The position and orientation of total knee replacement components
Abstract
We studied the intra- and interobserver reliability of measurements of the position of the components after total knee replacement (TKR) using a combination of radiographs and axial two-dimensional (2D) and three-dimensional (3D) reconstructed CT images to identify which method is best for this purpose.
A total of 30 knees after primary TKR were assessed by two independent observers (an orthopaedic surgeon and a radiologist) using radiographs and CT scans. Plain radiographs were highly reliable at measuring the tibial slope, but showed wide variability for all other measurements; 2D-CT also showed wide variability. 3D-CT was highly reliable, even when measuring rotation of the femoral components, and significantly better than 2D-CT. Interobserver variability in the measurements on radiographs were good (intraclass correlation coefficient (ICC) 0.65 to 0.82), but rotational measurements on 2D-CT were poor (ICC 0.29). On 3D-CT they were near perfect (ICC 0.89 to 0.99), and significantly more reliable than 2D-CT (p < 0.001).
3D-reconstructed images are sufficiently reliable to enable reporting of the position and orientation of the components. Rotational measurements in particular should be performed on 3D-reconstructed CT images. When faced with a poorly functioning TKR with concerns over component positioning, we recommend 3D-CT as the investigation of choice.
Malpositioning and malorientation are two major causes of pain following total knee replacement (TKR).1–3 The assessment of the position of the components after TKR is generally performed on radiographs.4–12
Gross malposition, particularly in varus-valgus and flexion-extension, might be identified on radiographs, but accurate measurement is prone to error owing to variation in rotation of the leg and magnification.5,13 Measurement of rotation of the components is difficult.14–16
Routine assessment of component position can also be performed on CT scans.4,6 Depending on the CT imaging protocol, the postoperative sagittal, coronal and rotational alignment of the components can be determined.17,18 Several different methods of measurement have been described.4,19 For rotational measurement of the femoral component, most authors use transverse 2D-CT slices measuring the angle between the surgical epicondylar axis and the posterior prosthesis axis.4,6 Less frequently, CT images are reconstructed in 3D, with the images aligned to standardised frames of reference, and rotational alignment is assessed on these images.20 For this purpose, specific protocols have been described that include not only the knee but also the head of the femur and the mid-ankle.21,22 This assessment of rotation may be more accurate and reproducible than measurements on 2D-CT slices. Only in 3D-CT can the measurements be adjusted for the variability of the position of the patient’s leg in the scanner, which may lead to a more reliable identification of the anatomical landmarks such as the medial and lateral femoral epicondyles.
The primary purpose of this study was to evaluate which of these imaging modalities best documented the position and orientation of the components of a TKR. We therefore set out to establish the intra- and interobserver reliability of tibial and femoral component measurements in patients after TKR using radiographs, axial 2D-CT images and 3D-CT.
Patients and Methods
A total of 30 knees were investigated in 29 patients with a mean age of 71 years (39 to 89) who had undergone primary TKR. The study was approved by our institutional review board.
The sagittal (γ and δ) and coronal (α and β) positions of the components were assessed according to Ewald23 on weight-bearing anteroposterior (AP), lateral and skyline radiographs of the knee (Fig. 1). For femoral sagittal alignment (flexion-extension, γ angle) an angle between the line of the midshaft of the femur and the neutral line of the femoral component was assessed. For tibial sagittal alignment (slope, δ angle) an angle between the line of the midshaft of the tibia and an AP tangent of the tibial component was determined. For femoral coronal alignment (varus-valgus, α angle) an angle between the line of the midshaft of the femur and a tangent of the distal condyles of the component was measured. For the tibial coronal alignment (varus-valgus, β angle) an angle between the line of the midshaft of the tibia and a tangent of the plateau of the component was measured.
