tuke M Murray D
Full Access

Introduction: Metal on metal hip resurfacing arthroplasty-induced pseudotumours are a serious complication, which occur in 4% of patients who undergo this procedure. The aim of this study was to measure the 3D in vivo wear on the surface of resurfacing components revised for pseudotumour, compared to a control group.

Method: Thirty-nine hip resurfacing implants were examined; these were sourced from our institutions prosthesis retrieval bank. They were divided into two groups; 22 patients with a clinical and histopathological diagnosis of pseudotumour and 17 controls. Patient demographics and time to revision were known. Three dimensional contactless metrology (Redlux™ Ltd) was used to scan the surface of the femoral and acetabular components, to a resolution of 20 nanometers. The location, depth and area of the wear scar was determined for each component. Volumetric wear was determined, along with the presence of absence of edge-loading. A separate blinded analysis to determine the presence of absence of impingement was performed by one of the authors. ANOVA was used to test for differences in wear and Fishers Exact test was used to compare the incidence of edge-loading between the groups.

Results: The volumetric wear rate for femoral component of the pseudotumour group was 4.7mm3/yr (SD3.5) and 1.7 mm3/yr (SD1.5) for the control group (p=0.03). In the pseudotumour group, the volumetric wear rate of the acetabular component was 3.5 mm3/yr (SD3.6) compared to 0.02 mm3/yr (SD0.07) for the control group (p=0.01). Edge-loading was detected in 74% of acetabular components in the pseudotumour group and 22% of those in the control group (p=0.01). Anterior or posterior edge-loading, consistent with impingement was present on the femoral components of 73% of patients in the pseudotumour group and 22% in the control group (p=0.01).

Discussion: This work demonstrates that implants revised for pseudotumour have significantly higher volumetric wear rates than controls. They also have a significantly higher incidence of edge-loading and impingement than controls. Edge-loading significantly increases wear. We suggest that pseudotumours are caused by high concentrations of metal wear debris, which have been shown to have a toxic effect on osteocytes and macrophages. This is the one of the first studies to demonstrate a clear link between pseudotumours and increased bearing surface wear. It is also the first to demonstrate that edge-loading, due to impingement, occurs in a significant number of patients who develop this condition. Improved implantation techniques and resurfacing designs may help avoid this serious complication of hip resurfacing.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 223 - 223
1 May 2011
Grammatopoulos G Langton D Kwon Y Pandit H Gundle R Mclardy-Smith P Whitwell D Murray D Gill H
Full Access

Introduction: The development of Inflammatory Pseudotumour (IP) is a recognised complication following Metal on Metal Hip Resurfacing Arthroplasty (MoMHRA), thought to occur secondary to wear and elevated ion levels. Studies have shown that acetabular component orientation influences the wear of metal-on-metal hip replacement bearings. The aims of this study were to investigate the significance of cup orientation in the development of IP, and to identify a ‘safe-zone’ for cup placement with lower-risk for IP development.

Methods: Twenty six patients (n=27 hips) with IP confirmed radiologically, intra-operatively and histologically were matched for sex, age, pre-operative diagnosis, component size and follow-up with a cohort of asymptomatic MoMHRA patients (Control n=58). Radiographic acetabular anteversion and inclination were measured using EBRA. We calculated the distance in degree space of each acetabular component from the optimum position of 40° inclination and 20° anteversion, recommended by the designers, and thus compared acetabular component position between the two groups. Three different zones were tested as possibly optimum for acetabular placement. These were Lewinneck’s Zone (LZ) (inclination/anteversion; 30–50°/5–25°), and two zones defined by ±5° (Zone 1) or ± 10° (Zone 2) about the suggested target of 40°/20°. An optimal placement zone was determined based on a significant difference in IP incidence between components in the zone versus those outside.

Results: There was a wide range in cup orientations; mean inclination and anteversion were similar in the two groups: IP 47.5° (10.1°–80.6°)/14.1° (4.1°–33.6°) Vs Control 46.1° (28.8°–59.8°)/15.6° (4.3°–32.9°). Acetabular components in the IP group were significantly further away from the optimum position of 40°/20° in comparison to the controls (p=0.023). There was no difference in IP incidence between cups positioned within (IP:13/27, Control:35/58) or out of LZ (p=0.09) and within (IP: 2/27, Control: 10/58) or out of Zone 1 (p=0.156). Cups placed in Zone 2 (IP:6/27, Control:27/58) had significantly lower IP incidence versus those outside this zone (p=0.01). The odd’s ratio of developing IP when the cup is positioned out-of Zone 2 was 3.7.

