Survivorship and long-term outcome of a consecutive series of 200 Scandinavian Total Ankle Replacement (STAR) implants



Download PDF

Abstract

Aims

We report the long-term clinical and radiological outcomes of a consecutive series of 200 total ankle arthroplasties (TAAs, 184 patients) at a single centre using the Scandinavian Total Ankle Replacement (STAR) implants.

Patients and Methods

Between November 1993 and February 2000, 200 consecutive STAR prostheses were implanted in 184 patients by a single surgeon. Demographic and clinical data were collected prospectively and the last available status was recorded for further survival analysis. All surviving patients underwent regular clinical and radiological review. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot scoring system. The principal endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan–Meier survival curves were generated with 95% confidence intervals and the rate of failure calculated for each year.

Results

A total of 84 patients (87 ankles) were alive by the end of this study. Of the surviving 84 patients (87 ankles; rheumatoid arthritis (RA), n = 40; OA, n = 47), 45 were women and 39 were men, with a mean age of 54 years (18 to 72 years) at the time of surgery. A total of 32 implants failed (16%), requiring revision surgery. The mean time to revision was 80 months (2 to 257). The implant survival at 15.8 years, using revision as an endpoint, was 76.16% (95% confidence interval (CI) 64.41 to 87.91). We found a steady but low decrease in survival over the study period. The mean AOFAS score improved from 28 (10 to 52) preoperatively to 61 (20 to 90) at long-term follow-up.

Conclusion

STAR prostheses in the United Kingdom have now been largely superseded by newer design TAAs, potentially with improved characteristics and surgical techniques. The long-term survivorship for the STAR prosthesis can provide a benchmark for these later designs of ankle arthroplasty.


Correspondence should be sent to H. Majeed; email: