PREDICTION OF CHARNLEY FEMORAL STEM ASEPTIC LOOSENING BY IMMEDIATE POST-OPERATIVE RADIOLOGICAL FEATURES



Abstract

Introduction: We describe the association between postoperative femoral stem radiological appearances and aseptic failure of THA (total hip arthroplasty).

Methods: A retrospective review of records and radiographs of all patients attending for follow-up between August 2002 and August 2003 who had a cemented Charnley femoral stem and either a cemented polyethylene acetabular cup inserted. Femoral stem aseptic loosening was defined either by findings at revision surgery, the definite radiographic loosening criteria of Harris or progressive endosteal cavitation across zones as described by Gruen. Well-fixed control THA’s were defined as those that demonstrated none of the radiographic features of aseptic loosening or ‘at risk’ signs as described by Wroblewski. Parameters measured were: Alignment, Barrack grade of cementation, cement mantle width of the cement mantle and the presence and width of any radiolucent lines

Results: 63 hips were entered into the aseptic failure group and 138 into the control group. The alignment of the femoral stem was not associated with failure (p=0.283). Thickness of the cement mantle was statistically associated with failure in Gruen zones 6 (p=0.040) and Gruen zone 7 (p=0.003). A significant association for the presence of radiolucent lines was found for Gruen zones 3 (p=0.0001) and 5 (p=0.0001). The grade of cementation as measured by the Barrack grade was strongly associated with failure for grades C (p=0.001) and D (p=0.001).

Discussion and conclusion: This study has demonstrated that easily applied radiological criteria can be used to identify ‘at risk’ Charnley THA’s from the immediate post-operative AP radiograph.

Editoral Secretary Mr Peter Howard. Correspondence should be addressed to BHS at the Royal College of Surgeons, 35 - 43 Lincoln’s Inn Fields, London WC2A 3PN.