Abstract

Background

Hip fractures cause significant morbidity and mortality, affecting 70,000 people in the UK each year. The dynamic hip screw (DHS) is used for the osteosynthesis of extracapsular neck of femur fractures, a procedure that requires complex psychomotor skills to achieve optimal lag screw positioning. The tip-apex distance (TAD) is a measure of the position of the lag screw from the apex of the femoral head, and is the most comprehensive predictor of cut-out (failure of the DHS construct). To develop these skills, trainees need exposure to the procedure, however with the European Working Time Directive, this is becoming harder to achieve. Simulation can be used as an adjunct to theatre learning, however it is limited. FluoroSim is a digital fluoroscopy simulator that can be used in conjunction with workshop bones to simulate the first step of the DHS procedure (guide-wire insertion) using image guidance. This study assessed the construct validity of FluoroSim. The null hypothesis stated that there would be no difference in the objective metrics recorded from FluoroSim between users with different exposure to the DHS procedure.

Methods

This multicentre study recruited twenty-six orthopaedic doctors. They were categorised into three groups based on the number of DHS procedures they had completed as the primary surgeon (novice <10, intermediate 10≤x<40 and experienced ≥40). Twenty-six participants completed a single DHS guide-wire attempt into a workshop bone using FluoroSim. The TAD, procedural time, number of radiographs, number of guide-wire retires and cut-out rate (COR) were recorded for each attempt.

Results

A significant construct effect was seen for TAD and COR between novice and other users (p < 0.05). The intermediate and experienced users were not significantly different for these metrics. For all other metrics, experienced users had the highest score, contrary to expectation.

Conclusion

FluoroSim was able to separate novice users from other cohorts for the two clinically significant outcome metrics. We can therefore partially reject the null hypothesis as construct validity was present for TAD and COR. We have demonstrated that FluoroSim has the potential to be a useful adjunct when learning the psychomotor skills needed for the DHS procedure away from theatre.