The rotational alignment of the femoral component was assessed on axial 2D-CT images (Fig. 2) using the method described by Berger et al.4 In addition, the sagittal, coronal and rotational alignment of both components was assessed on 3D reconstructed CT images (Fig. 3) and using customised software as previously described.20,21 The rotation of the femoral component (femoral posterior component axis) was measured in relation to the epicondylar axis. The rotation of the tibial component (tibial posterior component axis) was measured in relation to the posterior tibial plateau axis.
Measurements were performed three times by an orthopaedic surgeon (MTH) at an interval of two weeks, each in random order, for assessment of intra-observer reliability. For assessment of interobserver reliability, the measurements were repeated by an orthopaedic surgeon (MTH) and a radiologist (PK). The observers were not aware of any previous measurements at the time.
Statistical analysis.
Data were analysed using SPSS version 13.0 (SPSS Inc., Chicago, Illinois). Sample size was calculated according to the reported estimates for reliability studies using intraclass correlation coefficients (ICCs), a measure that takes into account not only the aspect of correlation but also systematic differences between the observations.24 An ICC value of 1 indicates perfect reliability, 0.81 to 1 very good, 0.61 to 0.80 good, 0.41 to 0.60 moderate and < 0.40 poor reliability.24 The sample size was calculated on the basis of three repetitions of the measurements. The α-error and β-error was 0.05 and 0.20, respectively. In order to describe interobserver reliability, the ICC and mean Pearson correlation r of the three observations was calculated. Intra-observer reliability was calculated comparing the mean of the three observations of the first observer with those of the second. The reliability between 2D and 3D imaging methods was compared by applying the same calculations using the mean values of observer 1 and 2. The mean, minimum and maximum differences of the observed values were calculated for all comparisons. In order to compare the strength of reliability between the ICCs of 2D versus 3D imaging methods, Fisher’s z- and t-values of the difference of the ICCs were calculated. A p-value < 0.05 was considered to be statistically significant.
Results
The ICCs representing intra- and interobserver reliability for measurement of the position of the components on radiographs, 2D-CT and 3D-CT images are presented in Tables I and II.
Intra-observer reliability of the measurements on radiographs were moderate or poor, with the exception of the posterior tibial slope (ICC 0.81). 2D-CT was only moderately reliable in measuring the rotation of the femoral component, but 3D-CT was significantly better (ICC 0.60 vs 0.73; p < 0.001). The ICC values for all other measurements on 3D-CT were highly reliable (ICC 0.96 to 0.99). Measurements on radiographs showed moderate to good interobserver reliability (0.65 to 0.82), but rotational measurements of the femoral component on 2D-CT were less reliable (ICC 0.29). The interobserver reliability for the measurements of all variables on 3D-CT slices were near perfect (ICC 0.89 to 0.97). Measurements of the rotation of the femoral components on 3D-CT showed a significantly higher interobserver agreement than measurements on 2D-CT images (ICC 0.91 vs 0.29, p < 0.001; Table III).
The ICCs and Pearson’s correlations comparing the different measurements and imaging modalities show that the only plain radiological variable that approaches reliable correlation with the 3D method is the posterior slope of the tibia (Table IV), but even this has a maximum difference of 8.1°; 2D measurements of femoral rotation correlate well with 3D-CT, yet the maximum difference here is 6.2° (Table IV).
Discussion
This study investigates the intra- and interobserver reliability of different methods (radiographs, 2D-CT, 3D-CT) of assessing the position of the components after TKR. Three important findings emerged. First, the measurements of rotation of the components using 3D-CT were highly reliable with regard to inter- and intra-observer variability. In contrast, the rotational measurements on 2D-CT slices showed variable results and less intra- and interobserver reliability. Secondly, all ICC values for intra- and inter-observer measurements of femoral varus-valgus, femoral flexion-extension, tibial rotation, tibia varus-valgus and tibial slope on 3D-CT were highly reliable. Thirdly, the measurements on radiographs showed good intra-and interobserver variability and low or only moderate ICC values for all positions and indicators of orientation. The sole exception was measurements of the sagittal tibial slope, which were highly reliable.