Discussion: This study highlights the importance of ace-tabular component orientation in IP development. On the whole, patients with pseudotumour had acetabular components that were further away from the optimum position in comparison to the controls. However, a small number of IP patients had well-placed components implying that additional factors, possibly patient and/or gender specific, are involved in the development of pseudotumour. Furthermore, we defined an optimum, ‘safe-zone’ of ±10° around the cup position of 40°/20°. Patients with acetabular components outside this safe zone have an increased risk of IP development.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 177 - 177
1 May 2011
Solayar G Walsh P Murray D Mulhall K
Full Access

Introduction: Low-molecular-weight heparin is commonly used for thromboprophylactic therapy post orthopaedic surgery. Studies in the past have suggested that it may have a negative effect on osteoblasts and some have implicated its use with the risk of developing osteoporosis. Recently, Rivaroxaban, an oral Factor Xa inhibitor is gaining impetus for antithrombotic therapy over the last year and has been recommended for licensing by the FDA for this purpose. The effect of Rivaroxa-ban on bone and osteoblasts, if any, remains to be seen.

Methods: In a standardized in vitro model, human osteoblasts were cultured and exposed to a range of Enoxaparin and Rivaroxaban concentrations including their therapeutic dose. We evaluated the effects of these drugs on osteoblastic proliferation and activity using CellTiter 96 AQueous non-radioactive cell proliferation (MTS) and alkaline phosphatase assays respectively. Gene expression of Runt-related transcription factor 2 (Runx2), osteocalcin and bone morphogenetic protein 2 (BMP-2) were evaluated using Real time-polymerase chain reaction (RT-PCR) studies. Statistical analyses (t-test) were conducted using Microsoft Excel 2007.

Results: Rivaroxaban and Enoxaparin significantly reduced alkaline phosphatase activity (p< 0.05) however, no negative effects on osteoblastic proliferation was seen at all concentrations of both drugs. Rivaroxaban decreased Osteocalcin and Runx2 mRNA expression levels at 24 hours at therapeutic concentrations (p< 0.05). This effect was similarly found at therapeutic levels of Enoxaparin. Both Rivaroxaban and Enoxaparin significantly reduced BMP-2 mRNA expression both at 24 hours and 7 days at therapeutic concentrations. (p< 0.05).

Conclusion: Our study suggests that Rivaroxaban has similar negative effects on osteoblasts compared to Enoxaparin in the early stages. The increased duration of recommended Rivaroxaban therapy (2 and 5 weeks) post arthroplasty compared to Enoxaparin therapy (around 1 week) may have a more pronounced effect on bone homeostasis.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 172 - 172
1 May 2011
Gill H Grammatopoulos G Pandit H Glyn-Jones S Whitwell D Mclardy-Smith P Taylor A Gundle R Murray D
Full Access

Introduction: Metal-on-metal hip resurfacing arthroplasty (MoMHRA) has gained popularity as an alternative to THR for younger patients with osteoarthritis. A growing concern has been the association of MoMHRA with the development of inflammatory pseudotumours (IP), especially in women. These have been linked to metal-on-metal wear, which can be related to metal ion concentrations. Elevated metal wear debris levels may result from impingement, rim contact and edge loading. Head-neck ratio (HNR) is a predetermining factor for range of movement and impingement. Neck thinning is a recognised phenomenon post-MoMHRA and we have found an association of IP with increased neck thinning based on a case control study. Our aims were to identify HNR changes a hip undergoes when resurfaced and at follow up; and whether greater neck thinning at follow-up could be associated with the presence of elevated metal ions.

Methods: A cohort of 91 patients (57M:34F) with unilateral MoMHRAs were included in this study. Blood tests were obtained at a mean follow up of 3.9 years (range 1.7–7 years) and serum (Co:Cr) ion levels were measured (ICPMS). High metal ion concentrations were defined as Co> 4.1ppb and Cr> 5.2ppb. For all patients, head-neck ratio (HNR) was measured on plain anterio-posterior pelvic radiographs pre-operatively, immediately post-operatively and at follow-up.

