EVALUATION OF IN VIVO WEAR OF VITAMIN E-STABILIZED HIGHLY CROSS-LINKED POLYETHYLENE AT FIVE YEARS: A MULTICENTRE RADIOSTEREOMETRIC ANALYSIS STUDY

International Hip Society (IHS) Closed Meeting, Gothenburg, Sweden, June 2018.



Abstract

The primary purpose of the current study was to evaluate and compare the wear properties of vitamin E-doped, highly-crosslinked PE (VEPE) and one formulation of moderately cross-linked and mechanically-annealed ultra-high molecular weight PE (ModXLPE) in patients five years after primary THA. We also sought to understand whether polyethylene wear is associated with radiographic evidence of bone resorption or with deterioration in patient-reported outcome measures (PROMs).

A total of 221 patients from four international centers were recruited into a prospective RSA and clinical outcomes study. Seventy percent (76%) of patients received VEPE (vs. ModXLPE) liners, and 36% received ceramic (vs. metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA radiographs were collected to measure PE wear.

We observed similar bedding in through the one-year interval and wear through the two-year interval between the two liner types. However, there was significantly more femoral head penetration in the ModXLPE cohort compared to the VEPE cohort at the five-year follow-up (p<0.001). The only variables independently predictive of increased wear were ModXLPE (vs VEPE) liner type (β=0.22, p=0.010) and metal (vs. ceramic) femoral head type (β=0.21, p=0.013). There was no association between increased wear and radiolucency development (p=0.866) or PROMs. No patients were found to have evidence of osteolysis.

At five-years postoperatively, patients treated with VEPE (vs. ModXLPE) and ceramic (vs. metal) femoral heads demonstrated decreased wear. At the longest follow-up (five years postoperatively), the wear rates for both liner groups were very low and have not led to any osteolysis or implant failures via aseptic loosening.