Extramedullary Femoral Reference Without Navigation for Total Knee Arthroplasty. a Multicenter Randomized Clinical Trial

The International Society for Technology in Arthroplasty (ISTA)



Abstract

The aim of tissue sparing surgery in total knee arthroplasty is to reduce surgical invasivity to the entire knee joint. Surgical invasion should not be limited only toward soft tissues but also toward bone. The classic technique for total knee arthroplasty implies intramedullary canal invasion for proper femoral component positioning. This phase is associated to fat embolism, activation of coagulation, and occult bleeding from the reamed canal. The purpose of our study was to validate a new extramedullary device which relies on templated data.

Two-hundred patients in four different orthopaedics centres were randomized to undergo primary total knee arthroplasty either using standard intramedullary femoral instruments (IM group) or using a new extramedullary device (EM group). A new set of instruments was developed to control the sagittal and coranl plane of the distal femoral resection. The extramedullary instrument was calibrated referencing to templated data obtained from the preoperative long-limb radiograph (Fig 1, 2). Varus-valgus orientation of the resection were established by moving the two paddles according to templated data. An L-shaped sliding tool (5 centimetres long) over the anterior cortex controls the flexion-extension parameter of the resection and is intended to allow a cut flush with the anterior cortex at 0° of angulation with the distal aspect of the femoral diaphysis on the sagittal plane

Femoral component coronal alignment was within 0±3° of the mechanical axis in 86% of the IM group and 88% of the EM group. Sagittal alignment of the femoral component was 0±3° in 80% of the IM group and 94% of the EM group. There was no difference in the average operative time between the two groups. The EM group showed a trend toward less postoperative blood loss

Extramedullary reference with careful preoperative templating can be safely utilized during total knee arthroplasty.