TKR WEAR MEASUREMENT IN 72 PATIENTS USING RADIOSTEREOMETRIC ANALYSIS

Canadian Orthopaedic Association (COA) and Canadian Orthopaedic Research Society (CORS) Annual Meeting, June 2016; PART 2.



Abstract

Articulation of the polyethylene (PE) insert between the metal femoral and tibial components in total knee replacements (TKR) results in wear of the insert which can necessitate revision surgery. Continuous PE advancements have improved wear resistance and durability increasing implant longevity. Keeping up with these material advancements, this study utilises model-based radiostereometric analysis (mbRSA) as a tool to measure in vivo short-term linear PE wear to thus predict long-term wear of the insert.

Radiographic data was collected from the QEII Health Sciences Centre in Halifax, NS. Data consisted of follow-up RSA examinations at post-operative, six-, 12-, and 24-month time periods for 72 patients who received a TKR. Implanted in all patients were Stryker Triathlon TKRs with a fixed, conventional PE bearing of either a cruciate retaining or posterior stabilised design. Computer-aided design (CAD) implant models were either provided by the manufacturer or obtained from 3D scanned retrieved implants. Tibial and femoral CAD models were used in mbRSA to capture pose data in the form of Cartesian coordinates at all follow-ups for each patient. Coordinate data was manually entered into a 3D modeling software (Geomagic Studio) to position the implant components in virtual space as presented in the RSA examinations. PE wear was measured over successive follow-ups as the linear change in joint space, defined as the shortest distance between the tibial baseplate and femoral component, independently for medial and lateral sides. A linear best-fit was applied to each patient's wear data; the slope of this line determined the annual wear rate per individual patient. Wear rates were averaged to provide a mean rate of in vivo wear for the Triathlon PE bearing.

Mean linear wear per annum across all 72 patients was 0.088mm/yr (SD: 0.271 mm/yr) for the medial condyle and 0.032 mm/yr (SD: 0.230 mm/yr) for the lateral condyle. Cumulative linear wear at the 2-year follow-up interval was 0.207mm (SD: 0.565mm) and 0.068mm (SD: 0.484mm) for the medial and lateral condyles, respectively.

Linear PE wear measurements using mbRSA and Geomagic Studio resulted in 0.056mm/yr additional wear on the medial condyle than the lateral condyle. Large standard deviations for yearly wear rates and cumulative measurements demonstrate this method does not yet exhibit the accuracy needed to provide short-term in vivo wear measurement. Inter-patient variability from RSA examinations is likely a source of error when dealing with such small units of measure. Further analysis on patient age and body mass index may eliminate some variability in the data to improve accuracy. Despite high standard deviations, the results from this research are in proximity to previously reported linear wear measurements 0.052mm/yr and 0.054mm/yr. Linear wear analysis will continue upon completion of >100 patients, in addition to volumetric PE wear over the entire articulating surface.