header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Spine

MAGEC MAGNETIC GROWING RODS IN THE TREATMENT OF EARLY ONSET SCOLIOSIS

British Scoliosis Society (BSS) Annual Meeting



Abstract

Aim:

To determine the efficacy and safety of the Magec system in early onset scoliosis (EOS).

Methods:

In 2011, 6 males and 2 females had Magec rods, with an average age of 8.5 years (2.9–12.7 years), 7 patients had dual rods, and 1 had single. The main cause of EOS was idiopathic scoliosis (n=6), followed by Congenital (n=1), and Syndromic (n=1). Average follow up was 19.4 months (14–26 months). 4 of these patients had their previous Paediatric Isola growing rods exchanged to Magec, and 1 patient had an exchange from single to Dual Magec rods.

Results:

In total, the cohort of patients underwent 65 extensions in clinic with an average 8 each (4–19). The first rod extension was performed at an average of 4.6 months following surgery. From there on extensions were done at two monthly intervals on an average. A total of 75 radiographs were taken after Magec insertion, ranging from 5 to 15 per patient. The mean pre-operative Cobb angle 59° improved to 41° at final follow up. This is a mean reduction of 31%. The mean pre-operative T1-S1 length was 305mm which improved to 349mm at final follow up. A mean increase in spinal length of 14%.m. During follow up, 5 complications were noted: 2 metalwork pullouts, 1 extension failure, 1 rod fracture and 1 patient had dermatitis at site of wound.

Conclusion:

Our early results using the Magec magnetic growing rod system are encouraging with maintenance of curve correction and in achieving trunk growth in EOS.

Conflict Of Interest Statement: No conflict of interest.