Results: Female patients had significantly bigger HNR pre-op (mean=1.35, range:1.22–1.64) compared to males(mean=1.22, range:1.05–1.38) (p< 0.01). Immediately post-op, female HNRs (mean: 1.26, range: 1.14–1.34) were not different to male patients(mean=1.24, range=1.11–1.38) (p=0.11). At follow-up HNR was once again significantly bigger (p< 0.01) in females (mean=1.35, range: 1.21–1.49), compare to males (mean=1.27, range:1.11–1.38). HNR alterations with operation (p=0.00) and at follow-up (p< 0.01) were significantly bigger in female patients. Furthermore, there was a significant correlation between high ion levels and HNR change at follow-up for both Co (p=0.02) and Cr (p< 0.01).

Conclusion: This study identified gender-specific changes in HNR that resurfaced hips undergo, not previously documented. Female hips have greater HNR pre-operatively, compared to male hips, and appear to be biomechanically disadvantaged when resurfaced. A decrease in HNR with resurfacing could result in impingement and lead to processes, known to be more prevalent in females, such as neck thinning, increased wear and IP development. In addition, we highlight a correlation between high ion levels and greater neck thinning at follow-up. Increased neck thinning in symptomatic MoMHRA hips could be secondary to increased wear and should be investigated further radiologically for the presence of IP.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 173 - 173
1 May 2011
Grammatopoulos G Pandit H Gill H Murray D
Full Access

Introduction: Metal on metal hip resurfacing arthroplasty (MoMHRA) has become an alternative option to THR in the treatment of young adults with OA. A recognised MoMHRA complication is the development of an inflammatory pseudotumour (IP). IPs can be cystic (predominantly posterio-laterally located), solid (mostly anteriorly located) or mixed in nature. Diagnosis is made with the aid of US and/or MRI. To-date, no radiographic aid in the diagnosis of IP has been identified. Neck thinning is a recognised phenomenon following MoMHRA, occurring in up to 90% of resurfaced hips, which has not been associated with any adverse clinical events. Its pathogenesis is considered multi-factorial secondary to stress shielding, impingement, pressure effect on cancellous femoral neck, bone necrosis secondary to femoral preparation and altered vascularity/AVN. Our aim was to establish whether neck thinning is associated with the presence of a pseudotumour.

Methods: Thirty-one hips (30 patients) with IP confirmed clinically, radiologically, intra-operatively and histologically were matched for sex, age, pre-operative diagnosis, component size and follow-up with an asymptomatic MoMHRA cohort without pseudotumour (Control n=60). Radiological and operative findings at the time of revision of all IP patients were reviewed regarding location of pseudotumour; 4 different locations were defined: anteriorly-extending, posteriorly-extending, anteriorly & posteriorly-extending and within joint only. For all patients, prosthesis-neck ratio (PNR) at follow-up was measured on plain AP pelvic radiographs as previously described and validated.

Results: All IP patients (6M:24F) and all (12M:48F) but two controls had a posterior approach at the time of MoMHRA. Mean femoral component size was 46 mm for both groups. At an average follow up of 3.5 years (0.7–8.3), IP patients (mean 1.26, 1.10–1.79) had a significantly higher (p< 0.0001) PNR in comparison to their controls (mean 1.14, 1.03–1.35). Greater neck thinning had occurred in both IP-males (p< 0.001) and IP-females (p=0.002) in comparison to their controls. Location of IP and hence nature did not appear to have an effect on the degree of neck thinning.

