NAVIO® ROBOTIC-ASSISTED SURGICAL SYSTEM FOR TOTAL KNEE ARTHROPLASTY USING JOURNEY® II GUIDED-MOTION TOTAL KNEE SYSTEM

International Society for Technology in Arthroplasty (ISTA) meeting, 32nd Annual Congress, Toronto, Canada, October 2019. Part 1 of 2.



Abstract

Background

The JOURNEY™ II Cruciate-Retaining Total Knee System (JIICR) and the JOURNEY™ II Bi-Cruciate Stabilized Total Knee System (JIIBCS) (both, Smith & Nephew, Memphis, TN, USA) are used for the treatment of end-stage degenerative knee arthritis. Belonging to the JOURNEY family of knee implants, the relatively new devices are designed to provide guided motion. Studies suggest that long-term outcomes of robotic-assisted navigation in total knee arthroplasty (TKA) are superior to the classical approach. This is the first report describing early postoperative outcomes of the NAVIO® robotic-assisted surgical navigation using the JOURNEY™ II family of knee implants.

Materials & Methods

In this ongoing study, six investigational sites in the US prospectively enrolled 122 patients (122 TKAs, 64 JIIBCS and 58 JIICR). Patients underwent TKA using the NAVIO system (Figure 1), a next-generation semi-autonomous tool that uses handheld miniaturized robotic-assisted instrumentation that the surgeon manipulates in 6 degrees of freedom, but restricts cutting to within the confines of the pre-designated resection area of the patient's bone. The primary outcome was postoperative mechanical alignment on long leg X-ray at one month postoperative compared to operative target alignment. Alignment within ±3 degrees of the target alignment was considered a success.

Results

Average age was 65.7 years (range, 39–79); 60.7% were females. All patients underwent patella resurfacing. Two patients had revision prior to the one-month follow-up visit; two patients withdrew from the study. 95% (112/118) attended the one-month follow-up. Four patients were missing either baseline or follow-up long leg X-ray, resulting in 108 evaluated TKAs. Overall, 92.6% (100/108) of TKAs were within 3 degrees of the target alignment. Of these, 24.1%, 39.8-, 19.4%, and 9.3% were at 0, 1, 2, and 3 degrees of the target alignment, respectively. There were two revisions, one at 18 days postoperative and the second at 27 days postoperative.

Discussion

At the one-month follow-up, the NAVIO™ Robotic Assisted TKA procedures resulted in a very high success rate of 92.6% in achieving planned mechanical alignment compared to standard instruments as historical control (73.4%) based on literature1. This demonstrates the improved accuracy and reliability of the NAVIO™ Robotic Assisted Surgical System for TKA procedures.

For any figures or tables, please contact the authors directly.