header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Impingement and loosening of the long posterior wall acetabular implant



Download PDF

Abstract

Long posterior wall (LPW) Charnley acetabular implants are widely used as it is believed that the LPW helps to prevent dislocation. This has, however, not been proven statistically. In a preliminary study of these implants removed at revision marked erosion of the LPW was frequently seen, indicating that repetitive impingement may occur. The influence of the long posterior wall was therefore investigated mathematically. LPW and standard sockets were found to be equally likely to dislocate provided that the standard socket was anteverted 5 degrees more than the LPW socket. With simulated external rotation, LPW sockets impinge 30% earlier than standard sockets. When impingement occurs a torque is applied to the components, which increases the shear stresses at the cement-bone interface. The torques, although not large enough to dislodge the socket immediately, are repetitive and so may contribute to loosening. The LPW socket can generate twice as much torque as the standard socket and therefore is more likely to loosen.

For access options please click here