Discussion: This study shows that IP patients had significantly narrower femoral necks at follow-up. Processes, such as impingement and increased wear that are thought to contribute to the process of neck narrowing are also thought to be factors in IP development. Furthermore, the presence of an IP, could lead to altered vascularity via a mass effect and further contribute to neck narrowing. Interestingly, nature of IP did not have a significantly affect PNR. Although one cannot be certain whether neck narrowing is a consequence or a contributing factor for IP development, their association is significant. Surgeons should consider the possibility of pseudotumour in symptomatic MoMHRA patients with neck narrowing.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 174 - 174
1 May 2011
Grammatopoulos G Kwon Y Langton D Pandit H Gundle R Whitwell D Mclardy-Smith P Murray D Gill H
Full Access

Introduction: Metal-on-metal hip resurfacing arthroplasty (MoMHRA) has gained popularity as an alternative to THR for younger patients with osteoarthritis. A growing concern has been the association of MoMHRA with the development of inflammatory pseudotumours (IP), especially in women. These have been linked to metal-on-metal wear, which can be related to metal ion concentrations. Although cup orientation has been shown to influence wear, the optimum cup position has not been clearly defined. We have identified an optimal cup orientation to minimise IP risk, based on a case controlled study, for inclination/anteversion within ±10° of 40°/20°. Our aim was to see if this optimal position results in lower metal ions, and to identify the boundary of an optimal placement zone for low wear.

Methods: A cohort of 104 patients (60M: 44F) with unilateral MoMHRA was included in this study. Blood tests were obtained at a mean follow up of 3.9 years (range 1.7–7 years) and serum Co and Cr ion levels were measured (ICPMS). High metal ion concentrations were defined as Co> 4.1ppb and Cr> 5.2ppb. Radiographic cup inclination and anteversion were measured using EBRA. The differences in ion levels between different cup orientation zones were investigated. Three orientation zones were defined centered on the target orientation of 40°/20°: Z1 within ±5°, Z2 outside ±5°/within ±10° and Z3: within ±10°.

Results: There was a wide range of cup placements; mean inclination/anteversion were 46.3°(21.5°–64.6°)/15°(2.7°–35.6°). Cr levels, but not Co, were higher in female patients (p=0.002) and those with small femoral components (< 50mm, p =0.03).

For the whole cohort, there was no significant difference in ion levels (Cr: p=0.092. Co=0.075) between cups positioned within Z3 (n=58) versus those outside (n=46 mean). Male patients with cups within Z3 (n=27) had lower ion levels in comparison to those outside Z3, which were significantly lower for Co (p=0.049) but not Cr (p=0.084). Female patients had similar levels within and out of Z3 for both ions (Cr: p=0.83, Co: p=0.84). However, patients with cups within Z1 (n=13) had significantly lower Co (p=0.005) and Cr (p=0.001) than those outside Z1 (n=95). Interestingly, Co levels were significantly lower in Z1 (n=13) in comparison to Z2 (n=33) (p=0.048) but Cr levels were not different (p=0.06).

Discussion: MoMHRA cups placed with ±5° of the ideal position of 40°/20°gave rise to significantly lower metal ions indicating lower wear within this narrow zone, in both sexes. This safe zone, could be extended to ±10° for male patients only. Gender specific factors, such as pelvic anatomy and joint flexibility, could be responsible for the narrower ‘safe’ zone seen in females. The narrower safe zone coupled with smaller components implanted are factors contributing to higher ion levels and hence the increased incidence of IP seen in females.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 19 - 19
1 Jan 2011
Pandit H Jenkins C Beard D Gill H Price A Dodd C Murray D
Full Access

The results of mobile bearing Oxford unicompartmental knee replacement (UKR) in the lateral compartment have been disappointing (five-year survival: 82%). Therefore, it is recommended that mobile bearings should not be used for lateral UKR. This low survivorship is primarily due to a high dislocation rate. A detailed analysis confirmed the elevated lateral tibial joint line to be a contributory factor to bearing dislocation. A new surgical technique was therefore introduced in which care was taken neither to remove too much bone from the distal femur nor to over tighten the knee and therefore ensure that the tibial joint line was not elevated. Other modifications included use of a domed tibial component.

The aim of this study is to compare the outcome of these iterations: the original series (series I), those with improved surgical technique (series II) and the domed tibial component (series III). The primary outcome measure was bearing dislocation at one year. One year was chosen as all the dislocations in the first series occurred within a year. In the original series (n=53), implanted using a standard open approach, there were six dislocations in the first year, the average flexion 110°, and 95% had no/mild pain on activity. In the second series (n=65), there were 3 dislocations, the average flexion was 117°, and 80% had no/mild pain on activity.

