Abstract
Abstract
Introduction
All-tissue quadriceps tendon (QT) is becoming an increasingly popular alternative to hamstrings tendon (HT) and bone-tendon-bone (BTB) autograft for anterior cruciate ligament (ACL) reconstruction. The relatively short graft length however dictates that one, or both, ends rely on suture fixation. The strength of this construct is therefore extremely important. This study evaluates whether the use of a novel fixation technique can improve the tensile properties of the construct compared to a Krackow suture, and a looped tendon (suture free) gold standard.
Methods
Eighteen porcine flexor tendons were tested, across three groups; suture-tape Krackow, looped tendon, and the novel ‘strain suture’. Biomechanical testing simulated the different stages of ACL graft preparation and loading (60N preload for 10 minutes, 10 cycles from 10N to 75N, and 1000 cycles from 100N to 400N). Elongation and load to failure were recorded, and stiffness calculated for each construct.
Results
The mean elongation was significantly improved for the strain suture compared to the suture tape Krackow for preload, 10 cycle and 1000 cycle testing protocols respectively (1.36mm vs 4.93mm, p<001; 0.60mm vs 2.72mm, p<0.001; 2.95mm vs 29.08mm, p<0.001). Compared with the looped tendon, the strain suture demonstrated similar results for preload and 10 cycle elongation, but greater elongation during the 1000 cycle stage. Stiffness of the latter two constructs was similar.
Conclusions
Augmentation of the suture fixation using this novel technique provides a construct that is significantly superior to currently practised suture techniques, and similar in elongation and stiffness to a looped graft.