header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Hip

Improving the detection of acetabular osteolysis using oblique radiographs



Download PDF

Abstract

Visualisation of periacetabular osteolysis by standard anteroposterior (AP) radiographs underestimates the extent of bone loss around a metal-backed acetabular component. We have assessed the effectiveness of standard radiological views in depicting periacetabular osteolysis, and recommend additional projections which make these lesions more visible. This was accomplished using a computerised simulation of radiological views and a radiological analysis of simulated defects placed at regular intervals around the perimeter of a cadaver acetabulum. The AP view alone showed only 38% of the defects over all of the surface of the cup and failed to depict a 3 mm lesion over 83% of the cup. When combined with the AP view, additional 45° obturator-oblique and iliac-oblique projections increased the depiction, showing 81% of the defects. The addition of the 60° obturator-oblique view further improved the visualisation of posterior defects, increasing the rate of detection to 94%. Based on this analysis, we recommend using at least three radiographic views when assessing the presence and extent of acetabular osteolysis.


Correspondence should be sent to Dr J. E. Bechtold.

For access options please click here