In the third series with the modified technique and a convex domed tibial plateau, there was one dislocation, average flexion was 125° and 94% had no/mild pain on activity. At four years the cumulative primary dislocation rates were 10%, 5% and 0% respectively, and were significantly different (p=0.04).

The improved surgical technique and implant design has reduced dislocation rate to an acceptable level so a mobile bearing can now be recommended for lateral UKR.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 26 - 26
1 Jan 2011
Gulati A Glyn-Jones S Simpson D Palan J Beard D Gill H McLardy-Smith P Gundle R Murray D
Full Access

Roentgen Stereophotogrammetric Analysis (RSA) can predict long-term outcome of prostheses by measuring migration over time. The Exeter femoral stem is a double-tapered highly polished implant and has been shown to subside within the cement mantle in 2 year RSA studies. It has a proven track record in terms of long-term survivorship and low revision rates. Several studies have demonstrated excellent clinical outcomes following its implantation but this is the first study to assess stem migration at 10 years, using RSA.

This is a single-centre study involving 20 patients (mean age: 63 years, SD=7) undergoing primary total hip replacement for degenerative osteoarthritis using the lateral (Hardinge) approach. RSA radiographs were taken with the patient bearing full weight post-operatively, at 3, 6, 12 months and at 2, 5 and 10 years follow-up. The three-dimensional migration of the Exeter femoral stem was determined.

The mean Oxford Hip Score at 10 years was 43.4 (SD=4.6) and there were no revisions. The stems subsided and rotated internally during a 10-year period. The mean migrations of the head and tip of the femoral stem in all three anatomic directions (antero-posterior, medio-lateral & supero-distal) were 0.69 mm posterior, 0.04 mm lateral and 1.67 mm distal for the head and 0.20 mm anterior, 0.02 mm lateral and 1.23 mm distal for the tip. The total migration at 10 years was 1.81 mm for the head and 1.25 mm for the tip.

The Exeter femoral stem exhibits migration which is a complex combination of translation and rotation in three dimensions. Comparing our 10 year with our previous 2 year migration results, the Exeter stems show continued, but slow distal migration and internal rotation. The subsidence continues to compress the cement and bone-cement interface which maintains secure fixation in the long term.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 19 - 19
1 Jan 2011
Pandit H Jenkins C Beard D Gill H Price A Dodd C Murray D
Full Access

About ten years ago we introduced sophisticated instrumentation and an increased range of component sizes for the Oxford unicompartmental knee replacement (UKR) to facilitate a minimally invasive surgical (MIS) approach. The device is now routinely implanted through an incision from the medial pole of the patella to the tibial tuberosity. This has resulted in a more rapid recovery and an improved functional result. As the access to the knee is limited there is a concern that the long term results may be compromised. The aim of this study was to determine the 10 year survival.

A prospective follow up of all Phase 3 minimally invasive Oxford UKR implanted by two senior authors (DWM & CAFD) has been undertaken. So far 1015 UKRs have been implanted for anteromedial osteoarthritis. All patients received a cemented implant through a MIS approach and were followed up prospectively by an independent observer. The data was collected prospectively regarding pre-operative status, complications and clinical as well as functional outcome at predetermined intervals.

The average age of patients was 66.4 years (range: 33 – 88) with mean Oxford Knee Score 41 (SD: 7.9) at the time of last follow up, Knee Society Score (objective) of 84 (SD: 13) and Knee Society Score (functional) of 83 (SD: 21). At ten years the survival of this cohort is 96%. There were 22 revisions including 7 for progression of arthritis, 5 for infection, 5 for bearing dislocation, 4 for unexplained pain and one for rupture of ACL secondary to trauma.