This study has two weaknesses. It was a small study, containing information from only 30 knees. However, the strength of the ICC values in the 3D group and the size of the difference in values gives some confidence that the findings are more widely applicable. We have not validated the 3D method using a standard such as digitised phantoms, which would have further supported the methods and conclusions. However, the focus of this paper is clinical practice, from which the images were obtained, so the data obtained have clinical relevance. The high reliability of the 3D-CT data suggests that there is no major problem with this dataset, nor with this method.
The outcome of TKR depends on a number of details, such as patient-related factors, restoration of the mechanical axis, adequate soft-tissue balancing and appropriate alignment of the components.1,2,4,6,7,11
Abnormal varus or valgus alignment has been reported to be a cause of loosening. Abnormal femoral extension or flexion may lead to notching or early loosening owing to gapping of the anterior femoral cortex-prosthesis interface. Rotational malalignment leads to patellar maltracking, anterior knee pain, femorotibial flexion instability, decreased movement and increased polyethylene wear.25–31 Higher revision rates have been found in patients with malaligned femoral or tibial components,1,2,4,6,32,33 and so identifying malalignment is crucial for optimal management of patients with a painful TKR.
A variety of different methods using conventional radiographs or axial 2D-CT images have been used to assess the position of the components after TKR.2,4,10,20,23,34,35 Most commonly the American Knee Society’s radiological assessment system, which is based on the anatomical axis is used.23
However, based on our findings, the inter- and intra-observer variability of plain radiographs is high, indicating low clinical utility and reproducibility. Another major limitation is that although the coronal and sagittal orientation can easily be assessed using plain radiographs, the assessment of rotation of the components is difficult.4,34 To our knowledge, only Eckhoff et al10,11 have described a technique for measuring rotation from plain radiographs. However, its use was limited by problems with standardisation of radiographs, and it could only be used with a specific type of TKR.
More recently Berger et al36 described a technique using axial 2D-CT images and the transepicondylar and posterior prosthetic component axes to assess rotation. They showed that excessive internal rotation of the femoral and tibial components was associated with an increased incidence of patellofemoral symptoms. As in their study, we used the commonly applied epicondylar axis for assessent of the femoral component. This is generally accepted as representing the functional flexion-extension axis of the knee joint.20,34 The posterior tibial axis, combined with a tangential line to the posterior aspect of the tibial component, was used for assessment of the tibial component.20,34
2D-CT is an accurate method for determining rotation, but reliable identification of the anatomical landmarks remains the biggest problem.10,11,20,34 Clearly, the visualisation of both the lateral and medial epicondylar landmarks on the same CT slice is dependent on the orientation of the patient’s legs at the time of the CT scan and the width of the CT slices. This major limitation could be partly overcome by using 3D reconstructed CT images for assessment of the component. This does not solve the problem of having reliable anatomical landmarks for orientation, but minimises the inconsistencies.
Based on our findings, we recommend the use of a low-dose 3D reconstructed CT for determining the rotational, sagittal and coronal orientation of the components after TKR to reduce measurement errors. This method is fit for the purpose of describing the position and orientation of the components, whereas neither plain radiographs nor 2D-CT is sufficiently reliable to be effective. For the poorly functioning TKR, this technique will enable the surgeon to describe the accuracy of placement of the components with confidence.
