MINOR REACTIONS DUE TO MODERN METAL



Download PDF

Abstract

1. Two of the three metals at present in use in orthopaedic surgery have been studied to assess their tendency to cause wound reactions.

2. Cobalt-chrome alloy proved to be the better, the incidence of obligatory plate removal being at most 3 per cent. Visible corrosion in this metal never occurred.

3. 18/8 Mo stainless steel proved to be the poorer, the incidence of obligatory plate removal being 20 per cent. Visible corrosion of the metal is estimated to have occurred in about 5 per cent of screws inserted.

4. Regardless of such accelerating factors as metallic transfer, 18/8 Mo steel may have to be accepted as inherently more susceptible to corrosion than is cobalt-chrome alloy.

5. The qualities of two other stainless steels also became evident. Previous work describing the very bad situation arising from the use of EMS was confirmed. By contrast, a steel that was in use before 1951 gave little trouble. This steel was probably FSL.