THE USE OF AUTOLOGOUS BLOOD RE-TRANSFUSION DRAINS IN LOWER LIMB ARTHROPLASTY: A METHOD FOR REDUCING HOMOLOGOUS BLOOD TRANSFUSION.



Abstract

Introduction: Blood conservation has rapidly moved into political and medical agendas. The ongoing shortage of blood in blood banks and the discovery of vCJD pose a threat to UK blood supply with ever rising costs. The use of blood conservation techniques is increasingly being used in surgery to help reduce the need for homologous blood.

We studied the use of Autologous blood transfusion drains (Bellovac ABT) in lower limb arthroplasty compared with standard closed suction drains. We studied 123 lower limb arthroplasty (61 TKR & 62 THR) to see if there was a significant reduction in the need for homologous blood transfusion when using re-transfusion drains and its cost effectiveness.

Methods: Retrospective analysis of 123 patients undergone lower limb arthroplasty from March 2002 to Dec 2004 under one surgeon using the same technique for TKR and THR. 61 TKR (30 ABT drains v 31 standard drain) and 62 THR (30 ABT drains v 32 standard drain). Data was collected on sex, age, pre & post op Hb, volumes drained, volumes re-transfused and the number of homologous blood transfusions.

Results: 30 THR with ABT drains: 14 male, 16 female, mean age 68.7, mean pre op Hb 13.67, mean post op Hb 10.55,mean volume re-transfused 324ml, mean volume drained 466ml. 7 patients(23%) required additional homologous blood transfusion.

32 THR with standard drains: 14 male, 18 female, mean age 68.4, mean pre op Hb 12.96, mean post op Hb 9.36, mean volume drained 579.5ml. 24 patients (75%) required homologous blood transfusion.

30 TKR with ABT drains: 14 male, 16 female, mean age 69.8, mean pre-op Hb13.4, mean post-op Hb 11.03, mean volume re-transfused 415ml, mean volume drained 580ml. 4 patients (13%) required additional homologous blood transfusion.

31 TKR with Standard drains: 13 male, 18 female. Mean age72.1, mean pre-op Hb13.33, mean post-op Hb10.4, mean volume drained 711.5ml. 14 patients (45%) required homologous blood transfusion.

No re-transfusion complications occurred in the ABT group. 2 patients requiring homologous blood had increasing pyrexia and transfusion hence stopped.

Discussion: 11 out of 60 patients (18%) using ABT drains required additional homologous blood compared with 38 out of 63 patients (60%) requiring homologous blood using standard drains. Pvalue< 0.001. We show a stastically significant reduction in the need for homologous blood transfusion using an autologous blood re-transfusion drain. One unit of blood costs approximately £120 the ABT drain less than half of this amount, there is a significant cost saving in using autologous blood re-transfusion drains. We conclude that using Autologous blood Re-transfusion drains is safe, cost effective and reduces the need for homologous blood transfusion. If drains are to be used then Re-transfusion drains should be used.