THE EXETER SHORT REVISION STEM FOR CEMENT-IN-CEMENT FEMORAL REVISION: A FIVE- TO 12-YEAR REVIEW

British Hip Society (BHS) Meeting, Derby, England, March 2018



Abstract

Cement-in-cement femoral revision is a proven technique in revision total hip arthroplasty, with excellent results when using standard sized Exeter stems. The Exeter 44/00 125 mm short revision stem was introduced in 2004 to facilitate cement-in-cement revision. The stem is 25mm shorter and has a slimmer body to allow greater flexibility to adjust depth of insertion and version of the stem. However, it is not known if this change in stem length and size effects its longer term performance. We therefore reviewed the clinical outcome and survival of the Exeter 44/00 short stem used for cement-in-cement revision in our unit, with a minimum of 5 years follow up.

166 cases were performed between 2004 and 2010. 103 hips were available for 5 year clinical and radiological follow up, with 91 hips surviving to final review in 2017. At 5 years, 43 hips had died, 13 were revised and 7 were too frail to attend clinical review. The fate of all 166 hips were known and included in the survival analysis.

Median clinical scores improved significantly. Sixteen hips required re-revision (infection 6, loose cup 3, periprosthetic fracture 3, instability 2, stem fracture with chronic infection 1 and pain 1). Kaplan-meier survival analysis revealed 100% survival for aseptic loosening, 96.8% survival for stem failure and 88.9% survival for all causes.

This is the largest series with the longest follow up of the Exeter 44/00 short revision stem. The stem performs equally well as standard Exeter stems with regards to aseptic loosening. The single stem fracture occurred secondary to bone loss in chronic infection, highlighting the importance of providing adequate proximal support for the stem. Periprosthetic fracture occurred in 2.4% of this series of revision cases.

Larger registry-based studies may provide additional information on the performance of this stem.