| Within R1 (3 measures) | ||||||
|---|---|---|---|---|---|---|
| Reliability | Difference | |||||
| Mean Pearson r | ICC | p-value | Mean | Minimum | Maximum | |
| Femoral varus-valgus - α | 0.45 | 0.40 | 0.000 | 3.3 | 0.7 | 8.0 |
| Tibial varus-valgus - β | 0.41 | 0.38 | 0.000 | 1.6 | 0.0 | 3.3 |
| Femoral flexion-extension - γ | 0.32 | 0.33 | 0.002 | 2.3 | 0.7 | 21.3 |
| Tibial slope - δ | 0.80 | 0.81 | 0.000 | 1.6 | 0.0 | 3.3 |
| 2D-femoral rotation | 0.63 | 0.60 | 0.000 | 3.1 | 0.8 | 6.0 |
| 3D-femoral rotation | 0.75 | 0.73 | 0.000 | 1.1 | 0.0 | 6.5 |
| 3D-femoral varus-valgus | 0.97 | 0.97 | 0.000 | 0.6 | 0.0 | 5.1 |
| 3D-femoral flexion-extension | 0.98 | 0.98 | 0.000 | 0.6 | 0.0 | 2.0 |
| 3D-tibial rotation | 0.99 | 0.99 | 0.000 | 0.9 | 0.0 | 2.1 |
| 3D-tibial varus-valgus | 0.96 | 0.96 | 0.000 | 0.5 | 0.0 | 1.7 |
| 3D-tibial slope | 0.97 | 0.97 | 0.000 | 0.7 | 0.0 | 1.7 |
| R1 (mean) - R2 | ||||||
|---|---|---|---|---|---|---|
| Reliability | Difference | |||||
| Pearson r | ICC | p-value | Mean | Minimum | Maximum | |
| Femoral varus-valgus - α | 0.83 | 0.65 | 0.000 | 3.1 | 0.3 | 8.0 |
| Tibial varus-valgus - β | 0.70 | 0.70 | 0.000 | 0.9 | 0.0 | 3.0 |
| Femoral flexion-extension - γ | 0.86 | 0.82 | 0.000 | 1.7 | 0.0 | 8.0 |
| Tibial slope - δ | 0.91 | 0.79 | 0.000 | 2.0 | 0.3 | 4.7 |
| 2D-femoral rotation | 0.32 | 0.29 | 0.061 | 3.4 | 0.3 | 9.8 |
| 3D-femoral rotation | 0.91 | 0.91 | 0.000 | 0.7 | 0.0 | 4.4 |
| 3D-femoral varus-valgus | 0.97 | 0.97 | 0.000 | 0.8 | 0.0 | 4.9 |
| 3D-femoral flexion-extension | 0.97 | 0.97 | 0.000 | 0.7 | 0.0 | 3.3 |
| 3D-tibial rotation | 0.89 | 0.99 | 0.000 | 1.0 | 0.0 | 2.8 |
| 3D-tibial varus-valgus | 0.92 | 0.89 | 0.000 | 0.9 | 0.0 | 1.7 |
| 3D-tibial slope | 0.96 | 0.96 | 0.000 | 0.9 | 0.0 | 3.5 |
| ICC | Difference ICC (Fisher’s test) | ||||
|---|---|---|---|---|---|
| 2D | 3D | z-value | t-value | p-value | |
| 2D-femoral rotation - 3D-femoral rotation | 0.29 | 0.91 | 1.23 | 4.43 | < 0.001 |
| Femoral varus-valgus - α | 0.65 | 0.97 | 1.32 | 4.75 | < 0.001 |
| Femoral flexion-extension - γ | 0.82 | 0.97 | 0.94 | 3.37 | < 0.001 |
| Tibial varus-valgus - β | 0.70 | 0.89 | 0.55 | 2.00 | 0.046 |
| Tibial slope - δ | 0.79 | 0.96 | 0.87 | 3.15 | 0.002 |
| Pearson r | ICC | p-value | Mean difference | Minimum difference | Maximum difference | |
|---|---|---|---|---|---|---|
| 2D-femoral rotation - 3D-femoral rotation | 0.52 | 0.44 | 0.007 | 2.5 | 0.0 | 6.2 |
| 3D-femoral varus-valgus - α | 0.15 | 0.04 | 0.405 | 5.5 | 0.0 | 13.7 |
| 3D-femoral flexion-extension - γ | −0.21 | negative | - | 4.0 | 0.3 | 21.6 |
| 3D-tibial varus-valgus - β | 0.31 | negative | - | 2.8 | 0.0 | 6.8 |
| 3D-tibial slope - δ | 0.59 | 0.58 | < 0.001 | 2.8 | 0.2 | 8.1 |

Fig. 1 Measurement of the position of the component on conventional radiographs according to Ewald.23

Fig. 2 Two-dimensional CT image showing measurement of rotation of the femoral component using transepicondylar and posterior prosthetic axes.

