IS A MODULAR DUAL MOBILITY ACETABULUM A VIABLE OPTION FOR THE YOUNG, ACTIVE TOTAL HIP ARTHROPLASTY PATIENT?

International Hip Society (IHS) Closed Meeting, Gothenburg, Sweden, June 2018.



Abstract

To assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular dual mobility acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design.

This was a prospective study of patients  65 years of age, with a BMI  35 kg/m2, and UCLA activity score > 6 who received a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. All patients had a minimum of 2-year clinical follow-up.

Patient reported outcome measures, whole blood metal ion levels (ug/L), and periprosthetic femur BMD were measured at baseline and at 1- and 2-years postoperatively.

43 patients (30 male, 13 female; mean age 52.6 ± 6.5 years) were enrolled. Harris Hip Scores improved from 54.1 ± 20.5 to 91.2 ± 10.8 at 2 years postoperatively (p<0.001). All patients had radiographically well-fixed components, no patients have sustained an instability event, and no patients have required a return to the operating room or revision procedure.

Mean cobalt levels increased from 0.065 ± 0.03 ug/L preoperatively to 0.30 ± 0.51 at 1-year postoperatively (p=0.01), but decreased at 2 years postoperatively to 0.16 ± 0.23 (p=0.2) (Table 1). Four patients (9.3%) had a cobalt level outside the reference range (0.03 to 0.29ug/L) at 2 years postoperatively with values from 0.32 to 0.94. None were symptomatic

The mean femoral BMD ratio was maintained in Gruen zones 2 thru 7 at both 1- and 2-years postoperatively using this stem design (Table 2). At 2 years postoperatively, BMD in the medial calcar was 101.5% of the baseline value.

Use of a modular dual mobility prosthesis and cementless, tapered femoral stem has shown encouraging results in young, active patients undergoing primary THA. Elevation in mean cobalt levels and the presence of four patients outside the reference range at 2 years postoperatively demonstrates the necessity of continued surveillance in this cohort.