header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Hip

The type of cement and failure of total hip replacements



Download PDF

Abstract

Using data from the Norwegian Arthroplasty Register, we have the assessed survival of 17 323 primary Charnley hip prostheses in patients with osteoarthritis based upon the type of cement used for the fixation of the implant.

Overall, 9.2% had been revised after follow-up for ten years; 71% of the failures involved aseptic loosening of the femoral component. We observed significantly increased rates of failure for prostheses inserted with CMW1 and CMW3 cements. Using implants fixed with gentamicin-containing Palacos cement as the reference, the adjusted Cox regression failure rate ratios were 1.1 (95% CI 0.9 to 1.4) for implants cemented with plain Palacos, 1.1 (95% CI 0.7 to 1.6) for Simplex, 2.1 (95% 1.5 to 2.9) for gentamicin-containing CMW1, 2.0 (95% CI 1.6 to 2.4) for plain CMW1 and 3.0 (95% CI 2.3 to 3.9) for implants fixed with CMW3 cement. The adjusted failure rate at ten years varied from 5.9% for implants fixed with gentamicin-containing Palacos to 17% for those fixed with CMW3.


Correspondence should be sent to Dr B. Espehaug.

For access options please click here