Fig. 3 Three-dimensional CT images showing measurement of the position of the components.
No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.
References
- 1 Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L. Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop 2001;392:46–55. Crossref, Google Scholar
- 2 Hofmann S, Romero J, Roth-Schiffl E, Albrecht T. Rotational malalignment of the components may cause chronic pain or early failure in total knee arthroplasty. Orthopade 2003;32:469–76 (in German). Crossref, Medline, ISI, Google Scholar
- 3 Torga-Spak R, Parikh SN, Stuchin SA. Anterior knee pain due to biplanar rotatory malalignment of the femoral component in total knee arthroplasty: case report. J Knee Surg 2004;17:113–16. Medline, Google Scholar
- 4 Berger RA, Rubash HE, Seel MJ, Thompson WH, Crossett LS. Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop 1993;286:40–7. Google Scholar
- 5 Skyttä ET, Lohman M, Tallroth K, Remes V. Comparison of standard anteroposterior knee and hip-to-ankle radiographs in determining the lower limb and implant alignment after total knee arthroplasty. Scand J Surg 2009;98:250–3. Crossref, Medline, ISI, Google Scholar
- 6 Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R. Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 2009;24:570–8. Crossref, Medline, ISI, Google Scholar
- 7 Han HS, Seong SC, Lee S, Lee MC. Rotational alignment of femoral components in total knee arthroplasty: nonimage-based navigation system versus conventional technique. Orthopedics 2006;29(10 Suppl):S148–51. Medline, ISI, Google Scholar
- 8 Kanekasu K, Kondo M, Kadoya Y. Axial radiography of the distal femur to assess rotational alignment in total knee arthroplasty. Clin Orthop 2005;434:193–7. Google Scholar
- 9 Lam LO, Shakespeare D. Varus/valgus alignment of the femoral component in total knee arthroplasty. Knee 2003;10:237–41. Crossref, Medline, ISI, Google Scholar
- 10 Eckhoff DG, Piatt BE, Gnadinger CA, Blaschke RC. Assessing rotational alignment in total knee arthroplasty, Clin Orthop 1995;318:176–81. Google Scholar
- 11 Eckhoff DG, Metzger RG, Vandewalle MV. Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop 1995;321:28–31. Google Scholar
- 12 Oswald MH, Jakob RP, Schneider E, Hoogewoud HM. Radiological analysis of normal axial alignment of femur and tibia in view of total knee arthroplasty. J Arthroplasty 1993;8:419–26. Crossref, Medline, Google Scholar
- 13 Bathis H, Perlick L, Tingart M, et al. Alignment in total knee arthroplasty: a comparison of computer-assisted surgery with the conventional technique. J Bone Joint Surg [Br] 2004;86-B:682–7. Link, Google Scholar
- 14 Dennis DA. Evaluation of painful total knee arthroplasty. J Arthroplasty 2004;19(4 Suppl 1):35–40. Google Scholar
- 15 Mandalia V, Eyres K, Schranz P, Toms AD. Evaluation of patients with a painful total knee replacement. J Bone Joint Surg [Br] 2008;90-B:265–71. Link, Google Scholar
- 16 Toms AD, Mandalia V, Haigh R, Hopwood B. The management of patients with painful total knee replacement. J Bone Joint Surg [Br] 2009;91-B:143–50. Link, Google Scholar
- 17 Suter T, Zanetti M, Schmid M, Romero J. Reproducibility of measurement of femoral component rotation after total knee arthroplasty using computer tomography. J Arthroplasty 2006;21:744–8. Crossref, Medline, ISI, Google Scholar
- 18 Victor J. Rotational alignment of the distal femur: a literature review. Orthop Traumatol Surg Res 2009;95:365–72. Crossref, Medline, ISI, Google Scholar
