header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

PULLOUT STRENGTH OF SUTURE ANCHOR FIXATION IN DIFFERENT ANCHOR INTERVALS

The European Orthopaedic Research Society (EORS) 2018 Meeting, PART 2, Galway, Ireland, September 2018.



Abstract

Suture anchor have been used in surgical procedure of tendon or ligament repair. Recently, there has been developed an all suture anchor (soft anchor) which can be used even when the insertion area is narrow. But, the stability of soft anchors due to narrow zone has not been elucidated. This purpose of this study was to investigate stability of soft anchors with respect to their fixation intervals. Polyurethane foams with two different bone densities (10 pcf; 0.16g / cm³, 20 pcf; 0.32g / cm³) were used. All suture anchors and conventional suture anchors were fixed at 10mm, 5mm, and 2.5mm intervals. The failure load was measured using a mechanical testing machine. The average load to failure of conventional suture anchor were 200.4N, 200.2N, 184.7N in the 10mm, 5mm and 2.5mm interval with 10pcf foam bones and 200.4 N, 200.2 N and 184.7 N with the 20 pcf foam bone respectively. Average load to failure load of soft anchor was 97.3N, 93.9N and 76.9N with 10pcf foam bones and 200.4 N, 200.2 N and 184.7 N with 20 pcf foam bone. Suture screw spacing and bone density are important factors in anchor pullout strength. In osteoporotic bone density, insertion of the suture screw interval of 5 mm might be necessary.


Email: