header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

Eradication of Negative Bone Remodelling Using the C-Stem Polished, Triple-Tapered Femoral Implant

The International Society for Technology in Arthroplasty (ISTA)



Abstract

Negative remodelling of the femoral cortex in the form of calcar resorption due to stress-shielding, and femoral cortical hypertrophy at the level of the tip of the implant due to distal load transfer, is frequenly noted following cemented total hip replacement, most commonly with composite beam implants, but also with polished double tapered components.

The C-stem polished femoral component was designed with a third taper running from lateral to medial across and along the entire length of the implant, with the aim of achieving more proximal and therefore more natural loading of the femur. The implant is designed to subside within the femoral cement mantle utilising the cement property of creep, generating hoop stresses, which are transferred more proximally to the femoral bone, starting at the level of the medial calcar. The intention is to load the proximal femur minimising stress-shielding and calcar resorption, as well as reducing distal load transfer as signified by the lack of distal femoral cortical hypertrophy.

We present the results of a consecutive series of 500 total hip replacements using C-stem femoral components, performed between March 2000 and December 2005 at a single institution. Data was collected prospectively and all patients remain under annual follow-up by a Specialist Arthroplasty Practitioner. The operations were performed using a standard surgical technique with third generation cementing using Palacos-R antibiotic loaded cement.

500 arthroplasties were performed on 455 patients with an average age at the time of surgery of 68.3 years (23-92). There were 282 (62%) female and 173 (38%) male patients with osteoarthritis being the predominant diagnosis. 77 patients have died (73 hips) and the average duration of follow-up for the entire series is 81 months (52-124).

Only 2 femoral implants have been revised - one for deep sepsis and the other as part of a revision procedure for a loose acetabulum, although the femoral component itself was not loose. One implant is currently loose following a periprosthetic fracture treated by internal fixation, but none of the remaining implants demonstrates any progressive radiolucencies in any Gruen zones or any features suggestive of current or future loosening. Calcar rounding has been observed, but there have been no cases with obvious loss of calcar height and no cases of distal femoral cortical hypertrophy.

The C-stem femoral component has therefore performed well in clinical practice and the objective of eradicating negative bone remodelling has been achieved. The study is ongoing.


Email: