10-17 Years Survivorship of CAD-CAM Primary and Revision Hip Stems

The International Society for Technology in Arthroplasty (ISTA)



Abstract

Total hip replacements have shown great benefits to patients through relief of pain and restoration of function. However, because of the extensive variation in the size and shape of the femoral canal, especially for the situation encountered in the revision hip arthroplasty, standard uncemented hip systems with a limited number of sizes are unable to provide an accurate fit in every case. This study showed clinical results of 112 primary total hip replacements and 158 revision total hip replacements, using custom made CAD-CAM (Computer Aided Design-Computer Aided Manufactured) hip prostheses inserted between 1992 and 1998.

For primary hip replacements, the implants were designed to produce proximal line-to-line fit with the femoral bone and to provide optimal biomechanical environment of the hip. The stem was HA coated, 53 males and 58 females were included. Mean age was 46.2 years (range 24.6yrs - 62.2 yrs). The average duration of the follow up was 24 years (10 – 17 years). The mean Harris Hip Score (HHS) was improved from 42.4 to 90.3, mean Oxford Hip Score (OHS) was improved from 43.1 to 18.2 and the mean WOMAC hip score was improved from 57.0 to 11.9. There was 1 revision due to failure of the acetabular components but there were no failures of the femoral components. In the whole follow-up period, the survival of the femoral stem alone was 100%.

For revision hip replacements, the implants were designed using our design strategies of graduate approach to different revision situation based on Paprosky's classification of femoral bone defect. The implants were HA coated; some of them had distal cutting flutes. A total of 158 patients (97 males and 61 females) who had operation between 1991 and 1998 were followed up, among them 138 cases were due to aseptic loosening, 6 cases were periprosthetic fractures and 14 cases were infection. The average age was 63.1 years (range 34.6 – 85.9 years). The minimum follow up was 10 years (range 10 – 12 years). The mean Harris Hip Score was improved from 44.2 to 89.3, mean Oxford Hip Score was improved from 41.1 to 18.2 and the mean WOMAC hip score was improved from 52.4 to 12.3 respectively. 6 cases required further revision surgery, among them 3 were due to aseptic implant loosening, the overall survivorship at ten years was 97%.

The CAD-CAM hip stems are able to provide optimal implant fixation and restore hip function for every patient regardless their original femoral shape, bone condition and biomechanics of the hip. The excellent medium to long term clinical results justifies the use of CAD-CAM custom hip stems.