header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

PREVENTION OF IMMOBILIZATION RELATED MUSCULAR ATROPHY USING THE MYOSPARE DEVICE: A CONTROLLED, RANDOMIZED, OPEN STUDY TO INVESTIGATE THE FEASIBILITY, SAFETY AND EFFICACY OF ELECTRICAL GASTROCNEUMIUS STIMULATION IN ANKLE FRACTURES



Abstract

Introduction Musculoskeletal injuries, especially fractures, cause reduced limb mobilization. The diminished limb activity promotes muscular atrophy, leading to a slower return to function. Attempts to prevent this atrophy using electrical stimulation have been described after knee reconstruction.

The Myospare percutaneous electrical stimulator has been developed to prevent immobilization related atrophy. We undertook this pilot study to assess feasibility, safety, and efficacy of applying electrical stimulation under a cast after ankle fractures.

Patients and Methods Between May and December 2004, patients who sustained closed ankle fractures requiring surgery, were recruited to participate in this study. 24 patients took part in the study, sixteen male and eight female. Age range was 18 to 62 years (average 40). All patients underwent open reduction and internal fixation using standard AO technique. A short walking cast was applied after surgery. Patients were randomized into a treatment and a control group. The experimental device was applied in the treatment group for 6 weeks. Patients were examined at 2, 6 and 12 weeks.

Evaluation included measurement of calf and ankle circumference, dorsiflexion and plantiflexion, and calculation of the ratio between the injured and uninjured side. At each visit pain intensity was assessed using a visual analog score, and patients filled out a function assessment questionnaire. Analysis was performed using chi square, t-test and repeated measures analysis.

Results All patients tolerated the stimulator well. No adverse effects were encountered. There is a trend toward improvement in calf diameter, dorsiflexion and plantarflexion. However, with the small number of patients in this study, no significant difference was apparent. Functional recovery and VAS scores were borderline higher in the treatment group at 12 weeks (p=0.043 and p=.049) when compared to baseline.

Discussion The use of the Myospare device under a cast in patients after surgical fixation of ankle fractures has been demonstrated as feasible and safe. In this pilot study a trend toward enhanced recovery was apparent in the treatment group.

The abstracts were prepared by Orah Naor, IOA Co-ordinator and Secretary. Correspondence should be addressed to Israel Orthopaedic Association, PO Box 7845, Haifa 31074, Israel.