TWO-YEAR RESULTS OF THE MOXIMED KINESPRING KNEE IMPLANT SYSTEM

Scottish Committee for Orthopaedics and Trauma (SCOT) meeting, Dunblane, Scotland, August 2019.



Abstract

MOXIMED KineSpring® Knee Implant System is an Orthopaedic device designed for younger or highly active patients with osteoarthritis. The device is placed under the skin, is attached to the tibia and femur, and contains springs which help limit some of the forces that are transmitted through the knee during activities such as walking or running and thereby relieve pain that may be experienced by patients with early arthritis of the knee. The aim of this study is to determine the long term safety and efficacy of the KineSpring knee implant system.

This is a prospective case series involving two centres in Glasgow. 29 patients (mean age of 45.1 years and range 18-65 years) were recruited into the study between 2011 and 2016. The Primary outcome measure was Oxford knee score (OKS) at 2, 5 and 10 years post-operatively. Secondary outcome measures include device related complications and survival, patient reported functional outcome measures, patient satisfaction, pain levels and change in radiographic classification of osteoarthritis

At 2-year follow-up, 7 implants were removed (74.1% survival). Complications include deep infection, requiring removal in 1 patient, 2 implant failures requiring removal and one spring breakage. In comparison to pre-operative measures there was an improvement in the pain (3.58 vs. 5.20, p=0.02), stiffness (4.16 vs. 4.47, p=0.6) and OKS (32.4 vs. 36.9, p=0.03).

The KineSpring improves overall pain, stiffness and functional outcome at 2 years following surgery, however there was a high rate of removal and further long-term follow up analysis is required regarding its effectiveness.