header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Hip

The Exeter Universal cemented femoral component at 15 to 17 years

AN UPDATE ON THE FIRST 325 HIPS



Download PDF

Abstract

The first 325 Exeter Universal stems (309 patients) implanted at the originating centre were inserted between March 1988 and February 1990 by a group of surgeons with differing experience. In this report we describe the clinical and radiological results at a mean of 15.7 years (14.7 to 17.3) after operation with no loss to follow-up. There were 97 patients (108 hips) with replacements still in situ and 31 (31 hips) who had undergone a further procedure. With an endpoint of revision for aseptic loosening, the survivorship at 17 years was 100% and 90.4% for the femoral and acetabular component, respectively. The mean Merle D’Aubigné and Postel scores at review were 5.4 (sd 0.97) for pain and 4.5 (sd 1.72) for function. The mean Oxford score was 38.4 (sd 9.8) (0 to 48 worst-to-best scale) and the mean combined Harris pain and function score was 73.2 (sd 16.9). Radiological review showed excellent preservation of bone stock in the proximal femur and no failures of the femoral component.


Correspondence should be sent to Mr J. R. Howell; e-mail: jrhowell@doctors.org.uk

For access options please click here