PREVENTION OF DISLOCATION WITH THE DUAL MOBILITY CONCEPT DURING THA REVISION WITH SEVERE BONE LOSS, ABOUT 123 CASES

Canadian Orthopaedic Association (COA) and Canadian Orthopaedic Research Society (CORS) Annual Meeting, June 2016; PART 2.



Abstract

The main causes of total hip arthroplasty (THA) revisions are loosening and instability. Use of a dual mobility cup cemented in a acetabular reconstruction cage device limits the risk of instability and does not hinder the acetabular fixation during THA revisions. The objective of this study was to analyse a retrospective series of 123 THA revisions with antiprotusio cage and dual mobility socket.

Patients and methods: At a mean follow-up of 10 years, we analysed a continuous series of 123 revisions using a reconstruction device (87 Kerboull cross-plates, 12 Burch-Schneider antiprotrusio cages, 24 custom-fit Novae ARM cages associated in all cases with a Novae Stick dual mobility cup cemented into the cage). There were 80 women and 43 males. The mean age at the surgery was 69.2 years old.

PMA score increased from 9.6 +/− 3.06 preoperatively to 14.2 +/− 2. at the follow-up. 9 early dislocations occurred and one late dislocation. At the last follow-up, the X-rays showed nine hardware failures, including one cross-plate fracture, one hook fracture, and one flange fracture. Analysis of the radiological position of the cup showed a mean lowering of 13 mm and a 7 mm lateralisation compared to the preoperative position. 2 revisions for aseptic loosening and 3 for septic loosening were performed.

This study confirms the advantage of dual mobility cups during acetabular reconstruction cemented in antiprotrusio cages as a way to limit, without eliminating, the risk of dislocation. Therefore cemented fixation of dual mobility cups in cages appears to be a reliable short-term option.