We conclude that the Oxford Knee can be implanted reliably through a minimally invasive approach, giving excellent long term results.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 66 - 67
1 Jan 2011
Kwon Y Thomas P Summer B McLardy-Smith P Ostlere S Gundle R Whitwell D Gibbons C Athanasou N Gill H Murray D
Full Access

Introduction: Symptomatic abnormal soft-tissue masses relating to the hip joint, such as those described as pseudotumours, are being increasingly reported following metal-on-metal hip resurfacing arthroplasty (MoMHRA). These were found to be locally destructive, requiring revision surgery in a high proportion (75%) of patients. Lymphocyte infiltrations seen in pseudotumours were similar to aseptic lymphocyte vascular associated lesion (ALVAL), which is thought to represent a T-lymphocyte-mediated delayed type hypersensitivity. Therefore, a delayed hypersensitivity reaction to nickel (Ni), chromium (Cr) or cobalt (Co) has been suggested to play a role in pseudotumour aetiology. In patients with bilateral MoMHRA who presented with symptoms on one side, subsequent scans have demonstrated pseudotumours both on the symptomatic and asymptomatic side. Thus, there are concerns that there may be an appreciable number of asymptomatic pseudotumours that surgeons are unaware of and these may eventually become symptomatic.

Aim: The aims of this study were:

to determine the prevalence of asymptomatic pseudotumours after MoMHRA; and

to measure Co and Cr ion levels as well as lymphocyte proliferation responses to Ni, Co and Cr (the principal elements in the CoCr alloy used in MoMHRA) in MoMHRA patients with and without asymptomatic pseudotumours.

Methods: A total of 201 MoMHRA implanted hips in 158 patients (97 male, 61 female) with a mean age of 56 years (range 33–73 years) were evaluated. The mean follow-up was 61 months (range 13–88 months). Resurfacing devices implanted included 128 Birmingham Hip Resurfacing, 66 Conserve Plus and seven ReCap. The control groups included additional 20 patients, 10 male and 10 female (a mean age 68 years, range 57–80 years) with metal-on-polyethylene total hip arthroplasty and a further 22 age-matched patients (a mean age 55 years) without any metal implants. Ultrasound was used as the initial imaging modality and MRI was used to assess the extent of the identified masses. Patients with a soft-tissue mass had ultrasound-guided aspiration or core biopsy performed. Venous blood samples were collected in all patients for serum cobalt and chromium ion levels analysis using Inductively-Coupled Plasma Mass Spectrometer and lymphocyte transformation tests (LTT). The Oxford Hip Score (OHS) was used to measure the functional outcomes of patients. Acetabular component abduction angle was measured from standardised anteroposterior pelvis radiographs.

Results: Prevalence – Pseudotumours were found in 7 patients (6 female and 1 male). The overall prevalence of asymptomatic pseudotumours was 4%, with a relatively very high (30%) prevalence in females with bilateral implants. Histological examinations showed extensive necrosis of connective tissue, in which there were scattered aggregates of metal particles and a diffuse lymphocyte infiltrate.

Metal Ion Levels – The presence of pseudotumour was associated with significantly higher median serum cobalt levels (9.2mg/L vs. 1.9mg/L, p< 0.001), chromium levels (12.0mg/L vs. 2.1mg/L, p< 0.001), hip aspirate cobalt levels (1182 mg/L vs. 86.2mg/L, p=0.003), and aspirate chromium levels (883mg/L vs. 114.8mg/ L, p=0.006), as well as with inferior functional scores (OHS 41 vs. 47 p< 0.001). There was no significant difference in acetabular cup inclination angle (p=0.51). Lymphocyte Reactivity: A higher incidence and level of enhanced lymphocyte reactivity to Ni (p=0.001), but not to Co or Cr (the principal elements in the CoCr alloy used in metal-on-metal hip resurfacing implants), was found in patients with MoMHRA compared to the patients without MoM implants. However, lymphocyte reactivity to Co, Cr and Ni did not significantly differ in patients with pseudotumours compared to those patients without pseudotumours.

Conclusion: The prevalence of asymptomatic pseudotumours in females was high, especially in females with bilateral MoMHRA implants (30%). The patients with ‘asymptomatic’ pseudotumours were in fact mildly symptomatic. Lymphocyte reactivity to Co, Cr and Ni did not differ in patients with pseudotumour compared to those patients without pseudotumours, suggesting that systemic hypersensitivity type IV reactions, mediated by lymphocyte reactivity to these metals, is not the dominant mechanism in pathogenesis of the soft tissue pseudotumours. Furthermore, pseudotumours were not detected in those patients who had normal levels of cobalt and chromium ions. This suggests that pseudotumours do not occur if MoM articulations are well functioning. Therefore, pseudotumours are likely to be a biological consequence of the large amount of metal debris generated in vivo due to excessive wear.