Arthroplasty

Mismatch ‘never events’ in hip and knee arthroplasty

a cohort and intervention study



Download PDF

Abstract

Aims

The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.

Materials and Methods

Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).

Results

A total of 26 524 arthroplasties were analyzed. From 12 307 procedures that were undertaken before implementation of the compatibility check, 21 class A incompatibilities were identified (real- or high-risk combinations; 0.17%; 95% confidence interval (CI) 0.11 to 0.26). From 5692 hip and 6615 knee procedures prior to implementation of the compatibility check, we found rates of class A-I incompatibility (real- or high-risk combinations unknowingly inserted) of 0.14% (95% CI 0.06 to 0.28) and 0.17% (95% CI 0.08 to 0.30), respectively. From 14 217 procedures after the introduction of compatibility checking (7187 hips and 7030 knees), eight class A incompatibilities (0.06%; 95% CI 0.02 to 0.11) were identified. This difference was statistically significant (p = 0.008).

Conclusion

Our data presents validated estimates of the baseline incidence of incompatibility events for hip and knee arthroplasty procedures and shows that a significant reduction in class A incompatibility events is possible using a web-based recording system.

Cite this article: Bone Joint J 2019;101-B:960–969.


Correspondence should be sent to A. Odgaard; email: