MAKO ROBOTIC REVISION TOTAL HIP ARTHROPLASTY

International Hip Society (IHS) Closed Meeting, Berlin, Germany, 3–5 November 2021.



Abstract

With the approval of our institute, we reviewed all the robot-assisted hip revision during October 2019 and August 2021. MAKO joint arthroplasty system was used to perform the hip revision surgery.

Seventy-one robot-assisted hip revision cases were included. Cup revisions were carried out in 68 patients while stem revisions were also carried out in 68 patients. Three types of registration techniques (extra acetabular bone surface based, liner based, metal shell based or cage surface based) on the acetabular side. The extra acetabular bone surface was the commonest used for registration (48/70, 68.6%, mean accuracy 0.37mm), followed by liner surface (11/70, 15.7%, mean accuracy 0.36mm), acetabulum cup (10/70, 14.3%, mean accuracy 0.37mm), and cage surface (1/70, 1.4%, accuracy 0.40mm). We succeeded cup registration and robotic arm guided cup insertion in all the cases. The average cup inclination and anteversion after revision were 40.87°±4.39° and 13.87°±4.24°, respectively. Cups in 62 cases (62/68, 91.2%) were within the Lewinnek safe zone while in 55 cases (55/68, 80.9%) were within the Callanan safe zone.

The Mako robot-assisted system could bring favorable cup reconstruction in hip revision with acceptable surgical time and blood loss. Accurate registration could be achieved by different methods.