Radiological humeral adaptative changes five years after anatomical total shoulder arthroplasty using a standard-length cementless hydroxyapatite-coated humeral component



Download PDF

Abstract

Aims

The purpose of this study was to report bone adaptive changes after anatomical total shoulder arthroplasty (TSA) using a standard-length hydroxyapatite (HA)-coated humeral component, and to report on a computer-based analysis of radiographs to determine changes in peri-implant bone density objectively.

Methods

A total of 44 TSAs, performed between 2011 and 2014 using a cementless standard-length humeral component proximally coated with HA, were included. There were 23 males and 21 females with a mean age of 65 years (17 to 65). All shoulders had good quality radiographs at six weeks and five years postoperatively. Three observers graded bone adaptive changes. All radiographs were uploaded into a commercially available photographic software program. The grey value density of humeral radiological areas was corrected to the grey value density of the humeral component and compared over time.

Results

Stress shielding was graded as mild in 14 shoulders and moderate in three; the greater tuberosity was the predominant site for stress shielding. The mean metaphyseal and diaphyseal fill-fit ratios were 0.56 (SD 0.1) and 0.5 (SD 0.07), respectively. For shoulders with no radiologically visible stress shielding, the mean decrease in grey value in zones 1 and 7 was 20%, compared with 38% in shoulders with radiologically visible stress shielding.

Conclusion

The rate of moderate stress shielding was 7%, five years after implantation of a cementless standard-length HA-coated humeral component. Clinical observation of stress shielding identified on radiographs seems to represent a decrease in grey value of 25% or more.

Cite this article: Bone Joint J 2021;103-B(5):958–963.