header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Hip

Long-term results of uncemented total hip arthroplasty with the Taperloc femoral component in patients with Dorr type C proximal femoral morphology



Download PDF

Abstract

Aims

To investigate the longevity of uncemented fixation of a femoral component in total hip arthroplasty (THA) in patients with Dorr type C proximal femoral morphology.

Patients and Methods

A total of 350 consecutive uncemented THA in 320 patients were performed between 1983 and 1987, by a single surgeon using the Taperloc femoral component. The 63 patients (68 hips) with Dorr type C proximal femoral morphology were the focus of this review. The mean age of the patients was 69 years (24 to 88) and mean follow-up was 16.6 years (ten to 29). Survival analysis included eight patients (eight hips) who died without undergoing revision surgery prior to obtaining ten years follow-up. All 55 surviving patients (60 hips) were available for clinical assessment and radiographic review. As a comparator group, the survival and implant fixation in the remaining 282 THAs (257 patients) with Dorr type A and B morphology were evaluated. The mean age of these patients was 52 years (20 to 82).

Results

In the Dorr C patient group the mean Harris hip score improved from 51 points (21 to 69 points) pre-operatively to 89 (74 to 100) at final follow-up. No femoral component was loose by radiological criteria and osteolysis was only identified around two stems (3.3%). There was one revision (1.6%) of a well-fixed femoral component for sepsis at 11 years. The survival of the Taperloc femoral component at 20 years with revision for any reason as the endpoint was 98% (95% confidence interval; 87 to 99).

A total of ten (3.5%) of the Dorr A and B patient group of 282 THAs required revision surgery. Only one (0.4%) for aseptic loosening. A total of two hips (1%) were loose by radiographic criteria and osteolysis occurred around 12 hips (4%).

Conclusion

This study demonstrates that excellent fixation can be achieved using the Taperloc stem in patients with Dorr type A and B, and Dorr type C bone.

Take home message: The Taperloc stem demonstrated equivalent results in Dorr type A and B and Dorr type C bone.

Cite this article: Bone Joint J 2016;98-B:595–600.


Correspondence should be sent to Dr J. R. McLaughlin; e-mail:

For access options please click here