OPTIMUM ACETABULAR COMPONENT POSITION FOR REDUCED ION LEVELS IN HIP RESURFACING



Abstract

Introduction: Metal-on-metal-hip-resurfacing-arthroplasty (MoMHRA) has been associated with the development of inflammatory pseudotumours(IP), especially in females. IPs have been linked to wear debris, which can be related to metal-ion blood levels. Acetabular component position has been shown to influence wear. We have identified an optimum component orientation minimising IP risk around an inclination/anteversion of 40°/20°±10°. Our aim was to see if this optimal position results in lower metal ions and to identify the boundary of an optimal placement zone for low wear.

Methods: A cohort of 104 patients(60M:44F) with unilateral MoMHRA was studied. Blood tests were obtained at a mean follow up of 3.9 years and serum Co/Cr levels were measured(ICPMS). High metal ion concentrations were defined as Co> 4.1ppb and Cr> 5.2ppb. Radiographic cup inclination/anteversion were measured using EBRA. The differences in ion levels between different orientation zones were investigated. Three orientation zones were defined centered on 40°/20°: Z1-within ±5°, Z2-outside ±5°/within ±10° and Z3-within ±10°.

Results: There was a wide range of cup placements. Females had significantly (p< 0.001) smaller components(mean:51, 44–60) than males(mean: 56, 52–64). Cr levels, but not Co, were higher in females(p=0.002) and those with small femoral components(< 50mm, p =0.03). Patients with cups within Z1 (n=13) had significantly lower Co(p=0.005) and Cr(p=0.001). Males with cups within Z3(n=27) had lower ion levels in comparison to those outside, which were significantly lower for Co(p=0.049) but not Cr(p=0.084). Females had similar ion levels within and out of Z3(Cr/Co: p=0.83/0.84). Co levels were significantly lower in Z1(n=13) in comparison to Z2(n=33)(p=0.048) but Cr levels were not different (p=0.06).

Discussion: MoMHRA cups placed within ±5° of the optimum(40°/20°) had significantly lower metal ions indicating lower wear within this narrow zone. This safe zone, could extend to ±10° for males only. The narrower safe zone coupled with smaller components implanted are possible factors contributing to the increased IP incidence seen in females.