Introduction

Current standard cups of metal on metal resurfacing hip arthroplasty (RHA) have no dome holes and it is very difficult for surgeons to confirm full seating of these cups. This sometimes results in gap formation between the cup and acetabular floor. Although the incidence of initial gaps using modular press-fit cups with dome screw holes has been reported to range from 20 to 35%, few studies have reported the incidence of gap formation with monoblock metal cups and its clinical consequences in RHA. The purpose of this study was to investigate retrospectively the incidence of initial gap formation and whether the initial gap influences the clinical results in RHA.

Material and Method

RHA was performed on 166 hips of 146 patients using the Birmingham Hip Resurfacing (BHR) (MMT, UK) between 1998 and 2007. Mean age at operation was 48.7 years (range, 19-85 years). Mean duration of follow-up was 6.9 years (2.0-10.6). Acetabular reaming was performed with the use of hemispherical reamers and the reamer size was increased up to an odd number diameter which provided tight rim fit in the antero-posterior direction. The same size hemispherical provisional cup with dome holes and slits was used to check the cavity for complete seating. If the provisional cup could not be seated on the floor, reaming was repeated with the same reamer to remove the rim bump until full seating was achieved. Acetabular cups of 1mm larger diameter were impacted into the acetabulum by a press-fit technique. After press-fit fixation, the stability of the cups was confirmed with a synchronized movement of the pelvis and the cup inserter by applying a gentle torque. Clinical evaluation was performed using WOMAC at the latest follow-up. Radiographic assessments were performed using radiographs immediately after the surgery, at 3 weeks, 3 months, 1 year, and then annually thereafter. We evaluated the height of the gap between the cup and acetabular surface, cup inclination angle, cup migration and the time to gap filling. To investigate the relationship between the magnitude of the gap and the radiographic results, the patients were divided into two groups according to the height of the initial gap; the cases with a gap of less than 3 mm on the initial radiograph were grouped into a small gap group, the cases with a gap of 3mm or more were grouped into a large gap group. We compared the changes in the height of the gap, in the cup inclination angle and the cup migration between the groups.

Results

Gaps were identified in 21 of 166 hips (13%) on the postoperative radiograph. The average height of the gaps was 2.4 mm (0.56-4.5mm). Of the 21 hips with gaps, there was no revision during the follow up period. 12 of the 21 hips were classified into the small gap group, 9 of the 21 hips were classified into the large gap group. In the small gap group, there were no changes of cup inclination angle of more than 3 degrees nor was there cup migration of more than 3mm. On the other hand, in the large gap group, 6 of the 9 hips showed reductions of the cup inclination angle of more than 3 degrees with 3mm or more of migration during the initial 3 months (P<0.05). After 3 months, neither progressive angle change nor migration of the cup were observed. All acetabular components were judged to be bone ingrown at the last follow up.

Conclusions

The incidence of initial postoperative gaps (13%) of this series is similar to that of modular cementless cups fixed with press-fit technique. There were no serious clinical consequences of the initial gap during the 6.9 year follow up. However, gaps of 3mm or more led to early migration of the acetabular component and change in inclination angle.