Hip

Intra-operative digital imaging

assuring the alignment of components when undertaking total hip arthroplasty



Download PDF

Abstract

Aims

The aims of this study were to examine the rate at which the positioning of the acetabular component, leg length discrepancy and femoral offset are outside an acceptable range in total hip arthroplasties (THAs) which either do or do not involve the use of intra-operative digital imaging.

Patients and Methods

A retrospective case-control study was undertaken with 50 patients before and 50 patients after the integration of an intra-operative digital imaging system in THA. The demographics of the two groups were comparable for body mass index, age, laterality and the indication for surgery. The digital imaging group had more men than the group without. Surgical data and radiographic parameters, including the inclination and anteversion of the acetabular component, leg length discrepancy, and the difference in femoral offset compared with the contralateral hip were collected and compared, as well as the incidence of altering the position of a component based on the intra-operative image.

Results

Digital imaging took a mean of five minutes (2.3 to 14.6) to perform. Intra-operative changes with the use of digital imaging were made for 43 patients (86%), most commonly to adjust leg length and femoral offset. There was a decrease in the incidence of outliers when using intra-operative imaging compared with not using it in regard to leg length discrepancy (20% versus 52%, p = 0.001) and femoral offset inequality (18% versus 44%, p = 0.004). There was also a difference in the incidence of outliers in acetabular inclination (0% versus 7%, p = 0.023) and version (0% versus 4%, p = 0.114) compared with historical results of a high-volume surgeon at the same centre.

Conclusion

The use of intra-operative digital imaging in THA improves the accuracy of the positioning of the components at THA without adding a substantial amount of time to the operation.

Cite this article: Bone Joint J 2018;100B(1 Supple A):36–43.