A biomechanical evaluation of the Gamma nail



Download PDF

Abstract

We examined the effect of the Gamma nail on strain distribution in the proximal femur, using ten cadaver femora instrumented with six unidirectional strain gauges along the medial and lateral cortices. The femora were loaded to 1800 N and strains were determined with or without distal interlocking screws before and after experimentally created two-part and four-part fractures. Motion of the sliding screw and the nail was also determined. Strain patterns and screw motion were compared with previously obtained values for a sliding hip screw device (SHS). The Gamma nail was shown to transmit decreasing load to the calcar with decreasing fracture stability, such that virtually no strain on the bone was seen in four-part fractures with the posteromedial fragment removed; increasing compression was noted, however, at the proximal lateral cortex. Conversely, the SHS showed increased calcar compression with decreasing fracture stability. The insertion of distal interlocking screws did not change the pattern of proximal femoral strain. The Gamma nail imparts non-physiological strains to the proximal femur, probably because of its inherent stiffness. These strains may alter bone remodelling and interfere with healing. Distal interlocking screws may not be necessary for stable intertrochanteric fractures.

For access options please click here