header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Knee

VALIDITY OF MAKO PLANNING CT SCANS: ACCURATE UP TO ONE YEAR

The British Association for Surgery of the Knee (BASK) May 2022 Meeting, Newport, Wales, 17–18 May 2022.



Abstract

Abstract

Introduction

Mako robotic assisted knee arthroplasty requires a planning CT scan within 8 weeks of surgery according to the supplier's protocol. This is often impractical, therefore we evaluated whether CT scans remain valid for an extended period.

Methodology

Patients undergoing Mako partial (PKA) and total (TKA) knee arthroplasty were identified from our hospital database. The hospital PACS system was used to define the time interval between the initial planning CT scan and surgery, and whether further imaging was required prior to surgery.

Results

443 consecutive Mako cases (225 TKA and 218 PKA) were undertaken between November 2019 and December 2021 (33 cases to March 2020, and 410 cases from August 2020). CT scans were done within 8 weeks of surgery in 229 patients (51.7%); between 8 and 24 weeks in 148 patients (33.4%); between 24 and 48 weeks in 53 patients (12.0%); and more than 48 weeks in 13 patients (2.9%). Repeat pre-operative radiographs were done in the first 43 patients with a delay to surgery of more than 8 weeks following their CT scan. No gross anatomical changes were identified, and this practice was therefore discontinued. No patients required a repeat CT scan. There were no intra-operative registration errors in any patient in this series.

Conclusion

Planning CT scans were valid for up to one year in a large series of patients undergoing Mako PKA and TKA. This may allow for more cost-effective use of resources, while minimising irradiation to patients.