- 19 Nicoll D, Rowley DF. Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg [Br] 2010;92-B:1238–44. Link, Google Scholar
- 20 Hirschmann MT, Iranpour F, Konala P, et al. A novel standardized algorithm for evaluating patients with painful total knee arthroplasty using combined single photon emission tomography and conventional computerized tomography. Knee Surg Sports Traumatol Arthrosc 2010;18:939–44. Crossref, Medline, ISI, Google Scholar
- 21 Henckel J, Richards R, Lozhkin K, et al. Very low-dose computed tomography for planning and outcome measurement in knee replacement: the imperial knee protocol. J Bone Joint Surg [Br] 2006;88-B:1513–18. Link, Google Scholar
- 22 Chauhan SK, Clark GW, Lloyd S, et al. Computer-assisted total knee replacement: a controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol). J Bone Joint Surg [Br] 2004;86-B:818–23. Link, Google Scholar
- 23 Ewald FC. The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop 1989;248:9–12. Google Scholar
- 24 Walter SD, Eliasziw M, Donner A. Sample size and optimal designs for reliability studies. Stat Med 1998;17:101–10. Crossref, Medline, ISI, Google Scholar
- 25 Akagi M, Matsusue Y, Mata T, et al. Effect of rotational alignment on patellar tracking in total knee arthroplasty. Clin Orthop 1999;366:155–63. Crossref, Google Scholar
- 26 Arima J, Whiteside LA, McCarthy DS, White SE. Femoral rotational alignment, based on the anteroposterior axis, in total knee arthroplasty in a valgus knee: a technical note. J Bone Joint Surg [Am] 1995;77-A:1331–4. ISI, Google Scholar
- 27 Bargren JH, Blaha JD, Freeman MA. Alignment in total knee arthroplasty: correlated biomechanical and clinical observations. Clin Orthop 1983;173:178–83. Google Scholar
- 28 Benjamin J. Component alignment in total knee arthroplasty. Instr Course Lect 2006;55:405–12. Medline, Google Scholar
- 29 Boldt JG, Stiehl JB, Hodler J, Zanetti M, Munzinger U. Femoral component rotation and arthrofibrosis following mobile-bearing total knee arthroplasty. Int Orthop 2006;30:420–5. Crossref, Medline, ISI, Google Scholar
- 30 Fang DM, Ritter MA, Davis KE. Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 2009;24(6 Suppl):39–43. Crossref, Medline, ISI, Google Scholar
- 31 Jeffcote B, Shakespeare D. Varus/valgus alignment of the tibial component in total knee arthroplasty. Knee 2003;10:243–7. Crossref, Medline, ISI, Google Scholar
- 32 Kinzel V, Scaddan M, Bradley B, Shakespeare D. Varus/valgus alignment of the femur in total knee arthroplasty: can accuracy be improved by pre-operative CT scanning. Knee 2004;11:197–201. Crossref, Medline, ISI, Google Scholar
- 33 Lemaire P, Pioletti DP, Meyer FM, et al. Tibial component positioning in total knee arthroplasty: bone coverage and extensor apparatus alignment. Knee Surg Sports Traumatol Arthrosc 1997;5:251–7. Crossref, Medline, Google Scholar
- 34 Jazrawi LM, Birdzell L, Kummer FJ, Di Cesare PE. The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation. J Arthroplasty 2000;15:761–6. Crossref, Medline, ISI, Google Scholar
- 35 Kawano T, Miura H, Nagamine R, et al. Alignment in total knee arthroplasty following failed high tibial osteotomy. J Knee Surg 2003;16:168–72. Medline, Google Scholar
- 36 Berger RA, Crossett LS, Jacobs JJ, Rubash HE. Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop 1998;356:144–53. Crossref, Google Scholar