The Acclaim total elbow replacement is a modular system which allows implantation in both unlinked and linked modes. The results of the use of this implant in primary total elbow replacement in 36 patients, operated on between July 2000 and August 2002, are presented at a mean follow-up of 36 months (24 to 49). Only one patient did not have good relief of pain, but all had improved movement and function.

No implant showed clinical or radiological loosening, although one had a lucent area in three of seven humeral zones. The short-term results of the Acclaim total elbow replacement are encouraging. However, 11 patients (30.5%) suffered an intra-operative fracture of the humeral condyle. This did not affect the outcome, or the requirement for further surgery, except in one case where the fracture failed to unite. This problem has hopefully been addressed by redesigning the humeral resection guide. Other complications included three cases of ulnar neuropathy (8.3%) and one of deep infection (2.8%).


Correspondence should be sent to Mr I. A. Trail; e-mail: upperlimb@wrightington.org.uk